Myopia
Conditions
Brief summary
The main objective of the study was to evaluate the clinical acceptance of soft contact lenses for Myopia control.
Detailed description
Participants were randomized to wear each study contact lenses for 10 days with investigator and participant masking. The participants who are not contact lens wearer at enrollment wore single vision contact lenses for two weeks during the pre-investigation phase to get adapted to contact lens wear.
Interventions
Participants were randomized to wear omafilcon A control soft contact lens for 10 days.
Participants were randomized to wear omafilcon A test soft contact lens 1 for 10 days.
Participants were randomized to wear omafilcon A test soft contact lens 2 for 10 days.
Sponsors
Study design
Eligibility
Inclusion criteria
* Age 10 to 16 years; * Spectacle refraction: -0.75 to -6.00D spherical equivalent, maximum anisometropia 1.25D, cylinder up to -1.00DC * Best corrected visual acuity of at least 20/25 in each eye. * Parents/guardians and participant have read and understood the Participant Information Sheet; * Parents/guardians and participant have read, signed and dated the Informed Consent; * Have normal eyes with the exception of the need for visual correction; * Be willing and able to adhere to the instructions set in the clinical protocol and maintain the appointment schedule.
Exclusion criteria
* Ocular anterior segment infection, inflammation, abnormality, or active disease that would contraindicate contact lens wear; * Newly prescribed use of some systemic or ocular medications for which contact lens wear could be contraindicated as determined by the investigator; * Monocular participants (only one eye with functional vision) or participants fit with only one lens; * Subjects with slit lamp findings greater than grade 1 (e.g. edema, infiltrates, corneal neovascularization, corneal staining, tarsal abnormalities, conjunctival, anterior segment inflammation) as per ISO 11980, any previous history or signs of a contact lens related corneal inflammatory event (past corneal ulcers), or any other ocular abnormality that may contraindicate contact lens wear at the enrolment visit; * History of herpetic keratitis, ocular surgery or irregular cornea; * Enrolment of the family members of the investigator, family members of the investigator's staff, or individuals living in the households of these individuals.
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| Overall Vision Satisfaction | 10 days | Overall vision satisfaction recorded on a visual analog scale of 0 to100 (0 - Not Happy, 100 - Very Happy) |
| Habitual Daily Wearing Time | 10 days | Habitual daily wearing time in hours |
Secondary
| Measure | Time frame | Description |
|---|---|---|
| Visual Satisfaction During the Use of Ipad With White Screen- How Clear is the Close-up Vision. | 10 Days | Visual satisfaction during specific visual tasks on visual analog scale of 0 to100 (0 - Not Clear at All, 100 - Very Clear) |
| Visual Acuity | 10 days | Visual acuity measured using logMAR |
| Visual Satisfaction During the Use of Ipad With Black Screen- How Clear is the Close-up the Vision is. | 10 Days | Visual satisfaction during specific visual tasks on visual analog scale of 0 to100 (0 - Not Clear at All, 100 - Very Clear) |
| Visual Satisfaction During the Use of iPad With Black Screen- How Happy the Vision is. | 10 Days | Visual satisfaction during specific visual tasks on visual analog scale of 0 to100 (0 - Not Happy, 100 - Very Happy) |
| Overall Visual Satisfaction During the Use of iPad With White Screen | 10 days | Visual satisfaction during specific visual tasks on visual analog scale of 0 to100 (0 - Not Happy, 100 - Very Happy) |
Countries
United Kingdom
Participant flow
Participants by arm
| Arm | Count |
|---|---|
| Overall Study Nineteen total participants were analyzed out of the 20 participants who completed the study. | 19 |
| Total | 19 |
Withdrawals & dropouts
| Period | Reason | FG000 | FG001 | FG002 |
|---|---|---|---|---|
| Overall Study | Lost to Follow-up | 1 | 0 | 0 |
Baseline characteristics
| Characteristic | Overall Study | — |
|---|---|---|
| Age, Continuous | 14.4 Years STANDARD_DEVIATION 1.9 | — |
| Race and Ethnicity Not Collected | — | — Participants |
| Sex: Female, Male Female | 11 Participants | — |
| Sex: Female, Male Male | 8 Participants | — |
Adverse events
| Event type | EG000 affected / at risk | EG001 affected / at risk | EG002 affected / at risk |
|---|---|---|---|
| deaths Total, all-cause mortality | 0 / 19 | 0 / 19 | 0 / 19 |
| other Total, other adverse events | 2 / 19 | 0 / 19 | 0 / 19 |
| serious Total, serious adverse events | 0 / 19 | 0 / 19 | 0 / 19 |
Outcome results
Habitual Daily Wearing Time
Habitual daily wearing time in hours
Time frame: 10 days
Population: 19 total participants were analyzed out of the 20 participants. One participant was not included in the study because of the compliance issue (Did not complete required hours with lens).
