Early Feasibility Study Using IOPCL MAG to Improve Near Vision in ARMD Subjects

Early Feasibility Study to Evaluate the Safety and Effectiveness of the Intraoperative Pseudophakic Capsular Lens Magnifier for Age-Related Macular Degeneration (IOPCL AMD-MAG) for Secondary Implantation in the Capsular Bag to Improve Near Vision in Subjects With Age-Related Macular Degeneration After Cataract Surgery

Status
Completed
Phases
Unknown
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT04984590
Enrollment
9
Registered
2021-07-30
Start date
2022-01-21
Completion date
2024-03-20
Last updated
2025-01-31

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Age Related Macular Degeneration, Low Vision

Brief summary

This early feasibility study aims to improve near vision in subjects 55 years or older who have a clinical diagnosis of Age-Related Macular Degeneration. Subjects must have previously been implanted with either the Alcon Model SN60WF or Model SA60AT intraocular lens at least 6-months prior to receiving the IOPCL (intraocular pseudophakic capsular lens). Subject will be followed for a period of 12-months.

Detailed description

This will be a 12-month study, in which a maximum of 10 pseudophakic subjects from up to three clinical sites with a diagnosis of age-related macular degeneration will be enrolled. The IOPCL MAG (intraocular pseudophakic capsular lens magnifier) consists of a 4.5mm diameter optic. The IOPCL MAG is centered above the existing intraocular lens. The IOPCL MAG is designed to improve near vision of the AMD subject by providing magnification of 10 diopters in the central 1.8mm zone of the lens optic. The primary objective of this study is to determine the stability of the IOPCL-AMD-MAG to successfully adhere to a pseudophakic intraocular lens (PCIOL) without slippage or rotation. The secondary objective of this study is to determine if the IOPCL-AMD-MAG can improve uncorrected near vision in subjects previously implanted with the Alcon Models SN60WF or SA60AT. Outcome simulation testing will be performed prior to the subject receiving the IOPCL. To qualify for the study the subject must demonstrate an improvement in uncorrected near visual acuity of 10 letters over baseline.

Interventions

DEVICEIOPCL AMD MAG

The IOPCL MAG (intraocular pseudophakic capsular lens magnifier) consists of a 4.5 mm diameter optic. The central 1.8mm of the anterior surface has a power of 10.0 diopters. This IOPCL MAG will be implanted over an Alcon Model SN60WF or SA60AT intraocular lens to improve near vision.

Sponsors

OnPoint Vision Inc
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
TREATMENT
Masking
NONE

Intervention model description

Treatment with IOPCL AMD MAG at least 6-months after cataract surgery with Alcon Model SN60WF or SA60AT

Eligibility

Sex/Gender
ALL
Age
55 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

1. Subjects aged 55 years or older. 2. Subjects who have already had cataract surgery with an Alcon SN60WF or SA60AT monofocal intraocular lens with a lens power between 17.0 diopters and 24.0 diopters clearly evidenced by photographic documentation by one of the following: * patient medical record * clinic chart with labeling attached * surgical record with labeling attached, or * patient identification card with make, model and serial number. 3. Subjects who have already had cataract surgery at least 6 months from the planned date of the IOPCL surgery. 4. Subjects who have had a Nd: YAG capsulotomy at least 1 month prior to the planned date of the IOPCL surgery. 5. Subjects with non-neovascular dry AMD meeting the following criteria: * No change in Amsler Grid test within the past 12 months * No subretinal fluid or macular hemorrhage confirmed by optical coherence tomography (OCT) imaging. * No fundoscopic changes within the past 12 months (i.e., changes in retinal pigment epithelium). 6. Subjects with best corrected distance visual acuity from 20/80 to 20/800. 7. Subjects with a preoperative manifest refraction spherical equivalent (MRSE) of +1.0D to -1.0D. 8. Subjects with ≤1.0D of corneal cylinder determined by keratometry readings. 9. Subjects who are unsatisfied with their current near vision correction provided by either spectacles or external magnifier. 10. Subjects who demonstrate at least 10 letters of near visual acuity improvement with the simulation/tolerance test (using manifest refraction with +7.0D add) compared to conventional near vision testing (using manifest refraction with +3.0D add). 11. Subjects with a minimum endothelial cell count of 1800 cells/mm2. 12. Subjects willing to abstain from pursuing any other surgical vision-correcting procedures for the duration of the study. 13. Subjects who are willing and able to complete all required postoperative visits. 14. Subjects who are able to comprehend and sign a statement of informed consent.

