Accurate DCE-MRI Measurement of Glioblastoma Using Point-of-care Portable Perfusion Phantom

Accurate DCE-MRI Measurement of Glioblastoma Using Point-of-care Portable Perfusion Phantom

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT05140902
Enrollment
12
Registered
2021-12-02
Start date
2022-03-28
Completion date
2024-08-25
Last updated
2025-02-14

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Glioblastoma

Brief summary

The goal of this study is to test whether a new device developed at the University of Alabama at Birmingham (UAB) can decrease the error in calculating blood flow of a brain tumor, leading to better prognosis. UAB radiological research team has been studying a cutting-edge imaging technique named dynamic contrast enhanced (DCE) magnetic resonance imaging (MRI) , or DCE-MRI, over 10 years. This technique has been globally used to calculate blood flow of various tissues including tumors. Blood flow often serves as a critical indicator showing a disease status. For example, a brain tumor has typically high blood flow, so the magnitude of blood flow can be used as an indicator to identify the presence and aggressiveness of a brain tumor. In addition, an effective therapy can result in the alteration of the blood flow in a brain tumor. Therefore, the investigators may be able to determine whether the undergoing therapy is effective or not by measuring the blood flow in the brain tumor, and decide whether they need to continue the therapy or try a different one. However, unfortunately, the measurement of blood flow using DCE-MRI is often inaccurate. MRI scanners may use different hardware and software thus the measurement may be different across scanners. The measurement may also be different over time due to hardware instability. Therefore, the investigators propose to use an artificial tissue, named phantom, together with a patient. The phantom has a constant blood flow thus it can serve as a standard. Errors, if it occurs, will affect the images of both the patient and the phantom. Therefore, the investigators will be able to correct the errors in the patient image using the phantom image. UAB radiological research team invented a new device for this purpose named point-of-care portable perfusion phantom, or shortly P4. The team recently demonstrated the utility of the P4 phantom for accurate measurement of blood flow in pancreatic cancer and prostate cancer. In this study, they will test whether the P4 phantom will improve the measurement accuracy in brain cancer.

Detailed description

Glioblastoma is the most common primary malignant type of brain tumor in adults. Surgical tumor resection followed by chemoradiation therapy is the standard of care for patients with glioblastoma, but its prognosis is still fairly dismal (median survival time = 15 months). One major concern that prevents effective treatment management is the difficulty of differentiating between pseudo-progression and true-progression. Pseudo-progression occurs in about 20-30% of glioblastoma patients typically within 3 months after chemoradiation therapy has been completed. Pseudo-progression is a local inflammatory reaction caused by irradiation and enhanced by concurrent chemotherapy, which leads to a transient increase of blood brain barrier (BBB) permeability. The BBB, however, is also disrupted by new cancer occurrence. Therefore, both pseudo- and true-progressions appear with an increased contrast enhancement in MRI, and there are currently no established techniques to differentiate between them. Pseudo-progression is typically known to be associated with better clinical outcomes, so pseudo-progression mistaken for true-progression results in the discontinuation of an effective therapy, while true-progression mistaken for pseudo-progression leads to the continuation of an ineffective therapy that may induce adverse side effects. DCE-MRI has potential to differentiate between pseudo- and true-progressions of glioblastoma. The enhancing lesions of pseudo-progression are due to inflammation, whereas those of true-progression are caused by cancer growing. Thus, true-progression typically presents higher perfusion than pseudo-progression does. DCE-MRI can quantitatively assess the tissue perfusion by monitoring the dynamic change of MRI contrast agent concentration. Several investigators have demonstrated the potential of quantitative DCE-MRI to differentiate between pseudo- and true-progressions. However, the variability in quantitative DCE-MRI measurement across different MRI scanners remains a major concern, as it hinders data comparison among institutes to retrieve a reliable threshold for accurate prognosis and subsequent treatment optimization. A point-of-care perfusion phantom may allow high reproducibility and accurate comparison of quantitative DCE-MRI data across MRI platforms. The UAB radiological research team recently developed the P4 phantom, which is small enough to be imaged concurrently with a patient for real-time quality assurance, but large enough not to suffer from the partial volume effect. The P4 phantom creates constant contrast enhancement curves with very robust repeatability, and thus the contrast agent concentration time-course in a tumor, which is a major source of error in quantitating DCE-MRI parameters, can be accurately calculated in reference to the values observed in the phantom. In our previous study, the variability in quantitating the volume transfer constant of various human tissues across two different MRI scanners was reduced fivefold after P4-based error correction. The investigators hypothesize that the variability in quantitative DCE-MRI measurement of glioblastoma across different scanners will be significantly reduced when the P4 is used for error correction, leading to better differentiation between pseudo- and true-progressions. The goal of this study is to test this hypothesis.

