Skip to content

A Comparative Study of an Innovative 3-lead Wireless Water Resistant Holter System and a Conventional Holter System

A Comparative Study of the Findings of Dynamic Electrocardiography (Holter System) Using the an Innovative 3-lead Wireless Water Resistant Device and a Conventional Device

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT04723355
Enrollment
231
Registered
2021-01-25
Start date
2021-01-26
Completion date
2021-11-16
Last updated
2024-10-24

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Cardiac Arrhythmia

Brief summary

Holter monitoring is one of the most widely used diagnostic methods to detect cardiac arrhythmias. Newer Holter monitors may provide some advantages over the more traditional ones. This study will compare the findings of a Holter exam using an innovative 3-lead wireless water resistant device with a conventional device.

Detailed description

The 24-hour ambulatory electrocardiography, commonly known as Holter monitoring, is a diagnostic method to correlate clinical symptoms with cardiac arrhythmias. The Holter monitor records electrical signals from the heart using electrodes attached to the chest, which are connected via cables to a digital recorder. In addition to the device to record the cardiac electrical signals, the Holter system also has a software for review and analysis of the record. The software of the Holter system has an integrated automatic analysis process to detect different sorts of heart beats, rhythms, etc, which are then validated by a technician and a physician. The success of the automatic analysis depends of the quality of the captured electrical signals, which are dependent on the appropriate attachment of the electrodes/cables to the patient body. Therefore, inappropriately attached electrodes/cables can result in electromagnetic disturbance that can interfere with the ECG signal, resulting in a very noisy record that is hard to be analysed. Newer Holter systems that eliminate the cables and attach the recorder directly to the electrodes positioned in the patient´s chest might reduce the electromagnetic disturbance in the ECG signal capture and, therefore, improve the quality of the exam. This study will compare the findings of a Holter exam using an innovative 3-lead wireless water resistant device with a conventional device.

Interventions

DEVICEAn innovative 3-lead wireless water resistant Holter device

The participant will use the innovative 3-lead wireless water resistant Holter device simultaneously with the conventional Holter device for 24 hours.

DEVICEA conventional device

The participant will use the innovative 3-lead wireless water resistant Holter device simultaneously with the conventional Holter device for 24 hours.

Sponsors

Instituto Dante Pazzanese de Cardiologia
CollaboratorOTHER
Quoretech
CollaboratorINDUSTRY
Hospital Israelita Albert Einstein
Lead SponsorOTHER

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
DIAGNOSTIC
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

* Any individual who has an indication of 24 hours Holter monitoring and is referred to the arrhythmia clinic of the Instituto Dante Pazzanese de Cardiologia

Exclusion criteria

* Refusal to provide written informed consent

Design outcomes

Primary

MeasureTime frameDescription
Detection of Cardiac Arrhythmias24 hoursNumber of patients with cardiac arrhythmias detected in both Holter monitors, including a) atrial fibrillation or atrial flutter; b) supraventricular tachycardia (any tachycardia with ≥ 3 beats arising outside the sinus node and above the His-Purkinje bifurcation, including atrial tachycardia, atrioventricular nodal reentry tachycardia, junctional tachycardia and accessory pathway reentry tachycardia, except atrial fibrillation and atrial flutter); c) ventricular tachycardia (any tachycardia with ≥ 3 beats arising from the ventricles, except polymorphic ventricular tachycardia or ventricular fibrillation); d) polymorphic ventricular tachycardia or ventricular fibrillation; e) atrioventricular (AV) block (high-degree AV block, in which there is an interruption in the transmission of a heart beat from the atria to the ventricles, including those classified as 2nd degree AV block Mobitz type I and II, AV block 2:1 or complete AV block); or f) heart beat pauses (≥ 2,5 seconds)

