Cardiac Output, LIDCO Rapid, Septic Shock
Conditions
Keywords
cardiac output , LIDCO . Septic shock
Brief summary
Transthoracic echocardiography (TTE) has been used as a reference technique for CO measurement and its accuracy has been adequate for many clinical uses. TTE is painless, safe and non-invasive. Some drawbacks of TTE are being expensive, bulky and needs advanced training. Currently, due to increased interest in minimally invasive hemodynamic monitoring, multiple new methods have become commercially available to assess CO: the arterial pulse aortic flow, bioreactance, and bioimpedance. One of these new devices is LiDCOrapid (LiDCO Ltd, Cambridge, UK). It is a new, minimally invasive monitor which estimates beat-by-beat CO and fluid responsiveness from the arterial waveform. It has been validated after vascular, urological and thoracic surgery and shows good agreement with the PA catheter in swine, in patients with normal left ventricular systolic function and in patients with vasodilatation. However, LiDCO has not yet been validated in patients with septic shock. Therefore, the aim of this study is to validate LiDCOrapid parameters in septic shock patients using TTE derived parameters as the non-invasive reference.
Detailed description
Obtaining a patient's cardiac output (CO) could contribute to optimized, safe hemodynamic control. Accurate CO measurements can serve as a guide for resuscitation therapy, catecholamine use, differential diagnosis, and intervention during a circulatory failure. Although the thermodilution technique via a pulmonary artery catheter (PAC) has an invasive and intermittent, it remains the gold standard for CO measurements. Previously monitoring of CO required invasive and costly devices limiting their use to intensive care setting. More recently ultrasound is increasingly being used and offers a range of non-invasive tools to estimate CO. Transthoracic echocardiography (TTE) has been used as a reference technique for CO measurement and its accuracy has been adequate for many clinical uses. TTE is painless, safe and non-invasive. Some drawbacks of TTE are being expensive, bulky and needs advanced training. Currently, due to increased interest in minimally invasive hemodynamic monitoring, multiple new methods have become commercially available to assess CO: the arterial pulse aortic flow, bioreactance, and bioimpedance. One of these new devices is LiDCOrapid (LiDCO Ltd, Cambridge, UK). It is a new, minimally invasive monitor which estimates beat-by-beat CO and fluid responsiveness from the arterial waveform. It has been validated after vascular, urological and thoracic surgery and shows good agreement with the PA catheter in swine, in patients with normal left ventricular systolic function and in patients with vasodilatation. However, LiDCO has not yet been validated in patients with septic shock. Therefore, the aim of this study is to validate LiDCOrapid parameters in septic shock patients using TTE derived parameters as the non-invasive reference.
Interventions
give two-volume of fluids initial 250ml then another 250 ml
Sponsors
Study design
Eligibility
Inclusion criteria
1. Age \> 18 years. 2. Patients admitted to ICU with septic shock
Exclusion criteria
1. Patients with poor Echo window. 2. If they presented with a history of heart failure, valvular disease or arrhythmias.
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| primary outcome | on patient admission to the icu for 1 hour | To measure cardiac output difference between LiDCOrapid measurement and echocardiographic measurement . |
Secondary
| Measure | Time frame | Description |
|---|---|---|
| secondary outcome | on patient admission to the icu for 1 hour | to measure dose of vasopressors and inotropic agents |
Countries
Egypt