Skip to content

Safety and Efficacy of SinocrownTM Transcatheter Aortic Valve Replacement System

A Prospective,Multicenter, Single-arm Clinical Trial to Evaluate Safety and Efficacy of SinocrownTM Transcatheter Aortic Valve Replacement System With Accessories in Patients With Symptomatic, Calcified and Severe Aortic Stenosis

Status
Enrolling by invitation
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT05202977
Enrollment
10
Registered
2022-01-24
Start date
2021-04-06
Completion date
2026-08-31
Last updated
2022-01-24

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Transcatheter Aortic Valve Replacement

Keywords

aortic stenosis

Brief summary

This study was designed to preliminary verify the safety and of efficacy of SinocrownTM Transcatheter Aortic Valve Replacement System With accessories in patients with symptomatic, calcified and severe aortic stenosis using objective performance criteria

Detailed description

This prospective, single-center, single-arm clinical trial was designed to preliminary confirm the safety and efficacy of SinocrownTM transcatheter aortic valve system with accessories. According to the inclusion and exclusion criteria, 10 patients were enrolled and implanted with SinocrownTM .Patients will been followed up 30 days, 6 months, 12 months, 2 years, 3 years, 4 years and 5 years after valve implantation. The main follow-up included clinical symptoms and signs, cardiac ultrasound, etc. The primary endpoint was device success after TAVR, Secondary endpoints included procedural success, all-cause mortality, myocardial infarction, disabling stroke, severe bleeding, acute kidney injury, permanent pacemaker implantation, vascular complications, other complication, MACCEs, valvular function , cardiac function improvement, quality of life of patients.

Interventions

DEVICESinocrownTM Transcatheter Aortic Valve Replacement System

Subjects receiving SinocrownTM Transcatheter Aortic Valve Replacement System of Lepu Medical

Sponsors

Lepu Medical Technology (Beijing) Co., Ltd.
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
65 Years to No maximum
Healthy volunteers
No

Inclusion criteria

1. Age ≥ 65 years ; 2. Autologous degenerative aortic stenosis(evaluated by echocardiography: transaortic valve pressure gradient ≥40mmHg , or transaortic valve blood flow rate ≥4m/s, or aortic Valve area\<1.0cm2, or AVA \<0.5cm2 /m2) 3. Patients with aortic annulus size≥17.5mm and ≤29mm (echocardiography measurement); 4. The diameter of the ascending aorta of the patient was \< 45mm. 5. Appear obvious symptoms caused by aortic stenosis, NYHA class Ⅱ to IV; 6. Subject or subject's legal representative have been informed of the nature of the study, agrees to participate and has signed the approved consent form. 7. Subjects are willing to accept all clinical follow-up. 8. The cardiologist and cardiac surgeon evaluated the subject as unsuitable for surgical treatment (the risk of death or irreversible complications within 30 days after operation is expected to be above 50% , or contraindicated comorbidities such as prior chest radiotherapy, liver failure, diffuse severe calcification of the aorta, extreme weakness, etc.). Or Moderate or higher surgical risk (STS score ≥ 4%).

Exclusion criteria

1. defined as Q-wave myocardial infarction or non-Q-wave myocardial infarction with CK-MB level ≥ 2 times the upper limit of normal value and / or Tn increase. 2. Subjects with congenital unicuspid, without calcification. 3. Any therapeutic or traumatic cardiac surgery within 30 days (within 3 months in case of drug-eluting stents implantation) 4. The patient's heart has been implanted with other artificial heart valves, artificial rings, or severe mitral annulus calcification (MAC), severe (\> 3 +) mitral insufficiency. 5. Hematological abnormalities, including leukopenia (WBC\<3×109/L), acute anemia (HB \<90g/L), thrombocytopenia (PLT\<50×109/L), bleeding constitution and coagulopathy ; 6. Subjects with hemodynamic instability, requiring continuous mechanical heart assistance. 7. Subjects who need emergency surgery for any reason; 8. Hypertrophic cardiomyopathy with or without obstruction; 9. Severe left ventricular dysfunction, left ventricular ejection fraction (LVEF) \<20%; 10. The echocardiogram indicates the presence of a thrombus or vegetation; 11. A history of active peptic ulcer or upper gastrointestinal bleeding within 3 months; 12. Are allergic to nickel titanium alloys or contrast agents, and cannot tolerate anticoagulation and antiplatelet therapy; 13. Cerebrovascular accident occurred within 6 months, including transient ischemic attack; 14. renal insufficiency(creatinine\> 3.0mg/dL) and/or end-stage renal disease requiring Renal dialysis; 15. Life expectancy\< 12 months; 16. Significant aortic diseases, including abdominal aortic or thoracic aortic aneurysm(defined as maximum lumen diameter \> 50mm, obvious distortion), atherosclerosis of the aortic arch(especially bulge with thickness \> 5mm or ulceration), narrowing of the abdominal or thoracic aorta(especially calcification and irregular surface), and severe curvature of the thoracic aorta; 17. The iliac-femoral artery cannot be inserted with a guide sheath of 22F or above, such as severely obstructed with calcification, severely tortuous or vessel diameter less than 6mm; 18. Screening participant who is participating in other drug or medical device clinical trials; 19. Active infective endocarditis or other active infections; 20. Anatomically, the approach or aortic root is not suitable for TAVR (such as a high risk of coronary artery occlusion). 21. Other conditions considered by the investigator to be inappropriate for participation in this clinical trial.

