Electrocardiographic Diagnostic Performance of the Apple Watch Augmented With an Artificial Intelligence Algorithm

Confirmatory Clinical Investigation of Electrocardiographic Diagnostic Performance of the Apple Watch Augmented With an Artificial Intelligence Algorithm

Status
Completed
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT05167058
Enrollment
220
Registered
2021-12-22
Start date
2022-04-21
Completion date
2023-03-31
Last updated
2023-10-03

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Atrial Fibrillation, Bradycardia, Premature Supraventricular Beat, Premature Ventricular Contraction, Tachycardia

Brief summary

The SWAF study will compare the performance of a smartwatch combined with Cardiologs Platform algorithm in the detection of Atrial Fibrillation and other arrhythmias with that measured on a manually read 12-lead ECG in subjects hospitalized for cardioversion or AF ablation.

Detailed description

The SWAF study is a prospective, non-significant risk, non-randomized, multicentric, open, comparative, confirmatory study. Under subject consent, subjects hospitalized for cardioversion or AF ablation will have a smartwatch ECG recording done simultaneously with 12-lead ECG measurement right before the intervention. If a subject is found in Normal Sinus Rhythm he/she will be discharged otherwise the patient will undergo cardioversion and will have simultaneous recordings done a second time after the intervention. All the measurements will be done in accordance with the existing subject monitoring protocol.

Interventions

Smartwatch recordings interpreted by Cardiologs AI done simultaneously with each 12-lead ECG

Sponsors

Cardiologs Technologies
Lead SponsorINDUSTRY

Study design

Observational model
CASE_ONLY
Time perspective
PROSPECTIVE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Subjects over 18, able and willing to participate in the study * Subjects who are admitted to the hospital for a cardioversion or AF ablation procedure. * Subjects having read the patient information letter and provided his/her consent to participate in writing, by dating and signing the informed consent prior to any trial-related procedure being conducted

Exclusion criteria

* Subjects with paced rhythm or implanted electronic devices * Pregnant or breast-feeding subjects

Design outcomes

Primary

MeasureTime frameDescription
Evaluation of the performance of smartwatch ECG interpreted by Cardiologs Artificial Intelligence in detecting AF (Atrial Fibrillation or Flutter) as identified by the physician on the 12-lead ECGReadings taken simultaneously right before and right after the cardioversion according to the monitoring protocol at the hospitalEvaluation of the performance of smartwatch ECG interpreted by Cardiologs Artificial Intelligence in detecting AF (Atrial Fibrillation or Flutter) as identified by the physician on the 12-lead ECG in the independent annotation center, providing the ground truth from the 12-lead ECG

Secondary

MeasureTime frameDescription
Evaluation of the performance of smartwatch ECG interpreted by Cardiologs Artificial Intelligence in detecting AF as identified by the physician on the smartwatch ECGReadings taken simultaneously right before and right after the cardioversion according to the monitoring protocol at the hospitalEvaluation of the performance of smartwatch ECG interpreted by Cardiologs Artificial Intelligence in detecting AF as identified by the physician on the smartwatch ECG in the independent annotation center, providing the ground truth from the smartwatch ECG.
Evaluation of the performance of smartwatch ECG interpreted by Cardiologs Artificial Intelligence in detecting Sinus Rhythm as identified by the physician on the smartwatch ECGReadings taken simultaneously right before and right after the cardioversion according to the monitoring protocol at the hospitalEvaluation of the performance of smartwatch ECG interpreted by Cardiologs Artificial Intelligence in detecting Sinus Rhythm as identified by the physician on the smartwatch ECG in the independent annotation center, providing the ground truth from the smartwatch ECG.
Evaluation of the performance of smartwatch ECG interpreted by Cardiologs Artificial Intelligence in detecting Tachycardia as identified by the physician on the smartwatch ECGReadings taken simultaneously right before and right after the cardioversion according to the monitoring protocol at the hospitalEvaluation of the performance of smartwatch ECG interpreted by Cardiologs Artificial Intelligence in detecting Tachycardia as identified by the physician on the smartwatch ECG in the independent annotation center, providing the ground truth from the smartwatch ECG.
Assessment of the proportion of smartwatch ECGs identified as inconclusive by Cardiologs Artificial Intelligence and by the physicianReadings taken simultaneously right before and right after the cardioversion according to the monitoring protocol at the hospitalAssessment of the proportion of smartwatch ECGs identified as inconclusive by Cardiologs Artificial Intelligence and by the physician. Inconclusive may mean that there may have been too much artefact or noise to acquire a good signal or that the rhythm is unclassifiable or contains other abnormal rhythm.
Evaluation of the performance of smartwatch ECG interpreted by Cardiologs Artificial Intelligence in detecting Premature Supraventricular Complexes as identified by the physician on the smartwatch ECGReadings taken simultaneously right before and right after the cardioversion according to the monitoring protocol at the hospitalEvaluation of the performance of smartwatch ECG interpreted by Cardiologs Artificial Intelligence in detecting Premature Supraventricular Complexes as identified by the physician on the smartwatch ECG in the independent annotation center, providing the ground truth from the smartwatch ECG.
Evaluation of the performance of smartwatch ECG interpreted by Cardiologs Artificial Intelligence in detecting Premature Ventricular Complexes as identified by the physician on the smartwatch ECGReadings taken simultaneously right before and right after the cardioversion according to the monitoring protocol at the hospitalEvaluation of the performance of smartwatch ECG interpreted by Cardiologs Artificial Intelligence in detecting Premature Ventricular Complexes as identified by the physician on the smartwatch ECG in the independent annotation center, providing the ground truth from the smartwatch ECG.
Evaluation of the performance of smartwatch ECG interpreted by Cardiologs Artificial Intelligence in detecting Bradycardia as identified by the physician on the smartwatch ECGReadings taken simultaneously right before and right after the cardioversion according to the monitoring protocol at the hospitalEvaluation of the performance of smartwatch ECG interpreted by Cardiologs Artificial Intelligence in detecting Bradycardia as identified by the physician on the smartwatch ECG in the independent annotation center, providing the ground truth from the smartwatch ECG.