Secondary

MeasureTime frameDescription
Detection of Supraventricular Tachycardia24 hoursNumber of patients with supraventricular tachycardia (any tachycardia with ≥ 3 beats arising outside the sinus node and above the His-Purkinje bifurcation, including atrial tachycardia, atrioventricular nodal reentry tachycardia, junctional tachycardia and accessory pathway reentry tachycardia, except atrial fibrillation and atrial flutter) detected in both Holter monitors
Detection of Ventricular Tachycardia24 hoursNumber of patients with ventricular tachycardia (any tachycardia with ≥ 3 beats arising from the ventricles, except polymorphic ventricular tachycardia or ventricular fibrillation) detected in both Holter monitors
Detection of Polymorphic Ventricular Tachycardia or Ventricular Fibrillation24 hoursNumber of patients with polymorphic ventricular tachycardia or ventricular fibrillation detected in both Holter monitors
Detection of Atrioventricular Block24 hoursNumber of patients with atrioventricular block (high-degree AV block, in which there is an interruption in the transmission of a heart beat from the atria to the ventricles, including those classified as 2nd degree AV block Mobitz type I and II, AV block 2:1 or complete AV block) detected in both Holter monitors
Detection of Heart Beat Pauses24 hoursNumber of patients with heart beat pauses ≥ 2.5 seconds detected in both Holter monitors
Duration of the Holter Exam24 hoursDuration of the Holter exam in both Holter monitors
Mean Heart Rate24 hoursMean heart rate measured in both Holter monitors
Detection of Atrial Fibrillation or Atrial Flutter24 hoursNumber of patients with atrial fibrillation or atrial flutter detected in both Holter monitors
Minimum Heart Rate24 hoursMinimum heart rate measured in both Holter monitors
Atrial Ectopic Heart Beats24 hoursTotal number of atrial ectopic heart beats measured in both Holter monitors
Ventricular Ectopic Heart Beats24 hoursTotal number of ventricular ectopic heart beats measured in both Holter monitors
Episodes of Ventricular Tachycardia24 hoursTotal number of ventricular tachycardia episodes detected in both Holter monitors
Episodes of Supraventricular Tachycardia24 hoursTotal number of supraventricular tachycardia episodes detected in both Holter monitors
Episodes of Heart Beat Pauses24 hoursTotal number of heart beat pauses ≥ 2.5 seconds detected in both Holter monitors
Holter System Convenience of Use24 hoursNumber of Participants Reporting the convenience of use of the Holter systems
Maximum Heart Rate24 hoursMaximum heart rate measured in both Holter monitors

Countries

Brazil

Participant flow

Recruitment details

Individuals attending the study inclusion criteria were consecutively approached to participate in the study between 26 January 2021 and 16 November 2021 at the arrhythmia clinic of the Instituto Dante Pazzanese de Cardiologia in Sao Paulo, Brazil

Participants by arm

ArmCount
Holter Device
In this single arm of the study, the participant will use 2 Holter systems simultaneously for 24 hours, an innovative 3-lead wireless water resistant device and a conventional device. An innovative 3-lead wireless water resistant Holter device: The participant will use the innovative 3-lead wireless water resistant Holter device simultaneously with the conventional Holter device for 24 hours. A conventional device: The participant will use the innovative 3-lead wireless water resistant Holter device simultaneously with the conventional Holter device for 24 hours.
179
Total179

Withdrawals & dropouts

PeriodReasonFG000
Overall StudyErrors in the informed consent form2
Overall StudyIssues with leads off, substantial ECG signal artefacts, and ECG data upload errors50

