Investigation of a Supporting Ostomy Product Intended for Leakage Detection

Investigation of a Supporting Ostomy Product Intended for Leakage Detection

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT04894084
Enrollment
25
Registered
2021-05-20
Start date
2021-04-29
Completion date
2021-07-14
Last updated
2023-09-22

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Stoma Ileostomy

Brief summary

People with intestinal stomas (especially an ileostomy) can have, despite development of better ostomy products, problems with leakage which influence their quality of life negatively. To overcome this, Coloplast has developed a new supporting ostomy product, which has an adhesive sensor layer that should be place underneath the baseplate normally used by the subject. Single arm design. Open-labelled Study duration: 21 days + 3 days.

Detailed description

Test product is a Digital Leakage Notification System. It consist of a sensor layer (worn beneath the baseplate), a transmitter attached to the sensor layer, a charger for charging of transmitter and an app, to be installed on a smartphone, which notifies the user of the status of their baseplate such as everything looks okay or there is a possible leakage. Subjects are asked to use test product for 21 days. n=15 were using a phone with Android operating system and n=10 a phone with iOS operating system.

Interventions

DEVICECP321

The primary objective is to evaluate the system performance of the test product. This is done by asking subjects to answer questions throughout the study

Sponsors

Coloplast A/S
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
SUPPORTIVE_CARE
Masking
NONE

Intervention model description

Single arm, open label study. Subjects are allocated to use test product for 3 weeks. Subjects will complete questionnaires at baseline (V1) and again after 3 weeks use of test product (V2). Comparative tests included: Baseline vs. follow-up measures.

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

1. Have given written informed consent 2. Be at least 18 years of age and have full legal capacity 3. Have had a stoma for more than three months 4. Have intact skin on the area used in the evaluation meaning no broken skin and only minor discoloration of the skin (assessed by investigator) 5. Be able to use one of the five sensor layer sizes (i.e. Ø40, Ø50, Ø60, Ø70, Ø80 mm) 6. Ileo- or colostomists with liquid output (Bristol scale type 6-7). 7. Currently using a Coloplast product (1pc/2pc Flat/Convex/Concave) from e.g Assura/SenSura/SenSura Mio 8. Have self-reported problems with leakage\* (three times within 14 days) 9. Have worry of leakage 'to some, high or very high degree' 10. Is familiar with the use of a smartphone \*Leakage: Leakage is defined as output from the stoma on the backside of the baseplate (underneath the baseplate)

Exclusion criteria

1. Currently receiving or have within the past 2 month received radio- and/or chemotherapy 2. Currently receiving or have within the past month received topical steroid treatment in the peristomal skin area or systemic steroid (tablet/injection) treatment. 3. Are pregnant or breastfeeding 4. Participating in other interventional clinical investigations or have previously participated in this investigation. Exception: Participation in other Coloplast sponsored clinical investigations is accepted under the circumstances that the subject has paused the activities in the investigation and are otherwise complying with the inclusion and

Design outcomes

Primary

MeasureTime frameDescription
System Performance Accuracy of Pictures21 daysThe primary endpoint is system performance accuracy: * True Negative: No leakage observed on picture and no leakage detected by system * True Positive: Leakage observed in picture and leakage detected by system * False Negative: Leakage observed in picture, but no leakage detected by system * False Positive: No leakage observed in picture, but leakage detected by system