Exclusion criteria

1. Subjects who have already had cataract surgery with a toric or multifocal or accommodating Intraocular Lens. 2. Subjects who have already had cataract surgery with an Alcon SN60WF or SA60AT monofocal intraocular lens with a power below 17.0 diopters and greater than 24.0 diopters. 3. Subjects with neovascular (wet) AMD. 4. Subjects who have not had an Nd: YAG capsulotomy. 5. Subjects who were treated with an IOL off-label. 6. Subjects who have more than 1.0D of corneal cylinder determined by Keratometry readings. 7. Subjects whose continuous curvilinear capsulorhexis was less than 5mm or more than 6.0 mm in size at the time of IOL surgery. 8. Subjects who had cataract surgery less than 6 months from the planned date of the IOPCL surgery. 9. Subjects with anterior capsule fibrosis and phimosis that in the opinion of the investigator may confound the outcome or increase the risk to the subject. 10. Subjects who do not gain at least 10 letters of near visual acuity with the simulation/tolerance testing (using manifest refraction with +7.0D add) compared to conventional near vision testing (using manifest refraction with +3.0D add). 11. Subjects with a concomitant retinal or choroidal disorder other than AMD. 12. Subjects with any corneal abnormality, other than regular corneal astigmatism that in the opinion of the investigator would confound the outcome(s) of the study. 13. Subjects with clinically severe corneal dystrophy (e.g., epithelial, stromal, or endothelial dystrophy). 14. Subjects with microphthalmos. 15. Subjects with a previous retinal detachment. 16. Subjects with a recurrent severe anterior or posterior segment inflammation of unknown etiology. 17. Subjects with iris neovascularization. 18. Subjects with glaucoma. 19. Subjects with advanced visual field defects (e.g., central large scotoma where magnification would not help). 20. Subjects with a fundus not visible. 21. Subjects with aniridia. 22. Subjects with advanced optic nerve atrophy. 23. Subjects with damaged or incomplete zonules. 24. Subjects with a known history of pseudoexfoliation. 25. Subjects with acute, chronic or uncontrolled systemic or ocular disease that in the opinion of the investigator would increase the operative risk or confound the outcome(s) of the study. 26. Subjects with medications that, in the opinion of the investigator, may confound the outcome or increase the risk to the subject (tamsulosin hydrochloride (Flomax)) or other medications with similar side effects (floppy iris syndrome). 27. Subjects with cognitive impairment that may interfere with the ability to understand the potentially complex considerations underlying an informed decision on trial participation and/or interfere with the ability to neuro-adapt to this device in the postoperative period. \-

Design outcomes

Primary

MeasureTime frameDescription
Tilt of the PCIOL and IOPCL MAG Complex12 MonthsMeasuring postoperative tilt of the PCIOL and IOPCL MAG complex. Tilt was defined as less than or equal to 10 degrees.
IOPCL MAG Decentration12 MonthsMeasures the decentration of the IOPCL MAG against the PCIOL. Decentration was defined as the horizontal and vertical distances greater than or equal to 0.25mm or less than or equal to 0.75mm.
Uncorrected Near Visual Acuity at 14cm Change From Baseline (Letters) (≥ 75% Achieving ≥ 10 Letters)12 MonthsProportion of eyes (≥ 75%) able to achieve improvement of 10 letters or more of uncorrected near visual acuity (at 14 cm) at 12 months from baseline