Baseline characteristics

CharacteristicHolter Device
Age, Continuous58.4 years
STANDARD_DEVIATION 13.9
Body mass index27.6 kg/m^2
STANDARD_DEVIATION 4.8
Chronic obstructive pulmonary disease
Did not know
1 Participants
Chronic obstructive pulmonary disease
Missing
50 Participants
Chronic obstructive pulmonary disease
No
117 Participants
Chronic obstructive pulmonary disease
Yes
11 Participants
Diabetes
Did not know
2 Participants
Diabetes
Missing
8 Participants
Diabetes
No
130 Participants
Diabetes
Yes
39 Participants
Heart failure
Did not know
2 Participants
Heart failure
Missing
10 Participants
Heart failure
No
81 Participants
Heart failure
Yes
86 Participants
Hypertension
Missing
8 Participants
Hypertension
No
48 Participants
Hypertension
Yes
123 Participants
Hypothyroidism
Did not know
3 Participants
Hypothyroidism
Missing
50 Participants
Hypothyroidism
No
106 Participants
Hypothyroidism
Yes
20 Participants
Level of education
Complete postgraduate school
2 Participants
Level of education
Complete primary school
61 Participants
Level of education
Complete secondary school
57 Participants
Level of education
Complete technical school
2 Participants
Level of education
Complete undergraduate school
26 Participants
Level of education
Illiterate
1 Participants
Level of education
Incomplete primary school
21 Participants
Level of education
Incomplete undergraduate school
9 Participants
Marital status
Divorced
19 Participants
Marital status
Married
115 Participants
Marital status
Single
30 Participants
Marital status
Widowed
15 Participants
Previous myocardial infarction
Did not know
1 Participants
Previous myocardial infarction
Missing
8 Participants
Previous myocardial infarction
No
128 Participants
Previous myocardial infarction
Yes
42 Participants
Previous stroke
Did not know
1 Participants
Previous stroke
Missing
50 Participants
Previous stroke
No
92 Participants
Previous stroke
Yes
36 Participants
Race/Ethnicity, Customized
Race
Asian
2 Participants
Race/Ethnicity, Customized
Race
Black
29 Participants
Race/Ethnicity, Customized
Race
Mixed
35 Participants
Race/Ethnicity, Customized
Race
White
113 Participants
Region of Enrollment
Brazil
179 participants
Sex: Female, Male
Female
84 Participants
Sex: Female, Male
Male
95 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 179
other
Total, other adverse events
0 / 179
serious
Total, serious adverse events
0 / 179

Outcome results

Primary

Detection of Cardiac Arrhythmias

Number of patients with cardiac arrhythmias detected in both Holter monitors, including a) atrial fibrillation or atrial flutter; b) supraventricular tachycardia (any tachycardia with ≥ 3 beats arising outside the sinus node and above the His-Purkinje bifurcation, including atrial tachycardia, atrioventricular nodal reentry tachycardia, junctional tachycardia and accessory pathway reentry tachycardia, except atrial fibrillation and atrial flutter); c) ventricular tachycardia (any tachycardia with ≥ 3 beats arising from the ventricles, except polymorphic ventricular tachycardia or ventricular fibrillation); d) polymorphic ventricular tachycardia or ventricular fibrillation; e) atrioventricular (AV) block (high-degree AV block, in which there is an interruption in the transmission of a heart beat from the atria to the ventricles, including those classified as 2nd degree AV block Mobitz type I and II, AV block 2:1 or complete AV block); or f) heart beat pauses (≥ 2,5 seconds)

Time frame: 24 hours

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Holter DeviceDetection of Cardiac ArrhythmiasConventional Holter System104 Participants
Holter DeviceDetection of Cardiac ArrhythmiasInnovative 3-lead Wireless Water Resistant Holter System107 Participants
Comparison: The comparison of the two Holter monitoring devices was analyzed in terms of the accuracy of the Innovative 3-lead Wireless Water Resistant Holter System in detecting the arrhythmias compared to the Conventional Holter System as the gold standard.95% CI: [0.81, 0.92]
Comparison: Other measures of diagnostic accuracy, including the sensitivity, specificity, positive and negative predictive values were also analyzed.95% CI: [0.83, 0.95]
Comparison: Other measures of diagnostic accuracy, including the sensitivity, specificity, positive and negative predictive values were also analyzed.95% CI: [0.72, 0.9]
Comparison: Other measures of diagnostic accuracy, including the sensitivity, specificity, positive and negative predictive values were also analyzed.95% CI: [0.8, 0.93]
Comparison: Other measures of diagnostic accuracy, including the sensitivity, specificity, positive and negative predictive values were also analyzed.95% CI: [0.76, 0.93]
Comparison: We also did an additional analysis to further evaluate the concordance between both Holter monitoring devices using the Cohen´s Kappa coefficient.
Comparison: Exploratory analyses of positive and negative likelihood ratios were also conducted.95% CI: [3.17, 8.58]
Comparison: Exploratory analyses of positive and negative likelihood ratios were also conducted.95% CI: [0.06, 0.21]
Secondary

