Skip to content

A Prospective Study to Evaluate Clinical Performance of Thermalytix in Detecting Breast Cancers

A Prospective Study to Evaluate the Effectiveness of Thermalytix (AI-based Thermographic Solution Developed by Niramai) for Breast Cancer Screening as Compared to Standard Screening Modalities for Breast Cancer

Status
Completed
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT04688086
Enrollment
687
Registered
2020-12-29
Start date
2018-12-15
Completion date
2020-01-30
Last updated
2024-10-17

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Breast Cancer, Early Detection of Cancer

Keywords

Breast cancer, Thermography, Thermalytix, Artificial Intelligence, Machine Learning, Dense Breasts, Breast Density, Mammography, Breast Cancer Screening

Brief summary

The purpose of this study is to determine the clinical performance of AI-based Thermalytix with the current standard-of-care diagnostic modalities in women.

Detailed description

While mammography is the most accepted breast cancer screening test, it is less sensitive in women with dense breast tissue, who have a four fold high risk for developing breast cancer. Thermalytix is a novel automated breast cancer screening solution that uses artificial intelligence (AI) over thermal images to detect and localise breast cancer lesions in women. In this study the performance of Thermalytix over mammography is evaluated in women with dense and non-dense breast tissue who presented for a health check-up at a hospital. All women underwent Thermalytix and mammography, and participants who were reported as positive on either test were recommended for further investigations. Sensitivity, specificity, positive predictive value, and negative predictive value were calculated across age-groups, menopausal status, and breast densities against the current standard-of-care diagnostic modalities.

Interventions

DIAGNOSTIC_TESTThermalytix

Thermalytix is an Artificial intelligence based automated breast screening solution that analyzes thermal distribution on the breast to generate a breast health score automatically. Thermal imaging was performed by a trained technician to capture thermal images of the participant in five views. These thermal images were uploaded to Thermalytix software on the cloud where it was automatically analyzed by AI-based Thermalytix computer-aided detection (CADe) engine. This CADe engine analyzes uploaded thermal images and outputs an interpretation report for each participant with quantitative scores corresponding to computed probability of malignancy based on the structural, vascular, areolar, thermal properties of the observed abnormality. Thermalytix also generates annotated images with markings of abnormal regions and an overall Thermalytix score suggesting likelihood of breast malignancy. The locked AI model Thermalytix algorithm version 3, dated December 2018 was used for the analysis.

Sponsors

Max Healthcare Insititute Limited
CollaboratorOTHER
Niramai Health Analytix Private Limited
Lead SponsorINDUSTRY

Study design

Observational model
COHORT
Time perspective
PROSPECTIVE

Eligibility

Sex/Gender
FEMALE
Age
18 Years to 80 Years
Healthy volunteers
Yes

Inclusion criteria

1. Female subjects equal to and above 18 years 2. Subjects who are willing to give written informed consent for study participation 3. Subjects who are ready to comply with the study related visits and procedures

Exclusion criteria

1. Subjects who are pregnant 2. Subjects who are lactating 3. Subjects who have undergone either lumpectomy or mastectomy 4. Subjects who have undergone chemotherapy in the last 2 weeks at the time of study enrollment 5. Any active illness, psychological and/or pathological condition that would interfere with study participation in the opinion of the Investigator

Design outcomes

Primary

MeasureTime frameDescription
Sensitivity, Specificity, Positive Predictive Value and Negative Predictive Value of Thermalytix2 daysTo assess the clinical performance of Thermalytix as compared to standard screening modalities. Sensitivity, specificity, positive predictive value and negative predictive value of Thermalytix

Secondary

MeasureTime frameDescription
Sensitivity and Specificity of Thermalytix for Women With Dense Breast Tissue (ACR Category C or D)2 daysTo assess the influence of patient characteristics (Breast density) on the diagnostic accuracy of Thermalytix.
Sensitivity and Specificity of Thermalytix in Women With Breast Density ACR Category A or B2 daysTo assess the influence of patient characteristics (Breast density) on the diagnostic accuracy of Thermalytix. Sensitivity and Specificity of Thermalytix in Women with Fatty Breasts - Breast Density ACR category A or B