Secondary

MeasureTime frameDescription
Experience of Leakage Onto Clothes Compared to Only Using Usual ProductV2 (follow-up, Day 21)Subject own interpretation, Q: to which degree have you experienced leakage onto your clothes with test product? (5-point scale: very low degree/not at all, low degree, some degree, high degree, very high degree).
Reliability of NotificationsV2 (follow-up, Day 21)Subjects are asked to rate, Q: To which degree did you find notifications from the app reliable and trustworthy (5-point scale; very low degree, low degree, some degree, high degree, very high degree)
Ability to Move With Test ProductV2 (follow-up, Day 21)Q: Compared to your usual product, how was your freedom to move with test product? Much better - better - Same - Worse - Much worse
Subjects Worry of Leakage Before and After Use of Test ProductAt V1 (baseline) and V2 (follow-up, Day 21)Q: Think back on the last 3 weeks - to which degree have you worried about leakage? Very high degree - High degree - Some degree - Low degree - Very low degree/Not at all
Confidence to Avoid Full Leakage With Test Product Compared to Only Using Usual ProductV2 (follow-up, Day 21)Q: Compared to your usual product, did you feel confident that you could avoid leakage onto clothes? Yes -No
Confidence in Daily Life Compared to Only Using Usual ProductV2 (follow-up, Day 21)Q: Compared to your usual product, to which degree did the test product give you confidence in daily life? Much higher degree- Higher degree- Same degree - Lower degree- Much lower degree
Feeling of Control With Test Product Compared to Only Using Usual ProductV2 (follow-up, Day 21)Q: Compared to your usual product, did test product give you a feeling of leakage control? Yes, higher leakage control No, lower leakage control
Control of Leakage Progression With Test Product Compared to Only Using Usual ProductV2 (follow-up, Day 21)Q: Compared to your usual product, did the test product give you a feeling of control of leakage progression? Yes, higher control of leakage progression No, lower control of leakage progression
Feeling of Security With Test Product Compared to Only Using Usual ProductV2 (follow-up, Day 21)Q: Compared to your usual product, did you feel more or less secure with test product?
Frequency of Leakage Onto Clothes Before and After Use of Test ProductAt V1 (baseline) and V2 (follow-up, Day 21)Subjects are asked to recall numbers of leakage events outside baseplate (and/or onto clothes) the past 3 weeks (at baseline and follow-up). Subjects shall give a specific frequency.
Evaluation of Sleep With Test Product Compared to Only Using Usual ProductV2 (follow-up, Day 21)Q: Compared with your usual product, did the test product affect your sleep? Yes, to the better Yes, to the worse No, the same
Confidence at Night With Test Product Compared to Only Using Usual ProductV2 (follow-up, Day 21)Compared with your usual product, to which degree did the test product give you confidence at night? Much higher - Higher - Same - Lower - Much lower
Evaluation of Users' PreferenceV2 (follow-up, Day 21)Q: Which product solution do you prefer? Own product - Own product with test product
Reason for PreferenceV2 (follow-up, Day 21)Q: What is your reason for preference? (Please tick all that apply) * Body fit * Ability to bend and stretch * Feeling of security * Feeling of confidence * Less fear of leakage onto clothes * Discreet to wear * Discreetness in leakage control * Better sleep * Less embarrassment * Other
Evaluation of Users' Peace of Mind With Test ProductV2 (follow-up, Day 21)Q: To which degree did the test product give you peace of mind? Very high degree- High degree- Low degree- Very low degree
To Evaluate Users' Assessment of QoL With Test Product Compared to Usual ProductV2 (follow-up, Day 21)Overall, do you think the test product can improve your feeling of control and confidence to avoid leakage onto clothes and therefore improve your quality of life? Yes - No
Adhesive Residue to SkinV2 (follow-up, Day 21)To which degree did you experience adhesive residue to the skin in the area around the transmitter? Very low degree/Not at all - Low degree - Some degree - High degree -Very high degree
Adhesive Residue to TransmitterV2 (follow-up, Day 21)To which degree did you experience adhesive residue on the backside of the transmitter? Very low degree/Not at all - Low degree - Some degree - High degree -Very high degree
Evaluation of Confidence to Increase Social Activities With Test Product Compared to Only Using Usual ProductV2 (follow-up, Day 21)Q: Compared with your usual product, did the test product affect your confidence to engage in social activities such as seeing friends, and do physical activities? I felt more confident - I felt less confident

Countries

Denmark

Participant flow

Participants by arm

ArmCount
Test Arm, Intention to Treat
All subjects enrolled into the study (n=25) who have at least one measure on any of the outcomes measured.
25
Total25

