Implant Primary Stability and Ridge Dimensional Changes Utilizing the Osseodensification Protocol

Implant Primary Stability and Ridge Dimensional Changes Utilizing the Osseodensification Protocol

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT05283928
Enrollment
15
Registered
2022-03-17
Start date
2022-08-01
Completion date
2024-03-04
Last updated
2024-11-29

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Dental Implants

Keywords

osseodensification, ridge expansion

Brief summary

The purpose of this study is to compare ridge dimensional changes between densification and standard drilling protocols and to compare primary and secondary stability of implants placed by bone densification and standard drilling protocols

Interventions

DEVICEOsseoDensification (OD) protocol

Osseodensification is a drilling process by which bone is being compacted and auto-grafted in an outwardly expanding direction from the osteotomy. When the osseodensification drills are being rotated in a reversed, non-cutting direction a strong and dense layer of bone is formed along the walls and base of the osteotomy.

DEVICEstandard drilling (SD) protocol (NobelReplace® Conical)

osteotomy preparation through bone excavation is necessary in implant site development for the placement of dental implants. The process involves a forward cutting process with removal of bone tissue creating appropriate size osteotomy for the dental implant.

Sponsors

The University of Texas Health Science Center, Houston
Lead SponsorOTHER

Study design

Allocation
RANDOMIZED
Intervention model
PARALLEL
Primary purpose
TREATMENT
Masking
NONE

Intervention model description

Each participant served as their own control. Each participant had at least 1 control implant and at least 1 test implant.

Eligibility

Sex/Gender
ALL
Age
22 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

* Patients who understand and agree to this study. * Adequate Oral hygiene * One or more edentulous spaces, which are 10 weeks or longer after extractions

Exclusion criteria

* Pregnancy. * Smoking of more than 5 cigarettes/day. * History of alcoholism or drug abuse during the last 5 years. * Uncontrolled Hypertension or diabetes. * Patient with malignant tumor. * Patients on daily dose of steroids. * Patients with history of chemotherapy or radiation for the last 12 months.

Design outcomes

Primary

MeasureTime frameDescription
Change in Bone Ridge Width at Crestday of implant surgerybone ridge width will be measured using a Caliper
Change in Bone Ridge Width at 5mm Apical From Crestday of implant surgerybone ridge width will be measured using a Caliper
Change in Bone Ridge Width at 10mm Apical From Crestday of implant surgerybone ridge width will be measured using a Caliper

Secondary

MeasureTime frameDescription
Primary Stability of the Implants as Indicated by the Implant Stability Quotient (ISQ) as Measured by the Resonance Frequency Analysis Using the Ostell ISQ Meter 6 Weeks After Implant Placement6 weeks after implant placementThe Implant Stability Quotient (ISQ) value is a measurement used to assess the stability of dental implants. It is obtained by resonance frequency analysis and ranges from 1 to 100, with higher values indicating greater stability.
Primary Stability of the Implants as Indicated by the Implant Stability Quotient (ISQ) as Measured by the Resonance Frequency Analysis Using the Ostell ISQ Meter 12 Weeks After Implant Placement12 weeks after implant placementThe Implant Stability Quotient (ISQ) value is a measurement used to assess the stability of dental implants. It is obtained by resonance frequency analysis and ranges from 1 to 100, with higher values indicating greater stability.
Primary Stability of the Implants as Measured by the Insertion Torque Valuesday of implant surgeryInsertion torque refers to the force used to insert an implant into bone, and it is reported in units of Newton-Centimeters (N-cms).
Percent Change of Buccal Bone Thickness as Assessed by the Cone Beam Computer Tomography at 1.5 mm From the Implant Platform6 months post implantationThe buccal bone thickness (BBT) was measured using cone beam computed tomography (CBCT). The percentage change in BBT was calculated by comparing the post-implant measurements to baseline values, with positive percentages indicating an increase in thickness and negative percentages indicating a decrease in thickness.
Percent Change of Buccal Bone Thickness as Assessed by the Cone Beam Computer Tomography at 5 mm From the Implant Platform6 months post implantationThe buccal bone thickness (BBT) was measured using cone beam computed tomography (CBCT). The percentage change in BBT was calculated by comparing the post-implant measurements to baseline values, with positive percentages indicating an increase in thickness and negative percentages indicating a decrease in thickness.
Volume of the Ridge as Assessed by the Cone Beam Computer Tomography (CBCT)6 months post implantation
Primary Stability of the Implants as Indicated by the Implant Stability Quotient (ISQ) as Measured by the Resonance Frequency Analysis Using the Ostell ISQ Meter Baseline(After Implant Placement)baseline(after implant placement)The Implant Stability Quotient (ISQ) value is a measurement used to assess the stability of dental implants. It is obtained by resonance frequency analysis and ranges from 1 to 100, with higher values indicating greater stability.
Primary Stability of the Implants as Indicated by the Implant Stability Quotient (ISQ) as Measured by the Resonance Frequency Analysis Using the Ostell ISQ Meter 3 Weeks After Implant Placement3 weeks after implant placementThe Implant Stability Quotient (ISQ) value is a measurement used to assess the stability of dental implants. It is obtained by resonance frequency analysis and ranges from 1 to 100, with higher values indicating greater stability.

