Skip to content

Principles and Test Methods of Non-contact Body Thermometry

Investigation of the Effect of Environmental Temperatures on Non-contact Thermal Imaging for Core Body Temperature Measurement

Status
UNKNOWN
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT05247736
Enrollment
110
Registered
2022-02-21
Start date
2021-07-22
Completion date
2022-05-15
Last updated
2022-02-21

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Thermometry

Keywords

non-contact infrared thermometry, thermal imaging, elevated body temperature, febrile

Brief summary

Long-wavelength infrared (IR) detectors have a more than 20-year history in medical thermometry and have been used widely for febrile screening. However, over the past year and a half, public health entities, medical professionals, and the general public have begun to question the claimed accuracy of non-contact body thermometry. The standard assessment of a device's performance relies on clinical testing with febrile individuals, yet this practice may have inadvertently allowed the approval of IR systems that are unable to detect moderate fevers. The ability to test device performance without relying on febrile test participants would have important ramifications for public health, especially if this test discovered undisclosed differences in accuracy in widely used devices. The aim is to examine the effect of the local environment and the physiology of the human body on the relationship between core body temperature and inner canthi (region near tear duct) skin temperature measured using non-contact thermal imaging and to use of this relationship to test actual device performance at detecting simulated elevated temperatures, without requiring volunteers having actual elevated temperatures. The overall goal of this research study is to validate and improve the science of non-contact core body temperature measurement.

Detailed description

It is difficult to assess the performance of non-contact thermometry at detection of febrile subjects unless febrile subjects are included in the test, which severely limits development and independent testing of such devices. The underlying mechanisms of non-contact thermometry might be used to develop a test without requiring the inclusion of febrile subjects, and the development and demonstration of such a test is the object of this study. Literature review and pilot study revealed the offset between skin temperature and core (oral) body temperature is dependent on air temperature by a linear relationship. In a 70F room, the inner canthi temperature of a human with core (oral-reference) body temperature of 98.6F is 94.4F, a difference of 4.2F. This same individual in an 87F room (after waiting several minutes) will have an inner canthi temperature of 97.0F, a difference of only 1.6F (the actual values may depend on sensor confounds not addressed here). Therefore, one could use a several-minutes exposure to elevated air temperatures (equilibration environment) to simulate an elevated body temperature, by continuing to operate the device under test from a non-elevated air temperature environment (test environment), while taking care to limit the effects of temperature changes and gradients due to the mixing of these two environments. Subjects will be recruited from a college campus via a recruitment protocol to participate in an up to 2 hour study session. The study session will involve collection of demographic information on a non-identifiable form, five collections of oral thermometry, four 10 minute periods of equilibration inside 4 elevated temperature environments during repeat continuous non-contact thermometry, and measurements via several non-contact test devices operated from within a non-elevated environment through a window into each elevated environment.

Interventions

DIAGNOSTIC_TESTInfrared Thermography

Non-contact body thermometry

Sponsors

Thermal Diagnostics LLC
Lead SponsorINDUSTRY

Study design

Observational model
OTHER
Time perspective
PROSPECTIVE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

* Age \> 18. * Have received a clinical oral thermometry measurement in the past.

Exclusion criteria

* Contraindication to oral temperature measurement. * Inability to follow instructions. * Any diagnosis associated with abnormal core thermoregulation, e.g. autonomic dysfunction. * Obvious signs of tissue damage present in either inner canthus. * Facial artery abnormality if previously diagnosed. * Experienced discomfort while receiving most recent clinical oral thermometry measurement.

Design outcomes

Primary

MeasureTime frameDescription
Clinical Bias (CB)10 monthsAverage difference between device measurement and reference measurement as per ASTM E1965-98
Limits of Acceptability (LA)10 monthsTwo times standard deviation of device measurement as per ASTM E1965-98
Bland-Altman (BA) Plot10 monthsPlot of device minus reference output versus reference output (or versus average of reference and device outputs)

Countries

United States

Outcome results

None listed