Efficacy of the Nanodropper Device on Pupillary Dilation

Efficacy of the Nanodropper Device on Pupillary Dilation

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT05274321
Enrollment
50
Registered
2022-03-10
Start date
2021-07-29
Completion date
2021-11-19
Last updated
2023-09-25

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Dilation

Brief summary

Prescription eye drop bottles elute drops that exceed the capacity of the human eye by five times. This study describes performing in clinic dilation using a novel solution for combating medical waste with Nanodropper, an eye drop bottle adapter that creates smaller eye drops.

Interventions

Device used for administering eye drops

DEVICEStandard Eye Dropper

Standard Eye Dropper used for administering eye drops

Sponsors

Nanodropper, Inc.
CollaboratorINDUSTRY
University of California, San Francisco
Lead SponsorOTHER

Study design

Allocation
RANDOMIZED
Intervention model
SINGLE_GROUP
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
2 Years to 17 Years
Healthy volunteers
No

Inclusion criteria

* Participants under the age of 18 years old

Exclusion criteria

* Participants 18 years and older * Pupillary or anterior segment abnormality * Participants with pre-existing health conditions that would prevent pupillary dilation.

Design outcomes

Primary

MeasureTime frameDescription
Spherical EquivalentBefore dilation and 30 minutes after dilationCycloplegia is the paralysis of the ciliary muscle of the eye resulting in dilatation of the pupil and paralysis of accommodation. Cycloplegia is defined as the change score. Change scores were calculated by subtracting the spherical equivalent value after dilation from the value before dilation for each eye.
Maximum Pupil DiameterBefore dilation and 30 minutes after dilationThe pupil is the opening in the center of the iris (the structure that gives our eyes their color). The diameter is a straight line passing from side to side through the center of the pupil, which is shaped like a circle.
Constriction PercentageBefore and 30 minutes after dilationPupillary constriction percentage was calculated by the pupillometer in response to a 180-micro watt flash: (maximum - minimum) / maximum

Secondary

MeasureTime frameDescription
Change in Intraocular PressureBefore and 30 minutes After DilationIntraocular pressure is the pressure, or force, inside of your eyes. Specifically, it's a measurement of the fluid pressure in your aqueous humor.

Countries

United States

Participant flow

Recruitment details

Children 18 years of age or younger undergoing pupillary dilation as part of a routine eye examination at the University of California, San Francisco, at the Pediatric Ophthalmology Clinic between August 4, 2021, and November 19, 2021 were assessed for eligibility. Children who did not show up to appointments or did not meet inclusion criteria were not enrolled in the study.

Participants by arm

ArmCount
Pediatric Participants
Consenting participants will undergo standard pupillary dilation in one eye and pupillary dilation in the second eye using the Nanodropper attachment. Nanodropper: Device used for administering eye drops
50
Total50

Baseline characteristics

CharacteristicPediatric Participants
Age, Categorical
<=18 years
50 Participants
Age, Categorical
>=65 years
0 Participants
Age, Categorical
Between 18 and 65 years
0 Participants
Age, Continuous9 years
STANDARD_DEVIATION 3
Race and Ethnicity Not Collected— Participants
Region of Enrollment
United States
50 Participants
Sex: Female, Male
Female
21 Participants
Sex: Female, Male
Male
29 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
deaths
Total, all-cause mortality
0 / 500 / 50
other
Total, other adverse events
0 / 500 / 50
serious
Total, serious adverse events
0 / 500 / 50

Outcome results

Primary

Constriction Percentage

Pupillary constriction percentage was calculated by the pupillometer in response to a 180-micro watt flash: (maximum - minimum) / maximum

Time frame: Before and 30 minutes after dilation

Population: Participants received eye drops from the Nanodropper in one eye only and Standard of Care in the other eye.

ArmMeasureGroupValue (MEAN)
Eyes Recieving Eye Drops From Standard of CareConstriction PercentageBefore Dilation30.44 percent change
Eyes Recieving Eye Drops From Standard of CareConstriction PercentageAfter Dilation5.67 percent change
Eyes Recieving Eye Drops From Standard of CareConstriction PercentageChange Score-24.76 percent change
Eyes Recieving Eye Drops Using Nanodropper DeviceConstriction PercentageBefore Dilation31.07 percent change
Eyes Recieving Eye Drops Using Nanodropper DeviceConstriction PercentageAfter Dilation6.26 percent change
Eyes Recieving Eye Drops Using Nanodropper DeviceConstriction PercentageChange Score-24.81 percent change
Comparison: Absolute between-eye differences were calculated for each participant by subtracting the standard of care (SOC) change score from the Nanodropper change score. The confidence interval (CI) is 95%.p-value: 0.03Regression, Linear
Primary

Maximum Pupil Diameter

The pupil is the opening in the center of the iris (the structure that gives our eyes their color). The diameter is a straight line passing from side to side through the center of the pupil, which is shaped like a circle.