Secondary

MeasureTime frameDescription
Postoperative Best Corrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 75% Achieving ≥ 10 Letters)12 monthsProportion of eyes (≥ 75%) able to achieve improvement of 10 letters or more when comparing best corrected near visual acuity (BCNVA) (at 14 cm) at 12 months from baseline BCNVA (at 30 cm, wearing the appropriate preoperative distance refractive correction in addition to +3.0D add)
Uncorrected Near Visual Acuity at 14cm Change From Baseline (Letters) (≥ 50% Achieving ≥ 20 Letters)12 monthsProportion of eyes (≥ 50%) able to achieve improvement of 20 letters or more of uncorrected near visual acuity (at 14 cm) at 12 months from baseline
Postoperative Best Corrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 50% Achieving ≥ 20 Letters)12 monthsProportion of eyes (≥ 50%) able to achieve improvement of 20 letters or more when comparing best corrected near visual acuity (BCNVA) (at 14 cm) at 12 months from baseline BCNVA (at 30 cm, wearing the appropriate preoperative distance refractive correction in addition to +3.0D add)
Postoperative Uncorrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 75% Achieving ≥ 10 Letters)12 monthsProportion of eyes (≥ 75%) able to achieve improvement of 10 letters or more when comparing uncorrected near visual acuity (UCNVA) (at 14 cm) at 12 months from baseline BCNVA (at 30 cm, wearing the appropriate preoperative distance refractive correction in addition to +3.0D add)
Postoperative Uncorrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 50% Achieving ≥ 20 Letters)12 monthsProportion of eyes (≥ 50%) able to achieve improvement of 20 letters or more when comparing uncorrected near visual acuity (UCNVA) (at 14 cm) at 12 months from baseline BCNVA (at 30 cm, wearing the appropriate preoperative distance refractive correction in addition to +3.0D add)

Countries

United States

Participant flow

Participants by arm

ArmCount
IOPCL AMD MAG
The IOPCL MAG (intraocular pseudophakic capsular lens magnifier) consists of a 4.5 mm diameter optic. The central 1.8mm of the anterior surface has a power of 10.0 diopters. This IOPCL MAG will be implanted over an Alcon Model SN60WF or SA60AT intraocular lens to improve near vision.
9
IOPCL AMD MAG
The IOPCL MAG (intraocular pseudophakic capsular lens magnifier) consists of a 4.5 mm diameter optic. The central 1.8mm of the anterior surface has a power of 10.0 diopters. This IOPCL MAG will be implanted over an Alcon Model SN60WF or SA60AT intraocular lens to improve near vision.
9
Total18

Baseline characteristics

CharacteristicIOPCL AMD MAG
Age, Continuous83.2 Years
STANDARD_DEVIATION 5.78
Race/Ethnicity, Customized
Not of Hispanic, Latino/a, or Spanish origin
9 Participants
Race/Ethnicity, Customized
White
9 Participants
Region of Enrollment
United States
9 Participants
Sex: Female, Male
Female
1 Participants
Sex: Female, Male
Male
8 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 9
other
Total, other adverse events
1 / 9
serious
Total, serious adverse events
0 / 9

Outcome results

Primary

IOPCL MAG Decentration

Measures the decentration of the IOPCL MAG against the PCIOL. Decentration was defined as the horizontal and vertical distances greater than or equal to 0.25mm or less than or equal to 0.75mm.

Time frame: 12 Months

ArmMeasureGroupValue (COUNT_OF_UNITS)
IOPCL AMD MAGIOPCL MAG DecentrationNumber of eyes with IOPCL MAG horizontal and vertical distances < 0.25mm and >0.75mm0 Eyes
IOPCL AMD MAGIOPCL MAG DecentrationNumber of eyes with IOPCL MAG horizontal and vertical distances >= 0.25mm and <=0.75mm9 Eyes
Primary

Tilt of the PCIOL and IOPCL MAG Complex

Measuring postoperative tilt of the PCIOL and IOPCL MAG complex. Tilt was defined as less than or equal to 10 degrees.