Baseline characteristics

CharacteristicTest Arm, Intention to Treat
Age, Continuous56 years
Race and Ethnicity Not Collected— Participants
Region of Enrollment
Denmark
25 Participants
Sex: Female, Male
Female
10 Participants
Sex: Female, Male
Male
15 Participants
Type of stoma
Colostomy
10 Participants
Type of stoma
Ileostomy
15 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 25
other
Total, other adverse events
2 / 25
serious
Total, serious adverse events
0 / 25

Outcome results

Primary

System Performance Accuracy of Pictures

The primary endpoint is system performance accuracy: * True Negative: No leakage observed on picture and no leakage detected by system * True Positive: Leakage observed in picture and leakage detected by system * False Negative: Leakage observed in picture, but no leakage detected by system * False Positive: No leakage observed in picture, but leakage detected by system

Time frame: 21 days

ArmMeasureCategoryValue (COUNT_OF_UNITS)
Test Arm, Intention to TreatSystem Performance Accuracy of PicturesTrue positive122 Pictures
Test Arm, Intention to TreatSystem Performance Accuracy of PicturesTrue negative126 Pictures
Test Arm, Intention to TreatSystem Performance Accuracy of PicturesFalse positive24 Pictures
Test Arm, Intention to TreatSystem Performance Accuracy of PicturesFalse negative45 Pictures
Secondary

Ability to Move With Test Product

Q: Compared to your usual product, how was your freedom to move with test product? Much better - better - Same - Worse - Much worse

Time frame: V2 (follow-up, Day 21)

ArmMeasureCategoryValue (COUNT_OF_PARTICIPANTS)
Test Arm, Intention to TreatAbility to Move With Test ProductMuch worse0 Participants
Test Arm, Intention to TreatAbility to Move With Test ProductWorse1 Participants
Test Arm, Intention to TreatAbility to Move With Test ProductSame16 Participants
Test Arm, Intention to TreatAbility to Move With Test ProductBetter4 Participants
Test Arm, Intention to TreatAbility to Move With Test ProductMuch better2 Participants
Comparison: For secondary endpoints an exact test in the binomial distribution a 95% confidence interval was estimated for the proportion. For ability to move with test product it was tested if the proportion evaluating same or better was significantly different from 50% when using a 5% test level. The answers of 'same, better or much better' was grouped against the answers of 'worse or much worse' ability to move.p-value: <0.000195% CI: [78, 100]Exact test in the binomial distribution
Secondary

Adhesive Residue to Skin

To which degree did you experience adhesive residue to the skin in the area around the transmitter? Very low degree/Not at all - Low degree - Some degree - High degree -Very high degree

Time frame: V2 (follow-up, Day 21)

ArmMeasureCategoryValue (COUNT_OF_PARTICIPANTS)
Test Arm, Intention to TreatAdhesive Residue to SkinVery low degree/Not at all6 Participants
Test Arm, Intention to TreatAdhesive Residue to SkinLow degree7 Participants
Test Arm, Intention to TreatAdhesive Residue to SkinSome degree7 Participants
Test Arm, Intention to TreatAdhesive Residue to SkinHigh degree2 Participants
Test Arm, Intention to TreatAdhesive Residue to SkinVery high degree1 Participants
Secondary

Adhesive Residue to Transmitter

To which degree did you experience adhesive residue on the backside of the transmitter? Very low degree/Not at all - Low degree - Some degree - High degree -Very high degree

Time frame: V2 (follow-up, Day 21)

ArmMeasureCategoryValue (COUNT_OF_PARTICIPANTS)
Test Arm, Intention to TreatAdhesive Residue to TransmitterVery low degree/Not at all9 Participants
Test Arm, Intention to TreatAdhesive Residue to TransmitterLow degree5 Participants
Test Arm, Intention to TreatAdhesive Residue to TransmitterSome degree9 Participants
Test Arm, Intention to TreatAdhesive Residue to TransmitterHigh degree0 Participants
Test Arm, Intention to TreatAdhesive Residue to TransmitterVery high degree0 Participants
Secondary

Confidence at Night With Test Product Compared to Only Using Usual Product

Compared with your usual product, to which degree did the test product give you confidence at night? Much higher - Higher - Same - Lower - Much lower