Interventions

P4 is a perfusion phantom developed by Dr. Harrison Kim that can significantly reduce variation in quantitating perfusion of human abdominal tissues across MRI scanners.

Sponsors

National Center for Advancing Translational Sciences (NCATS)
CollaboratorNIH
University of Alabama at Birmingham
Lead SponsorOTHER

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
DIAGNOSTIC
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

1. Adult patients (age 18 years or older). 2. Patients treated with surgery, followed by chemoradiation therapy, and currently under chemotherapy. 3. Patients with a newly or enlarged enhancing lesion inside the radiation field at least three months after completion of radiation therapy. 4. Patients with signed informed consent.

Exclusion criteria

1. Participants with safety contraindications to MRI examination (determined by standard clinical screening). 2. Participants on hemodialysis or with acute renal failure. 3. Participants who are pregnant, lactating or are planning to become pregnant during the study. 4. Participants who are planning to farther a child during the study.

Design outcomes

Primary

MeasureTime frameDescription
To Measure the Reproducibility of qDCE-MRI Measurement of Glioblastoma.At the end of Cycle 2 of chemoradiation therapy (each cycle is 28 days)The goal is to measure the reproducibility of blood perfusion measurement in the glioblastoma using the two consecutive DCE-MRI scans with and without P4-based error correction. The pharmacokinetic (PK) parameter within the region of interest (ROI) will be averaged at each scan after P4-based error correction, and the mean values of two scans will be compared to calculate the coefficient of variation (%COV).
The Differentiation Between the Pseudo- and True-progressions of Glioblastoma Can be Improved Using qDCE-MRI After P4-based Error Correction.At the end of Cycle 2 of chemoradiation therapy (each cycle is 28 days)The PK parameter (e.g., Ktrans) in the tumor with pseudoprogression will be statistically compared with that with true-progression before and after P4-based error correction to determine whether the differentiation between the pseudo- and true-progressions of glioblastoma can be improved using qDCE-MRI after P4-based error correction. Each tumor will be classified into pseudo- or true-progression based on RANO criteria.

Countries

United States

Participant flow

Participants by arm

ArmCount
Glioblastoma Patients
glioblastoma patients with newly or enlarged enhancing lesion within 3 months after completing 6 weeks of adjuvant chemoradiation therapy Point-of-care Portable Perfusion Phantom (P4): P4 is a perfusion phantom developed by Dr. Harrison Kim that can significantly reduce variation in quantitating perfusion of human abdominal tissues across MRI scanners.
12
Total12

Baseline characteristics

CharacteristicGlioblastoma Patients
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
3 Participants
Age, Categorical
Between 18 and 65 years
9 Participants
Age, Continuous60 years
Ethnicity (NIH/OMB)
Hispanic or Latino
1 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
11 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
0 Participants
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants
Race (NIH/OMB)
Asian
0 Participants
Race (NIH/OMB)
Black or African American
0 Participants
Race (NIH/OMB)
More than one race
0 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants
Race (NIH/OMB)
Unknown or Not Reported
0 Participants
Race (NIH/OMB)
White
12 Participants
Region of Enrollment
United States
12 participants
Sex: Female, Male
Female
1 Participants
Sex: Female, Male
Male
11 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 12
other
Total, other adverse events
0 / 12
serious
Total, serious adverse events
0 / 12

Outcome results

Primary

The Differentiation Between the Pseudo- and True-progressions of Glioblastoma Can be Improved Using qDCE-MRI After P4-based Error Correction.

The PK parameter (e.g., Ktrans) in the tumor with pseudoprogression will be statistically compared with that with true-progression before and after P4-based error correction to determine whether the differentiation between the pseudo- and true-progressions of glioblastoma can be improved using qDCE-MRI after P4-based error correction. Each tumor will be classified into pseudo- or true-progression based on RANO criteria.

Time frame: At the end of Cycle 2 of chemoradiation therapy (each cycle is 28 days)

ArmMeasureValue (NUMBER)
Glioblastoma PatientsThe Differentiation Between the Pseudo- and True-progressions of Glioblastoma Can be Improved Using qDCE-MRI After P4-based Error Correction.100 Accuracy (%)
Primary

To Measure the Reproducibility of qDCE-MRI Measurement of Glioblastoma.

The goal is to measure the reproducibility of blood perfusion measurement in the glioblastoma using the two consecutive DCE-MRI scans with and without P4-based error correction. The pharmacokinetic (PK) parameter within the region of interest (ROI) will be averaged at each scan after P4-based error correction, and the mean values of two scans will be compared to calculate the coefficient of variation (%COV).

Time frame: At the end of Cycle 2 of chemoradiation therapy (each cycle is 28 days)

ArmMeasureValue (NUMBER)
Glioblastoma PatientsTo Measure the Reproducibility of qDCE-MRI Measurement of Glioblastoma.5 %COV