Time frame: 12 Months

ArmMeasureGroupValue (COUNT_OF_UNITS)
IOPCL AMD MAGTilt of the PCIOL and IOPCL MAG ComplexPCIOL Tilt <= 10 degrees9 Eyes
IOPCL AMD MAGTilt of the PCIOL and IOPCL MAG ComplexIOPCL MAG Tilt <=10 degrees9 Eyes
Primary

Uncorrected Near Visual Acuity at 14cm Change From Baseline (Letters) (≥ 75% Achieving ≥ 10 Letters)

Proportion of eyes (≥ 75%) able to achieve improvement of 10 letters or more of uncorrected near visual acuity (at 14 cm) at 12 months from baseline

Time frame: 12 Months

ArmMeasureGroupValue (COUNT_OF_UNITS)
IOPCL AMD MAGUncorrected Near Visual Acuity at 14cm Change From Baseline (Letters) (≥ 75% Achieving ≥ 10 Letters)Gain of >= 20 letters (>= 4 lines)8 Eyes
IOPCL AMD MAGUncorrected Near Visual Acuity at 14cm Change From Baseline (Letters) (≥ 75% Achieving ≥ 10 Letters)Gain of >= 15 letters (>= 3 lines)9 Eyes
IOPCL AMD MAGUncorrected Near Visual Acuity at 14cm Change From Baseline (Letters) (≥ 75% Achieving ≥ 10 Letters)Gain of >= 10 letters (>= 2 lines)9 Eyes
IOPCL AMD MAGUncorrected Near Visual Acuity at 14cm Change From Baseline (Letters) (≥ 75% Achieving ≥ 10 Letters)Gain of >= 5 letters (>= 1 line)9 Eyes
IOPCL AMD MAGUncorrected Near Visual Acuity at 14cm Change From Baseline (Letters) (≥ 75% Achieving ≥ 10 Letters)No change0 Eyes
IOPCL AMD MAGUncorrected Near Visual Acuity at 14cm Change From Baseline (Letters) (≥ 75% Achieving ≥ 10 Letters)Loss of >=10 letters (>= 2 lines)0 Eyes
IOPCL AMD MAGUncorrected Near Visual Acuity at 14cm Change From Baseline (Letters) (≥ 75% Achieving ≥ 10 Letters)Loss of >= 15 letters (>=3 lines)0 Eyes
IOPCL AMD MAGUncorrected Near Visual Acuity at 14cm Change From Baseline (Letters) (≥ 75% Achieving ≥ 10 Letters)Loss of >= 5 letters (>= 1 line)0 Eyes
Secondary

Postoperative Best Corrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 50% Achieving ≥ 20 Letters)

Proportion of eyes (≥ 50%) able to achieve improvement of 20 letters or more when comparing best corrected near visual acuity (BCNVA) (at 14 cm) at 12 months from baseline BCNVA (at 30 cm, wearing the appropriate preoperative distance refractive correction in addition to +3.0D add)

Time frame: 12 months

ArmMeasureGroupValue (COUNT_OF_UNITS)
IOPCL AMD MAGPostoperative Best Corrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 50% Achieving ≥ 20 Letters)Gain of >= 20 letters (>= 4 lines)5 Eyes
IOPCL AMD MAGPostoperative Best Corrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 50% Achieving ≥ 20 Letters)Gain of >= 10 letters (>= 2 lines)9 Eyes
IOPCL AMD MAGPostoperative Best Corrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 50% Achieving ≥ 20 Letters)Gain of >= 5 letters (>= 1 line)9 Eyes
IOPCL AMD MAGPostoperative Best Corrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 50% Achieving ≥ 20 Letters)No change0 Eyes
IOPCL AMD MAGPostoperative Best Corrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 50% Achieving ≥ 20 Letters)Loss of >= 5 letters (>= 1 line)0 Eyes
IOPCL AMD MAGPostoperative Best Corrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 50% Achieving ≥ 20 Letters)Loss of >= 10 letters (>= 2 lines)0 Eyes
IOPCL AMD MAGPostoperative Best Corrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 50% Achieving ≥ 20 Letters)Loss of >= 15 letters (>= 3 lines)0 Eyes
IOPCL AMD MAGPostoperative Best Corrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 50% Achieving ≥ 20 Letters)Gain of >= 15 letters (>= 3 lines)8 Eyes
Secondary

