Mindray Pediatric Hematology Reference Intervals Study

Establishment of Pediatric Reference Intervals for Mindray Hematology Parameters for Worldwide Application in Clinical Laboratories

Status
Completed
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT04758208
Enrollment
691
Registered
2021-02-17
Start date
2021-07-12
Completion date
2022-02-11
Last updated
2022-03-10

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Establish Hematology Reference Intervals

Brief summary

This study is a collaborative initiative between the CALIPER (Canadian Laboratory Initiative on Pediatric Reference Intervals) program at the Hospital for Sick Children (HSC) and Mindray Medical International Limited to establish a database of pediatric reference intervals for hematology parameters on BC-6800Plus.

Detailed description

The purpose of this study is to establish accurate and robust ethnic-, age- and sex-specific reference intervals for hematology parameters during childhood and adolescence on the Mindray analyzer (BC-6800Plus). Subjects will be recruited for this study through the CALIPER initiative and will include healthy children and adolescents in the community and in outpatient clinics at the Hospital for Sick Children (HSC). Since all participants will be recruited through the CALIPER initiative, the CALIPER consent forms will be used to obtain consent from each child and their parent/guardian. These forms obtain consent to use the samples to develop reference intervals for a range of tests. This study aims to measure hematology parameters in these samples and is therefore encompassed within the CALIPER consent. Recruitment will focus on subjects from three ethnic groups: Caucasians, Hispanics, and South Asians. A total of 960 subjects will be recruited for this study through the CALIPER initiative. The total sample size was determined by considering the requirement of 60-120 samples per partition for five age groups (0-\<1 years, 1-\<5 years, 5-\<10 years, 10-\<15 years, 15-\<19 years). Each age group will consist of an equal number of females and males. Additional samples may be required if data analysis suggests further age and/or sex partitioning for specific parameters. Each subject will provide a blood sample, which will be collected through the CALIPER initiative. These samples will be analyzed as part of this study in a secondary-use/retrospective fashion. Specifically, fresh whole blood specimens will be collected in K2EDTA tubes, placed in biohazard bags, stored on ice, and analyzed on the Mindray hematology platform within 8 hours of collection. All available hematology parameters for the BC-6800Plus will be measured (37 report and 48 research hematology parameters, as per the user manual). All procedures will be performed according to the user manual. Mindray quality control materials will be used to monitor the performance of the assays during the study. HSC will retain all quality controls data, maintenance record and reagents/calibrators lot numbers. Quality controls will be run daily and analysis of the CALIPER samples will proceed only after all the quality control (QC) data are acceptable. Data analysis will be performed by the CALIPER team in accordance with Clinical & Laboratory Standards Institute (CLSI) EP28-A3c guidelines on defining, establishing, and verifying reference intervals in the clinical laboratory. The statistical approach for reference interval establishment has been described in-depth previously. In brief, analyte concentrations will be plotted against age and colour-coded by sex. Following visual inspection, extreme outliers will be removed and suspected age- and sex-specific partitions will be confirmed using the Harris and Boyd statistical method. Normality of data will be qualitatively assessed using quantile-quantile plots and quantitatively assessed using the Shapiro-Wilk's test. If skewed, the Box-Cox transformation will be applied in an effort to transform the data to a normal distribution. Outliers will be subsequently identified and removed using the Tukey or adjusted Tukey method for Gaussian and non-Gaussian data distributions, respectively. Reference intervals for partitions with ≥120 participants will be calculated using the nonparametric method. Reference intervals for partitions with ≥40 and \<120 participants (60 for group 0-\<1 years and 1-\<5 years, 120 for group 5-\<10 years, 10-\<15 years and 15-\<19 years) will be calculated using the robust method. 90% percent confidence intervals will also be calculated for the lower and upper limit of each reference interval. Ethnic-specific reference intervals will be calculated if sample size for each ethnic group is adequate. Otherwise, reference value distributions for South Asian and Hispanic children will be compared to those of Caucasian children. Oversight Authority (additional information) * This study is regulated by 'Canada: Health Canada' * This study has been reviewed and approved by 'Canada: Ethics Review Committee'

Interventions

DIAGNOSTIC_TESTHematology Parameters

Measurement of hematology parameters on BC-6800Plus

Sponsors

MINDRAY
CollaboratorUNKNOWN
The Hospital for Sick Children
Lead SponsorOTHER

Study design

Observational model
COHORT
Time perspective
RETROSPECTIVE

Eligibility

Sex/Gender
ALL
Age
No minimum to 18 Years
Healthy volunteers
Yes

Inclusion criteria

* 0-\<19 years

Exclusion criteria

* History of chronic illness * Use of prescription medications within 2 weeks of collection * Acute illness within 7 days of collection * Pregnancy

