Performance of the Aktiia Optical Blood Pressure Monitoring Device Against Double Auscultation

Single-center Prospective Clinical Trial to Validate the Performance of the Aktiia Optical Blood Pressure Monitoring (OBPM) Device at the Wrist Against Double Auscultation

Status
Recruiting
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT05236348
Enrollment
225
Registered
2022-02-11
Start date
2022-01-11
Completion date
2025-03-31
Last updated
2024-08-19

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Blood Pressure, Heart Rate, Hypertension

Brief summary

This study with N = 225 participants (1 cohort of a minimum of 85 and maximum of 140 participants, and an additional cohort with 85 participants) over 9 visits has been designed to assess the accuracy of Aktiia.product-us systolic and diastolic blood pressure measurements against double-auscultation reference measurements. In order to induce blood pressure changes and to simulate ambulatory scenarios, the study participants will be asked to perform a set of activities, involving body position changes and physical exercises. Additionally, this study will compare the heart rate measured by Aktiia.product-us against the reference readings provided by a finger pulse oximeter (as per ISO80601-2-61). The reliability of the auscultation readings will be controlled with simultaneous volume-clamp measurements.

Detailed description

According to the World Health Organization, by 2025 hypertension will affect 1.5 billion adults worldwide. Half of the adult population suffering from hypertension is currently not diagnosed and half of the treated population is not at the defined target of BP. Widespread use of out-of-office Blood Pressure measurement is one of the proposed strategies to fight against hypertension worldwide. Aktiia S.A. has developed an intermittent automated non-invasive cuffless blood pressure monitor, determining the values of the systolic blood pressure, the diastolic blood pressure and the heart rate. This miniature device is comfortably positioned with a bracelet on a user's wrist. The goal of this study is to establish a reasonable assurance of safety and effectiveness with Aktiia OBPM system for monitoring blood pressure in home use. In particular, the study will collect data from subjects by means of the device under test and collect data from subjects by means of double-auscultation, pulse oximetry and volume-clamp BP reading while subjects wear the devices at different body positions and induce Blood Pressure changes, during eight visits that will be held over the span of seven days, and generate performance reports.

Interventions

DEVICEAktiia.product-us

The study will collect data from subjects by means of the device under test and collect data from subjects by means of double-auscultation, pulse oximetry and volume-clamp BP reading while subjects wear the devices at different body positions and induce BP changes, during eight visits that will be held over the span of seven days, and generate performance reports.

Sponsors

Aktiia SA
Lead SponsorINDUSTRY

Study design

Allocation
NON_RANDOMIZED
Intervention model
SINGLE_GROUP
Primary purpose
DIAGNOSTIC
Masking
NONE

Intervention model description

All study participants will wear the Aktiia.bracelet-us during 9 visits that will be held over the span of seven days. In order to induce blood pressure changes and to simulate ambulatory scenarios, the study participants will be asked to perform a set of activities, involving body position changes and physical exercises during these visits.

Eligibility

Sex/Gender
ALL
Age
21 Years to 85 Years
Healthy volunteers
Yes

Inclusion criteria

* Adult subjects (aged between 21 and 85 years old) * Subjects can read and speak French * Subjects can perform simple physical exercises * Subjects agreeing to attend the totality of 9 visits * Subjects that have signed the informed consent form

Exclusion criteria

* Clinical staff collaborating with study PI * Subjects with tachycardia (heart rate at rest \> 120bpm) * Subjects with atrial fibrillation * Subjects with diabetes * Subjects with renal dysfunctions * Subjects with hyper-/hypothyroidism * Subjects with pheochromocytoma * Subjects with Raynaud's disease * Subjects with trembling and shivering * Subjects with interarm systolic difference \> 15 mmHg * Subjects with interarm diastolic difference \> 10 mmHg * Subjects with arm paralysis * Women in known pregnancy (for ARM 1 only) * Subjects with an arteriovenous fistula * Subjects with arm amputations * Subjects with the upper arm circumference \< 22cm or \> 42 cm * Subjects with the wrist circumference \> 21 cm * Subjects with the exfoliative skin diseases * Subjects with lymphoedema

Design outcomes

Primary

MeasureTime frameDescription
Blood pressure mean value of differences1 weekThe mean value of the differences between Aktiia.product-us and double auscultation blood pressure measurements.
Blood pressure standard deviation of differences1 weekThe standard deviation of the differences between Aktiia.product-us and double auscultation blood pressure measurements.

Secondary

MeasureTime frameDescription
Heart rate root-mean-square error1 weekThe root-mean-square difference between the Aktiia.product-us heart rate determinations and the reference method.

Countries

Switzerland