Postoperative Best Corrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 75% Achieving ≥ 10 Letters)

Proportion of eyes (≥ 75%) able to achieve improvement of 10 letters or more when comparing best corrected near visual acuity (BCNVA) (at 14 cm) at 12 months from baseline BCNVA (at 30 cm, wearing the appropriate preoperative distance refractive correction in addition to +3.0D add)

Time frame: 12 months

ArmMeasureGroupValue (COUNT_OF_UNITS)
IOPCL AMD MAGPostoperative Best Corrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 75% Achieving ≥ 10 Letters)Gain of >= 20 letters (>= 4 lines)5 Eyes
IOPCL AMD MAGPostoperative Best Corrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 75% Achieving ≥ 10 Letters)Gain of >= 15 letters (>= 3 lines)8 Eyes
IOPCL AMD MAGPostoperative Best Corrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 75% Achieving ≥ 10 Letters)Gain of >= 10 letters (>= 2 lines)9 Eyes
IOPCL AMD MAGPostoperative Best Corrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 75% Achieving ≥ 10 Letters)Gain of >= 5 letters (>= 1 line)9 Eyes
IOPCL AMD MAGPostoperative Best Corrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 75% Achieving ≥ 10 Letters)No change0 Eyes
IOPCL AMD MAGPostoperative Best Corrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 75% Achieving ≥ 10 Letters)Loss of >= 10 letters (>= 2 lines)0 Eyes
IOPCL AMD MAGPostoperative Best Corrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 75% Achieving ≥ 10 Letters)Loss of >= 15 letters (>= 3 lines)0 Eyes
IOPCL AMD MAGPostoperative Best Corrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 75% Achieving ≥ 10 Letters)Loss of >= 5 letters (>= 1 line)0 Eyes
Secondary

Postoperative Uncorrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 50% Achieving ≥ 20 Letters)

Proportion of eyes (≥ 50%) able to achieve improvement of 20 letters or more when comparing uncorrected near visual acuity (UCNVA) (at 14 cm) at 12 months from baseline BCNVA (at 30 cm, wearing the appropriate preoperative distance refractive correction in addition to +3.0D add)

Time frame: 12 months

ArmMeasureGroupValue (COUNT_OF_UNITS)
IOPCL AMD MAGPostoperative Uncorrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 50% Achieving ≥ 20 Letters)Gain of >= 20 letters (>= 4 lines)5 Eyes
IOPCL AMD MAGPostoperative Uncorrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 50% Achieving ≥ 20 Letters)Gain of >= 15 letters (>= 3 lines)8 Eyes
IOPCL AMD MAGPostoperative Uncorrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 50% Achieving ≥ 20 Letters)Gain of >= 10 letters (>= 2 lines)9 Eyes
IOPCL AMD MAGPostoperative Uncorrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 50% Achieving ≥ 20 Letters)Gain of >= 5 letters (>= 1 line)9 Eyes
IOPCL AMD MAGPostoperative Uncorrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 50% Achieving ≥ 20 Letters)No change0 Eyes
IOPCL AMD MAGPostoperative Uncorrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 50% Achieving ≥ 20 Letters)Loss of >= 5 letters (>= 1 line)0 Eyes
IOPCL AMD MAGPostoperative Uncorrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 50% Achieving ≥ 20 Letters)Loss of >= 10 letters (>= 2 lines)0 Eyes
IOPCL AMD MAGPostoperative Uncorrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 50% Achieving ≥ 20 Letters)Loss of >= 15 letters (>= 3 lines)0 Eyes
Secondary

Postoperative Uncorrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 75% Achieving ≥ 10 Letters)