Time frame: V2 (follow-up, Day 21)

ArmMeasureCategoryValue (COUNT_OF_PARTICIPANTS)
Test Arm, Intention to TreatConfidence at Night With Test Product Compared to Only Using Usual ProductMuch lower degree0 Participants
Test Arm, Intention to TreatConfidence at Night With Test Product Compared to Only Using Usual ProductLower degree1 Participants
Test Arm, Intention to TreatConfidence at Night With Test Product Compared to Only Using Usual ProductSame degree11 Participants
Test Arm, Intention to TreatConfidence at Night With Test Product Compared to Only Using Usual ProductHigher degree7 Participants
Test Arm, Intention to TreatConfidence at Night With Test Product Compared to Only Using Usual ProductMuch higher degree4 Participants
p-value: 195% CI: [27, 69]Exact test in the binomial distribution
Secondary

Confidence in Daily Life Compared to Only Using Usual Product

Q: Compared to your usual product, to which degree did the test product give you confidence in daily life? Much higher degree- Higher degree- Same degree - Lower degree- Much lower degree

Time frame: V2 (follow-up, Day 21)

ArmMeasureCategoryValue (COUNT_OF_PARTICIPANTS)
Test Arm, Intention to TreatConfidence in Daily Life Compared to Only Using Usual ProductMuch lower degree1 Participants
Test Arm, Intention to TreatConfidence in Daily Life Compared to Only Using Usual ProductLower degree2 Participants
Test Arm, Intention to TreatConfidence in Daily Life Compared to Only Using Usual ProductSame degree5 Participants
Test Arm, Intention to TreatConfidence in Daily Life Compared to Only Using Usual ProductHigher degree10 Participants
Test Arm, Intention to TreatConfidence in Daily Life Compared to Only Using Usual ProductMuch higher degree5 Participants
p-value: 0.2195% CI: [43, 84]Exact test in the binomial distribution
Secondary

Confidence to Avoid Full Leakage With Test Product Compared to Only Using Usual Product

Q: Compared to your usual product, did you feel confident that you could avoid leakage onto clothes? Yes -No

Time frame: V2 (follow-up, Day 21)

ArmMeasureCategoryValue (COUNT_OF_PARTICIPANTS)
Test Arm, Intention to TreatConfidence to Avoid Full Leakage With Test Product Compared to Only Using Usual ProductNo5 Participants
Test Arm, Intention to TreatConfidence to Avoid Full Leakage With Test Product Compared to Only Using Usual ProductYes18 Participants
p-value: 0.010695% CI: [56, 93]Exact test in the binomial distribution
Secondary

Control of Leakage Progression With Test Product Compared to Only Using Usual Product

Q: Compared to your usual product, did the test product give you a feeling of control of leakage progression? Yes, higher control of leakage progression No, lower control of leakage progression

Time frame: V2 (follow-up, Day 21)

ArmMeasureCategoryValue (COUNT_OF_PARTICIPANTS)
Test Arm, Intention to TreatControl of Leakage Progression With Test Product Compared to Only Using Usual ProductYes, higher control of leakage progression16 Participants
Test Arm, Intention to TreatControl of Leakage Progression With Test Product Compared to Only Using Usual ProductNo, lower control of leakage progression7 Participants
p-value: 0.09395% CI: [47, 87]Exact test in the binomial distribution
Secondary

Evaluation of Confidence to Increase Social Activities With Test Product Compared to Only Using Usual Product

Q: Compared with your usual product, did the test product affect your confidence to engage in social activities such as seeing friends, and do physical activities? I felt more confident - I felt less confident

Time frame: V2 (follow-up, Day 21)

ArmMeasureCategoryValue (COUNT_OF_PARTICIPANTS)
Test Arm, Intention to TreatEvaluation of Confidence to Increase Social Activities With Test Product Compared to Only Using Usual ProductI felt less confident2 Participants
Test Arm, Intention to TreatEvaluation of Confidence to Increase Social Activities With Test Product Compared to Only Using Usual ProductI felt more confident21 Participants
p-value: <0.000195% CI: [72, 99]Exact test in the binomial distribution
Secondary

