10-year Follow-up of a Wide Diameter Bone Anchored Hearing Implant

Long-term Follow-up of a Wide Diameter Bone Anchored Hearing Implant: the 10-year Experience on Stability, Survival and Tolerability of the BIA300®

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT05058066
Enrollment
51
Registered
2021-09-27
Start date
2019-06-24
Completion date
2020-09-21
Last updated
2025-11-18

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Hearing Loss

Brief summary

A single centre, open, comparative, parallel group, prospective clinical investigation with a single 10 year follow up visit.

Detailed description

This study is a continuation of two previous conducted trials: 1. Stability, long term survival and tolerability of a novel Baha® implant system, a multi-centre investigation new BIA300 (test) implant vs previous generation (control) implant: 1. CAG5173: 6-month(1) and 3-year(2) data. 2. CBAS5562: 5-year data(3). 2. 3 weeks loading: stability, long term survival and tolerability of a novel Baha® implant system, a multi-centre investigation 3 week loading vs 6+ week loading (= same group as control group trial 1). 6- month(4) and 3-year(5) data. In this study the (5-year data for trial 2 and) 10-year data regarding stability, survival and soft tissue tolerability from the above mentioned trials will be evaluated and compared. The previous trials were multicentre, but the current study will be single-centred and therefore only include the participants of the Radboud site. Data will be collected at a visit at (5 years, for trial 2 and) 10 years (+/- 6 months) after implantation. Up to now, no RCT of these implants (with the loading of 3 weeks after implantation) has assessed stability, survival and tolerability over a 10-year follow-up period.

Interventions

DEVICEBaha® flange fixture (diameter 3.75mm; length 4mm) with 6mm abutment

The control implant included the previous generation as-machined titanium Baha® flange fixture (diameter 3.75mm; length 4mm) with 6mm conically shaped abutment.

DEVICEBaha® BIA300 Implant (diameter 4.5mm; length 4mm) with 6mm abutment

The test implant and the implant used for the 3-week loading trial included the wide diameter titanium implant (diameter 4.5mm; length 4mm) with 6mm rounded, apically converging titanium abutment developed by Cochlear Bone Anchored Solutions AB (Mölnlyncke, Sweden). This system with an additional minor change to the internal abutment connection design was later commercialized under the name Cochlear™ Baha® BIA300 Implant with abutment. In addition to the difference in abutment shape and the wider diameter, the test implant incorporates small-sized threads at the implant neck and the moderately rough TiOblast™ (Dentsply, Mölndal, Sweden) surface on the intraosseous part of the implant6-8.

PROCEDUREEarly loading

Loading the implant at 3 weeks after implantation (instead of 6 weeks after implantation)

Sponsors

Cochlear Bone Anchored Solutions
CollaboratorINDUSTRY
Radboud University Medical Center
Lead SponsorOTHER

Study design

Allocation
NON_RANDOMIZED
Intervention model
PARALLEL
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* To be included in the original trials, patients had to be at least 18 years old.

Exclusion criteria

*

Design outcomes

Primary

MeasureTime frameDescription
implant stability10 years after implantationImplant Stability Quotiënt (ISQ) will be measured at the 10y-visit. The resonance frequency analysis renders an ISQ value ranging from 1 to 100. Measurements shall be performed at the abutment level, as is done in the previous investigations. The highest and lowest ISQ value out of two perpendicular measurements shall be recorded as ISQ High and ISQ low. The measurement should preferably be done by the same Osstell instrument (Osstell, Gothenburg, Sweden) that was used in the previous completed study. SmartPeg Type 43 and 55 shall be used during the measurement (Osstell, Gothenburg, Sweden). The procedures for obtaining ISQ values is described in the manufacturer's instruction that will be handed out to the investigators.

Secondary

MeasureTime frameDescription
implant survival10 years after implantationAll patients will be asked if they have experienced any implant osseointegration problems. The time from implantation until implant loss or removal will be noted. In case of implant removal, reason for removal shall be recorded.
soft tissue reaction10 years after implantationSoft tissue reactions will be measured using the Holgers' classification & IPS-score. The Holgers' classification is designed to capture signs and symptoms of inflammation or infection at the site of implantation (scale 0-4). The IPS-score is a complete standardized assessment scale for skin complications for both percutaneous and transcutaneous implants for bone conduction devices (BCDs) with a proposed treatment advice based on the outcome. The IPS score has three subscales: inflammation (0-4); pain (0-2), and skin height (0-2).
device use10 years after surgeryTo evaluate the long-term device use of a Baha®.