Aerosol Particle Concentrations Among Different Oxygen Devices for Spontaneous Breathing Patients With Tracheostomy

Aerosol Particle Concentrations Among Different Oxygen Devices for Spontaneous Breathing Patients With Tracheostomy: a Randomized Cross-over Trial

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT04654754
Enrollment
12
Registered
2020-12-04
Start date
2020-12-23
Completion date
2021-07-16
Last updated
2022-12-21

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Transmission, Patient-Professional

Keywords

aerosol generating procedure, tracheostomy, oxygen therapy

Brief summary

For spontaneous breathing patients with tracheostomy, whose lower airway is directly opened to the room air, the aerosol particles generated by the patients would be directly dispersed into the room air, which might be an direct resource of virus transmission. However, the transmission risk has not been evaluated and the appropriate humidification therapy is unknown. Thus this study is aimed to investigate the aerosol particle concentrations among different oxygen devices for spontaneous breathing patients with tracheostomy, in order to reflect the transmission risk.

Detailed description

The transmission route of the SARS-CoV-2 virus remains controversial, and concerns persist of potentially increased virus transmission when utilizing high-flow devices and aerosol devices among COVID-19 patients. For spontaneous breathing patients with tracheostomy, whose lower airway is directly opened to the room air, the aerosol particles generated by the patients would be directly dispersed into the room air, which might be an direct resource of virus transmission. However, the transmission risk of tracheostomy during spontaneous breathing has not been evaluated and the appropriate humidification therapy is unknown. Thus this study is aimed to investigate the aerosol particle concentrations among different oxygen devices for spontaneous breathing patients with tracheostomy, in order to reflect the transmission risk.

Interventions

DEVICEhigh-flow high humidity oxygen device with tracheostomy adapter

This device can provide heat and humidified gas for spontaneous breathing patients with tracheostomy at a high gas flow rate.

Sponsors

Jie Li
Lead SponsorOTHER

Study design

Allocation
RANDOMIZED
Intervention model
CROSSOVER
Primary purpose
OTHER
Masking
SINGLE (Outcomes Assessor)

Masking description

the device utilized in the study will be labeled as 1,2,3,4 and 5, the investigator who performed the statistical analysis will be blinded for the device

Intervention model description

five oxygen and humidification different devices will be used for the enrolled patients, with 5 minutes each. Aerosol particle concentrations at 1 foot away from the patients will be measured.

Eligibility

Sex/Gender
ALL
Age
18 Years to 90 Years
Healthy volunteers
No

Inclusion criteria

* adults; * tracheostomy; * able to spontaneous breathing without ventilator support

Exclusion criteria

* confirmed diagnosis of COVID-19 within recent two weeks; * non-English speaking; * refuse to participate in the study; * palliative care; * receiving ECMO; * unable to connect with tracheostomy adapter, such as laryngectomy tube

Design outcomes

Primary

MeasureTime frameDescription
Aerosol Particle Concentrations With Size of 0.3-0.5 Micrometer at 1 Foot Away From the Patient5 minutes after using the deviceaerosol particle concentrations (the concentrations of particles inside the room air, the unit is particles per cubic meters) at 1 foot away from the patient
Aerosol Particle Concentrations With Size of 1-3 Micrometer at 1 Foot Away From the Patient5 minutes after using the deviceaerosol particle concentrations (the concentrations of particles inside the room air, the unit is particles per cubic meters) at 1 foot away from the patient
Aerosol Particle Concentrations With Size of <0.3 Micrometer at 1 Foot Away From the Patient5 minutes after using the deviceaerosol particle concentrations (the concentrations of particles inside the room air, the unit is particles per cubic meters) at 1 foot away from the patient
Aerosol Particle Concentrations With Size of 5-10 Micrometer at 1 Foot Away From the Patient5 minutes after using the deviceaerosol particle concentrations (the concentrations of particles inside the room air, the unit is particles per cubic meters) at 1 foot away from the patient
Aerosol Particle Concentrations With Size of 0.5-1 Micrometer at 1 Foot Away From the Patient5 minutes after using the deviceaerosol particle concentrations (the concentrations of particles inside the room air, the unit is particles per cubic meters) at 1 foot away from the patient
Aerosol Particle Concentrations With Size of 3-5 Micrometer at 1 Foot Away From the Patient5 minutes after using the deviceaerosol particle concentrations (the concentrations of particles inside the room air, the unit is particles per cubic meters) at 1 foot away from the patient

