BD MiniDraw™ Capillary System Clinical Equivalence Study

A Prospective Multi-center Study to Evaluate Clinical Equivalence (Method Comparison) Between the BD MiniDraw™ SST™ Capillary Blood Collection Tube and BD MiniDraw™ EDTA Capillary Blood Collection Tube to Respective Comparators

Status
Completed
Phases
Unknown
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT05186311
Enrollment
262
Registered
2022-01-11
Start date
2022-04-22
Completion date
2022-10-24
Last updated
2023-11-03

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Acute Disease, Disease, Chronic, Healthy, Sample Collection

Keywords

sample collection device

Brief summary

This study designed to evaluate the safety and effectiveness of a newly developed sample collection device. Participants will be enrolled and samples will be collected by representative future users of the device. Comparator venous and conventional capillary tubes will be collected by skilled healthcare professionals with phlebotomy experience. Samples will be tested and average differences between the investigational device compared to a marketed comparator will be evaluated. Visual observations will be captured and device safety will be assessed.

Detailed description

The current method for collection of a large volume of capillary blood (up to 600 µL) involves optional warming of the hand or finger prior to procedure, lancing of the fingertip, squeezing the fingertip by hand, and dripping the blood into an open non-sterile tube. The capillary blood collection process is considered cumbersome and lacks standardization in how much pressure is applied in squeezing of the finger which may lead to poor sample quality or insufficient blood volume. Common sample quality issues for capillary blood collection include hemolysis and dilution of sample with interstitial fluid, potentially causing bias in test results. Open collection also carries a risk of blood exposure for healthcare workers and potential risk of infection at the finger puncture site. A new device, BD Microtainer® Easy Collect Capillary Blood Collection System for the collection of capillary blood was created to standardize the current collection process and improve sample quality. This study designed to evaluate the safety and effectiveness of the BD Microtainer® Easy Collect Capillary Tube SST™ and BD Microtainer® Easy Collect Capillary Tube EDTA devices. This study will be conducted in two parts, one for the SST product and the other for the EDTA product. A minimum of 100 participants, and a maximum of 150 participants per part will be enrolled at a minimum of three various ancillary healthcare site settings representative of the intended use environment such as retail pharmacy site, patient service center, and/or urgent care clinic. Samples will be collected by representative future users of the BD Microtainer® Easy Collect Capillary Blood Collection System who may not have prior blood collection experience. Comparator venous and conventional capillary tubes will be collected by skilled healthcare professionals with phlebotomy experience. Samples will be collected, clotted (serum samples only) and centrifuged (serum samples only) before transportation to the laboratory where they will be tested. Average differences between the investigational device compared to a marketed comparator will be evaluated. Visual observations for the SST products will be captured which include hemolysis as measured by Plasma Free Hemoglobin (PFH); collection volume; and duration of sample collection. Visual Observations for the EDTA products will be captured which include frequency of occurrence of instrument flags; frequency of occurrence of platelet clump instrument flags; frequency of instrument platelet clump flags with no platelet clump; evaluate whole blood collection volume; and duration of sample collection. Lastly, device safety will be assessed by device and procedure related adverse events.

Interventions

DEVICECapillary Blood Collection System

Test Blood Sample Collection

Sponsors

Babson Diagnostics
CollaboratorINDUSTRY
Becton, Dickinson and Company
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
OTHER
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

* Participants ≥18 years of age. * Not currently pregnant (self-reported) * Adequate access to four fingers (ring and middle finger on each hand) and inner elbows for blood collection procedures * Willing and able to comply with all study procedures and evaluations * Ability to read, write, and understand English language * Provision of signed and dated informed consent form

Exclusion criteria

* Evidence of skin issues such as infections, ulcerations, blisters, peripheral vascular disease, inflammation, extensive scarring or calluses, or healed burns at the fingerstick site. Subject to post-study exclusion period for this or other related studies: 1. A study participant may enroll once in study part A and once in study part B of this clinical study. They may not enroll more than once in a single study part. 2. If a study participant has previously enrolled in one study part and would like to enroll in the second study part, they must wait 8 days from the time of the first study part collection procedures to enroll in the second study part collection procedures. * Any condition which, in the opinion of the Investigator, would preclude participation in this study.