Countries

United States

Participant flow

Pre-assignment details

15 participants were enrolled. Each participant served as their own control. Each participant had at least 1 control implant and at least 1 test implant

Participants by arm

ArmCount
All Study Participants
The purpose of this study is to compare ridge dimensional changes between densification and standard drilling protocols and to compare primary and secondary stability of implants placed by bone densification and standard drilling protocols.
15
Total15

Baseline characteristics

CharacteristicAll Study Participants
Age, Continuous50.67 years
STANDARD_DEVIATION 14.8
Race and Ethnicity Not Collected— Participants
Region of Enrollment
United States
15 participants
Sex: Female, Male
Female
9 Participants
Sex: Female, Male
Male
6 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
deaths
Total, all-cause mortality
0 / 150 / 15
other
Total, other adverse events
0 / 150 / 15
serious
Total, serious adverse events
0 / 150 / 15

Outcome results

Primary

Change in Bone Ridge Width at 10mm Apical From Crest

bone ridge width will be measured using a Caliper

Time frame: day of implant surgery

ArmMeasureValue (MEAN)Dispersion
OsseoDensification (OD) ProtocolChange in Bone Ridge Width at 10mm Apical From Crest0.29 millimetersStandard Deviation 0.76
Standard Drilling (SD) ProtocolChange in Bone Ridge Width at 10mm Apical From Crest0.01 millimetersStandard Deviation 0.46
Primary

Change in Bone Ridge Width at 5mm Apical From Crest

bone ridge width will be measured using a Caliper

Time frame: day of implant surgery

ArmMeasureValue (MEAN)Dispersion
OsseoDensification (OD) ProtocolChange in Bone Ridge Width at 5mm Apical From Crest0.62 millimetersStandard Deviation 0.65
Standard Drilling (SD) ProtocolChange in Bone Ridge Width at 5mm Apical From Crest0.11 millimetersStandard Deviation 0.64
Primary

Change in Bone Ridge Width at Crest

bone ridge width will be measured using a Caliper

Time frame: day of implant surgery

ArmMeasureValue (MEAN)Dispersion
OsseoDensification (OD) ProtocolChange in Bone Ridge Width at Crest0.84 millimetersStandard Deviation 0.59
Standard Drilling (SD) ProtocolChange in Bone Ridge Width at Crest0.01 millimetersStandard Deviation 0.45
Secondary

Percent Change of Buccal Bone Thickness as Assessed by the Cone Beam Computer Tomography at 1.5 mm From the Implant Platform

The buccal bone thickness (BBT) was measured using cone beam computed tomography (CBCT). The percentage change in BBT was calculated by comparing the post-implant measurements to baseline values, with positive percentages indicating an increase in thickness and negative percentages indicating a decrease in thickness.