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Omafilcon A Control Lens | Habitual Daily Wearing Time | 7.9 Hours/Day | Standard Deviation 2 |
| Omafilcon A Test Lens 1 | Habitual Daily Wearing Time | 8.1 Hours/Day | Standard Deviation 2.2 |
| Omafilcon A Test Lens 2 | Habitual Daily Wearing Time | 7.8 Hours/Day | Standard Deviation 2.3 |
Overall Vision Satisfaction
Overall vision satisfaction recorded on a visual analog scale of 0 to100 (0 - Not Happy, 100 - Very Happy)
Time frame: 10 days
Population: 19 total participants were analyzed out of the 20 participants. One participant was not included in the study because of the compliance issue (Did not complete required hours with lens).
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Omafilcon A Control Lens | Overall Vision Satisfaction | 87.7 units on a scale | Standard Deviation 12.1 |
| Omafilcon A Test Lens 1 | Overall Vision Satisfaction | 79.7 units on a scale | Standard Deviation 16.5 |
| Omafilcon A Test Lens 2 | Overall Vision Satisfaction | 71.8 units on a scale | Standard Deviation 27.3 |
Overall Visual Satisfaction During the Use of iPad With White Screen
Visual satisfaction during specific visual tasks on visual analog scale of 0 to100 (0 - Not Happy, 100 - Very Happy)
Time frame: 10 days
Population: 19 total participants were analyzed out of the 20 participants. One participant was not included in the study because of the compliance issue (Did not complete required hours with lens).
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Omafilcon A Control Lens | Overall Visual Satisfaction During the Use of iPad With White Screen | 90.8 units on a scale | Standard Deviation 9.3 |
| Omafilcon A Test Lens 1 | Overall Visual Satisfaction During the Use of iPad With White Screen | 90.8 units on a scale | Standard Deviation 10 |
| Omafilcon A Test Lens 2 | Overall Visual Satisfaction During the Use of iPad With White Screen | 85.1 units on a scale | Standard Deviation 16.2 |
Visual Acuity
Visual acuity measured using logMAR
Time frame: 10 days
Population: 19 total participants were analyzed out of the 20 participants. One participant was not included in the study because of the compliance issue (Did not complete required hours with lens).
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Omafilcon A Control Lens | Visual Acuity | 0.06 logMAR | Standard Deviation 0.07 |
| Omafilcon A Test Lens 1 | Visual Acuity | 0.08 logMAR | Standard Deviation 0.07 |
| Omafilcon A Test Lens 2 | Visual Acuity | 0.10 logMAR | Standard Deviation 0.09 |
Visual Satisfaction During the Use of Ipad With Black Screen- How Clear is the Close-up the Vision is.
Visual satisfaction during specific visual tasks on visual analog scale of 0 to100 (0 - Not Clear at All, 100 - Very Clear)
Time frame: 10 Days
Population: 19 total participants were analyzed out of the 20 participants. One participant was not included in the study because of the compliance issue (Did not complete required hours with lens).
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Omafilcon A Control Lens | Visual Satisfaction During the Use of Ipad With Black Screen- How Clear is the Close-up the Vision is. | 85.3 units on a scale | Standard Deviation 14 |
| Omafilcon A Test Lens 1 | Visual Satisfaction During the Use of Ipad With Black Screen- How Clear is the Close-up the Vision is. | 87.5 units on a scale | Standard Deviation 15.1 |
| Omafilcon A Test Lens 2 | Visual Satisfaction During the Use of Ipad With Black Screen- How Clear is the Close-up the Vision is. | 79.2 units on a scale | Standard Deviation 22.3 |
Visual Satisfaction During the Use of iPad With Black Screen- How Happy the Vision is.
Visual satisfaction during specific visual tasks on visual analog scale of 0 to100 (0 - Not Happy, 100 - Very Happy)
Time frame: 10 Days
Population: 19 total participants were analyzed out of the 20 participants. One participant was not included in the study because of the compliance issue (Did not complete required hours with lens).
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Omafilcon A Control Lens | Visual Satisfaction During the Use of iPad With Black Screen- How Happy the Vision is. | 85.4 units on a scale | Standard Deviation 16.1 |
| Omafilcon A Test Lens 1 | Visual Satisfaction During the Use of iPad With Black Screen- How Happy the Vision is. | 86.8 units on a scale | Standard Deviation 15.7 |
| Omafilcon A Test Lens 2 | Visual Satisfaction During the Use of iPad With Black Screen- How Happy the Vision is. | 80.0 units on a scale | Standard Deviation 20 |
Visual Satisfaction During the Use of Ipad With White Screen- How Clear is the Close-up Vision.
Visual satisfaction during specific visual tasks on visual analog scale of 0 to100 (0 - Not Clear at All, 100 - Very Clear)
Time frame: 10 Days
Population: 19 total participants were analyzed out of the 20 participants. One participant was not included in the study because of the compliance issue (Did not complete required hours with lens).
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| Omafilcon A Control Lens | Visual Satisfaction During the Use of Ipad With White Screen- How Clear is the Close-up Vision. | 90.7 units on a scale | Standard Deviation 9.5 |
| Omafilcon A Test Lens 1 | Visual Satisfaction During the Use of Ipad With White Screen- How Clear is the Close-up Vision. | 90.0 units on a scale | Standard Deviation 13.3 |
| Omafilcon A Test Lens 2 | Visual Satisfaction During the Use of Ipad With White Screen- How Clear is the Close-up Vision. | 84.4 units on a scale | Standard Deviation 15.4 |