Atrial Ectopic Heart Beats

Total number of atrial ectopic heart beats measured in both Holter monitors

Time frame: 24 hours

ArmMeasureGroupValue (MEAN)Dispersion
Holter DeviceAtrial Ectopic Heart BeatsInnovative 3-lead Wireless Water Resistant Holter System717.09 atrial ectopic heart beats in 24 hrStandard Deviation 2323.94
Holter DeviceAtrial Ectopic Heart BeatsConventional Holter System723.46 atrial ectopic heart beats in 24 hrStandard Deviation 2223.27
Comparison: The quantitative secondary outcomes were analyzed using Bland-Altman graphs as well as the Lin´s concordance correlation coefficient (CCC).95% CI: [-1391.46, 1378.72]
Comparison: The quantitative secondary outcomes were analyzed using Bland-Altman graphs as well as the Lin´s concordance correlation coefficient (CCC).95% CI: [0.94, 0.96]
Secondary

Detection of Atrial Fibrillation or Atrial Flutter

Number of patients with atrial fibrillation or atrial flutter detected in both Holter monitors

Time frame: 24 hours

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Holter DeviceDetection of Atrial Fibrillation or Atrial FlutterInnovative 3-lead Wireless Water Resistant Holter System23 Participants
Holter DeviceDetection of Atrial Fibrillation or Atrial FlutterConventional Holter System22 Participants
Comparison: The comparison of the two Holter monitoring devices was analyzed in terms of the accuracy of the Innovative 3-lead Wireless Water Resistant Holter System in detecting the arrhythmias compared to the Conventional Holter System as the gold standard.95% CI: [0.97, 1]
Comparison: Other measures of diagnostic accuracy, including the sensitivity, specificity, positive and negative predictive values were also analyzed.95% CI: [0.85, 1]
Comparison: Other measures of diagnostic accuracy, including the sensitivity, specificity, positive and negative predictive values were also analyzed.95% CI: [0.97, 1]
Comparison: Other measures of diagnostic accuracy, including the sensitivity, specificity, positive and negative predictive values were also analyzed.95% CI: [0.78, 1]
Comparison: Other measures of diagnostic accuracy, including the sensitivity, specificity, positive and negative predictive values were also analyzed.95% CI: [0.98, 1]
Comparison: We also did an additional analysis to further evaluate the concordance between both Holter monitoring devices using the Cohen´s Kappa coefficient.
Comparison: Exploratory analyses of positive and negative likelihood ratios were also conducted.95% CI: [22.25, 1107.61]
Comparison: Exploratory analyses of positive and negative likelihood ratios were also conducted.95% CI: [0, 0]
Secondary

Detection of Atrioventricular Block

Number of patients with atrioventricular block (high-degree AV block, in which there is an interruption in the transmission of a heart beat from the atria to the ventricles, including those classified as 2nd degree AV block Mobitz type I and II, AV block 2:1 or complete AV block) detected in both Holter monitors

