The SAVVY Guidewire in Transcatheter Aortic Valve Replacement Procedures

The SAVVY Guidewire in Transcatheter Aortic Valve Replacement Procedures: A Feasibility Study

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT05082337
Acronym
SAVVY
Enrollment
20
Registered
2021-10-18
Start date
2021-10-08
Completion date
2021-12-28
Last updated
2022-01-18

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Aortic Valve Disease, Aortic Valve Stenosis, Hemodynamic Measurement, Transcatheter Aortic Valve Replacement

Brief summary

TAVR patients exhibiting post-procedural residual AR had higher mortality and hospitalization rates due to heart failure, with the extent of this association increasing proportionally to the severity of the regurgitation. Optimizing transcatheter valve performance by intra-procedural hemodynamic evaluation of AR and residual transvalvular gradient remains of high clinical importance.

Detailed description

Continuous technological development and procedural refinements are important to further reduce TAVR peri-procedural complications, facilitate the minimalist strategy and improve clinical outcomes following TAVR. Residual aortic regurgitation (AR) has been one of the main drawbacks of TAVR. TAVR patients exhibiting post-procedural residual AR had higher mortality and hospitalization rates due to heart failure, with the extent of this association increasing proportionally to the severity of the regurgitation. Various actions could be undertaken to minimize AR and its consequences, either intraprocedural (balloon post-dilation, implantation of a second valve) or during follow-up (diuretic treatment, closest clinical follow-up or percutaneous leakage closure). Thus, an accurate diagnosis of the presence and severity of residual AR post-TAVR is key to implement the proper measures and optimize clinical outcomes. The use of cardiac imaging remains the gold standard for evaluating AR post-TAVR, limitations of contrast amount along with the subjectivity of AR evaluation by aortic angiography, and the challenges of echocardiography at the time of the TAVR procedure represent a significant drawback. Thus, alternative tools for evaluating the presence and severity of AR would be very helpful in this setting. Current data supports the implementation of hemodynamic measurements during TAVR procedures in order to improve the clinical decision-making process following valve implantation. However, these actions entail an inherent risk associated with the exchange of catheter-wires. The use of a support guidewire with pressure measurement capabilities would facilitate the hemodynamic evaluation of transcatheter valve performance in a safer and more rapid manner. The new SAVVY guidewire, with both dedicated pacing properties and allowing a continuous hemodynamic pressure monitoring during the procedure, is a unique system in the field and may represent an important step forward in the process of optimizing the TAVR procedure while facilitating procedural steps.

Interventions

DEVICESAVVY guidewire

The SAVVY guidewire will be used for rapid pacing runs, valve delivery and for left ventricular pressure measurements pre- and post-THV deployment

Sponsors

Institut de Cardiologie de Montréal
CollaboratorUNKNOWN
Institut universitaire de cardiologie et de pneumologie de Québec, University Laval
Lead SponsorOTHER

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
OTHER
Masking
NONE

Intervention model description

Patient with severe symptomatic aortic stenosis with a clinical indication for TAVR according to the Heart Team criteria will be treated with an approved aortic valve (balloon-expandable or self-expandable aortic valve) and the SAVVY guidewire will be used for the rapid pacing runs and ventricular pressure measurements.

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Patients with severe symptomatic AS undergoing a TAVR procedure.

Exclusion criteria

* Failure to provide signed informed consent. * Extremely calcified aortic valve (Agatston score \>3000 AU) * Extremely horizontal aorta. * Severe septal hypertrophy (interventricular septum \>15 mm as measured by TTE). * Extreme tortuosity at the level of the iliofemoral arteries, thoracic or abdominal aorta. * Prohibitive surgical risk precluding (according to the Heart Team) conversion to open heart surgery in case of a life threatening complication. * Young patients (\<18 Years) and pregnant women

Design outcomes

Primary

MeasureTime frameDescription
Presence of Major complicationsPeriproceduralPresence of major complications related to the SAVVY guidewire including (i) guidewire kink hindering or preventing the advancement of the transcatheter valve system
Presence of Major vascular complicationsPeriproceduralMajor vascular complications related to the SAVVY guidewire are defined as stroke, bleeding or left ventricular perforation
Number of transcatheter valve malpositioningPeriproceduralPacing capture failure translating into valve malpositioning. Valve malpositiong will be evaluated by a landing site to low or too high, leading to hemodynamically unfavorable results.
Number of valve embolizationPeriproceduralPacing capture failure translating into valve embolization
Number of effective rapid pacing runPeriproceduralEffective rapid pacing will be defined as an adequate ventricular pacing capture for a minimum of 10 seconds, with no capture loss, and leading to a reduction of aortic pressure of at least 50%, with/or a systolic pressure value \<60 mmHg. Efficacy will be assessed by the physician.
Number of accurate ventricular pressurePeriproceduralAccurate ventricular pressure will be defines as a pressure wire measurements similar (differences \<5 mmHg) to those obtained simultaneously with a pigtail catheter in the same cavity or vascular segment (differences \<5 mmHg in SBP, differences \<5 mmHg in LVEDP).