Evaluation of Sleep With Test Product Compared to Only Using Usual Product

Q: Compared with your usual product, did the test product affect your sleep? Yes, to the better Yes, to the worse No, the same

Time frame: V2 (follow-up, Day 21)

ArmMeasureCategoryValue (COUNT_OF_PARTICIPANTS)
Test Arm, Intention to TreatEvaluation of Sleep With Test Product Compared to Only Using Usual ProductYes, to the better8 Participants
Test Arm, Intention to TreatEvaluation of Sleep With Test Product Compared to Only Using Usual ProductYes, to the worse0 Participants
Test Arm, Intention to TreatEvaluation of Sleep With Test Product Compared to Only Using Usual ProductNo, the same15 Participants
p-value: 0.2195% CI: [16, 57]Exact test in the binomial distribution
Secondary

Evaluation of Users' Peace of Mind With Test Product

Q: To which degree did the test product give you peace of mind? Very high degree- High degree- Low degree- Very low degree

Time frame: V2 (follow-up, Day 21)

ArmMeasureCategoryValue (COUNT_OF_PARTICIPANTS)
Test Arm, Intention to TreatEvaluation of Users' Peace of Mind With Test ProductVery high degree6 Participants
Test Arm, Intention to TreatEvaluation of Users' Peace of Mind With Test ProductHigh degree10 Participants
Test Arm, Intention to TreatEvaluation of Users' Peace of Mind With Test ProductLow degree5 Participants
Test Arm, Intention to TreatEvaluation of Users' Peace of Mind With Test ProductVery low degree2 Participants
p-value: 0.093195% CI: [47, 87]Exact test in the binomial distribution
Secondary

Evaluation of Users' Preference

Q: Which product solution do you prefer? Own product - Own product with test product

Time frame: V2 (follow-up, Day 21)

ArmMeasureCategoryValue (COUNT_OF_PARTICIPANTS)
Test Arm, Intention to TreatEvaluation of Users' PreferenceOwn product10 Participants
Test Arm, Intention to TreatEvaluation of Users' PreferenceOwn product with test product13 Participants
p-value: 0.67895% CI: [34, 77]Exact test in the binomial distribution
Secondary

Experience of Leakage Onto Clothes Compared to Only Using Usual Product

Subject own interpretation, Q: to which degree have you experienced leakage onto your clothes with test product? (5-point scale: very low degree/not at all, low degree, some degree, high degree, very high degree).

Time frame: V2 (follow-up, Day 21)

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Test Arm, Intention to TreatExperience of Leakage Onto Clothes Compared to Only Using Usual ProductVery high degree0 Participants
Test Arm, Intention to TreatExperience of Leakage Onto Clothes Compared to Only Using Usual ProductHigh degree0 Participants
Test Arm, Intention to TreatExperience of Leakage Onto Clothes Compared to Only Using Usual ProductSome degree2 Participants
Test Arm, Intention to TreatExperience of Leakage Onto Clothes Compared to Only Using Usual ProductLow degree3 Participants
Test Arm, Intention to TreatExperience of Leakage Onto Clothes Compared to Only Using Usual ProductVery low degree / Not at all18 Participants
Comparison: For secondary endpoints where the possible answer is Yes or No, the proportion of subjects answering Yes was calculated. By use of an exact test in the binomial distribution a 95% confidence interval was estimated for the proportion. It was tested if the proportion was significantly different from 50% when using a 5% test level. For questions answered on a 5-point scale the answers was grouped in 2 (e.g. higher or much higher against the rest of the answers).p-value: <0.000195% CI: [72, 99]Exact test in the binomial distribution
Secondary

Feeling of Control With Test Product Compared to Only Using Usual Product

Q: Compared to your usual product, did test product give you a feeling of leakage control? Yes, higher leakage control No, lower leakage control

Time frame: V2 (follow-up, Day 21)

ArmMeasureCategoryValue (COUNT_OF_PARTICIPANTS)
Test Arm, Intention to TreatFeeling of Control With Test Product Compared to Only Using Usual ProductYes, higher leakage control18 Participants
Test Arm, Intention to TreatFeeling of Control With Test Product Compared to Only Using Usual ProductNo, lower leakage control5 Participants
p-value: 0.010695% CI: [56, 93]Exact test in the binomial distribution
Secondary

Feeling of Security With Test Product Compared to Only Using Usual Product

Q: Compared to your usual product, did you feel more or less secure with test product?