Time frame: 24 hours

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Holter DeviceDetection of Atrioventricular BlockConventional Holter System4 Participants
Holter DeviceDetection of Atrioventricular BlockInnovative 3-lead Wireless Water Resistant Holter System5 Participants
Comparison: The comparison of the two Holter monitoring devices was analyzed in terms of the accuracy of the Innovative 3-lead Wireless Water Resistant Holter System in detecting the arrhythmias compared to the Conventional Holter System as the gold standard.95% CI: [0.95, 1]
Comparison: Other measures of diagnostic accuracy, including the sensitivity, specificity, positive and negative predictive values were also analyzed.95% CI: [0.19, 0.99]
Comparison: Other measures of diagnostic accuracy, including the sensitivity, specificity, positive and negative predictive values were also analyzed.95% CI: [0.96, 1]
Comparison: Other measures of diagnostic accuracy, including the sensitivity, specificity, positive and negative predictive values were also analyzed.95% CI: [0.15, 0.95]
Comparison: Other measures of diagnostic accuracy, including the sensitivity, specificity, positive and negative predictive values were also analyzed.95% CI: [0.97, 1]
Comparison: We also did an additional analysis to further evaluate the concordance between both Holter monitoring devices using the Cohen´s Kappa coefficient.
Comparison: Exploratory analyses of positive and negative likelihood ratios were also conducted.95% CI: [14.8, 291.07]
Comparison: Exploratory analyses of positive and negative likelihood ratios were also conducted.95% CI: [0.05, 1.38]
Secondary

Detection of Heart Beat Pauses

Number of patients with heart beat pauses ≥ 2.5 seconds detected in both Holter monitors

Time frame: 24 hours

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Holter DeviceDetection of Heart Beat PausesInnovative 3-lead Wireless Water Resistant Holter System7 Participants
Holter DeviceDetection of Heart Beat PausesConventional Holter System11 Participants
Comparison: The comparison of the two Holter monitoring devices was analyzed in terms of the accuracy of the Innovative 3-lead Wireless Water Resistant Holter System in detecting the arrhythmias compared to the Conventional Holter System as the gold standard.95% CI: [0.94, 0.99]
Comparison: Other measures of diagnostic accuracy, including the sensitivity, specificity, positive and negative predictive values were also analyzed.95% CI: [0.31, 0.89]
Comparison: Other measures of diagnostic accuracy, including the sensitivity, specificity, positive and negative predictive values were also analyzed.95% CI: [0.98, 1]
Comparison: Other measures of diagnostic accuracy, including the sensitivity, specificity, positive and negative predictive values were also analyzed.95% CI: [0.59, 1]
Comparison: Other measures of diagnostic accuracy, including the sensitivity, specificity, positive and negative predictive values were also analyzed.95% CI: [0.94, 0.99]
Comparison: We also did an additional analysis to further evaluate the concordance between both Holter monitoring devices using the Cohen´s Kappa coefficient.
Comparison: Exploratory analyses of positive and negative likelihood ratios were also conducted.
Comparison: Exploratory analyses of positive and negative likelihood ratios were also conducted.95% CI: [0.17, 0.79]
Secondary

Detection of Polymorphic Ventricular Tachycardia or Ventricular Fibrillation

Number of patients with polymorphic ventricular tachycardia or ventricular fibrillation detected in both Holter monitors

Time frame: 24 hours

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Holter DeviceDetection of Polymorphic Ventricular Tachycardia or Ventricular FibrillationInnovative 3-lead Wireless Water Resistant Holter System14 Participants
Holter DeviceDetection of Polymorphic Ventricular Tachycardia or Ventricular FibrillationConventional Holter System12 Participants
Comparison: The comparison of the two Holter monitoring devices was analyzed in terms of the accuracy of the Innovative 3-lead Wireless Water Resistant Holter System in detecting the arrhythmias compared to the Conventional Holter System as the gold standard.95% CI: [0.89, 0.96]
Comparison: Other measures of diagnostic accuracy, including the sensitivity, specificity, positive and negative predictive values were also analyzed.95% CI: [0.28, 0.85]
Comparison: Other measures of diagnostic accuracy, including the sensitivity, specificity, positive and negative predictive values were also analyzed.95% CI: [0.92, 0.98]
Comparison: Other measures of diagnostic accuracy, including the sensitivity, specificity, positive and negative predictive values were also analyzed.95% CI: [0.23, 0.77]
Comparison: Other measures of diagnostic accuracy, including the sensitivity, specificity, positive and negative predictive values were also analyzed.95% CI: [0.93, 0.99]
Comparison: We also did an additional analysis to further evaluate the concordance between both Holter monitoring devices using the Cohen´s Kappa coefficient.
Comparison: Exploratory analyses of positive and negative likelihood ratios were also conducted.95% CI: [5.84, 33.17]
Comparison: Exploratory analyses of positive and negative likelihood ratios were also conducted.95% CI: [0.22, 0.85]
Secondary