Countries

India

Participant flow

Participants by arm

ArmCount
Women With no Personal History of Breast Cancer
Women who came in for a breast mammography between ages 30 and 80 years were invited to take part in the study. All the women included in the study underwent breast cancer screening first by Thermalytix, the AI-based thermal imaging test, followed by mammography. Thermalytix: Thermalytix is an Artificial intelligence based automated breast screening solution that analyzes thermal distribution on the breast to generate a breast health score automatically. Thermal imaging was performed by a trained technician to capture thermal images of the participant in five views. These thermal images were uploaded to Thermalytix software on the cloud where it was automatically analyzed by AI-based Thermalytix computer-aided detection (CADe) engine. This CADe engine analyzes uploaded thermal images and outputs an interpretation report for each participant with quantitative scores corresponding to computed probability of malignancy based on the structural, vascular, areolar, thermal properties of the observed abnormality. Thermalytix also generates annotated images with markings of abnormal regions and an overall Thermalytix score suggesting likelihood of breast malignancy. The locked AI model Thermalytix algorithm version 3, dated December 2018 was used for the analysis.
459
Total459

Withdrawals & dropouts

PeriodReasonFG000
Overall StudyImaging error;64
Overall StudyLost to Follow-up164

Baseline characteristics

CharacteristicWomen With no Personal History of Breast Cancer
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
18 Participants
Age, Categorical
Between 18 and 65 years
441 Participants
Race and Ethnicity Not Collected— Participants
Region of Enrollment
India
459 participants
Sex: Female, Male
Female
459 Participants
Sex: Female, Male
Male
0 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 459
other
Total, other adverse events
0 / 459
serious
Total, serious adverse events
0 / 459

Outcome results

Primary

Sensitivity, Specificity, Positive Predictive Value and Negative Predictive Value of Thermalytix

To assess the clinical performance of Thermalytix as compared to standard screening modalities. Sensitivity, specificity, positive predictive value and negative predictive value of Thermalytix

Time frame: 2 days

Population: All women who satisfied the inclusion criteria

ArmMeasureGroupValue (NUMBER)
Women With no Personal History of Breast CancerSensitivity, Specificity, Positive Predictive Value and Negative Predictive Value of ThermalytixSensitivity95.24 percentage
Women With no Personal History of Breast CancerSensitivity, Specificity, Positive Predictive Value and Negative Predictive Value of ThermalytixSpecificity88.58 percentage
Women With no Personal History of Breast CancerSensitivity, Specificity, Positive Predictive Value and Negative Predictive Value of ThermalytixNPV99.74 percentage
Women With no Personal History of Breast CancerSensitivity, Specificity, Positive Predictive Value and Negative Predictive Value of ThermalytixPPV28.57 percentage
Secondary

Sensitivity and Specificity of Thermalytix for Women With Dense Breast Tissue (ACR Category C or D)

To assess the influence of patient characteristics (Breast density) on the diagnostic accuracy of Thermalytix.

Time frame: 2 days

Population: Women with breast density ACR category C or D

ArmMeasureGroupValue (NUMBER)
Women With no Personal History of Breast CancerSensitivity and Specificity of Thermalytix for Women With Dense Breast Tissue (ACR Category C or D)Sensitivity100 percentage
Women With no Personal History of Breast CancerSensitivity and Specificity of Thermalytix for Women With Dense Breast Tissue (ACR Category C or D)Specificity81.65 percentage
Secondary

Sensitivity and Specificity of Thermalytix in Women With Breast Density ACR Category A or B

To assess the influence of patient characteristics (Breast density) on the diagnostic accuracy of Thermalytix. Sensitivity and Specificity of Thermalytix in Women with Fatty Breasts - Breast Density ACR category A or B

Time frame: 2 days

Population: Women with Breast Density ACR category A or B

ArmMeasureGroupValue (NUMBER)
Women With no Personal History of Breast CancerSensitivity and Specificity of Thermalytix in Women With Breast Density ACR Category A or BSensitivity90.9 percentage
Women With no Personal History of Breast CancerSensitivity and Specificity of Thermalytix in Women With Breast Density ACR Category A or BSpecificity92.5 percentage

Source: ClinicalTrials.gov · Data processed: Feb 15, 2026