A 3D Tablet Sensor Approach to the Measurement of a Lymphedema

A 3D Tablet Sensor Approach to the Measurement of an Upper and Lower Limb Volume in the Monitoring of a Lymphedema

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT04821440
Acronym
LO3D
Enrollment
42
Registered
2021-03-29
Start date
2021-06-16
Completion date
2021-09-23
Last updated
2025-02-04

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Lymphedema

Keywords

3D sensor, Lower Extremity, Upper Extremity, Body Regions, Measurement, Monitoring, Water displacement, Volume

Brief summary

Up to now, the diagnosis of lymphedema remains hard and delayed. It suffers from many limitations such as lack of coordination and formation of the health-care network. The diversity of used tools is another obstacle because all provided solutions on the market are either costless with poor accuracy, or highly expensive but with excellent accuracy. Therefore, we decided to evaluate an affordable and open-source 3D iPad sensor sharing the same technology used in Kinect sensors.

Interventions

DEVICEWater displacement volumetry

The examination will take place in several stages and by two operators. Initially, the patient will strip the limb concerned (leg or arm). Once the member is in the water column of the volumeter, a delimitation of the submerged area of the member will be carried out using a marking (black dermographic marker). Indeed, the level of acquisition in 3D is determined relative to the height of the water line marked on the skin with a black felt, this felt is usually used for venous mapping and is well tolerated The first operator will carry out two measurements one after the other with a change of water previously stabilized at room temperature.

DEVICEThree dimension volumetry

The 3D volumetric acquisition is done with the sensor (CE marking) attached to the touch pad and does not require invasive or cutaneous contact with the patient (minimum acquisition distance of 40 cm). The operator triggers the acquisition via a button in an application installed on the tablet and physically turns around the patient to carry out the acquisition of the affected member. At the end of the acquisition, the operator presses a stop button and chooses whether or not to calculate the volume using another button depending on the quality of the acquisition.

Sponsors

University Hospital, Brest
Lead SponsorOTHER

Study design

Allocation
NON_RANDOMIZED
Intervention model
PARALLEL
Primary purpose
HEALTH_SERVICES_RESEARCH
Masking
DOUBLE (Investigator, Outcomes Assessor)

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

\- Unilateral lymphedema

Exclusion criteria

* Minor patient, * Bilateral lymphedema, * Trophic disorders: Acute or recent skin infection (\<1 month after the end of antibiotic therapy), unhealed wounds. * Predictable difficulties in carrying out the measurements due to the patient's physical limitations (maintenance of orthostatic or sitting position \<5 minutes) * Single or bilateral edema with a different etiology than lymphedema.

Design outcomes

Primary

MeasureTime frameDescription
Accuracy of a 3D sensor for lymphedema volumetryOne hourIntra-class correlation coefficient (ICC) Mean bias and limits of agreement (LOA) derived from the Blant-Altman diagram Mean bias and limits of agreement (LOA) derived from the Blant-Altman diagram

Secondary

MeasureTime frameDescription
Reproducibility of measurements made with a 3D sensor for lymphedema volumetryOne hourIntra-class correlation coefficient (ICC)