Detection of Supraventricular Tachycardia

Number of patients with supraventricular tachycardia (any tachycardia with ≥ 3 beats arising outside the sinus node and above the His-Purkinje bifurcation, including atrial tachycardia, atrioventricular nodal reentry tachycardia, junctional tachycardia and accessory pathway reentry tachycardia, except atrial fibrillation and atrial flutter) detected in both Holter monitors

Time frame: 24 hours

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Holter DeviceDetection of Supraventricular TachycardiaInnovative 3-lead Wireless Water Resistant Holter System71 Participants
Holter DeviceDetection of Supraventricular TachycardiaConventional Holter System71 Participants
Comparison: The comparison of the two Holter monitoring devices was analyzed in terms of the accuracy of the Innovative 3-lead Wireless Water Resistant Holter System in detecting the arrhythmias compared to the Conventional Holter System as the gold standard.95% CI: [0.79, 0.9]
Comparison: Other measures of diagnostic accuracy, including the sensitivity, specificity, positive and negative predictive values were also analyzed.95% CI: [0.71, 0.9]
Comparison: Other measures of diagnostic accuracy, including the sensitivity, specificity, positive and negative predictive values were also analyzed.95% CI: [0.8, 0.93]
Comparison: Other measures of diagnostic accuracy, including the sensitivity, specificity, positive and negative predictive values were also analyzed.95% CI: [0.71, 0.9]
Comparison: Other measures of diagnostic accuracy, including the sensitivity, specificity, positive and negative predictive values were also analyzed.95% CI: [0.8, 0.93]
Comparison: We also did an additional analysis to further evaluate the concordance between both Holter monitoring devices using the Cohen´s Kappa coefficient.
Comparison: Exploratory analyses of positive and negative likelihood ratios were also conducted.95% CI: [4.03, 11.43]
Comparison: Exploratory analyses of positive and negative likelihood ratios were also conducted.95% CI: [0.13, 0.34]
Secondary

Detection of Ventricular Tachycardia

Number of patients with ventricular tachycardia (any tachycardia with ≥ 3 beats arising from the ventricles, except polymorphic ventricular tachycardia or ventricular fibrillation) detected in both Holter monitors

Time frame: 24 hours

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Holter DeviceDetection of Ventricular TachycardiaInnovative 3-lead Wireless Water Resistant Holter System26 Participants
Holter DeviceDetection of Ventricular TachycardiaConventional Holter System23 Participants
Comparison: The comparison of the two Holter monitoring devices was analyzed in terms of the accuracy of the Innovative 3-lead Wireless Water Resistant Holter System in detecting the arrhythmias compared to the Conventional Holter System as the gold standard.95% CI: [0.88, 0.96]
Comparison: Other measures of diagnostic accuracy, including the sensitivity, specificity, positive and negative predictive values were also analyzed.95% CI: [0.56, 0.93]
Comparison: Other measures of diagnostic accuracy, including the sensitivity, specificity, positive and negative predictive values were also analyzed.95% CI: [0.9, 0.98]
Comparison: Other measures of diagnostic accuracy, including the sensitivity, specificity, positive and negative predictive values were also analyzed.95% CI: [0.48, 0.86]
Comparison: Other measures of diagnostic accuracy, including the sensitivity, specificity, positive and negative predictive values were also analyzed.95% CI: [0.93, 0.99]
Comparison: We also did an additional analysis to further evaluate the concordance between both Holter monitoring devices using the Cohen´s Kappa coefficient.
Comparison: Exploratory analyses of positive and negative likelihood ratios were also conducted.95% CI: [7.51, 30.99]
Comparison: Exploratory analyses of positive and negative likelihood ratios were also conducted.95% CI: [0.11, 0.5]
Secondary