Design outcomes

Primary

MeasureTime frameDescription
Chemistry Analytes [Albumin (ALB) and Total Protein (TP)] in Serum Measured Using the Silica Clot Activator for Serum Sampleswithin 4 hours of sample collectionTo demonstrate clinical equivalence of the SST test device as compared to Venous Comparator Serum and Capillary Comparator for chemistry analytes. This comparison was focused on the analytes with a unit of g/dL
White Blood Cell Differential Counts, Hemoglobin A1c, Hematocrit, and Red Blood Cell Distribution Width in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)within 2 hours of sample collectionTo demonstrate clinical equivalence of the EDTA test device as compared to their relevant comparators venous comparator and capillary comparator. This comparison was focused on the analytes with a unit of percentage
Thyroid Stimulating Hormone (TSH) in Serum Measured Using the Silica Clot Activator for Serum Sampleswithin 4 hours of sample collectionTo demonstrate clinical equivalence of the SST test device as compared to Venous Comparator Serum and Capillary Comparator for chemistry analytes. This comparison was focused on the analytes with a unit of µIU/mL
Lipid Analytes and Metabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum Sampleswithin 4 hours of sample collectionTo demonstrate clinical equivalence of the SST test device as compared to Venous Comparator Serum and Capillary Comparator for chemistry analytes. This comparison was focused on the analytes with a unit of mg/dL
Metabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum Sampleswithin 4 hours of sample collectionTo demonstrate clinical equivalence of the SST test device as compared to Venous Comparator Serum and Capillary Comparator for chemistry analytes. This comparison was focused on the analytes with a unit of mmol/L
Vitamin D (Vit D), in Serum Measured Using the Silica Clot Activator for Serum Sampleswithin 4 hours of sample collectionTo demonstrate clinical equivalence of the SST test device as compared to Venous Comparator Serum and Capillary Comparator for chemistry analytes. This comparison was focused on the analytes with a unit of ng/mL
Alkaline Phosphatase (ALP), Alanine Transaminase (ALT), and Aspartate Aminotransferase (AST), in Serum Measured Using the Silica Clot Activator for Serum Sampleswithin 4 hours of sample collectionTo demonstrate clinical equivalence of the SST test device as compared to Venous Comparator Serum and Capillary Comparator for chemistry analytes. This comparison was focused on the analytes with a unit of U/L
White Blood Cells (WBC) and Platelets (PLT) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)within 2 hours of sample collectionTo demonstrate clinical equivalence of the EDTA test device as compared to their relevant comparators Venous Comparator and Capillary Comparator. This comparison was focused on the analytes with a unit of 10\^3/µL
Red Blood Cells (RBC) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)within 2 hours of sample collectionTo demonstrate clinical equivalence of the EDTA test device as compared to their relevant comparators Venous Comparator and Capillary Comparator. This comparison was focused on the analytes with a unit of 10\^6/µL
Mean Corpuscular Volume (MCV) and Red Blood Cell Distribution Width (RDW) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)within 2 hours of sample collectionTo demonstrate clinical equivalence of the EDTA test device as compared to their relevant comparators Venous Comparator and Capillary Comparator. This comparison was focused on the analytes with a unit of fL
Hemoglobin (HgB) and Mean Corpuscular Hemoglobin Concentration (MCHC) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)within 2 hours of sample collectionTo demonstrate clinical equivalence of the EDTA test device as compared to their relevant comparators Venous Comparator and Capillary Comparator. This comparison was focused on the analytes with a unit of g/dL
Mean Corpuscular Hemoglobin (MCH) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)within 2 hours of sample collectionTo demonstrate clinical equivalence of the EDTA test device as compared to their relevant comparators Venous Comparator and Capillary Comparator. This comparison was focused on the analytes with a unit of pg

Other

MeasureTime frameDescription
Plasma Free Hemoglobin (PFH) in Serum Measured Using the Silica Clot Activator for Serum Sampleswithin 4 hours of sample collectionTo evaluate sample hemolysis as measured by Plasma Free Hemoglobin (PFH) for BD Microtainer® Easy Collect Capillary Tube SST™ and all comparator tube types: capillary and venous
Serum Collection Volume of the Test Device SST Tubewithin 4 hours of sample collectionTo evaluate serum collection volume of the Test Device SST tube
Duration of Sample Collection for Test Device SST Tube and Comparator Tubewithin 4 hours of sample collectionTo evaluate duration of sample collection for Test Device SST tube tube and capillary comparator.
Frequency of Occurrence of All Instrument Flags (Both Asterisks and IP Messages) in the Test Device EDTA and All Comparator Tube Typeswithin 4 hours of sample collectionEvaluation of the frequency of occurrence of all instrument flags (both asterisks and IP messages) in the Test Device EDTA and all comparator tube types: venous and capillary tubes
Whole Blood Collection Volume for the Test Device EDTAwithin 4 hours of sample collectionTo evaluate whole blood collection volume for the Test Device EDTA
Duration of Sample Collection for the Test Device EDTA and Capillary Comparatorwithin 4 hours of sample collectionTo evaluate duration of sample collection for the Test Device EDTA and capillary comparator

Countries

United States

Participant flow

Participants by arm

ArmCount
Test Product
BD MiniDraw™ Capillary Blood Collection System Capillary Blood Collection System: Test Blood Sample Collection
262
Total262

