CP1110 Sound Processor Feasibility

A Pre-Marketing, Prospective, Multi-Site, Open-Label, Within-Subject, Feasibility, Interventional Study of Speech Perception With Experienced Adult Cochlear Implant Recipients Using the CP1110 Sound Processor and Compared With the CP1000 Sound Processor

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT04898673
Enrollment
21
Registered
2021-05-24
Start date
2021-10-25
Completion date
2021-12-16
Last updated
2025-02-11

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Adult Cochlear Implant Recipients

Brief summary

The clinical study aims to investigate the speech performance with the CP1110 Sound Processor, compared with the CP1000 Sound Processor, and inclusion of a noise reduction feature in the Automatic Scene Classifier..

Detailed description

The study will build on the evidence previously collected on behind-the-ear sound processors and noise reduction, with particular focus on the speech perception performance of the CP1110 Sound Processor when compared to the CP1000 Sound Processor. To assess the primary and secondary speech perception objectives, the study incorporates a within-subject repeated-measures design in which each subject will undergo in-booth speech perception testing with all combinations of hardware and signal processing settings in a sound booth. The average difference scores will indicate the performance difference for each of the paired comparisons.

Interventions

DEVICECP1110

Sound Processor

DEVICECP1000

Sound Processor

Sponsors

Avania
CollaboratorINDUSTRY
Cochlear
Lead SponsorINDUSTRY

Study design

Allocation
NON_RANDOMIZED
Intervention model
CROSSOVER
Primary purpose
TREATMENT
Masking
NONE

Intervention model description

prospective, pre-market, Multi-site, non-randomized.

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

1. Aged 18 years or older 2. Post lingually deafened 3. Implanted with the CI600 Series (CI612, CI632, CI622, CI624), CI500 Series (CI512, CI532, CI522) or Freedom Series (CI24RE(CA), CI24RE(ST), CI422) 4. At least 6 months experience with a cochlear implant. 5. At least 3 months experience with a Nucleus 6 (CP910/920), Kanso (CP950), Kanso 2 (CP1150) or Nucleus 7 (CP1000) Sound Processor 6. Able to score 30% or more at +15 SNR with CI alone on a sentence in babble test 7. Willingness to participate in and to comply with all requirements of the protocol. 8. Fluent speaker in English as determined by the investigator 9. Willing and able to provide written informed consent

Exclusion criteria

1. Additional disabilities that would prevent participation in evaluations. 2. Unrealistic expectations on the part of the subject, regarding the possible benefits, risks and limitations that are inherent to the procedures. 3. Unable or unwilling to comply with the requirements of the clinical investigation as determined by the Investigator. 4. Investigator site personnel directly affiliated with this study and/or their immediate families; immediate family is defined as a spouse, parent, child, or sibling. 5. Cochlear employees or employees of Contract Research Organisations or contractors engaged by Cochlear for the purposes of this investigation. 6. Currently participating, or participated in another interventional clinical study/trial in the past 30 days, or (if less than 30 days) the prior investigation was Cochlear sponsored and determined by the investigator to not impact clinical findings of this investigation.

Design outcomes

Primary

MeasureTime frameDescription
Speech Reception Thresholds (SRT) Assessed Via Sentence Scores in Spatially Separated Adaptive NoiseOne dayPaired difference in dB SRT between treatment and control, with their applicable noise reduction feature enabled. The speech recognition threshold (SRT) examines at which level the participant can repeat 50% of the speech material correctly. These scores may be negative or positive numeric values. Lower dB SRT scores indicate a better outcome.

Secondary

MeasureTime frameDescription
Percentage of Correct Words Assessed by CNC (Consonant Nucleus Consonant) Monosyllabic Word Scores in QuietOne dayPaired differences in percentage CNC Words correct in quiet (50 dB) between treatment and control, with various configurations of signal processing settings being enabled or disabled for each measure. Word recognition is a score of the number of words correctly repeated, expressed as a percentage. Higher percentage scores indicate a better outcome. Secondary Endpoint 1: CNC Words correct in quiet (50 dB) with both CP1000 and CP1110 sound processors using Signal-to-Noise Ratio - Noise Reduction (SNR-NR) turned on.
Percentage of Correct Assessed by CNC (Consonant Nucleus Consonant) Monosyllabic Word Scores in QuietOne dayPaired differences in percentage CNC Words correct in quiet (50 dB) between treatment and control, with various configurations of signal processing settings being enabled or disabled for each measure. Word recognition is a score of the number of words correctly repeated, expressed as a percentage. Higher percentage scores indicate a better outcome. CNC Words in Quiet Score using both CP1110 and CP1000 sound processors Signal-to-Noise Ratio - Noise Reduction (SNR-NR) off with expander on.

