To Evaluate Safety and Effectiveness of Transcatheter Aortic Valve System in Patients With Severe Aortic Stenosis

A Prospective, Multi-center, Single-arm Clinical Trial to Evaluate the Safety and Effectiveness of Transcatheter Aortic Valve System in Patients With Severe Aortic Stenosis With or Without Moderate or Lower Degree Regurgitation

Status
Recruiting
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT04893603
Enrollment
120
Registered
2021-05-19
Start date
2020-12-16
Completion date
2026-12-31
Last updated
2021-05-19

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Aortic Valve Stenosis

Brief summary

To Evaluate the Safety and Effectiveness of Transcatheter Aortic Valve System ( Chengdu Silara Medtech Inc. ,Chengdu, China)in Patients with Severe Aortic Stenosis with or without Moderate or Lower Degree Regurgitation

Detailed description

Transcatheter Aortic Valve System (Chengdu Silara Medtech Inc. ,Chengdu, China) will be used for this study.

Interventions

Procedure: Transcatheter aortic valve replacement

Sponsors

Chengdu Silara Meditech Inc.
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
70 Years to No maximum
Healthy volunteers
No

Inclusion criteria

1. Age ≥70 years old; 2. Symptomatic severe aortic stenosis determined by echocardiogram, defined as: mean gradient ≥40 mmHg or peak jet velocity ≥4.0 m/s, or an aortic valve area \<0.8 cm2 or aortic valve area index \<0.5 cm2/m2. 3. NYHA Functional Class ≥II; 4. Life expectancy after aortic valve implantation thought to be \>1 year; 5. Native valvular or peripheral vascular anatomy is appropriate for TAVR; 6. Patient is assessed by at least two cardiothoracic surgeons and recorded as not suitable for surgery; 7. Patient can understand the purpose of the study, voluntarily participates and signs the informed consent form and is willing to accept the relevant examination and clinical follow-up visits.

Exclusion criteria

1. Anatomy is not appropriate for percutaneous valve implantation. Native valve annulus diameter is not in the scope of application(native TAV diameter ≤21mm, or ≥30mm); 2. Evidence of an acute myocardial infarction within 30 days prior to the study procedure, defined as: Q-wave myocardial infarction, or non-Q-wave myocardial infarction, with CK-MB≥2 ULN/or elevated TN (WHO definition); 3. Any treatment for traumatic cardiac surgery within 30 day prior to the study procedure(except coronary revascularization); 4. Pre-existing prosthetic valve in any position(except severely dysfunctional aortic bioprosthesis), or severe (\>3+) mitral insufficiency, or Gorlin syndrome; 5. Hematological abnormality, defined as: Leukopenia (WBC \<3x10\^9/L), acute anemia (Hb \<90g/L), or thrombocytopenia (platelet count\<50×109/L),history of bleeding diathesis or coagulopathy; 6. Hemodynamic unstable requiring myocardial contractile support or mechanical cardiac assistance; 7. Severe ventricular insufficiency. Left ventricular ejection fraction (LVEF) \<20%; 8. Echocardiographic evidence of intra-cardiac thrombus or vegetation etc.; 9. Active pepticulcer or upper GI bleeding within 3 months prior to the study procedure; 10. Cerebral Vascular Accident (CVA) within 3 months prior to the study procedure, including TIA; 11. A known hypersensitivity or contraindication to aspirin, heparin, ticlopidine or clopidogrel, or sensitivity to contrast media, which cannot be adequately pre-medicated; 12. Patients with infective endocarditis or other active stage of infection; 13. Currently participating in an investigational drug or another device trial.

Design outcomes

Primary

MeasureTime frameDescription
Rate of All-cause Mortality12 monthsPercentage of subjects who died from all causes in this population.

Secondary

MeasureTime frameDescription
Rate of Procedure SuccessImmediate post- procedureProcedure Success is defined as no death, stroke, myocardial infarction and renal failure occurred within 72 hours after the operation on the basis of the final device success.
Delivery System PerformanceImmediate post- procedureDelivery System Performance wil be evaluated subjectively by the researchers participating in the operation, which is generally classified as excellent, good ,average and poor.
Retrieval System Performance(if using)Immediate post- procedureRetrieval System Performance wil be evaluated subjectively by the researchers participating in the operation, which is generally classified as excellent, good ,average and poor.
Exchange Systerm EvaluationImmediate post- procedureExchange Systerm Evaluation wil be evaluated subjectively by the researchers participating in the operation, which is generally classified as excellent, good ,average and poor.
Valvular function evaluation, including valve stenosis, AR, valve function(e.g. EOA, gradient) and PVLImmediate post-procedure, 7days or discharge, 30 days, 6 months ,12months and 2-5 yearsThe evaluation criteria refer to the 2012 edition of the Association for Valvular Academic Research (VARC-2) consensus document
Quality of Life Assessment30 days, 6 months,and 12 monthsSF-12
NYHA function7 days or discharge, 30 days, 6 months ,12 months and 2-5 years
Rate of All-cause MortalityImmediate post- procedure ,7 days /discharge, 30 days, 6 months and 2-5 years.
Rate of Device SuccessImmediate post- procedureDevice Success is defined as a composite of : 1. Absence of procedural mortality AND 2. Successful vascular access, delivery and deployment of the device, and successful retrieval of the delivery system AND 3. Correct positioning of prosthetic heart valve into the proper anatomical location AND 4. Intended performance of the prosthetic heart valve (mean aortic valve gradient\<20 mmHg or peak velocity\<3 m/s, and no severe prosthetic valve regurgitation or PVL)
Incidence of MACCE(including mortality, stroke, MI, re-procedure, Conduction disturbances and cardiac arrhythmias)Immediate post- procedure, 7 days /discharge, 30 days, 6 months,12months and 2-5 years.
Incidence of Bleeding (life-threatening or disabling and major)Immediate post- procedure, 7 days /discharge, 30 days, 6 months,12months and 2-5 years.
Incidence of Stroke7 days /discharge, 30 days, 6 months, 12 months and 2-5 years.
Incidence of AKI (stage 2 and 3,or renal replacement therapy (RRT: Hemodialysis, abdomen Cavity dialysis, hemofiltration))7 days /discharge, 30 days, 6 months, 12 months and 2-5 years.
Incidence of Permanent Pacemaker Implantation7 days /discharge, 30 days, 6 months, 12 months and 2-5 years.
Incidence of major vascular complicationsImmediate post- procedure, 7 days /discharge, 30 days, 6 months, 12 months.
Incidence of Conduction disturbances and cardiac arrhythmiasImmediate post- procedure, 7 days /discharge, 30 days, 6 months, 12 month and 2-5years
Incidence of Other TAVI-related ComplicationsImmediate post- procedure, 7 days /discharge, 30 days, 6 months, 12 month and 2-5yearsincluding transfer to surgery , Accidental heart-lung machine, coronary obstruction, ventricular septal rupture, Mitral valve damage or dysfunction,cardiac tamponade,endocarditis ,valvular thrombus,Valve migration(shifting, detachment,embolism,error deployment),valve in valve
Incidence of Myocardial InfarctionImmediate post- procedure, 7 days /discharge, 30 days, 6 months ,12months and 2-5 years.