Withdrawals & dropouts

PeriodReasonFG000
Part A: SerumWithdrawal by Subject1

Baseline characteristics

CharacteristicTest Product
Age, Customized
Age over 18
262 Participants
Race and Ethnicity Not Collected— Participants
Sex/Gender, Customized
Sex/Gender
Female
149 Participants
Sex/Gender, Customized
Sex/Gender
Male
96 Participants
Sex/Gender, Customized
Sex/Gender
not reported
17 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
deaths
Total, all-cause mortality
0 / 2620 / 262
other
Total, other adverse events
0 / 2620 / 262
serious
Total, serious adverse events
0 / 2620 / 262

Outcome results

Primary

Alkaline Phosphatase (ALP), Alanine Transaminase (ALT), and Aspartate Aminotransferase (AST), in Serum Measured Using the Silica Clot Activator for Serum Samples

To demonstrate clinical equivalence of the SST test device as compared to Venous Comparator Serum and Capillary Comparator for chemistry analytes. This comparison was focused on the analytes with a unit of U/L

Time frame: within 4 hours of sample collection

Population: This was a test for an analyte result. If there was no measurable value reported, it was not included in the analysis.

ArmMeasureGroupValue (MEAN)
Test ProductAlkaline Phosphatase (ALP), Alanine Transaminase (ALT), and Aspartate Aminotransferase (AST), in Serum Measured Using the Silica Clot Activator for Serum SamplesALKP Test Device SST™ SST™98.6 U/L
Test ProductAlkaline Phosphatase (ALP), Alanine Transaminase (ALT), and Aspartate Aminotransferase (AST), in Serum Measured Using the Silica Clot Activator for Serum SamplesALKP Capillary Comparator SST™98.9 U/L
Test ProductAlkaline Phosphatase (ALP), Alanine Transaminase (ALT), and Aspartate Aminotransferase (AST), in Serum Measured Using the Silica Clot Activator for Serum SamplesALT Test Device SST™50.2 U/L
Test ProductAlkaline Phosphatase (ALP), Alanine Transaminase (ALT), and Aspartate Aminotransferase (AST), in Serum Measured Using the Silica Clot Activator for Serum SamplesALT Capillary Comparator SST™51 U/L
Test ProductAlkaline Phosphatase (ALP), Alanine Transaminase (ALT), and Aspartate Aminotransferase (AST), in Serum Measured Using the Silica Clot Activator for Serum SamplesAST Test Device SST™47.4 U/L
Test ProductAlkaline Phosphatase (ALP), Alanine Transaminase (ALT), and Aspartate Aminotransferase (AST), in Serum Measured Using the Silica Clot Activator for Serum SamplesAST Capillary Comparator SST™49.5 U/L
Test ProductAlkaline Phosphatase (ALP), Alanine Transaminase (ALT), and Aspartate Aminotransferase (AST), in Serum Measured Using the Silica Clot Activator for Serum SamplesALKP Venous comparator Serum100.4 U/L
Test ProductAlkaline Phosphatase (ALP), Alanine Transaminase (ALT), and Aspartate Aminotransferase (AST), in Serum Measured Using the Silica Clot Activator for Serum SamplesALT Venous comparator Serum50.4 U/L
Test ProductAlkaline Phosphatase (ALP), Alanine Transaminase (ALT), and Aspartate Aminotransferase (AST), in Serum Measured Using the Silica Clot Activator for Serum SamplesAST Venous comparator Serum45.7 U/L
Primary

Chemistry Analytes [Albumin (ALB) and Total Protein (TP)] in Serum Measured Using the Silica Clot Activator for Serum Samples

To demonstrate clinical equivalence of the SST test device as compared to Venous Comparator Serum and Capillary Comparator for chemistry analytes. This comparison was focused on the analytes with a unit of g/dL

Time frame: within 4 hours of sample collection

Population: This was a test for an analyte result. If there was no measurable value reported, it was not included in the analysis.