Proportion of eyes (≥ 75%) able to achieve improvement of 10 letters or more when comparing uncorrected near visual acuity (UCNVA) (at 14 cm) at 12 months from baseline BCNVA (at 30 cm, wearing the appropriate preoperative distance refractive correction in addition to +3.0D add)

Time frame: 12 months

ArmMeasureGroupValue (COUNT_OF_UNITS)
IOPCL AMD MAGPostoperative Uncorrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 75% Achieving ≥ 10 Letters)Gain of >= 20 letters (>= 4 lines)5 Eyes
IOPCL AMD MAGPostoperative Uncorrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 75% Achieving ≥ 10 Letters)Gain of >= 15 letters (>= 3 lines)8 Eyes
IOPCL AMD MAGPostoperative Uncorrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 75% Achieving ≥ 10 Letters)Gain of >= 10 letters (>= 2 lines)9 Eyes
IOPCL AMD MAGPostoperative Uncorrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 75% Achieving ≥ 10 Letters)Gain of >= 5 letters (>= 1 line)9 Eyes
IOPCL AMD MAGPostoperative Uncorrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 75% Achieving ≥ 10 Letters)No change0 Eyes
IOPCL AMD MAGPostoperative Uncorrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 75% Achieving ≥ 10 Letters)Loss of >= 5 letters (>= 1 line)0 Eyes
IOPCL AMD MAGPostoperative Uncorrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 75% Achieving ≥ 10 Letters)Loss of >= 10 letters (>= 2 lines)0 Eyes
IOPCL AMD MAGPostoperative Uncorrected Near Visual Acuity at 14cm Compared to Preoperative Best Corrected Near Visual Acuity at 30cm With +3.0D Add (≥ 75% Achieving ≥ 10 Letters)Loss of >= 15 letters (>= 3 lines)0 Eyes
Secondary

Uncorrected Near Visual Acuity at 14cm Change From Baseline (Letters) (≥ 50% Achieving ≥ 20 Letters)

Proportion of eyes (≥ 50%) able to achieve improvement of 20 letters or more of uncorrected near visual acuity (at 14 cm) at 12 months from baseline

Time frame: 12 months

ArmMeasureGroupValue (COUNT_OF_UNITS)
IOPCL AMD MAGUncorrected Near Visual Acuity at 14cm Change From Baseline (Letters) (≥ 50% Achieving ≥ 20 Letters)Gain of >= 20 letters (>= 4 lines)8 Eyes
IOPCL AMD MAGUncorrected Near Visual Acuity at 14cm Change From Baseline (Letters) (≥ 50% Achieving ≥ 20 Letters)Gain of >= 15 letters (>= 3 lines)9 Eyes
IOPCL AMD MAGUncorrected Near Visual Acuity at 14cm Change From Baseline (Letters) (≥ 50% Achieving ≥ 20 Letters)Loss of >= 10 letters (>= 2 lines)0 Eyes
IOPCL AMD MAGUncorrected Near Visual Acuity at 14cm Change From Baseline (Letters) (≥ 50% Achieving ≥ 20 Letters)Loss of >= 15 letters (>= 3 lines)0 Eyes
IOPCL AMD MAGUncorrected Near Visual Acuity at 14cm Change From Baseline (Letters) (≥ 50% Achieving ≥ 20 Letters)Gain of >= 10 letters (>= 2 lines)9 Eyes
IOPCL AMD MAGUncorrected Near Visual Acuity at 14cm Change From Baseline (Letters) (≥ 50% Achieving ≥ 20 Letters)Gain of >= 5 letters (>= 1 line)9 Eyes
IOPCL AMD MAGUncorrected Near Visual Acuity at 14cm Change From Baseline (Letters) (≥ 50% Achieving ≥ 20 Letters)No change0 Eyes
IOPCL AMD MAGUncorrected Near Visual Acuity at 14cm Change From Baseline (Letters) (≥ 50% Achieving ≥ 20 Letters)Loss of >= 5 letters (>= 1 line)0 Eyes

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026