Skip to content

A Performance Evaluation of the LumiraDx Point of Care CRP Assay

A Performance Evaluation of the LumiraDx Point of Care CRP Assay.

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT05180110
Acronym
REACT
Enrollment
79
Registered
2022-01-06
Start date
2021-11-25
Completion date
2023-05-30
Last updated
2023-12-19

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Inflammation, Infections, Tissue Injury

Brief summary

In patients with symptoms of infection, tissue injury and inflammatory disorders the study will evaluate agreement between the CRP measurements from the LumiraDx POC CRP assay and the Siemens Dimension Xpand Plus CRP assay, as an aid in evaluation and detection of infection, tissue injury and inflammatory disorders.

Detailed description

This is a performance evaluation study designed to assess the precision and accuracy of the LumiraDx point of Care (POC) C-reactive Protein (CRP) test when used in patients presenting with symptoms of infection, tissue injury or inflammatory disorder. The LumiraDx POC CRP test is a quantitative immunoassay providing results in under 10 minutes. The accuracy of the LumiraDx POC CRP test will be assessed using capillary whole blood, venous blood, and plasma samples, by comparison to the CRP results obtained from the same individuals as analysed by trained laboratory professionals using the reference device, Siemens Dimension Xpand Plus CRP assay. The results of this study are intended to be used for regulatory filings for use as an in vitro diagnostic test in the UK, European Union (EU) and other relevant geographies. Adults (aged 18 and over at time of consent) presenting to the study sites with symptoms of infection, tissue injury or inflammatory disorder will be included in the study. Approximately 250 patients will be recruited to ensure that sufficient suitable patients are enrolled and to allow for a small level of exclusions. After obtaining informed consent, blood tube samples will be drawn and fingerstick samples of capillary blood will be taken from the subject, these will be applied directly onto unique test strips for immediate measurement of CRP on the LumiraDx POC instrument. The CRP results obtained via the LumiraDx CRP test will be used in a set of comparative analyses designed to determine the performance characteristics of the test. This study is an observational, cross-sectional design. As an observational study of an investigative device, results obtained during the course of the study will not inform or alter patient management or treatment decisions in any way.

Interventions

DIAGNOSTIC_TESTVenepuncture

Venous blood draw

DIAGNOSTIC_TESTFingerstick

Capillary blood draw

Sponsors

LumiraDx UK Limited
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
DIAGNOSTIC
Masking
NONE

Intervention model description

IVD Performance Evaluation Study

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Patients 18 years and over at the time of consent * Willing and able to provide written informed consent and comply with study procedures * Presenting with symptoms of infection, tissue injury, inflammatory disorders. This includes but not limited to : respiratory tract infection (upper and lower), rheumatoid arthritis, Lupus, burns, trauma, inflammatory bowel disease.

Exclusion criteria

* The patient is currently receiving or has received within the past thirty (30) days of the study visit an experimental biologic or drug including either treatment or therapy. * Skin lesions or conditions that would preclude a fingerstick and/or a venous blood draw. * Patient has previously participated in this study * Any patient with critical illness or requiring a time critical intervention * Patient with end of life or palliative care * Patient suffering from the following conditions: Myeloma, monoclonal gammopathy, extreme lipaemia * Patient is deemed medically unfit to participate.

Design outcomes

Primary

MeasureTime frameDescription
To determine the performance of the LumiraDx POC CRP assay when compared to the Siemens Dimension Xpand Plus CRP in patients with symptoms of infection, inflammation, or injury.2 monthsMeasurement of blood samples from patients with symptoms of inflammation, infection or injury in a reference method and in the LumiraDx method to assess accuracy of the LumiraDx method

Countries

United Kingdom

Outcome results

None listed

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026