Skip to content

A RCT to Evaluate the Efficacy and Safety of Automated Needle Targeting (ANT-X) System Compared to Traditional Free Hand Puncture for Renal Access in Percutaneous Nephrolithotomy (PCNL)

A Randomized Controlled Trial to Evaluate the Efficacy and Safety of Automated Needle Targeting (ANT-X) System Compared to Traditional Free Hand Puncture for Renal Access in Percutaneous Nephrolithotomy (PCNL) Performed by Urologists in Training.

Status
UNKNOWN
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT05213702
Enrollment
60
Registered
2022-01-28
Start date
2022-01-25
Completion date
2022-12-31
Last updated
2022-01-28

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Kidney Stone

Brief summary

Objectives: This study aims to determine if the device provides an advantage over traditional free hand puncture by comparing the number of attempts required to obtain access, time to achieve access, radiation exposure to patient and surgeon; and complications.

Detailed description

This is a single centre, participant randomized, open label study to evaluate the efficacy and safety of automated needle targeting (ANT-X) system compared to traditional free hand puncture for achieving renal access in Percutaneous lithotomy (PCNL) performed by urologists in training. Prior to performing kidney puncture on patients, all participating surgeons will undergo a simulation using Medical Simulation 12 balls phantom device with and without use of the ANT-X. Four urologic trainees (with similar little previous PCNL experience) will be recruited for the simulation. Two end points will be assessed: Number of attempts to achieve desired renal puncture and total fluoroscopy time. This simulation via 12 balls phantom device will involve fluoroscopic screening of calyx and successful puncture by urologic trainees under consultant supervision. Trainees will have 30 attempts of free hand puncture and 30 attempts using the ANT-X. The trainee will only move on to patient's renal puncture once they are able to achieve more than 80% of satisfactory renal puncture on simulator. For human renal puncture, we will aim for 60 cases of PCNL, 30 to be done using ANT-X puncture and 30 using traditional free hand puncture performed by 4 urologic trainees (15 cases per urologic trainee).

Interventions

DEVICERenal Calyx punture with traditional free hand or ANT-X device

ANT-X device is a software-controlled automated electromechanical arm. It is developed to provide needle guidance for surgeon in MIS. This aims to reduce the surgery time taken and also minimize the radiation withstand by both surgeon and patient. ANT-X device is actually a Delta parallel robot implementing inverse kinematics to achieve effector's movement. Parallel structures excel in their great stiffness and high positioning accuracy. Most of the main components of ANT-X robot are made up of medical grade Polyether ether ketone (PEEK) material. PEEK is preferred to build the robot because it is radiolucent and lightweight while being able to provide relatively high material strength. The device is intended for use together with C-arm fluoroscope (fluoroscopy/CT technology). The fluoroscopy technology is used to provide image guidance for the needle

Sponsors

NDR Medical Technology Pte Ltd
CollaboratorINDUSTRY
University of Malaya
Lead SponsorOTHER

Study design

Allocation
RANDOMIZED
Intervention model
PARALLEL
Primary purpose
TREATMENT
Masking
NONE

Intervention model description

Male/non-pregnant female subjects between age 18-75 years, both inclusive, who will undergo Percutaneous Nephrolithotomy (PCNL).

Eligibility

Sex/Gender
ALL
Age
18 Years to 75 Years
Healthy volunteers
Yes

Inclusion criteria

* Subjects willing to sign informed consent form before initiation of any study specific procedures. * Subjects diagnosed with kidney stones and planned for prone PCNL. * Male/ non- pregnant female subjects between 18-75 years (both inclusive) of age at the time of informed consent. * Subjects with renal stones of ≥1 cm or not managed by other stone treatments (e.g. shock wave lithotripsy, ureteroscopy). * Serum creatinine of \< 1.2 mg/dl.

Exclusion criteria

* Subjects with bleeding disorders. * Recent infection of the kidney or urinary tract which leads to pus accumulation in the kidney, i.e., Pyonephrosis. * Subjects not able to have prone positions for surgical procedure due to comorbid conditions. * Renal stones \< 1 cm and can managed by another technique. * Patients with moderate to severe renal failure. * Patients with congenital urinary tract anomalies. * Patients on anticoagulation or antiplatelet therapy. * Evidence or history of any other disease that in the opinion of the investigator would put the patient at risk through study participation, or would affect the study analyses if the disease exacerbated during the study. * Legal incapacity or other circumstances that render the subject unable to understand the nature, scope and possible consequences of the study. * Clinically significant abnormal findings or condition (other than kidney stone), which might, in the opinion of the Investigator, interfere with study evaluations or pose a risk to subject safety during the study.

Design outcomes

Primary

MeasureTime frameDescription
Number of attempts required to obtain accessDay 0 surgeryAn access attempt is defined as each separate passage of the needle into the kidney. Number of attempts by Investigator/surgeon will be recorded.
Time for the accessDay 0 surgeryTime to successful access into the collecting system will be calculated in minutes starting from the placement of the needle on the skin surface to the identification of urine efflux from the end of the needle or reach of target. Start time of needle placement and time of urine efflux from the end of the needle or reach of target will be recorded in subject CRF in hh:mm:ss format.

Secondary

MeasureTime frameDescription
Radiation exposure to patient and surgeonDay 0 surgery
Assessment of AEsAdmission day till discharge within 7 daysAdvere events associated with the procedure

Contacts

Primary ContactWS Yeoh, MD
yeoh_uro@yahoo.com+60379494499

Outcome results

None listed