Time frame: V2 (follow-up, Day 21)

ArmMeasureCategoryValue (COUNT_OF_PARTICIPANTS)
Test Arm, Intention to TreatFeeling of Security With Test Product Compared to Only Using Usual ProductI felt less secure1 Participants
Test Arm, Intention to TreatFeeling of Security With Test Product Compared to Only Using Usual ProductI felt more secure22 Participants
p-value: <0.000195% CI: [78, 100]Exact test in the binomial distribution
Secondary

Frequency of Leakage Onto Clothes Before and After Use of Test Product

Subjects are asked to recall numbers of leakage events outside baseplate (and/or onto clothes) the past 3 weeks (at baseline and follow-up). Subjects shall give a specific frequency.

Time frame: At V1 (baseline) and V2 (follow-up, Day 21)

ArmMeasureGroupValue (LEAST_SQUARES_MEAN)
Test Arm, Intention to TreatFrequency of Leakage Onto Clothes Before and After Use of Test ProductV12.84 Leakage episodes onto clothes
Test Arm, Intention to TreatFrequency of Leakage Onto Clothes Before and After Use of Test ProductV20.51 Leakage episodes onto clothes
p-value: <0.00195% CI: [0.07, 0.49]Poisson loglinear model
Secondary

Reason for Preference

Q: What is your reason for preference? (Please tick all that apply) * Body fit * Ability to bend and stretch * Feeling of security * Feeling of confidence * Less fear of leakage onto clothes * Discreet to wear * Discreetness in leakage control * Better sleep * Less embarrassment * Other

Time frame: V2 (follow-up, Day 21)

Population: The 13 subjects with a preference for test product were asked for the reason(s) for this preference and the the 10 subjects with a preference for own product were asked for the reason(s) for this preference.

ArmMeasureGroupValue (NUMBER)
Test Arm, Intention to TreatReason for PreferencePreference for test product : Body fit2 Participants
Test Arm, Intention to TreatReason for PreferencePreference for test product : Ability to bend and stretch1 Participants
Test Arm, Intention to TreatReason for PreferencePreference for test product : Feeling of security10 Participants
Test Arm, Intention to TreatReason for PreferencePreference for test product : Feeling of confidence2 Participants
Test Arm, Intention to TreatReason for PreferencePreference for test product : Less fear of leakage onto clothes9 Participants
Test Arm, Intention to TreatReason for PreferencePreference for test product : Discreet to wear2 Participants
Test Arm, Intention to TreatReason for PreferencePreference for test product : Discreetness in leakage control10 Participants
Test Arm, Intention to TreatReason for PreferencePreference for test product : Better sleep5 Participants
Test Arm, Intention to TreatReason for PreferencePreference for test product : Less embarrassment1 Participants
Test Arm, Intention to TreatReason for PreferencePreference for test product : Other0 Participants
Test Arm, Intention to TreatReason for PreferencePreference for own product : Body fit5 Participants
Test Arm, Intention to TreatReason for PreferencePreference for own product : Ability to bend and stretch3 Participants
Test Arm, Intention to TreatReason for PreferencePreference for own product : Feeling of security3 Participants
Test Arm, Intention to TreatReason for PreferencePreference for own product : Feeling of confidence0 Participants
Test Arm, Intention to TreatReason for PreferencePreference for own product : Less fear of leakage onto clothes1 Participants
Test Arm, Intention to TreatReason for PreferencePreference for own product : Discreet to wear3 Participants
Test Arm, Intention to TreatReason for PreferencePreference for own product : Discreetness in leakage control4 Participants
Test Arm, Intention to TreatReason for PreferencePreference for own product : Better sleep0 Participants
Test Arm, Intention to TreatReason for PreferencePreference for own product : Less embarrassment0 Participants
Test Arm, Intention to TreatReason for PreferencePreference for own product : Other1 Participants
Secondary