Secondary

MeasureTime frameDescription
Patient Comfort With Different Oxygen Devices5 minutes after using the devicePatients would scale their comfort on a 5-point Likert scale, 1 was the most uncomfortable, and 5 was the most comfortable.

Countries

United States

Participant flow

Recruitment details

This is a randomized crossover study, all the 12 patients received the five treatments in a random order.

Participants by arm

ArmCount
Spontaneous Breathing Patients With Tracheostomy
Patients who had tracheostomy and spontaneous breathing, without ventilator support
12
Total12

Baseline characteristics

CharacteristicSpontaneous Breathing Patients With Tracheostomy
Age, Continuous50.5 years
STANDARD_DEVIATION 16.6
Race and Ethnicity Not Collected— Participants
Sex: Female, Male
Female
6 Participants
Sex: Female, Male
Male
6 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 12
other
Total, other adverse events
0 / 12
serious
Total, serious adverse events
0 / 12

Outcome results

Primary

Aerosol Particle Concentrations With Size of 0.3-0.5 Micrometer at 1 Foot Away From the Patient

aerosol particle concentrations (the concentrations of particles inside the room air, the unit is particles per cubic meters) at 1 foot away from the patient

Time frame: 5 minutes after using the device

ArmMeasureGroupValue (MEDIAN)
Spontaneous Breathing Patients With TracheostomyAerosol Particle Concentrations With Size of 0.3-0.5 Micrometer at 1 Foot Away From the PatientHigh-flow High Humidity Oxygen Device With Tracheostomy Adapter610000 particles/m3
Spontaneous Breathing Patients With TracheostomyAerosol Particle Concentrations With Size of 0.3-0.5 Micrometer at 1 Foot Away From the PatientLarge-volume Nebulizer (Cool Aerosol) With Trach Collar730000 particles/m3
Spontaneous Breathing Patients With TracheostomyAerosol Particle Concentrations With Size of 0.3-0.5 Micrometer at 1 Foot Away From the PatientVenturi-adapter With Trach Collar610000 particles/m3
Spontaneous Breathing Patients With TracheostomyAerosol Particle Concentrations With Size of 0.3-0.5 Micrometer at 1 Foot Away From the PatientLarge-volume Nebulizer (Cool Aerosol) With T-piece and a Filter590000 particles/m3
Spontaneous Breathing Patients With TracheostomyAerosol Particle Concentrations With Size of 0.3-0.5 Micrometer at 1 Foot Away From the PatientHigh-flow High Humidity Device With a Scavenger or a Surgical Mask680000 particles/m3
Primary

Aerosol Particle Concentrations With Size of <0.3 Micrometer at 1 Foot Away From the Patient

aerosol particle concentrations (the concentrations of particles inside the room air, the unit is particles per cubic meters) at 1 foot away from the patient

Time frame: 5 minutes after using the device

ArmMeasureGroupValue (MEDIAN)
Spontaneous Breathing Patients With TracheostomyAerosol Particle Concentrations With Size of <0.3 Micrometer at 1 Foot Away From the PatientHigh-flow High Humidity Oxygen Device With Tracheostomy Adapter10500000 particles/m3
Spontaneous Breathing Patients With TracheostomyAerosol Particle Concentrations With Size of <0.3 Micrometer at 1 Foot Away From the PatientLarge-volume Nebulizer (Cool Aerosol) With Trach Collar15200000 particles/m3
Spontaneous Breathing Patients With TracheostomyAerosol Particle Concentrations With Size of <0.3 Micrometer at 1 Foot Away From the PatientVenturi-adapter With Trach Collar12100000 particles/m3
Spontaneous Breathing Patients With TracheostomyAerosol Particle Concentrations With Size of <0.3 Micrometer at 1 Foot Away From the PatientLarge-volume Nebulizer (Cool Aerosol) With T-piece and a Filter11900000 particles/m3
Spontaneous Breathing Patients With TracheostomyAerosol Particle Concentrations With Size of <0.3 Micrometer at 1 Foot Away From the PatientHigh-flow High Humidity Device With a Scavenger or a Surgical Mask13000000 particles/m3
Primary