Time frame: Before dilation and 30 minutes after dilation

Population: Participants received eye drops from the Nanodropper in one eye only and Standard of Care in the other eye.

ArmMeasureGroupValue (MEAN)
Eyes Recieving Eye Drops From Standard of CareMaximum Pupil DiameterBefore Dilation5.95 mm
Eyes Recieving Eye Drops From Standard of CareMaximum Pupil DiameterAfter Dilation7.91 mm
Eyes Recieving Eye Drops From Standard of CareMaximum Pupil DiameterChange Score1.95 mm
Eyes Recieving Eye Drops Using Nanodropper DeviceMaximum Pupil DiameterBefore Dilation6.01 mm
Eyes Recieving Eye Drops Using Nanodropper DeviceMaximum Pupil DiameterAfter Dilation7.81 mm
Eyes Recieving Eye Drops Using Nanodropper DeviceMaximum Pupil DiameterChange Score1.77 mm
Comparison: Absolute between-eye differences were calculated for each participant by subtracting the standard of care (SOC) change score from the Nanodropper change score. The confidence interval (CI) is 95%.p-value: 0.02Regression, Linear
Primary

Spherical Equivalent

Cycloplegia is the paralysis of the ciliary muscle of the eye resulting in dilatation of the pupil and paralysis of accommodation. Cycloplegia is defined as the change score. Change scores were calculated by subtracting the spherical equivalent value after dilation from the value before dilation for each eye.

Time frame: Before dilation and 30 minutes after dilation

Population: Participants received eye drops from the Nanodropper in one eye only and Standard of Care in the other eye.

ArmMeasureGroupValue (MEAN)
Eyes Recieving Eye Drops From Standard of CareSpherical EquivalentBefore Dilation-2.62 Diopters
Eyes Recieving Eye Drops From Standard of CareSpherical EquivalentAfter Dilation-0.86 Diopters
Eyes Recieving Eye Drops From Standard of CareSpherical EquivalentChange Score1.76 Diopters
Eyes Recieving Eye Drops Using Nanodropper DeviceSpherical EquivalentBefore Dilation-2.28 Diopters
Eyes Recieving Eye Drops Using Nanodropper DeviceSpherical EquivalentAfter Dilation-0.55 Diopters
Eyes Recieving Eye Drops Using Nanodropper DeviceSpherical EquivalentChange Score1.73 Diopters
Comparison: Absolute between-eye differences were calculated for each participant by subtracting the standard of care (SOC) change score from the Nanodropper change score. The confidence interval (CI) is 95%.p-value: 0.22Regression, Linear
Secondary

Change in Intraocular Pressure

Intraocular pressure is the pressure, or force, inside of your eyes. Specifically, it's a measurement of the fluid pressure in your aqueous humor.

Time frame: Before and 30 minutes After Dilation

ArmMeasureGroupValue (MEAN)
Eyes Recieving Eye Drops From Standard of CareChange in Intraocular PressureBefore Dilation17.36 mmHg
Eyes Recieving Eye Drops From Standard of CareChange in Intraocular PressureAfter Dilation18.42 mmHg
Eyes Recieving Eye Drops From Standard of CareChange in Intraocular PressureChange Score1.06 mmHg
Eyes Recieving Eye Drops Using Nanodropper DeviceChange in Intraocular PressureChange Score-0.17 mmHg
Eyes Recieving Eye Drops Using Nanodropper DeviceChange in Intraocular PressureBefore Dilation18.13 mmHg
Eyes Recieving Eye Drops Using Nanodropper DeviceChange in Intraocular PressureAfter Dilation17.96 mmHg
Comparison: Absolute between-eye differences were calculated for each participant by subtracting the standard of care (SOC) change score from the Nanodropper change score. The confidence interval (CI) is 95%.p-value: 0.25Regression, Linear