Skip to content

CP1150 Sound Processor Speech Perception Compared With the Next Generation of Signal Processing Technology

A Pre-Marketing, Prospective, Single-Site, Open-Label, Within-Subject, Pilot, Interventional Study of Adult Cochlear Implant Speech Perception With the CP1150 Sound Processor Compared With the Next Generation of Signal Processing Technology

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT05286385
Enrollment
10
Registered
2022-03-18
Start date
2022-05-30
Completion date
2022-06-29
Last updated
2025-02-07

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Hearing Impairment, Sensorineural

Brief summary

This clinical study aims to investigate speech performance in quiet with an OTE Sound Processor with modified firmware compared with the commercially available CP1150. The study also investigates CP1110 and CP1150 with Forward Focus.

Detailed description

This study will build on the evidence previously collected on OTE and BTE Sound Processors and will support the design goals for access to the same sound processing algorithms across future OTE and BTE variants, including evidence required on the automation of FF and future implant compatibility.

Interventions

Speech perception tests whilst the subject is fitted with the CP1150 Sound Processor

DEVICECP1150 NF

Speech perception tests whilst the subject is fitted with the CP1150 modified firmware (the addition of notch filters at 978Hz, 1956Hz, 2934Hz, 3912Hz).

DEVICECP1150 + FF

Speech perception tests whilst the subject is fitted with the CP1150 Sound Processor with ForwardFocus.

DEVICECP1110

Speech perception tests whilst the subject is fitted with the CP1110 Sound Processor.

Sponsors

Avania
CollaboratorINDUSTRY
Cochlear
Lead SponsorINDUSTRY

Study design

Allocation
NON_RANDOMIZED
Intervention model
CROSSOVER
Primary purpose
OTHER
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

1. Aged 18 years or older 2. Post lingually deafened 3. Implanted with the CI600 Series, CI500 Series or Freedom Series 4. At least 6 months experience with a cochlear implant. 5. At least 3 months experience with a CP910/920, CP950, CP1150, or CP1000 Sound Processor 6. MAP Total Stimulation Rate of 7.2kHz or greater 7. Able to score 30% or more with CI alone on a monosyllabic words in quiet test 8. Willingness to participate in and to comply with all requirements of the protocol 9. Fluent speaker in English as determined by the investigator 10. Willing and able to provide written informed consent

Exclusion criteria

1. Additional disabilities that would prevent participation in evaluations. 2. Implant location that would result in undesirable hearing performance or discomfort with an off-the-ear sound processor, as determined by the investigator. 3. Unable or unwilling to comply with the requirements of the clinical investigation, as determined by the Investigator. 4. Investigator site personnel directly affiliated with this study and/or their immediate families; immediate family is defined as a spouse, parent, child, or sibling. 5. Cochlear employees or employees of Contract Research Organisations or contractors engaged by Cochlear for the purposes of this investigation. 6. Currently participating or participated in another interventional clinical study/trial in the past 30 days unless (if less than 30 days) the prior investigation was Cochlear sponsored and determined by the investigator to not impact clinical findings of this investigation. 7. Implanted with other active implantable medical devices (e.g. pacemaker, defibrillator).

Design outcomes

Primary

MeasureTime frameDescription
Speech Perception in Quiet Using an CP1150 Sound ProcessorOne day (in booth testing)Paired difference in percentage CNC Words correct in quiet with the CP1150 Sound Processor and CP1150 Sound Processor with modified firmware.

Secondary

MeasureTime frameDescription
Speech Perception in Quiet Using CP1150 Sound Processor Using Forward Focus Combined With Standard Microphone Directionality.One day (in booth testing)Paired difference in percentage CNC Words correct in quiet with the CP1150 Sound Processor and CP1150 Sound Processor with forward focus
Speech Perception in Quiet With CP1150 and CP1110 Sound ProcessorsOne day (in booth testing)Paired difference in percentage CNC Words correct in quiet with the CP1150 Sound Processor and CP1110 Sound Processor

Countries

Australia

Participant flow

Participants by arm

ArmCount
All Study Participants (CP1150 SP Compared With Next Generation Signal Processing Technology)
All study participants attended a single study visit. At the study visit, subjects underwent hearing assessments. Speech perception tests were carried out in all study participants in a randomized manner with 4 interventions involving the CP1150 Sound Processor, CP1150 with modified firmware, CP1150 Sound Processor with Forward Focus and the CP1110 Sound Processor.
10
Total10

Baseline characteristics

CharacteristicAll Study Participants (CP1150 SP Compared With Next Generation Signal Processing Technology)
Age at hearing implant59.2 years
STANDARD_DEVIATION 7.64
Age, Continuous70.6 years
Race and Ethnicity Not Collected— Participants
Region of Enrollment
Australia
10 participants
Sex: Female, Male
Female
5 Participants
Sex: Female, Male
Male
5 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 10
other
Total, other adverse events
0 / 10
serious
Total, serious adverse events
0 / 10

Outcome results

Primary

Speech Perception in Quiet Using an CP1150 Sound Processor

Paired difference in percentage CNC Words correct in quiet with the CP1150 Sound Processor and CP1150 Sound Processor with modified firmware.

Time frame: One day (in booth testing)

ArmMeasureValue (MEAN)Dispersion
CP1150 NFSpeech Perception in Quiet Using an CP1150 Sound Processor13.82 percentage of correct wordsStandard Deviation 10
CP1150Speech Perception in Quiet Using an CP1150 Sound Processor11.53 percentage of correct wordsStandard Deviation 10
95% CI: [-6.29, 0.29]
Secondary

Speech Perception in Quiet Using CP1150 Sound Processor Using Forward Focus Combined With Standard Microphone Directionality.

Paired difference in percentage CNC Words correct in quiet with the CP1150 Sound Processor and CP1150 Sound Processor with forward focus

Time frame: One day (in booth testing)

ArmMeasureValue (MEAN)Dispersion
CP1150 NFSpeech Perception in Quiet Using CP1150 Sound Processor Using Forward Focus Combined With Standard Microphone Directionality.11.24 percentage of correct wordsStandard Deviation 10
CP1150Speech Perception in Quiet Using CP1150 Sound Processor Using Forward Focus Combined With Standard Microphone Directionality.11.53 percentage of correct wordsStandard Deviation 10
95% CI: [-8.04, 3.44]
Secondary

Speech Perception in Quiet With CP1150 and CP1110 Sound Processors

Paired difference in percentage CNC Words correct in quiet with the CP1150 Sound Processor and CP1110 Sound Processor

Time frame: One day (in booth testing)

ArmMeasureValue (MEAN)Dispersion
CP1150 NFSpeech Perception in Quiet With CP1150 and CP1110 Sound Processors11.53 percentage of correct wordsStandard Deviation 10
CP1150Speech Perception in Quiet With CP1150 and CP1110 Sound Processors11.97 percentage of correct wordsStandard Deviation 10
95% CI: [-2.95, 3.55]

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026