Time frame: 6 months post implantation

Population: Noting that some participants had multiple implants, data for 3 implants in the OsseoDensification (OD) protocol arm were not collected, which includes data for 2 participants. Noting that some participants had multiple implants, data for 3 implants in the standard drilling (SD) protocol arm were not collected, which includes data for 2 participants

ArmMeasureValue (MEAN)Dispersion
OsseoDensification (OD) ProtocolPercent Change of Buccal Bone Thickness as Assessed by the Cone Beam Computer Tomography at 1.5 mm From the Implant Platform6.24 percent changeStandard Deviation 31.8
Standard Drilling (SD) ProtocolPercent Change of Buccal Bone Thickness as Assessed by the Cone Beam Computer Tomography at 1.5 mm From the Implant Platform-21.3 percent changeStandard Deviation 20.3
Secondary

Percent Change of Buccal Bone Thickness as Assessed by the Cone Beam Computer Tomography at 5 mm From the Implant Platform

The buccal bone thickness (BBT) was measured using cone beam computed tomography (CBCT). The percentage change in BBT was calculated by comparing the post-implant measurements to baseline values, with positive percentages indicating an increase in thickness and negative percentages indicating a decrease in thickness.

Time frame: 6 months post implantation

Population: Noting that some participants had multiple implants, data for 3 implants in the OsseoDensification (OD) protocol arm were not collected, which includes data for 2 participants. Noting that some participants had multiple implants, data for 3 implants in the standard drilling (SD) protocol arm were not collected, which includes data for 2 participants

ArmMeasureValue (MEAN)Dispersion
OsseoDensification (OD) ProtocolPercent Change of Buccal Bone Thickness as Assessed by the Cone Beam Computer Tomography at 5 mm From the Implant Platform8.61 percent changeStandard Deviation 25.3
Standard Drilling (SD) ProtocolPercent Change of Buccal Bone Thickness as Assessed by the Cone Beam Computer Tomography at 5 mm From the Implant Platform-15.1 percent changeStandard Deviation 21.6
Secondary

Primary Stability of the Implants as Indicated by the Implant Stability Quotient (ISQ) as Measured by the Resonance Frequency Analysis Using the Ostell ISQ Meter 12 Weeks After Implant Placement

The Implant Stability Quotient (ISQ) value is a measurement used to assess the stability of dental implants. It is obtained by resonance frequency analysis and ranges from 1 to 100, with higher values indicating greater stability.

Time frame: 12 weeks after implant placement

Population: Data for 1 implant in the OsseoDensification (OD) protocol arm were not collected, including 1 participant for whom no implant data were collected. Data for 1 implant in the standard drilling (SD) protocol arm were not collected, including 1 participant for whom no implant data were collected.

ArmMeasureValue (MEAN)Dispersion
OsseoDensification (OD) ProtocolPrimary Stability of the Implants as Indicated by the Implant Stability Quotient (ISQ) as Measured by the Resonance Frequency Analysis Using the Ostell ISQ Meter 12 Weeks After Implant Placement76.8 score on a scaleStandard Deviation 5.18
Standard Drilling (SD) ProtocolPrimary Stability of the Implants as Indicated by the Implant Stability Quotient (ISQ) as Measured by the Resonance Frequency Analysis Using the Ostell ISQ Meter 12 Weeks After Implant Placement73.3 score on a scaleStandard Deviation 7.31
Secondary

Primary Stability of the Implants as Indicated by the Implant Stability Quotient (ISQ) as Measured by the Resonance Frequency Analysis Using the Ostell ISQ Meter 3 Weeks After Implant Placement

The Implant Stability Quotient (ISQ) value is a measurement used to assess the stability of dental implants. It is obtained by resonance frequency analysis and ranges from 1 to 100, with higher values indicating greater stability.

