Skip to content

Validation of Point Partial User Needs With Partial Finger Amputees

Validation of Point Partial User Needs With Partial Finger Amputees

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT05012657
Enrollment
5
Registered
2021-08-19
Start date
2020-11-03
Completion date
2022-12-31
Last updated
2024-11-08

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Amputation; Traumatic, Hand

Brief summary

The objective of the device feasibility study will be to validate the user needs of the Point Partial system. This study will be a single group intervention model where one group of 5 partial finger amputees will be asked to perform several tasks. Successful completion of a task results in a fulfilled user need. Failure to complete a task results in an unfulfilled user need.

Interventions

The Point Partial partial finger prosthetic system consists of 1-4 ratcheting mechanical partial digits and mounting brackets

Sponsors

University of Colorado, Denver
CollaboratorOTHER
Point Designs
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
DEVICE_FEASIBILITY
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Partial finger amputees with partial absence of index and/or middle fingers and presence of thumb. Additional finger loss is acceptable if all other criterion are met. * Fluent in English * Individuals aged 18 or greater * Patients may present with either fused or mobile MCP joints.

Exclusion criteria

* Significant cognitive deficits as determined upon clinical evaluation * Significant neurological deficits as determined upon clinical evaluation * Significant physical deficits of the residual limb impacting full participation in the study as determined upon clinical evaluation * Uncontrolled pain or phantom pain impacting full participation in the study as determined upon OT evaluation * Serious uncontrolled medical problems as judged by the project therapist.

Design outcomes

Primary

MeasureTime frameDescription
VAL-1: System Comfort Validation TestDay 1 (8 hours)User wears Point Partial(s) for a full day (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall be comfortable for a full day's use. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.
VAL-2: Unilateral Function Validation TestDay 1 (10 minutes)User flexes and extends the Point Partial(s) without using contralateral hand (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall allow the user to flex and extend finger without their second hand. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.
VAL-3.1: System Robustness Validation Test With 25 lb BagDay 1 (10 minutes)User lifts a 25 lb. bag with Point Partial(s) (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall be robust enough to operate in challenging environments. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.
VAL-3.2: Robustness Validation Test With HammerDay 1 (10 minutes)User grips a hammer with Point Partial(s) and drives a nail into a piece of wood (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall be robust enough to operate in challenging environments. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.
VAL-4: Object Release Validation TestDay 1 (10 minutes)User releases Point Partial(s) while performing an active grasp by pressing the dorsal button (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall be releasable while performing active grasps. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.
VAL-5: Position Function Validation TestDay 1 (10 minutes)User positions Point Partial(s) into each of the locking positions (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall have a smooth ratcheting mechanism. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.
VAL-6: Extension Function Validation TestDay 1 (10 minutes)User activates spring-back mechanism by using (1) contralateral hand, (2) tabletop, and (3) side or thigh throughout day without failed extension (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall have a reliable auto spring-back mechanism. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.
VAL-7: Grip Functions Validation TestDay 1 (10 minutes)User grips a 5 lb. cylindrical smooth object using Point Partial(s) for 10 seconds (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall have high grip surface friction. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.
VAL-8: Fine Motor Function Validation TestDay 1 (10 minutes)User picks up 1 coin from a smooth tabletop (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall have a pronounced fingernail. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.
VAL-9: Overall Function & Robustness Validation TestDay 1 (8 hours)User performs VAL-1 through VAL-8 (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall be robustly attached to mounting system. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

Countries

United States

Participant flow

Participants by arm

ArmCount
Prosthesis
Patient is temporarily fit with Point Partial partial finger prosthetic system Point Partial: The Point Partial partial finger prosthetic system consists of 1-4 ratcheting mechanical partial digits and mounting brackets
5
Total5