ArmMeasureGroupValue (MEAN)
Test ProductChemistry Analytes [Albumin (ALB) and Total Protein (TP)] in Serum Measured Using the Silica Clot Activator for Serum SamplesALB Test Device SST™3.746 g/dL
Test ProductChemistry Analytes [Albumin (ALB) and Total Protein (TP)] in Serum Measured Using the Silica Clot Activator for Serum SamplesALB Capillary Comparator SST™3.787 g/dL
Test ProductChemistry Analytes [Albumin (ALB) and Total Protein (TP)] in Serum Measured Using the Silica Clot Activator for Serum SamplesTP Test Device SST™6.77 g/dL
Test ProductChemistry Analytes [Albumin (ALB) and Total Protein (TP)] in Serum Measured Using the Silica Clot Activator for Serum SamplesTP Capillary Comparator SST™6.83 g/dL
Test ProductChemistry Analytes [Albumin (ALB) and Total Protein (TP)] in Serum Measured Using the Silica Clot Activator for Serum SamplesALB Venous comparator Serum3.791 g/dL
Test ProductChemistry Analytes [Albumin (ALB) and Total Protein (TP)] in Serum Measured Using the Silica Clot Activator for Serum SamplesTP Venous comparator Serum6.78 g/dL
Comparison: One hundred (100) participants (with paired numerical data) per each analyte provide \> 90% power to ensure that mean biases and confidence intervals are within the clinical acceptance limit (CAL), at medically relevant points, assuming no true bias between the tube types, residual standard deviation (SD) or coefficient of variation (CV) = CAL and collected data cover medically relevant points.
Primary

Hemoglobin (HgB) and Mean Corpuscular Hemoglobin Concentration (MCHC) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)

To demonstrate clinical equivalence of the EDTA test device as compared to their relevant comparators Venous Comparator and Capillary Comparator. This comparison was focused on the analytes with a unit of g/dL

Time frame: within 2 hours of sample collection

Population: This was a test for an analyte result. If there was no measurable value reported, it was not included in the analysis.

ArmMeasureGroupValue (MEAN)
Test ProductHemoglobin (HgB) and Mean Corpuscular Hemoglobin Concentration (MCHC) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)HGB Capillary Comparator EDTA13.33 g/dL
Test ProductHemoglobin (HgB) and Mean Corpuscular Hemoglobin Concentration (MCHC) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)HGB Venous comparator EDTA13.13 g/dL
Test ProductHemoglobin (HgB) and Mean Corpuscular Hemoglobin Concentration (MCHC) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)HGB Test Device EDTA13.45 g/dL
Test ProductHemoglobin (HgB) and Mean Corpuscular Hemoglobin Concentration (MCHC) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)MCHC Capillary Comparator EDTA33.81 g/dL
Test ProductHemoglobin (HgB) and Mean Corpuscular Hemoglobin Concentration (MCHC) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)MCHC Venous comparator EDTA33.4 g/dL
Test ProductHemoglobin (HgB) and Mean Corpuscular Hemoglobin Concentration (MCHC) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)MCHC Test Device EDTA33.69 g/dL
Primary

Lipid Analytes and Metabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum Samples

To demonstrate clinical equivalence of the SST test device as compared to Venous Comparator Serum and Capillary Comparator for chemistry analytes. This comparison was focused on the analytes with a unit of mg/dL

Time frame: within 4 hours of sample collection

Population: This was a test for an analyte result. If there was no measurable value reported, it was not included in the analysis.

ArmMeasureGroupValue (MEAN)
Test ProductLipid Analytes and Metabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum Samplesblood urea nitrogen (BUN) Test Device SST™21.9 mg/dL
Test ProductLipid Analytes and Metabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum SamplesBUN Capillary Comparator SST™22 mg/dL
Test ProductLipid Analytes and Metabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum SamplesCalcium (Ca) Test Device SST™9.43 mg/dL
Test ProductLipid Analytes and Metabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum SamplesCa Capillary Comparator SST™9.52 mg/dL
Test ProductLipid Analytes and Metabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum Samplescholesterol (Chol) Test Device SST™190.2 mg/dL
Test ProductLipid Analytes and Metabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum SamplesChol Capillary Comparator SST™192.8 mg/dL
Test ProductLipid Analytes and Metabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum SamplesCreatinine (Creat) Test Device SST™2.418 mg/dL
Test ProductLipid Analytes and Metabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum SamplesCreat Capillary Comparator SST™2.432 mg/dL
Test ProductLipid Analytes and Metabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum SamplesGlucose (GLUC) Test Device SST™158.2 mg/dL
Test ProductLipid Analytes and Metabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum SamplesGLUC Capillary Comparator SST™159.1 mg/dL
Test ProductLipid Analytes and Metabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum Sampleshigh-density lipoprotein (HDL) Test Device SST™51.28 mg/dL
Test ProductLipid Analytes and Metabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum SamplesHDL Capillary Comparator SST™52.54 mg/dL
Test ProductLipid Analytes and Metabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum Sampleslow-density lipoprotein (LDL) Test Device SST™131.07 mg/dL
Test ProductLipid Analytes and Metabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum SamplesLDL Capillary Comparator SST™132.76 mg/dL
Test ProductLipid Analytes and Metabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum SamplesTotal Bilirubin (TBIL) Test Device SST™1.8604 mg/dL
Test ProductLipid Analytes and Metabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum SamplesTBIL Capillary Comparator SST™1.8763 mg/dL
Test ProductLipid Analytes and Metabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum SamplesTriglycerides (Trig) Test Device SST™155 mg/dL
Test ProductLipid Analytes and Metabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum SamplesTrig Capillary Comparator SST™158.6 mg/dL
Test ProductLipid Analytes and Metabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum SamplesBUN Venous comparator Serum20.4 mg/dL
Test ProductLipid Analytes and Metabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum SamplesCa Venous comparator Serum9.51 mg/dL
Test ProductLipid Analytes and Metabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum SamplesChol Venous comparator Serum195.8 mg/dL
Test ProductLipid Analytes and Metabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum SamplesCreat Venous comparator Serum2.11 mg/dL
Test ProductLipid Analytes and Metabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum SamplesGLUC Venous comparator Serum149.4 mg/dL
Test ProductLipid Analytes and Metabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum SamplesHDL Venous comparator Serum54.83 mg/dL
Test ProductLipid Analytes and Metabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum SamplesLDL Venous comparator Serum140.78 mg/dL
Test ProductLipid Analytes and Metabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum SamplesTBIL Venous comparator Serum1.7931 mg/dL
Test ProductLipid Analytes and Metabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum SamplesTrig Venous comparator Serum157.9 mg/dL
Primary