Reliability of Notifications

Subjects are asked to rate, Q: To which degree did you find notifications from the app reliable and trustworthy (5-point scale; very low degree, low degree, some degree, high degree, very high degree)

Time frame: V2 (follow-up, Day 21)

ArmMeasureCategoryValue (COUNT_OF_PARTICIPANTS)
Test Arm, Intention to TreatReliability of NotificationsVery high degree5 Participants
Test Arm, Intention to TreatReliability of NotificationsHigh degree7 Participants
Test Arm, Intention to TreatReliability of NotificationsSome degree6 Participants
Test Arm, Intention to TreatReliability of NotificationsLow degree3 Participants
Test Arm, Intention to TreatReliability of NotificationsVery low degree / Not at all2 Participants
Comparison: For secondary endpoints an exact test in the binomial distribution a 95% confidence interval was estimated for the proportion. It was tested if the proportion was significantly different from 50% when using a 5% test level. For reliability answered on a 5-point scale the answers some, high, or very high degree was considered acceptable and grouped against answers of 'low or very low' degree of reliability.p-value: 0.010695% CI: [56, 93]Exact test in the binomial distribution
Secondary

Subjects Worry of Leakage Before and After Use of Test Product

Q: Think back on the last 3 weeks - to which degree have you worried about leakage? Very high degree - High degree - Some degree - Low degree - Very low degree/Not at all

Time frame: At V1 (baseline) and V2 (follow-up, Day 21)

Population: Due to technical reasons, data from the questionnaire at V2 for two subjects could not be retrieved from the phones and therefore have been omitted from statistical analyses. Data from the two subjects provided at V1 were included, whenever possible. Therefore, the total number of subjects in the statistical analyses at V1 was n=25, while at V2 the number was n=23.

ArmMeasureGroupCategoryValue (COUNT_OF_PARTICIPANTS)
Test Arm, Intention to TreatSubjects Worry of Leakage Before and After Use of Test ProductV1Very high degree1 Participants
Test Arm, Intention to TreatSubjects Worry of Leakage Before and After Use of Test ProductV1High degree11 Participants
Test Arm, Intention to TreatSubjects Worry of Leakage Before and After Use of Test ProductV1Some degree11 Participants
Test Arm, Intention to TreatSubjects Worry of Leakage Before and After Use of Test ProductV1Low degree2 Participants
Test Arm, Intention to TreatSubjects Worry of Leakage Before and After Use of Test ProductV1Very low degree / Not at all0 Participants
Test Arm, Intention to TreatSubjects Worry of Leakage Before and After Use of Test ProductV2Very high degree0 Participants
Test Arm, Intention to TreatSubjects Worry of Leakage Before and After Use of Test ProductV2High degree1 Participants
Test Arm, Intention to TreatSubjects Worry of Leakage Before and After Use of Test ProductV2Some degree6 Participants
Test Arm, Intention to TreatSubjects Worry of Leakage Before and After Use of Test ProductV2Low degree8 Participants
Test Arm, Intention to TreatSubjects Worry of Leakage Before and After Use of Test ProductV2Very low degree / Not at all8 Participants
p-value: <0.001Proportional odds ratio model
Secondary

To Evaluate Users' Assessment of QoL With Test Product Compared to Usual Product

Overall, do you think the test product can improve your feeling of control and confidence to avoid leakage onto clothes and therefore improve your quality of life? Yes - No

Time frame: V2 (follow-up, Day 21)

ArmMeasureCategoryValue (COUNT_OF_PARTICIPANTS)
Test Arm, Intention to TreatTo Evaluate Users' Assessment of QoL With Test Product Compared to Usual ProductYes20 Participants
Test Arm, Intention to TreatTo Evaluate Users' Assessment of QoL With Test Product Compared to Usual ProductNo3 Participants
p-value: 0.000595% CI: [66, 97]Exact test in the binomial distribution