Time frame: 3 weeks after implant placement

Population: Noting that some participants had multiple implants, data for 6 implants in the OsseoDensification (OD) protocol arm were not collected, which includes data for 4 participants. Noting that some participants had multiple implants, data for 6 implants in the standard drilling (SD) protocol arm were not collected, which includes data for 4 participants

ArmMeasureValue (MEAN)Dispersion
OsseoDensification (OD) ProtocolPrimary Stability of the Implants as Indicated by the Implant Stability Quotient (ISQ) as Measured by the Resonance Frequency Analysis Using the Ostell ISQ Meter 3 Weeks After Implant Placement75 score on a scaleStandard Deviation 6.87
Standard Drilling (SD) ProtocolPrimary Stability of the Implants as Indicated by the Implant Stability Quotient (ISQ) as Measured by the Resonance Frequency Analysis Using the Ostell ISQ Meter 3 Weeks After Implant Placement69.8 score on a scaleStandard Deviation 7.99
Secondary

Primary Stability of the Implants as Indicated by the Implant Stability Quotient (ISQ) as Measured by the Resonance Frequency Analysis Using the Ostell ISQ Meter 6 Weeks After Implant Placement

The Implant Stability Quotient (ISQ) value is a measurement used to assess the stability of dental implants. It is obtained by resonance frequency analysis and ranges from 1 to 100, with higher values indicating greater stability.

Time frame: 6 weeks after implant placement

Population: Noting that some participants had multiple implants, data for 6 implants in the OsseoDensification (OD) protocol arm were not collected, which includes data for 4 participants. Noting that some participants had multiple implants, data for 6 implants in the standard drilling (SD) protocol arm were not collected, which includes data for 4 participants

ArmMeasureValue (MEAN)Dispersion
OsseoDensification (OD) ProtocolPrimary Stability of the Implants as Indicated by the Implant Stability Quotient (ISQ) as Measured by the Resonance Frequency Analysis Using the Ostell ISQ Meter 6 Weeks After Implant Placement75.3 score on a scaleStandard Deviation 6.16
Standard Drilling (SD) ProtocolPrimary Stability of the Implants as Indicated by the Implant Stability Quotient (ISQ) as Measured by the Resonance Frequency Analysis Using the Ostell ISQ Meter 6 Weeks After Implant Placement70.8 score on a scaleStandard Deviation 9.25
Secondary

Primary Stability of the Implants as Indicated by the Implant Stability Quotient (ISQ) as Measured by the Resonance Frequency Analysis Using the Ostell ISQ Meter Baseline(After Implant Placement)

The Implant Stability Quotient (ISQ) value is a measurement used to assess the stability of dental implants. It is obtained by resonance frequency analysis and ranges from 1 to 100, with higher values indicating greater stability.

Time frame: baseline(after implant placement)

ArmMeasureValue (MEAN)Dispersion
OsseoDensification (OD) ProtocolPrimary Stability of the Implants as Indicated by the Implant Stability Quotient (ISQ) as Measured by the Resonance Frequency Analysis Using the Ostell ISQ Meter Baseline(After Implant Placement)73.7 score on a scaleStandard Deviation 7.23
Standard Drilling (SD) ProtocolPrimary Stability of the Implants as Indicated by the Implant Stability Quotient (ISQ) as Measured by the Resonance Frequency Analysis Using the Ostell ISQ Meter Baseline(After Implant Placement)67.0 score on a scaleStandard Deviation 8.26
Secondary

Primary Stability of the Implants as Measured by the Insertion Torque Values

Insertion torque refers to the force used to insert an implant into bone, and it is reported in units of Newton-Centimeters (N-cms).

Time frame: day of implant surgery

ArmMeasureValue (MEAN)Dispersion
OsseoDensification (OD) ProtocolPrimary Stability of the Implants as Measured by the Insertion Torque Values57.7 Newton-centimetersStandard Deviation 23.4
Standard Drilling (SD) ProtocolPrimary Stability of the Implants as Measured by the Insertion Torque Values45 Newton-centimetersStandard Deviation 25.5
Secondary

Volume of the Ridge as Assessed by the Cone Beam Computer Tomography (CBCT)

Time frame: 6 months post implantation

Population: Data were not collected for this outcome measure.