Duration of the Holter Exam

Duration of the Holter exam in both Holter monitors

Time frame: 24 hours

ArmMeasureGroupValue (MEAN)Dispersion
Holter DeviceDuration of the Holter ExamInnovative 3-lead Wireless Water Resistant Holter System21.79 hoursStandard Deviation 1.4
Holter DeviceDuration of the Holter ExamConventional Holter System21.74 hoursStandard Deviation 1.95
Secondary

Episodes of Heart Beat Pauses

Total number of heart beat pauses ≥ 2.5 seconds detected in both Holter monitors

Time frame: 24 hours

ArmMeasureGroupValue (MEAN)Dispersion
Holter DeviceEpisodes of Heart Beat PausesInnovative 3-lead Wireless Water Resistant Holter System1.74 heart beat pauses in 24hrStandard Deviation 12.13
Holter DeviceEpisodes of Heart Beat PausesConventional Holter System1.84 heart beat pauses in 24hrStandard Deviation 12.43
Comparison: The quantitative secondary outcomes were analyzed using Bland-Altman graphs as well as the Lin´s concordance correlation coefficient (CCC).95% CI: [-1.75, 1.56]
Comparison: The quantitative secondary outcomes were analyzed using Bland-Altman graphs as well as the Lin´s concordance correlation coefficient (CCC).95% CI: [0.997, 0.998]
Secondary

Episodes of Supraventricular Tachycardia

Total number of supraventricular tachycardia episodes detected in both Holter monitors

Time frame: 24 hours

ArmMeasureGroupValue (MEAN)Dispersion
Holter DeviceEpisodes of Supraventricular TachycardiaConventional Holter System6.37 supraventricular tachycardia episodesStandard Deviation 41.1
Holter DeviceEpisodes of Supraventricular TachycardiaInnovative 3-lead Wireless Water Resistant Holter System5.71 supraventricular tachycardia episodesStandard Deviation 38.59
Comparison: The quantitative secondary outcomes were analyzed using Bland-Altman graphs as well as the Lin´s concordance correlation coefficient (CCC).95% CI: [-13.38, 12.07]
Comparison: The quantitative secondary outcomes were analyzed using Bland-Altman graphs as well as the Lin´s concordance correlation coefficient (CCC).95% CI: [0.98, 0.99]
Secondary

Episodes of Ventricular Tachycardia

Total number of ventricular tachycardia episodes detected in both Holter monitors

Time frame: 24 hours

ArmMeasureGroupValue (MEAN)Dispersion
Holter DeviceEpisodes of Ventricular TachycardiaInnovative 3-lead Wireless Water Resistant Holter System1.12 ventricular tachycardia episodes in 24hrStandard Deviation 6.29
Holter DeviceEpisodes of Ventricular TachycardiaConventional Holter System1.24 ventricular tachycardia episodes in 24hrStandard Deviation 7.86
Comparison: The quantitative secondary outcomes were analyzed using Bland-Altman graphs as well as the Lin´s concordance correlation coefficient (CCC).95% CI: [-5.45, 5.2]
Comparison: The quantitative secondary outcomes were analyzed using Bland-Altman graphs as well as the Lin´s concordance correlation coefficient (CCC).95% CI: [0.91, 0.94]
Secondary

