Hearing Aid Quality and Reliability Study

Hearing Aid Quality Check and Reliability Study

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT04995666
Enrollment
19
Registered
2021-08-09
Start date
2021-06-08
Completion date
2021-10-31
Last updated
2023-05-17

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Hearing Loss

Brief summary

Quality and reliability study of hearing aid. Adults with mild to severe hearing loss will wear hearing aids daily for approximately 3 months.

Interventions

Rechargeable hearing aid

Sponsors

Sonova AG
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to 90 Years
Healthy volunteers
Yes

Inclusion criteria

* Adults with mild to moderate severe sensorineural hearing loss * Adults who were able to incorporate some type of physical activity into their day (i.e walking, excercise, gardening, etc.)

Exclusion criteria

* Unable or unwilling to wear study devices during a home trial * Unable or unwilling to come to lab for study visits

Design outcomes

Primary

MeasureTime frameDescription
Electroacoustic Evaluation of Hearing Aids in Accordance With Manufacturer Specifications, Day 1Day 1 of studyTest box electroacoustic measurements of hearing aid done at full on settings must meet manufacturer product specifications per manufacturer data sheet. This includes maximum and high frequency average output sound pressure level, maximum and high frequency average acoustic gain, frequency response range, total harmonic distortion and equivalent input noise level. This is done using an Audioscan Verifit2 electroacoustic analyzer. Electroacoustic measurements listed above will also be taken with hearing aid set to user settings, as a baseline measure for all future appointments.
Electroacoustic Evaluation of Hearing Aids in Accordance With Manufacturer Specifications, Day 45Day 45 of studyTest box electroacoustic measurements (using the Audioscan Verifit 2 Electroacoustic Analyzer) of hearing aid done at user gain settings after daily use for extended period of time. Results will be compared to baseline electroacoustic measures taken on day 1 of study. A pass means that devices were within tolerance levels (per ANSI S3.22-2014) of baseline measures. A fail means that devices were not within tolerance levels of one or more baseline measures (i.e. output, gain, frequency range, total harmonic distortion, or equivalent input noise level).
Electroacoustic Evaluation of Hearing Aids in Accordance With Manufacturer Specifications, Day 90Day 90 of studyTest box electroacoustic measurements of hearing aid done at user gain settings after daily use for extended period of time. Test box electroacoustic measurements of hearing aid done at user gain settings after daily use for extended period of time. Results will be compared to baseline electroacoustic measures taken on day 1 of study. A pass means that devices were within tolerance levels (per ANSI S3.22-2014) of baseline measures. A fail means that devices were not within tolerance levels of one or more baseline measures (i.e. output, gain, frequency range, total harmonic distortion, or equivalent input noise level).

Countries

United States

Participant flow

Participants by arm

ArmCount
Hearing Aid
Participants will wear Phonak rechargeable Audeo hearing aids Phonak Audeo hearing aid: Rechargeable hearing aid
19
Total19

Baseline characteristics

CharacteristicHearing Aid
Age, Continuous70 years
Race and Ethnicity Not Collected— Participants
Region of Enrollment
United States
19 participants
Sex: Female, Male
Female
9 Participants
Sex: Female, Male
Male
10 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 19
other
Total, other adverse events
1 / 19
serious
Total, serious adverse events
0 / 19

Outcome results

Primary

Electroacoustic Evaluation of Hearing Aids in Accordance With Manufacturer Specifications, Day 1

Test box electroacoustic measurements of hearing aid done at full on settings must meet manufacturer product specifications per manufacturer data sheet. This includes maximum and high frequency average output sound pressure level, maximum and high frequency average acoustic gain, frequency response range, total harmonic distortion and equivalent input noise level. This is done using an Audioscan Verifit2 electroacoustic analyzer. Electroacoustic measurements listed above will also be taken with hearing aid set to user settings, as a baseline measure for all future appointments.

Time frame: Day 1 of study

Population: All devices dispensed to study participants on day 1 of study tested on electroacoustic verification equipment

ArmMeasureGroupValue (COUNT_OF_UNITS)
Hearing AidElectroacoustic Evaluation of Hearing Aids in Accordance With Manufacturer Specifications, Day 1Number of units that PASS manufacturer specifications38 hearing aid devices
Hearing AidElectroacoustic Evaluation of Hearing Aids in Accordance With Manufacturer Specifications, Day 1Number of units that FAIL manufacturer specifications0 hearing aid devices
Primary

Electroacoustic Evaluation of Hearing Aids in Accordance With Manufacturer Specifications, Day 45

Test box electroacoustic measurements (using the Audioscan Verifit 2 Electroacoustic Analyzer) of hearing aid done at user gain settings after daily use for extended period of time. Results will be compared to baseline electroacoustic measures taken on day 1 of study. A pass means that devices were within tolerance levels (per ANSI S3.22-2014) of baseline measures. A fail means that devices were not within tolerance levels of one or more baseline measures (i.e. output, gain, frequency range, total harmonic distortion, or equivalent input noise level).

Time frame: Day 45 of study

Population: All devices dispensed to study participants on day 1 of study tested on electroacoustic verification equipment

ArmMeasureGroupValue (COUNT_OF_UNITS)
Hearing AidElectroacoustic Evaluation of Hearing Aids in Accordance With Manufacturer Specifications, Day 45Number of devices that PASS manufacturer specifications specifications38 hearing aid devices
Hearing AidElectroacoustic Evaluation of Hearing Aids in Accordance With Manufacturer Specifications, Day 45Number of devices that FAIL manufacturer specifications0 hearing aid devices
Primary

Electroacoustic Evaluation of Hearing Aids in Accordance With Manufacturer Specifications, Day 90

Test box electroacoustic measurements of hearing aid done at user gain settings after daily use for extended period of time. Test box electroacoustic measurements of hearing aid done at user gain settings after daily use for extended period of time. Results will be compared to baseline electroacoustic measures taken on day 1 of study. A pass means that devices were within tolerance levels (per ANSI S3.22-2014) of baseline measures. A fail means that devices were not within tolerance levels of one or more baseline measures (i.e. output, gain, frequency range, total harmonic distortion, or equivalent input noise level).

Time frame: Day 90 of study

Population: All devices dispensed to study participants on day 1 of study tested on electroacoustic verification equipment. One participant lost a device, therefore total number of devices analyzed on Day 90 is 37.

ArmMeasureGroupValue (COUNT_OF_UNITS)
Hearing AidElectroacoustic Evaluation of Hearing Aids in Accordance With Manufacturer Specifications, Day 90Number of devices that PASS manufacturer specifications35 hearing aid devices
Hearing AidElectroacoustic Evaluation of Hearing Aids in Accordance With Manufacturer Specifications, Day 90Number of devices that FAIL manufacturer specifications2 hearing aid devices

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026