Countries

Australia

Participant flow

Recruitment details

21 subjects were recruited.

Participants by arm

ArmCount
All Study Participants
Baseline characteristics are combined across the 2 arms.
20
Total20

Withdrawals & dropouts

PeriodReasonFG000FG001
Overall StudyScreenfail10

Baseline characteristics

CharacteristicAll Study Participants
Age, Customized64.5 years
Race and Ethnicity Not Collected— Participants
Region of Enrollment
Australia
20 participants
Sex: Female, Male
Female
13 Participants
Sex: Female, Male
Male
7 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
deaths
Total, all-cause mortality
0 / 200 / 20
other
Total, other adverse events
0 / 200 / 20
serious
Total, serious adverse events
0 / 200 / 20

Outcome results

Primary

Speech Reception Thresholds (SRT) Assessed Via Sentence Scores in Spatially Separated Adaptive Noise

Paired difference in dB SRT between treatment and control, with their applicable noise reduction feature enabled. The speech recognition threshold (SRT) examines at which level the participant can repeat 50% of the speech material correctly. These scores may be negative or positive numeric values. Lower dB SRT scores indicate a better outcome.

Time frame: One day

ArmMeasureValue (MEAN)
CP1000Speech Reception Thresholds (SRT) Assessed Via Sentence Scores in Spatially Separated Adaptive Noise-6.07 decibels
CP1110Speech Reception Thresholds (SRT) Assessed Via Sentence Scores in Spatially Separated Adaptive Noise-11.30 decibels
p-value: <0.00195% CI: [-6.08, -4.37]ANOVA
Secondary

Percentage of Correct Assessed by CNC (Consonant Nucleus Consonant) Monosyllabic Word Scores in Quiet

Paired differences in percentage CNC Words correct in quiet (50 dB) between treatment and control, with various configurations of signal processing settings being enabled or disabled for each measure. Word recognition is a score of the number of words correctly repeated, expressed as a percentage. Higher percentage scores indicate a better outcome. CNC Words in Quiet Score using both CP1110 and CP1000 sound processors Signal-to-Noise Ratio - Noise Reduction (SNR-NR) off with expander on.

Time frame: One day

ArmMeasureValue (MEAN)
CP1000Percentage of Correct Assessed by CNC (Consonant Nucleus Consonant) Monosyllabic Word Scores in Quiet60.90 percentage of correct words
CP1110Percentage of Correct Assessed by CNC (Consonant Nucleus Consonant) Monosyllabic Word Scores in Quiet57.15 percentage of correct words
p-value: 0.02695% CI: [-7, -0.5]ANOVA
Secondary

Percentage of Correct Words Assessed by CNC (Consonant Nucleus Consonant) Monosyllabic Word Scores in Quiet

Paired differences in percentage CNC Words correct in quiet (50 dB) between treatment and control, with various configurations of signal processing settings being enabled or disabled for each measure. Word recognition is a score of the number of words correctly repeated, expressed as a percentage. Higher percentage scores indicate a better outcome. Secondary Endpoint 1: CNC Words correct in quiet (50 dB) with both CP1000 and CP1110 sound processors using Signal-to-Noise Ratio - Noise Reduction (SNR-NR) turned on.

Time frame: One day

ArmMeasureValue (MEAN)
CP1000Percentage of Correct Words Assessed by CNC (Consonant Nucleus Consonant) Monosyllabic Word Scores in Quiet57.5 percentage of correct words
CP1110Percentage of Correct Words Assessed by CNC (Consonant Nucleus Consonant) Monosyllabic Word Scores in Quiet58.10 percentage of correct words
p-value: 0.6795% CI: [-2.31, 3.51]ANOVA
Secondary

Percentage of Correct Words Assessed by CNC (Consonant Nucleus Consonant) Monosyllabic Word Scores in Quiet

Paired differences in percentage CNC Words correct in quiet (50 dB) between treatment and control, with various configurations of signal processing settings being enabled or disabled for each measure. Word recognition is a score of the number of words correctly repeated, expressed as a percentage. Higher percentage scores indicate a better outcome. CNC Words in Quiet Score of CP1110 sound processor ForwardFocus ON (Moderate) vs CP1000 sound processor Signal-to-Noise Ration-Noise Reduction) On.

Time frame: One Day

ArmMeasureValue (MEAN)
CP1000Percentage of Correct Words Assessed by CNC (Consonant Nucleus Consonant) Monosyllabic Word Scores in Quiet57.50 percentage of correct words
CP1110Percentage of Correct Words Assessed by CNC (Consonant Nucleus Consonant) Monosyllabic Word Scores in Quiet56.35 percentage of correct words
p-value: 0.41195% CI: [-4.02, 1.72]ANOVA