Evaluation of Ability to Detect Bowel Gas During Laparoscopic Right Colectomy With Intracorporeal Anastomosis

Evaluation of Ability to Detect Bowel Gas During Laparoscopic Right Colectomy With Intracorporeal Anastomosis

Status
Terminated
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT04964297
Enrollment
3
Registered
2021-07-16
Start date
2021-08-09
Completion date
2024-09-16
Last updated
2025-06-06

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Crohn's Disease of Large Intestine, Malignant Neoplasm of Colon

Keywords

Colon cancer, Crohn's disease, Perf-AlertTM device, Hydrogen gas(H2), gas Methane (CH4)

Brief summary

Undetected bowel perforation is a rare but dangerous complication of laparoscopic surgery. If the injury is not detected and treated at the time of the surgical procedure, the patient can suffer severe complications, including septic shock and eventually death. The investigator's goal is to test a novel device that can detect bowel gas leakage from perforation and alert the surgeon during the operation by evaluating the gases present in the insufflated abdomen during surgery. This study will determine the ability of the device to be attached to a standard trocar during the operation and periodically draw small samples or aliquots of gas from the abdomen to evaluate the gas and accurately detect gaseous content from the bowel. Before the device can be used to detect bowel perforations, the investigators must first ensure that it can accurately detect bowel gas in an insufflated abdomen.

Detailed description

The goal of this study is to test a novel device that can detect bowel gas leakage from perforation and alert the surgeon during the operation by evaluating the gases present in the insufflated abdomen during surgery. During laparoscopic surgery, CO2 is inserted into the abdominal cavity in order to perform the operation. This is a dynamic process as insufflation is constant during the entire procedure to maintain constant pressure and compensate for any small leaks due to the insertion and retrieval of instruments. This study will determine the ability of a novel device to be attached to a standard trocar during the operation and periodically draw a small amount of gas from the abdomen to evaluate the gas and accurately detect gaseous content from the bowel. Before the device can be used to detect bowel perforations, first the investigators must ensure that it can accurately detect bowel gas in an insufflated abdomen. The Investigational device, Perf-AlertTM, consists of a 15 x 13 x 7 unit containing gas sensors and valves, a small box containing a one way, ultra-low pressure pump, user controls (buttons), and software algorithms, which control the operation of the device ( Image1). It is used in conjunction with a single-use, disposable kit consisting of sterile tubing and filters used for sample collection and transport (Image 2). The sensing unit will be connected via a sterile tube/filter set to any trocar port in use during the procedure. At specific time points during the procedure, a one-way valve will be opened and the unit's pump turned on pulling small aliquots of gas from the abdominal cavity to the sensing unit. The system's architecture is configured such that sample collection, transport, analysis, and feedback occur in a single step such that sample collection and sensor feedback occur in real-time. A standard laptop running an analytical software program is connected to the unit to record and log sensor readings. No changes to the device are anticipated during the course of the study.

Interventions

DEVICEObtaining bowel gas samples

The Perf-AlertTM prototype consists of a 15 x 13 x 7 unit containing gas sensors and valves, a small box containing a one-way, ultra-low pressure pump, user controls (buttons), and software algorithms, which control the operation of the device. It is used in conjunction with a single-use, disposable kit of sterile tubing and filters used for sample collection. The sensing unit will be connected via a sterile tube/filter set to any trocar port in use during the procedure. At specific time points during surgery, a one-way valve will be opened, and the unit's pump turned on and withdrawing a small amount of gas from the abdominal cavity. The system is configured that sample collection, transport, analysis, and feedback occur in a single step such that sample collection and sensor feedback occur in real-time. A standard laptop running an analytical software program is connected to the unit to record and log sensor readings.

Sponsors

Sentire Medical Systems
CollaboratorUNKNOWN
Weill Medical College of Cornell University
Lead SponsorOTHER

Study design

Allocation
NON_RANDOMIZED
Intervention model
SEQUENTIAL
Primary purpose
DIAGNOSTIC
Masking
NONE

Intervention model description

Prospective pilot, single-center, single-arm, open-label, phase 1 investigator-initiated study to determine device ability to detect bowel gas during laparoscopic right colectomy with intracorporeal anastomosis using Perf-AlertTM gas detection system on 20 subjects into 2 Cohorts. Cohort 1: 10 subjects with measurement of bowel gas at 8 predetermined time points during right laparoscopic colectomy as follow: 1. Initiation of surgery/laparoscopy start Insufflation 2. Abdominal exploration 3. Completion of colon mobilization 4. Colon transection 5. At Colotomy 6. At Enterotomy 7. Anastomosis completion 8. End of surgery- after re-insufflation before closure Cohort 2: 10 subjects with continuous monitoring of bowel gases through the surgery. The level of H2 and CH4 gases will be noted at the 8 predetermined time points during the continuous monitoring as well.

