HORIZON Prospective Clinical Investigation

HORIZON. Clinical Investigation of the Qmedics EXIST NiTi Stent Type FLEX & PULL in Adults With Peripheral Artery Disease (PAD)

Status
Terminated
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT05234164
Enrollment
69
Registered
2022-02-10
Start date
2021-12-15
Completion date
2023-11-17
Last updated
2024-02-14

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Peripheral Arterial Disease

Keywords

stenotic lesions, Peripheral Arterial Disease, occluded lesions, Superficial Femoral Artery, nitinol stent, self-expanding stent, Chronic Limb-Threatening Ischemia

Brief summary

The objective of this clinical investigation is to demonstrate and provide long term clinical data on safety and performance of the Exist 6F NiTi stent system type FLEX & PULL in a prospective investigation for the treatment of adult patients with de novo or re-stenotic symptomatic atherosclerotic lesions in Peripheral Artery Disease (PAD) requiring treatment of the Superficial Femoral Artery (SFA) or Proximal Popliteal Artery (P1 segment).

Detailed description

This is a multicenter, prospective study with clinical and radiographic follow-up for 24 months months post-procedure. Approximately two hundred thirty subjects will be enrolled (115 subjects will receive the FLEX type and 115 subjects the PULL type) in several centers, minimum 10 patients per investigational center. All patients will be evaluated at 30 days and 6-, 12- and 24-months post-index procedure.

Interventions

DEVICEEXIST 6F NiTi Stent System FLEX

Subjects requiring treatment of the P1 segment of the proximal popliteal artery will be treated with the Exist 6F NiTi Stent FLEX type. Subjects requiring treatment of the SFA will undergo an additional calcification review based on the Fanelli classification. Subjects graded with Fanelli 1 or 2 will be treated with the Exist 6F NiTi Stent FLEX type.

DEVICEEXIST 6F NiTi Stent System PULL

Subjects requiring treatment of the SFA will undergo a calcification review based on the Fanelli classification. Subjects graded with Fanelli grade 3 and 4 will be treated with the Exist 6F NiTi Stent PULL type.

Sponsors

Policlinico di Monza
CollaboratorUNKNOWN
University Clinical Hospital of Bialystok
CollaboratorUNKNOWN
University Hospital, Catania
CollaboratorOTHER
Erasmus Medical Center
CollaboratorOTHER
University Hospital, Geneva
CollaboratorOTHER
Qmedics AG
Lead SponsorINDUSTRY

Study design

Allocation
NON_RANDOMIZED
Intervention model
PARALLEL
Primary purpose
TREATMENT
Masking
NONE

Intervention model description

Subjects requiring treatment of the SFA will undergo a calcification review based on the Fanelli classification. Subjects graded with Fanelli 1 or 2, and subjects requiring treatment of the P1 segment of the proximal popliteal artery will be treated with the Exist 6F NiTi Stent type FLEX. Subject graded with Fanelli 3 and 4 will be treated with the Exist 6F NiTi Stent PULL.

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

\- Clinical: 1. Patient age 18 years or older 2. Subject is willing and able to provide consent before any study specific test or procedure is performed, signs the consent form, and agrees to attend all required follow-up visits 3. Symptoms of peripheral arterial disease classified as Rutherford Category 2, 3 or 4 or Fontaine Class IIb or III 4. The stenotic or occlusive lesion in Superficial Femoral Artery (SFA) and proximal popliteal artery, P1 segment, is considered suitable for stenting 5. No underlying medical condition is present which would prevent the subject from performing the required testing or from completing the study. 6. Stable medical condition * Anatomical criteria: 7. Included TASC II, A-B-C measured on pre- angio CT-scan (if CT-scan is standard of care) 8. Lesions must be one or multiple that can be treated with maximum two stents, maximum one overlapping and maximum length of the stent 25 cm 9. Patent ipsilateral iliac, popliteal (P2 and P3) and at least one tibial vessel

Exclusion criteria

\- Clinical criteria: 1. Subjects pregnant, breastfeeding or planning to become pregnant during the trial participation 2. Documented life expectancy less than 24 months due to other medical co-morbid condition(s) 3. Thrombophlebitis or deep vein thrombosis within the past 30 days. 4. Unable to assume DAPT (Dual Antiplatelet Therapy) 5. Concomitant renal failure with serum creatinine level \> 2.5 mg/dL (or \> 220 µmol/L) or GFR \< 30 ml/min/1,73 m2 6. Unresolved neutropenia (white blood cell count \< 3,000 / µL) or thrombocytopenia (platelet count \< 80,000 / µL) at the time of the index procedure 7. Unresolved bleeding disorder (INR ≥ 1.2) at the time of the index procedure 8. Active gastrointestinal bleeding 9. Anticoagulation therapy for other medical condition * Anatomical criteria: 10. Target lesion(s) received previous treatment within 30 days prior to enrolment (point of enrolment is defined as the time when the trial device enters the body) 11. Previously stented ipsilateral SFA 12. Prior peripheral vascular bypass surgery involving the target limb(s) 13. Target lesion is located within an aneurysm or associated with an aneurysm in the vessel segment either proximal or distal to the target lesion 14. Target lesion requires the use of cutting balloons, atherectomy or drug coated balloons (DCB) during the intervention

Design outcomes

Primary

MeasureTime frameDescription
Major Adverse Events1 monthFreedom from procedure- or stent-related Major Adverse Events (MAEs) at 30-days post index-procedure will be reported. MAE is defined as all causes of death and target limb major amputation, defined as amputation of the lower limb at the ankle level or above.
Primary patency12 monthsThe primary patency rate (defined as freedom from more than ≥ 50% restenosis) at 12 months post-procedure as measured by Peak Systolic Velocity ratio (PSVr) assessed with Duplex ultrasound (DUS). A PSVr ≥ 2,5 suggests a reduction in the luminal diameter \>50%.

Secondary

MeasureTime frameDescription
Primary patency12-monthsCompare the Exist 6F NiTi stent primary patency with \>50% of subjects with peripheral artery disease with literature data.
Rate of stent fracture12- and 24-monthsStent fracture rate is evaluated with x-ray. Stent fracture is defined according to classification on x-ray.
Freedom from Target Lesion Revascularization1 month, 6-, 12- and 24-monthsDefined as the absence of revascularization (by any means) of the target lesion (fTLR).
Patency rate24 monthsPatency rate as measured by Peak Systolic Velocity ratio (PSVr) assessed with Duplex ultrasound (DUS). A PSVr ≥ 2,5 suggests a reduction in the luminal diameter \>50%.
Major Adverse Events24 monthsProcedure- or stent-related Major Adverse Events (MAEs) post index-procedure will be reported. MAE is defined as all causes of death and target limb major amputation, defined as amputation of the lower limb at the ankle level or above.
Walking and mobility1 month, 6-, 12- and 24-monthsImprovement of walking and mobility is assessed by change in the 36-Item Short Form Survey (SF-36) score from baseline.
Improvement of Rutherford / Fontaine classification6- and 12-monthsClinical success Defined as improvement of Rutherford / Fontaine classification of one class or more as compared to the pre-procedure and an ankle-brachial index improvement (ABI) by ≥ 0.15.

Other

MeasureTime frameDescription
Self-reported Quality of life1 month, 6-, 12- and 24-monthsImprovement of quality of life is assessed by change in the 36-Item Short Form Survey (SF-36) score from baseline.

Countries

Italy, Poland

Outcome results

None listed