Skip to content

Evaluation of a Low-cost CPAP Device on Hospitalized COVID-19 Patients

Safety and Efficacy Evaluation of a Low-cost CPAP Device for Hypoxemic COVID-19 Patients: A Pilot Study

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT04681859
Acronym
OxyJet-CPAP
Enrollment
45
Registered
2020-12-23
Start date
2021-04-17
Completion date
2021-08-09
Last updated
2022-03-25

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Covid19, Hypoxemia

Brief summary

The aim of this study is to evaluate the preliminary safety and performance of a low-cost locally-made Venturi-based Non-invasive Positive Pressure Ventilator (NIPPV) device for hypoxemic COVID-19 patients. The device administers Continuous Positive Airway Pressure (CPAP) therapy using the jet-mixing or Venturi effect to increase the volume flow rate of oxygenated air from a pressurized cylinder by entraining the atmospheric air. To provide CPAP therapy, this high flow of oxygenated air is delivered to the patient via a low-cost non-vented mask with a tight seal with a High-Efficiency Particulate Air (HEPA) filter connected to the exhalation limb. The tight seal and HEPA filter ensures a minimal risk of aerosol generation and thus the device can be used without a negative pressure room. The system consists of the developed Venturi-based flow-generator, a standard 22mm breathing tube, a standard Y-connector, a non-vented CPAP mask (e.g., snorkel mask, helmet), a HEPA filter, and a Positive End Expiratory Pressure (PEEP) valve. The bench-top testing of the device is done in the laboratories of BUET and was verified that the device performs within the CPAP guidelines provided by the Medicines and Healthcare products Regulatory Agency (MHRA), UK. This study aims to assess the safety of and efficacy of the device in three different steps: (1) design validation, (2) clinical feasibility and (3) pilot clinical trial for safety and efficacy evaluation. Only if the device successfully passes the parts 1 and 2, the investigators will proceed to the final clinical trial in step 3. In this final step, the investigators aim to conduct a randomized controlled trial (RCT) evaluating for non-inferiority of the CPAP intervention compared to standard HFNO treatment. The number of ventilator-free days will be used as the primary outcome for efficacy, while patient recovery, death, or need of intubation and other adverse events will be used as secondary outcomes.

Interventions

DEVICEHigh Flow Nasal Oxygen (HFNO) treatment

Administration of High Flow Nasal Oxygen (HFNO) to patients who did not respond to standard low-flow oxygen therapy (non-rebreather mask on 15L/min). Adjust flow-rate and FiO2 as per standard HFNO treatment protocol.

DEVICEContinuous Positive Airway Pressure (CPAP) therapy using OxyJet

Administer CPAP to patients who did not respond to standard low-flow oxygen therapy (non-rebreather mask on 15L/min). CPAP initiated with FiO2 at 40% and a 10cm PEEP. Titrate PEEP to 15 cm if required. Deliver additional oxygen via secondary port if a higher FiO2 is required.

Sponsors

Dhaka Medical College
CollaboratorOTHER
Bangladesh University of Engineering and Technology
Lead SponsorOTHER

Study design

Allocation
RANDOMIZED
Intervention model
PARALLEL
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to 65 Years
Healthy volunteers
No

Inclusion criteria

* Age ≥ 18 years * Confirmed or suspected COVID-19 patient (by RT-PCR) having severe pneumonia and hypoxemia (SpO2 ≤90%) who did not respond to standard oxygen therapy (non-rebreather mask on 15L/min at 100% FiO2).

Exclusion criteria

* Severely hypoxemic patients (SpO2≤85%) * Patients with low respiratory drive or requiring cardiopulmonary resuscitation * Patients with contraindications for CPAP * Pregnant status * Age \> 65 years

Design outcomes

Primary

MeasureTime frameDescription
Number of ventilator-free days10 daysThe total number of days the patient was able to avoid being placed under a mechanical ventilator.

Secondary

MeasureTime frameDescription
Recovery of the patient30 daysThe event that the patient has recovered and released from the hospital within 30 days.
Death or need of intubation10 daysThe event that the patient dies or requires to be placed under a mechanical ventilator.
Oxygen toxicity or other adverse events10 daysOxygen toxicity or other adverse event rating according to the CTCAE scale (1-5)

Countries