Mean Corpuscular Hemoglobin (MCH) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)

To demonstrate clinical equivalence of the EDTA test device as compared to their relevant comparators Venous Comparator and Capillary Comparator. This comparison was focused on the analytes with a unit of pg

Time frame: within 2 hours of sample collection

Population: This was a test for an analyte result. If there was no measurable value reported, it was not included in the analysis.

ArmMeasureGroupValue (MEAN)
Test ProductMean Corpuscular Hemoglobin (MCH) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)MCH Capillary Comparator EDTA30.22 pg
Test ProductMean Corpuscular Hemoglobin (MCH) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)MCH Venous comparator EDTA30.12 pg
Test ProductMean Corpuscular Hemoglobin (MCH) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)MCH Test Device EDTA30.21 pg
Primary

Mean Corpuscular Volume (MCV) and Red Blood Cell Distribution Width (RDW) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)

To demonstrate clinical equivalence of the EDTA test device as compared to their relevant comparators Venous Comparator and Capillary Comparator. This comparison was focused on the analytes with a unit of fL

Time frame: within 2 hours of sample collection

Population: This was a test for an analyte result. If there was no measurable value reported, it was not included in the analysis.

ArmMeasureGroupValue (MEAN)
Test ProductMean Corpuscular Volume (MCV) and Red Blood Cell Distribution Width (RDW) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)MCV Capillary Comparator EDTA89.38 fL
Test ProductMean Corpuscular Volume (MCV) and Red Blood Cell Distribution Width (RDW) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)MCV Venous comparator EDTA90.16 fL
Test ProductMean Corpuscular Volume (MCV) and Red Blood Cell Distribution Width (RDW) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)MCV Test Device EDTA89.66 fL
Test ProductMean Corpuscular Volume (MCV) and Red Blood Cell Distribution Width (RDW) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)RDW-SD Capillary Comparator EDTA44.14 fL
Test ProductMean Corpuscular Volume (MCV) and Red Blood Cell Distribution Width (RDW) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)RDW-SD Venous comparator EDTA44.97 fL
Test ProductMean Corpuscular Volume (MCV) and Red Blood Cell Distribution Width (RDW) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)RDW-SD Test Device EDTA44.32 fL
Primary

Metabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum Samples

To demonstrate clinical equivalence of the SST test device as compared to Venous Comparator Serum and Capillary Comparator for chemistry analytes. This comparison was focused on the analytes with a unit of mmol/L

Time frame: within 4 hours of sample collection

Population: This was a test for an analyte result. If there was no measurable value reported, it was not included in the analysis.

ArmMeasureGroupValue (MEAN)
Test ProductMetabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum Sampleschloride (CL) Test Device SST™107.37 mmol/L
Test ProductMetabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum SamplesCL Capillary Comparator SST™107.11 mmol/L
Test ProductMetabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum SamplesCarbon Dioxide (CO2) Test Device SST™21.72 mmol/L
Test ProductMetabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum SamplesCO2 Capillary Comparator SST™24.02 mmol/L
Test ProductMetabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum Samplespotassium (K) Test Device SST™4.436 mmol/L
Test ProductMetabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum SamplesK Capillary Comparator SST™4.506 mmol/L
Test ProductMetabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum SamplesSodium (Na) Test Device SST™138.29 mmol/L
Test ProductMetabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum SamplesNa Capillary Comparator SST™139.08 mmol/L
Test ProductMetabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum SamplesCL Venous comparator Serum104.94 mmol/L
Test ProductMetabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum SamplesCO2 Venous comparator Serum28.2 mmol/L
Test ProductMetabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum SamplesK Venous comparator Serum4.228 mmol/L
Test ProductMetabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum SamplesNa Venous comparator Serum139.33 mmol/L
Primary

Red Blood Cells (RBC) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)

To demonstrate clinical equivalence of the EDTA test device as compared to their relevant comparators Venous Comparator and Capillary Comparator. This comparison was focused on the analytes with a unit of 10\^6/µL

Time frame: within 2 hours of sample collection

Population: This was a test for an analyte result. If there was no measurable value reported, it was not included in the analysis.