Design outcomes

Primary

MeasureTime frameDescription
Immature Granulocyte percentageWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Reticulocyte percentageWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Reticulocyte numberWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Reticulocyte Hemoglobin ExpressionWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
White Blood Cell countWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Basophil numberWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Basophil percentageWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Neutrophil numberWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Neutrophil percentageWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Eosinophil numberWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Eosinophil percentageWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Lymphocyte numberWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Lymphocyte percentageWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Monocyte numberWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Monocyte percentageWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Immature Granulocyte numberWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Immature reticulocyte fractionWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Low fluorescent ratioWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Middle fluorescent ratioWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
High fluorescent ratioWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Red blood cell countWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Hemoglobin ConcentrationWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Mean Corpuscular VolumeWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Mean Corpuscular HemoglobinWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Mean Corpuscular Hemoglobin ConcentrationWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Red Blood Cell Distribution Width - Coefficient of VariationWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Red Blood Cell Distribution Width - Standard DeviationWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
HematocritWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Nucleated red blood cell (NRBC) numberWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Nucleated red blood cell percentageWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Platelet countWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Mean Platelet VolumeWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Platelet Distribution WidthWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
PlateletcritWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Immature Platelet FractionWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Platelet-large cell ratioWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Platelet-large cell countWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer

Secondary

MeasureTime frameDescription
DIFF scattergram, mean monocyte distribution-forward scatter intensityWithin 8 hours of sample collectionScattergram of laboratory values derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Dimorphic population, smaller distribution RBC countWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Dimorphic population, larger distribution RBC countWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Dimorphic population, smaller distribution mean corpuscular volumeWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Dimorphic population, larger distribution mean corpuscular volumeWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
White blood cell counts - Differential (DIFF)Within 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Mean Hemoglobin Distribution Width of Red Blood CellWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Mean Platelet ConcentrationWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Mean Platelet Matter ContentWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Hyperchromic Red Blood Cell percentage (Hyper%)Within 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Hypochromic Red Blood Cell percentageWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Mean Corpuscular Hemoglobin of ReticulocyteWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
High fluorescent Cell numberWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
High fluorescent Cell percentageWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Optical Red Blood Cell countWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Optical Platelet countWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Platelet count - ImpedanceWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Optical White Blood Cell countWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Total nucleated cell counts - DIFFWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Immature Eosinophil percentageWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Immature Eosinophil numberWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
High forward scatter NRBC ratioWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Low forward scatter NRBC ratioWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Neutrophil-to-lymphocyte ratioWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Platelet-to-lymphocyte ratioWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
White blood cell count - white cell nucleated basophil channel (WNB)Within 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Total nucleated cell countsWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Infected Red Blood Cell countWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Infected Red Blood Cell permillageWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Microcyte countWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Microcyte percentageWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Macrocyte countWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Macrocyte percentageWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Mean Reticulocyte VolumeWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Red Blood Cell Fragment CountWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Red Blood Cell Fragment PercentageWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Platelet Distribution Width Standard DeviationWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Reticulocyte Production IndexWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
High fluorescent Immature Platelet FractionWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
Immature Platelet CountWithin 8 hours of sample collectionLaboratory value derived from the Mindray BC-6800Plus Auto Hematology Analyzer
DIFF scattergram, mean neutrophil distribution-side scatter intensityWithin 8 hours of sample collectionScattergram of laboratory values derived from the Mindray BC-6800Plus Auto Hematology Analyzer
DIFF scattergram, mean neutrophil distribution-side fluorescent light intensityWithin 8 hours of sample collectionScattergram of laboratory values derived from the Mindray BC-6800Plus Auto Hematology Analyzer
DIFF scattergram, mean neutrophil distribution-forward scatter intensityWithin 8 hours of sample collectionScattergram of laboratory values derived from the Mindray BC-6800Plus Auto Hematology Analyzer
DIFF scattergram, mean lymphocyte distribution-side scatter intensityWithin 8 hours of sample collectionScattergram of laboratory values derived from the Mindray BC-6800Plus Auto Hematology Analyzer
DIFF scattergram, mean lymphocyte distribution-side fluorescent intensityWithin 8 hours of sample collectionScattergram of laboratory values derived from the Mindray BC-6800Plus Auto Hematology Analyzer
DIFF scattergram, mean lymphocyte distribution-forward scatter intensityWithin 8 hours of sample collectionScattergram of laboratory values derived from the Mindray BC-6800Plus Auto Hematology Analyzer
DIFF scattergram, mean monocyte distribution-side scatter intensityWithin 8 hours of sample collectionScattergram of laboratory values derived from the Mindray BC-6800Plus Auto Hematology Analyzer
DIFF scattergram, mean monocyte distribution-side fluorescent light intensityWithin 8 hours of sample collectionScattergram of laboratory values derived from the Mindray BC-6800Plus Auto Hematology Analyzer

Countries

Canada

Outcome results

None listed