Dynamic Electronic Filter for Arterial Waveform Distortion

Arterial Pressure Waveform Distortion: a New Dynamic Electronic Filter for Automatic Correction of Underdamping/Resonance Artifact. A Prospective Study

Status
Completed
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT05166993
Enrollment
70
Registered
2021-12-22
Start date
2017-03-15
Completion date
2018-09-15
Last updated
2022-01-12

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Underdamping/Resonance Artifacts

Brief summary

Invasive blood pressure (IBP) is the gold standard for arterial pressure (AP) monitoring in critically ill patients. Nevertheless, IBP may be affected by underdamping/resonance artifacts, that eventually lead to AP overestimation. In these cases, when a pulse contour method is applied, wrong hemodynamic data are also delivered by the monitor. In order to overtake this issue, MostCareUp (a PCM powered by Pressure Recording Analytical Method, PRAM), has been implemented with a dynamic electronic filter (EFMC) that automatically corrects the resonant AP waveform. In order to test the EFMC, the electronically corrected AP and the derived cardiac output were compared with respective raw data corrected with the Accudynamic, an adjustable damping device specifically manufactured for normalizing the pressure waveform (in our study: the reference method).

Interventions

DEVICEMostCareUp

Pressure transducers have been connected to the MostCareUp and the hemodynamic parameters compared as follows: no filter, electronic filter, no electronic filter plus Accudynamic (our reference system). Invasive pressure values have been then compared with the non-invasive ones.

Sponsors

University of Florence
Lead SponsorOTHER

Study design

Observational model
COHORT
Time perspective
PROSPECTIVE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

\- ICU patients with invasive blood pressure (radial artery) affected by underdamping/resonance artifacts;

Exclusion criteria

\- Arrhythmias

Design outcomes

Primary

MeasureTime frameDescription
The present study was aimed to test the MostCareUp electronic filter15 minutesTo test the electronic filter comparing the electronically corrected arterial pressure and the derived cardiac output values with those corrected with the Accudynamic, a mechanical adjustable damping device commonly used in clinical practice.