ArmMeasureGroupValue (MEAN)
Test ProductRed Blood Cells (RBC) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)RBC Capillary Comparator EDTA4.43 10^6 cells per µL
Test ProductRed Blood Cells (RBC) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)RBC Venous comparator EDTA4.375 10^6 cells per µL
Test ProductRed Blood Cells (RBC) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)RBC Test Device EDTA4.471 10^6 cells per µL
Primary

Thyroid Stimulating Hormone (TSH) in Serum Measured Using the Silica Clot Activator for Serum Samples

To demonstrate clinical equivalence of the SST test device as compared to Venous Comparator Serum and Capillary Comparator for chemistry analytes. This comparison was focused on the analytes with a unit of µIU/mL

Time frame: within 4 hours of sample collection

Population: This was a test for an analyte result. If there was no measurable value reported, it was not included in the analysis.

ArmMeasureGroupValue (MEAN)
Test ProductThyroid Stimulating Hormone (TSH) in Serum Measured Using the Silica Clot Activator for Serum SamplesTSH Test Device SST™4.7599 µIU/mL
Test ProductThyroid Stimulating Hormone (TSH) in Serum Measured Using the Silica Clot Activator for Serum SamplesTSH Capillary Comparator SST™5.4529 µIU/mL
Test ProductThyroid Stimulating Hormone (TSH) in Serum Measured Using the Silica Clot Activator for Serum SamplesTSH Venous comparator Serum5.4211 µIU/mL
Primary

Vitamin D (Vit D), in Serum Measured Using the Silica Clot Activator for Serum Samples

To demonstrate clinical equivalence of the SST test device as compared to Venous Comparator Serum and Capillary Comparator for chemistry analytes. This comparison was focused on the analytes with a unit of ng/mL

Time frame: within 4 hours of sample collection

Population: This was a test for an analyte result. If there was no measurable value reported, it was not included in the analysis.

ArmMeasureGroupValue (MEAN)
Test ProductVitamin D (Vit D), in Serum Measured Using the Silica Clot Activator for Serum SamplesVit D Test Device SST™41.784 ng/mL
Test ProductVitamin D (Vit D), in Serum Measured Using the Silica Clot Activator for Serum SamplesVit D Capillary Comparator SST™36.786 ng/mL
Test ProductVitamin D (Vit D), in Serum Measured Using the Silica Clot Activator for Serum SamplesVit D Venous comparator Serum37.603 ng/mL
Primary

White Blood Cell Differential Counts, Hemoglobin A1c, Hematocrit, and Red Blood Cell Distribution Width in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)

To demonstrate clinical equivalence of the EDTA test device as compared to their relevant comparators venous comparator and capillary comparator. This comparison was focused on the analytes with a unit of percentage

Time frame: within 2 hours of sample collection

Population: This was a test for an analyte result. If there was no measurable value reported, it was not included in the analysis.