Holter System Convenience of Use

Number of Participants Reporting the convenience of use of the Holter systems

Time frame: 24 hours

Population: Only 98 participants of the total of 179 responded the survey to assess the convenience of use of the Holter systems

ArmMeasureGroupValue (COUNT_OF_PARTICIPANTS)
Holter DeviceHolter System Convenience of UseInnovative 3-lead Wireless Water Resistant Holter System54 Participants
Holter DeviceHolter System Convenience of UseConventional Holter System14 Participants
Holter DeviceHolter System Convenience of UseNo difference30 Participants
p-value: <0.0001Chi-squared
Secondary

Maximum Heart Rate

Maximum heart rate measured in both Holter monitors

Time frame: 24 hours

ArmMeasureGroupValue (MEAN)Dispersion
Holter DeviceMaximum Heart RateInnovative 3-lead Wireless Water Resistant Holter System113.59 beats per minuteStandard Deviation 23.17
Holter DeviceMaximum Heart RateConventional Holter System116.30 beats per minuteStandard Deviation 24.76
Comparison: The quantitative secondary outcomes were analyzed using Bland-Altman graphs as well as the Lin´s concordance correlation coefficient (CCC).95% CI: [-17.09, 11.67]
Comparison: The quantitative secondary outcomes were analyzed using Bland-Altman graphs as well as the Lin´s concordance correlation coefficient (CCC).95% CI: [0.93, 0.96]
Secondary

Mean Heart Rate

Mean heart rate measured in both Holter monitors

Time frame: 24 hours

ArmMeasureGroupValue (MEAN)Dispersion
Holter DeviceMean Heart RateInnovative 3-lead Wireless Water Resistant Holter System69.73 beats per minuteStandard Deviation 12.1
Holter DeviceMean Heart RateConventional Holter System70.67 beats per minuteStandard Deviation 12.24
Comparison: The quantitative secondary outcomes were analyzed using Bland-Altman graphs as well as the Lin´s concordance correlation coefficient (CCC).95% CI: [-6.2, 4.31]
Comparison: The quantitative secondary outcomes were analyzed using Bland-Altman graphs as well as the Lin´s concordance correlation coefficient (CCC).95% CI: [0.96, 0.98]
Secondary

Minimum Heart Rate

Minimum heart rate measured in both Holter monitors

Time frame: 24 hours

ArmMeasureGroupValue (MEAN)Dispersion
Holter DeviceMinimum Heart RateInnovative 3-lead Wireless Water Resistant Holter System49.56 beats per minuteStandard Deviation 9.83
Holter DeviceMinimum Heart RateConventional Holter System49.17 beats per minuteStandard Deviation 8.87
Comparison: The quantitative secondary outcomes were analyzed using Bland-Altman graphs as well as the Lin´s concordance correlation coefficient (CCC).95% CI: [-5.43, 6.22]
Comparison: The quantitative secondary outcomes were analyzed using Bland-Altman graphs as well as the Lin´s concordance correlation coefficient (CCC).95% CI: [0.93, 0.96]
Secondary

Ventricular Ectopic Heart Beats

Total number of ventricular ectopic heart beats measured in both Holter monitors

Time frame: 24 hours

ArmMeasureGroupValue (MEAN)Dispersion
Holter DeviceVentricular Ectopic Heart BeatsInnovative 3-lead Wireless Water Resistant Holter System1198.89 ventricular ectopic heart beats in 24 hrStandard Deviation 2651.33
Holter DeviceVentricular Ectopic Heart BeatsConventional Holter System1187.85 ventricular ectopic heart beats in 24 hrStandard Deviation 2696.69
Comparison: The quantitative secondary outcomes were analyzed using Bland-Altman graphs as well as the Lin´s concordance correlation coefficient (CCC).95% CI: [-2089.58, 2111.65]
Comparison: The quantitative secondary outcomes were analyzed using Bland-Altman graphs as well as the Lin´s concordance correlation coefficient (CCC).95% CI: [0.89, 0.94]

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026