Assessment of Coronary Pressure Using the Combined Guidewire and Catheter Wirecath

Assessment of Coronary Pressure Using the Combined Guidewire and Catheter Wirecath

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT04776577
Enrollment
45
Registered
2021-03-01
Start date
2021-02-25
Completion date
2021-10-25
Last updated
2024-01-09

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Coronary Disease

Keywords

coronary disease, fractional flow reserve, coronary flow reserve, pressure wire

Brief summary

In summary, the purpose of the study is to evaluate whether the Wirecath device can be used as a standard pressure wire in a safe and efficient way.

Detailed description

Simultaneous measurements study (group 1) This group will be recruited and assessed in parallel with recruitment to group 2 at the discretion of the treating physician, based on the resource availability in the cath lab. In the simultaneous measurements study group, measurements will be collected from a Wirecath® and a regular sensor-tipped pressure wire in the same position in the coronary artery at the same time (simultaneously). The regular wire to be used is the market leading Abbott PressureWire®. By comparing the pressure measurements (Pd) from Wirecath® and from a regular sensor-tipped wire, we intend to examine if the pressure reported by the two wires deviate in any way. If a deviation exists, the hydrostatic error, caused by the height difference between the positions of Pa and Pd when using a sensor-tipped wire, will be assessed. In FFR-negative patients, after removal of the Wirecath®, the Abbott PressureWire® is used in LAD * first to measure CFR by saline bolus injections and thermodilution during rest and hyperemia (Bolus-thermo CFR) and * second together with a Hexacath/Rayflow@ infusion catheter and a saline infusion pump to measure absolute flow during rest and hyperemia that will be used to calculate CFR (Absolute-flow CFR). Wirecath will be used for PCI when indicated and performance documented by follow-up questions to the users. Regular study group (group 2) In the regular use study group, the Wirecath® is to be used as a regular pressure measurement tool, replacing the regularly used pressure wire. Pressure measurement results will be collected. Wirecath® will be used for PCI when indicated and performance documented by follow-up questions to the users. In FFR-negative patients, after removal of the Wirecath®, the Abbott PressureWire® is used in LAD * first to measure CFR by saline bolus injections and thermodilution during rest and hyperemia (Bolus-thermo CFR) and * second together with a Hexacath/Rayflow® infusion catheter and a saline infusion pump to measure absolute flow during rest and hyperemia that will be used to calculate CFR (Absolute-flow CFR). Echocardiography-CFR group (group 3) Patients from group 1 and group 2 that have undergone Bolus-thermo and Absolute-flow CFR measurements, will be recruited to group 3. In this group echocardiography-CFR (Echo CFR) will be measured within 14 days. The echocardiography is performed according to the standard procedure of the hospital. The correlation between the methods will be investigated.

Interventions

DEVICEWirecath@ and Abbott PressureWire® (20 patients)

Simultaneous FFR measurements with Wirecath@ and Abbott PressureWire®

DEVICEWirecath® (20 patients)

FFR measurement with Wirecath® only

Pressure derived CFR with Wirecath®, Thermodilution CFR with Abbott PressureWire®, Absolut flow CFR with Abbott PressureWire® and Hexacath Rayflow catheter®, Non-invasive echo derived CFR

Sponsors

Cavis Technologies AB
CollaboratorINDUSTRY
Vastra Gotaland Region
Lead SponsorOTHER_GOV

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
DIAGNOSTIC
Masking
NONE

Intervention model description

Results from simultaneous measurements with two wires will be graphically showed using the Bland-Altman plot and reported in terms of intraclass correlation coefficient (ICC). Also, whether a theoretical correction for the expected hydrostatic pressure error (0.77 mmHg per cm height difference) at the distal pressure measurement site (Pd) accounts for the difference in FFR during simultaneous measurements will be explored, and a corrected ICC reported. A two-sided p-value of \<0.05 will be considered to be significant, when applicable. The Pearson's correlation coefficient will be also calculated considering the Wirecath FFR value in comparison to the FFR value made with the regular wire with and without correction for hydrostatic error. For the CFR values, the Pearson's correlation coefficient will be used to determine the relation between the bolus-thermo CFR, absolute-flow CFR, Echo CFR and the pressure-derived CFR in the LAD.

Eligibility

Sex/Gender
ALL
Age
19 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Adults (\>18yrs age) * Informed consent given * Patients with result from diagnostic angiography * Indication for coronary pressure assessment

Exclusion criteria

* Patients with known Heparin-Induced Thrombocytopenia (HIT) or allergy to heparin. * Inability to tolerate Adenosine * Any condition, e.g. unstable health, which in the opinion of the investigator makes the patient unsuitable for inclusion (Decision documented in inclusion log).

Design outcomes

Primary

MeasureTime frameDescription
Safety of the Wirecath® will be assessed through the frequency and severity of all adverse events30-60 minutesThe safety of the Wirecath@ will be assessed through the frequency and severity of all adverse events reported for study subjects after the completion of the pressure measurement and PCI. The primary endpoint will be achieved when the device has been used on enrolled patients and required measurements have been taken.

Secondary

MeasureTime frameDescription
Comparison with Abbott pressure wire30-60 minutescomparing the Wirecath measurement values with the measurement values of a regular sensor-tipped pressure guidewire during simultaneous measurements in the same vessel.
Comparing pressure-derived coronary flow reserve (CFR) with bolus-thermo CFR, absolute flow CFR and non-invasive doppler derived CFR14 daysIn FFR negative LADs, comparing the Wirecath® pressure-derived CFR values with CFR derived from thermodilution, absolute flow and echocardiography-CFR values. This secondary endpoint will be achieved if these values correlate