ArmMeasureGroupValue (MEAN)
Test ProductWhite Blood Cell Differential Counts, Hemoglobin A1c, Hematocrit, and Red Blood Cell Distribution Width in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)Basophils (BA)% Test Device EDTA0.73 % of analyte in the sample
Test ProductWhite Blood Cell Differential Counts, Hemoglobin A1c, Hematocrit, and Red Blood Cell Distribution Width in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)BA% Capillary Comparator EDTA0.71 % of analyte in the sample
Test ProductWhite Blood Cell Differential Counts, Hemoglobin A1c, Hematocrit, and Red Blood Cell Distribution Width in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)BA% Venous comparator EDTA0.71 % of analyte in the sample
Test ProductWhite Blood Cell Differential Counts, Hemoglobin A1c, Hematocrit, and Red Blood Cell Distribution Width in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)Eosinophils (EO)% Capillary Comparator EDTA2.97 % of analyte in the sample
Test ProductWhite Blood Cell Differential Counts, Hemoglobin A1c, Hematocrit, and Red Blood Cell Distribution Width in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)EO% Venous comparator EDTA3.16 % of analyte in the sample
Test ProductWhite Blood Cell Differential Counts, Hemoglobin A1c, Hematocrit, and Red Blood Cell Distribution Width in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)EO% Test Device EDTA2.89 % of analyte in the sample
Test ProductWhite Blood Cell Differential Counts, Hemoglobin A1c, Hematocrit, and Red Blood Cell Distribution Width in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)hemoglobin A1c (HbA1c) Capillary Comparator EDTA5.938 % of analyte in the sample
Test ProductWhite Blood Cell Differential Counts, Hemoglobin A1c, Hematocrit, and Red Blood Cell Distribution Width in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)HbA1c Venous comparator EDTA5.965 % of analyte in the sample
Test ProductWhite Blood Cell Differential Counts, Hemoglobin A1c, Hematocrit, and Red Blood Cell Distribution Width in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)HbA1c Test Device EDTA5.934 % of analyte in the sample
Test ProductWhite Blood Cell Differential Counts, Hemoglobin A1c, Hematocrit, and Red Blood Cell Distribution Width in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)HCT Capillary Comparator EDTA39.41 % of analyte in the sample
Test ProductWhite Blood Cell Differential Counts, Hemoglobin A1c, Hematocrit, and Red Blood Cell Distribution Width in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)Hematocrit (HCT) Venous comparator EDTA39.26 % of analyte in the sample
Test ProductWhite Blood Cell Differential Counts, Hemoglobin A1c, Hematocrit, and Red Blood Cell Distribution Width in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)HCT Test Device EDTA39.9 % of analyte in the sample
Test ProductWhite Blood Cell Differential Counts, Hemoglobin A1c, Hematocrit, and Red Blood Cell Distribution Width in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)Lymphocytes (LY)% Capillary Comparator EDTA27.93 % of analyte in the sample
Test ProductWhite Blood Cell Differential Counts, Hemoglobin A1c, Hematocrit, and Red Blood Cell Distribution Width in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)LY% Venous comparator EDTA27.85 % of analyte in the sample
Test ProductWhite Blood Cell Differential Counts, Hemoglobin A1c, Hematocrit, and Red Blood Cell Distribution Width in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)LY% Test Device EDTA27.69 % of analyte in the sample
Test ProductWhite Blood Cell Differential Counts, Hemoglobin A1c, Hematocrit, and Red Blood Cell Distribution Width in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)Monocytes (MO)% Capillary Comparator EDTA8.4 % of analyte in the sample
Test ProductWhite Blood Cell Differential Counts, Hemoglobin A1c, Hematocrit, and Red Blood Cell Distribution Width in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)MO% Venous comparator EDTA8.49 % of analyte in the sample
Test ProductWhite Blood Cell Differential Counts, Hemoglobin A1c, Hematocrit, and Red Blood Cell Distribution Width in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)MO% Test Device EDTA8.6 % of analyte in the sample
Test ProductWhite Blood Cell Differential Counts, Hemoglobin A1c, Hematocrit, and Red Blood Cell Distribution Width in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)Neutrophils (NE)% Capillary Comparator EDTA59.51 % of analyte in the sample
Test ProductWhite Blood Cell Differential Counts, Hemoglobin A1c, Hematocrit, and Red Blood Cell Distribution Width in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)NE% Venous comparator EDTA59.39 % of analyte in the sample
Test ProductWhite Blood Cell Differential Counts, Hemoglobin A1c, Hematocrit, and Red Blood Cell Distribution Width in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)NE% Test Device EDTA59.7 % of analyte in the sample
Test ProductWhite Blood Cell Differential Counts, Hemoglobin A1c, Hematocrit, and Red Blood Cell Distribution Width in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)Red Blood Cell Distribution Width (RDW)-CV Capillary Comparator EDTA13.54 % of analyte in the sample
Test ProductWhite Blood Cell Differential Counts, Hemoglobin A1c, Hematocrit, and Red Blood Cell Distribution Width in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)RDW-CV Venous comparator EDTA13.67 % of analyte in the sample
Test ProductWhite Blood Cell Differential Counts, Hemoglobin A1c, Hematocrit, and Red Blood Cell Distribution Width in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)RDW-CV Test Device EDTA13.55 % of analyte in the sample
Primary

White Blood Cells (WBC) and Platelets (PLT) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)

To demonstrate clinical equivalence of the EDTA test device as compared to their relevant comparators Venous Comparator and Capillary Comparator. This comparison was focused on the analytes with a unit of 10\^3/µL

Time frame: within 2 hours of sample collection

Population: This was a test for an analyte result. If there was no measurable value reported, it was not included in the analysis.