Baseline characteristics

CharacteristicProsthesis
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
0 Participants
Age, Categorical
Between 18 and 65 years
5 Participants
Ethnicity (NIH/OMB)
Hispanic or Latino
0 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
2 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
3 Participants
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants
Race (NIH/OMB)
Asian
0 Participants
Race (NIH/OMB)
Black or African American
0 Participants
Race (NIH/OMB)
More than one race
0 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants
Race (NIH/OMB)
Unknown or Not Reported
3 Participants
Race (NIH/OMB)
White
2 Participants
Sex: Female, Male
Female
1 Participants
Sex: Female, Male
Male
4 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 5
other
Total, other adverse events
0 / 5
serious
Total, serious adverse events
0 / 5

Outcome results

Primary

VAL-1: System Comfort Validation Test

User wears Point Partial(s) for a full day (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall be comfortable for a full day's use. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

Time frame: Day 1 (8 hours)

ArmMeasureValue (NUMBER)
ProsthesisVAL-1: System Comfort Validation Test100 % of participants who Passed test
Primary

VAL-2: Unilateral Function Validation Test

User flexes and extends the Point Partial(s) without using contralateral hand (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall allow the user to flex and extend finger without their second hand. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

Time frame: Day 1 (10 minutes)

ArmMeasureValue (NUMBER)
ProsthesisVAL-2: Unilateral Function Validation Test100 % of participants who Passed test
Primary

VAL-3.1: System Robustness Validation Test With 25 lb Bag

User lifts a 25 lb. bag with Point Partial(s) (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall be robust enough to operate in challenging environments. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

Time frame: Day 1 (10 minutes)

ArmMeasureValue (NUMBER)
ProsthesisVAL-3.1: System Robustness Validation Test With 25 lb Bag100 % of participants who Passed test
Primary

VAL-3.2: Robustness Validation Test With Hammer

User grips a hammer with Point Partial(s) and drives a nail into a piece of wood (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall be robust enough to operate in challenging environments. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

Time frame: Day 1 (10 minutes)

ArmMeasureValue (NUMBER)
ProsthesisVAL-3.2: Robustness Validation Test With Hammer100 % of participants who Passed test
Primary

VAL-4: Object Release Validation Test

User releases Point Partial(s) while performing an active grasp by pressing the dorsal button (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall be releasable while performing active grasps. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

Time frame: Day 1 (10 minutes)

ArmMeasureValue (NUMBER)
ProsthesisVAL-4: Object Release Validation Test100 % of participants who Passed test
Primary

VAL-5: Position Function Validation Test

User positions Point Partial(s) into each of the locking positions (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall have a smooth ratcheting mechanism. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

Time frame: Day 1 (10 minutes)

ArmMeasureValue (NUMBER)
ProsthesisVAL-5: Position Function Validation Test100 % of participants who Passed test
Primary

VAL-6: Extension Function Validation Test

User activates spring-back mechanism by using (1) contralateral hand, (2) tabletop, and (3) side or thigh throughout day without failed extension (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall have a reliable auto spring-back mechanism. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

Time frame: Day 1 (10 minutes)

ArmMeasureValue (NUMBER)
ProsthesisVAL-6: Extension Function Validation Test100 % of participants who Passed test
Primary

VAL-7: Grip Functions Validation Test

User grips a 5 lb. cylindrical smooth object using Point Partial(s) for 10 seconds (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall have high grip surface friction. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

Time frame: Day 1 (10 minutes)

ArmMeasureValue (NUMBER)
ProsthesisVAL-7: Grip Functions Validation Test100 % of participants who Passed test
Primary

VAL-8: Fine Motor Function Validation Test

User picks up 1 coin from a smooth tabletop (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall have a pronounced fingernail. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

Time frame: Day 1 (10 minutes)

ArmMeasureValue (NUMBER)
ProsthesisVAL-8: Fine Motor Function Validation Test100 % of participants who Passed test
Primary

VAL-9: Overall Function & Robustness Validation Test

User performs VAL-1 through VAL-8 (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall be robustly attached to mounting system. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

Time frame: Day 1 (8 hours)

ArmMeasureValue (NUMBER)
ProsthesisVAL-9: Overall Function & Robustness Validation Test100 % of participants who Passed test

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026