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Male or female ≥ 18 years of age. * Documentation of a required elective right laparoscopic hemicolectomy with intracorporeal anastomosis * Subject signed inform consent

Exclusion criteria

* Less than 18 years old * Pregnant or breastfeeding patients * Patients undergoing emergency laparotomy for perforated right colon or trauma * Patients with Intraabdominal abscess, peritonitis, or enteric fistula * Patients who are on peritoneal dialysis * Subjects do not speak English

Design outcomes

Primary

MeasureTime frameDescription
Methane (CH4) Gasduring surgery(from start of laparoscopy to the end of surgical procedure)Methane (CH4) gas measured in ppm or mg/L (parts per million or milligrams per liter (mg/L) at 8-time points
Hydrogen (H2) Gasduring surgery(from start of laparoscopy to the end of surgical procedure)Hydrogen (H2) gas measured in ppm or mg/L (parts per million or milligrams per liter (mg/L) at 8-time points

Secondary

MeasureTime frameDescription
Response Time of the Perf-AlertTM Device.during surgery(from start of laparoscopy to the end of surgical procedure)The response time of the Perf-AlertTM device measured in seconds from obtaining the sample (from opening the valve to obtain sample) to first detecting a gas presence at the device electronic sensors

Countries

United States

Participant flow

Participants by arm

ArmCount
Cohort 1: Predetermined Points Measurement
Cohort 1: Predetermined points measurement: 10 subjects with measurement of bowel gas at 8 predetermined time points during right laparoscopic colectomy as follow: 1. Initiation of surgery/laparoscopy start Insufflation 2. Abdominal exploration 3. Completion of colon mobilization 4. Colon transection 5. At Colotomy 6. At Enterotomy 7. Anastomosis completion 8. End of surgery- after re-insufflation before closure Obtaining bowel gas samples: The Perf-AlertTM prototype consists of a 15 x 13 x 7 unit containing gas sensors and valves, a small box containing a one-way, ultra-low pressure pump, user controls (buttons), and software algorithms, which control the operation of the device. It is used in conjunction with a single-use, disposable kit of sterile tubing and filters used for sample collection. The sensing unit will be connected via a sterile tube/filter set to any trocar port in use during the procedure. At specific time points during surgery, a one-way valve will be opened, and the unit's pump turned on and withdrawing a small amount of gas from the abdominal cavity. The system is configured that sample collection, transport, analysis, and feedback occur in a single step such that sample collection and sensor feedback occur in real-time. A standard laptop running an analytical software program is connected to the unit to record and log sensor readings.
3
Cohort 2: Continuous Monitoring
Cohort 2: Continuous monitoring: 10 subjects with continuous monitoring of bowel gases through the surgery. The level of H2 and CH4 gases will be noted at the 8 predetermined time points during the continuous monitoring as well. Obtaining bowel gas samples: The Perf-AlertTM prototype consists of a 15 x 13 x 7 unit containing gas sensors and valves, a small box containing a one-way, ultra-low pressure pump, user controls (buttons), and software algorithms, which control the operation of the device. It is used in conjunction with a single-use, disposable kit of sterile tubing and filters used for sample collection. The sensing unit will be connected via a sterile tube/filter set to any trocar port in use during the procedure. At specific time points during surgery, a one-way valve will be opened, and the unit's pump turned on and withdrawing a small amount of gas from the abdominal cavity. The system is configured that sample collection, transport, analysis, and feedback occur in a single step such that sample collection and sensor feedback occur in real-time. A standard laptop running an analytical software program is connected to the unit to record and log sensor readings.
0
Total3

Baseline characteristics

CharacteristicCohort 1: Predetermined Points MeasurementTotal
Age, Categorical
<=18 years
0 Participants0 Participants
Age, Categorical
>=65 years
2 Participants2 Participants
Age, Categorical
Between 18 and 65 years
1 Participants1 Participants
Age, Continuous68.3 years
STANDARD_DEVIATION 10.4
68.3 years
STANDARD_DEVIATION 10.4
Ethnicity (NIH/OMB)
Hispanic or Latino
0 Participants0 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
3 Participants3 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
0 Participants0 Participants
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants0 Participants
Race (NIH/OMB)
Asian
0 Participants0 Participants
Race (NIH/OMB)
Black or African American
1 Participants1 Participants
Race (NIH/OMB)
More than one race
0 Participants0 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants0 Participants
Race (NIH/OMB)
Unknown or Not Reported
0 Participants0 Participants
Race (NIH/OMB)
White
2 Participants2 Participants
Region of Enrollment
United States
3 participants3 participants
Sex: Female, Male
Female
0 Participants0 Participants
Sex: Female, Male
Male
3 Participants3 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
deaths
Total, all-cause mortality
0 / 30 / 0
other
Total, other adverse events
0 / 30 / 0
serious
Total, serious adverse events
0 / 30 / 0

Outcome results

Primary

Hydrogen (H2) Gas

Hydrogen (H2) gas measured in ppm or mg/L (parts per million or milligrams per liter (mg/L) at 8-time points

Time frame: during surgery(from start of laparoscopy to the end of surgical procedure)

Population: Zero participants were enrolled in Cohort 2.

ArmMeasureValue (MEAN)Dispersion
Cohort 1: Predetermined Points MeasurementHydrogen (H2) Gas20.69 ppmStandard Deviation 30.53
Primary

Methane (CH4) Gas

Methane (CH4) gas measured in ppm or mg/L (parts per million or milligrams per liter (mg/L) at 8-time points

Time frame: during surgery(from start of laparoscopy to the end of surgical procedure)

Population: Zero participants were enrolled in Cohort 2.

ArmMeasureValue (MEAN)Dispersion
Cohort 1: Predetermined Points MeasurementMethane (CH4) Gas2.62 ppmStandard Deviation 0.27
Secondary

Response Time of the Perf-AlertTM Device.

The response time of the Perf-AlertTM device measured in seconds from obtaining the sample (from opening the valve to obtain sample) to first detecting a gas presence at the device electronic sensors

Time frame: during surgery(from start of laparoscopy to the end of surgical procedure)

Population: Zero participants were enrolled in Cohort 2.

ArmMeasureValue (MEAN)Dispersion
Cohort 1: Predetermined Points MeasurementResponse Time of the Perf-AlertTM Device.12 secondsStandard Deviation 20.78