ArmMeasureGroupValue (MEAN)
Test ProductWhite Blood Cells (WBC) and Platelets (PLT) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)PLT Capillary Comparator EDTA275.3 10^3 cells per µL
Test ProductWhite Blood Cells (WBC) and Platelets (PLT) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)PLT Venous comparator EDTA297 10^3 cells per µL
Test ProductWhite Blood Cells (WBC) and Platelets (PLT) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)PLT Test Device EDTA262.1 10^3 cells per µL
Test ProductWhite Blood Cells (WBC) and Platelets (PLT) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)WBC Capillary Comparator EDTA12.647 10^3 cells per µL
Test ProductWhite Blood Cells (WBC) and Platelets (PLT) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)WBC Venous comparator EDTA12.453 10^3 cells per µL
Test ProductWhite Blood Cells (WBC) and Platelets (PLT) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)WBC Test Device EDTA12.746 10^3 cells per µL
Other Pre-specified

Duration of Sample Collection for Test Device SST Tube and Comparator Tube

To evaluate duration of sample collection for Test Device SST tube tube and capillary comparator.

Time frame: within 4 hours of sample collection

Population: time to fill tube

ArmMeasureGroupValue (MEAN)
Test ProductDuration of Sample Collection for Test Device SST Tube and Comparator TubeCapillary Comparator SST™108.7 seconds
Test ProductDuration of Sample Collection for Test Device SST Tube and Comparator TubeTest Device SST™128.0 seconds
Other Pre-specified

Duration of Sample Collection for the Test Device EDTA and Capillary Comparator

To evaluate duration of sample collection for the Test Device EDTA and capillary comparator

Time frame: within 4 hours of sample collection

Population: samples from tubes filled

ArmMeasureGroupValue (MEAN)
Test ProductDuration of Sample Collection for the Test Device EDTA and Capillary ComparatorCapillary Comparator EDTA95.8 s
Test ProductDuration of Sample Collection for the Test Device EDTA and Capillary ComparatorTest Device EDTA81 s
Other Pre-specified

Frequency of Occurrence of All Instrument Flags (Both Asterisks and IP Messages) in the Test Device EDTA and All Comparator Tube Types

Evaluation of the frequency of occurrence of all instrument flags (both asterisks and IP messages) in the Test Device EDTA and all comparator tube types: venous and capillary tubes

Time frame: within 4 hours of sample collection

Population: samples from tubes collected from participants.

ArmMeasureGroupValue (COUNT_OF_UNITS)
Test ProductFrequency of Occurrence of All Instrument Flags (Both Asterisks and IP Messages) in the Test Device EDTA and All Comparator Tube TypesTest Device EDTA30 Blood samples
Test ProductFrequency of Occurrence of All Instrument Flags (Both Asterisks and IP Messages) in the Test Device EDTA and All Comparator Tube TypesCapillary Comparator EDTA25 Blood samples
Test ProductFrequency of Occurrence of All Instrument Flags (Both Asterisks and IP Messages) in the Test Device EDTA and All Comparator Tube TypesBD Venous comparator EDTA5 Blood samples
Test ProductFrequency of Occurrence of All Instrument Flags (Both Asterisks and IP Messages) in the Test Device EDTA and All Comparator Tube TypesVenous comparator EDTA5 Blood samples
Other Pre-specified

Plasma Free Hemoglobin (PFH) in Serum Measured Using the Silica Clot Activator for Serum Samples

To evaluate sample hemolysis as measured by Plasma Free Hemoglobin (PFH) for BD Microtainer® Easy Collect Capillary Tube SST™ and all comparator tube types: capillary and venous

Time frame: within 4 hours of sample collection

ArmMeasureGroupValue (MEAN)
Test ProductPlasma Free Hemoglobin (PFH) in Serum Measured Using the Silica Clot Activator for Serum SamplesTest Device SST™20.2659 mg/dL
Test ProductPlasma Free Hemoglobin (PFH) in Serum Measured Using the Silica Clot Activator for Serum SamplesCapillary Comparator SST™44.9719 mg/dL
Test ProductPlasma Free Hemoglobin (PFH) in Serum Measured Using the Silica Clot Activator for Serum SamplesVenous comparator SST™11.2487 mg/dL
Test ProductPlasma Free Hemoglobin (PFH) in Serum Measured Using the Silica Clot Activator for Serum SamplesVenous comparator Serum9.8032 mg/dL
Other Pre-specified

Serum Collection Volume of the Test Device SST Tube

To evaluate serum collection volume of the Test Device SST tube

Time frame: within 4 hours of sample collection

Population: samples for collection volume

ArmMeasureValue (MEAN)
Test ProductSerum Collection Volume of the Test Device SST Tube236.41521 uL
Other Pre-specified

Whole Blood Collection Volume for the Test Device EDTA

To evaluate whole blood collection volume for the Test Device EDTA

Time frame: within 4 hours of sample collection

Population: samples from tubes collected

ArmMeasureValue (MEAN)
Test ProductWhole Blood Collection Volume for the Test Device EDTA299.37786 uL