Design outcomes

Primary

MeasureTime frameDescription
device successimmediate post-surgicalImmediate postoperative device success was defined as : successfully loading the valve into the delivery system, sending it to the target site through the delivery system, accurately releasing the artificial aortic valve and completely withdrawing the delivery system (valve orifice area \> 1.2cm2, mean aortic gradient\< 20mmhg or maximal aortic valve velocity \< 3 m/ s, and there is no moderate or severe valve stenosis)

Secondary

MeasureTime frameDescription
All-cause mortalityimmediate post-surgical, 7 days after procedure/prior to discharge, 30 days, 6 months, 12 months, 2 years, 3 years, 4 years and 5 yearsincluding cardiac, non-cardiac and unexplained deaths
Myocardial infarctionimmediate post-surgical, 7 days after procedure/prior to discharge, 30 days, 6 months, 12 months, 2 years, 3 years, 4 years and 5 yearsRate of patients with myocardial infarction
Rate of disabling stroke7 days after procedure/prior to discharge, 30 days, 6 months, 12 months, 2 years, 3 years, 4 years and 5 yearsDisabling stroke was defined as mRS ≥ 2, and the mRS score increased by more than 1 point compared with the last visit
Rate of Severe bleeding(life-threatening or disabling)7 days after procedure/prior to discharge, 30 days, 6 months, 12 months, 2 years, 3 years, 4 years and 5 yearsSevere degree was defined as type 3a or above(according to BARC)
acute kidney injury7 days after procedure/prior to discharge, 30 days after procedureRate of patients with acute kidney injury
permanent pacemaker implantation7 days after procedure/prior to discharge, 30 days, 6 months, 12 months, 2 years, 3 years, 4 years and 5 yearsRate of patients implanted with permanent pacemaker
Procedural successimmediate post-surgicalDevice success without death and aortic valve reintervention during operation
other operative complicationimmediate post-surgical, 7 days after procedure/prior to dischargeincluding conversion to surgery, unexpected cardiopulmonary mechanical assistance, coronary artery occlusion, ventricular septal perforation, mitral valve injury or loss of function, pericardial tamponade, endocarditis, valve thrombosis, valve ectopic (displacement, embolization, misrelease), and so on.
The incidence of major adverse cardiovascular and cerebrovascular events during the trial(MACCEs)immediate post-surgical, 7 days after procedure/prior to discharge, 30 days, 6 months, 12 months, 2 years, 3 years, 4 years and 5 yearsIncidence of MACCEs (including mortality, stroke, myocardial infarction, reoperation, arrhythmias and conduction blocks) during the trial.
Valve and heart functionimmediate post-surgical, 7 days after procedure/prior to discharge, 30 days, 6 months, 12 monthsDoppler echocardiography was used to evaluate the valve performance at each follow-up point, including valve orifice area, mean aortic gradient, degree of valve regurgitation, left ventricular ejection fraction LVEF, end diastolic left ventricular inner diameter LVEDd, moderate or above perivalvular leakage PVL, and degree of aortic stenosis.
Cardiac function improvement7 days after procedure/prior to discharge, 30 days, 6 months, 12 monthsCardiac function improvement is accessed by NYHA cardiac function classification increases from baseline to discharge, 30 days, 6 months, 12 months
Quality of life of patients30 days, 6 months, 12 monthsQuality of Life of patients accessed by EuroQol-5 Dimensions (EQ-5D) Questionnaire. The higher scores mean a better health state. Recording changes from baseline to 30 days, 6 months, 12 months
vascular complicationsimmediate post-surgical, 7 days after procedure/prior to dischargeVascular complications mainly include dissection, occlusion, rupture and bleeding of access vessels.

Countries

China

Outcome results

None listed

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026