Aerosol Particle Concentrations With Size of 0.5-1 Micrometer at 1 Foot Away From the Patient

aerosol particle concentrations (the concentrations of particles inside the room air, the unit is particles per cubic meters) at 1 foot away from the patient

Time frame: 5 minutes after using the device

ArmMeasureGroupValue (MEDIAN)
Spontaneous Breathing Patients With TracheostomyAerosol Particle Concentrations With Size of 0.5-1 Micrometer at 1 Foot Away From the PatientLarge-volume Nebulizer (Cool Aerosol) With Trach Collar55000 particles/m3
Spontaneous Breathing Patients With TracheostomyAerosol Particle Concentrations With Size of 0.5-1 Micrometer at 1 Foot Away From the PatientHigh-flow High Humidity Oxygen Device With Tracheostomy Adapter53000 particles/m3
Spontaneous Breathing Patients With TracheostomyAerosol Particle Concentrations With Size of 0.5-1 Micrometer at 1 Foot Away From the PatientVenturi-adapter With Trach Collar46000 particles/m3
Spontaneous Breathing Patients With TracheostomyAerosol Particle Concentrations With Size of 0.5-1 Micrometer at 1 Foot Away From the PatientLarge-volume Nebulizer (Cool Aerosol) With T-piece and a Filter52000 particles/m3
Spontaneous Breathing Patients With TracheostomyAerosol Particle Concentrations With Size of 0.5-1 Micrometer at 1 Foot Away From the PatientHigh-flow High Humidity Device With a Scavenger or a Surgical Mask50000 particles/m3
Primary

Aerosol Particle Concentrations With Size of 1-3 Micrometer at 1 Foot Away From the Patient

aerosol particle concentrations (the concentrations of particles inside the room air, the unit is particles per cubic meters) at 1 foot away from the patient

Time frame: 5 minutes after using the device

ArmMeasureGroupValue (MEDIAN)
Spontaneous Breathing Patients With TracheostomyAerosol Particle Concentrations With Size of 1-3 Micrometer at 1 Foot Away From the PatientHigh-flow High Humidity Oxygen Device With Tracheostomy Adapter13000 particles/m3
Spontaneous Breathing Patients With TracheostomyAerosol Particle Concentrations With Size of 1-3 Micrometer at 1 Foot Away From the PatientLarge-volume Nebulizer (Cool Aerosol) With Trach Collar11000 particles/m3
Spontaneous Breathing Patients With TracheostomyAerosol Particle Concentrations With Size of 1-3 Micrometer at 1 Foot Away From the PatientVenturi-adapter With Trach Collar10000 particles/m3
Spontaneous Breathing Patients With TracheostomyAerosol Particle Concentrations With Size of 1-3 Micrometer at 1 Foot Away From the PatientLarge-volume Nebulizer (Cool Aerosol) With T-piece and a Filter12000 particles/m3
Spontaneous Breathing Patients With TracheostomyAerosol Particle Concentrations With Size of 1-3 Micrometer at 1 Foot Away From the PatientHigh-flow High Humidity Device With a Scavenger or a Surgical Mask13000 particles/m3
Primary

Aerosol Particle Concentrations With Size of 3-5 Micrometer at 1 Foot Away From the Patient

aerosol particle concentrations (the concentrations of particles inside the room air, the unit is particles per cubic meters) at 1 foot away from the patient

Time frame: 5 minutes after using the device

ArmMeasureGroupValue (MEDIAN)
Spontaneous Breathing Patients With TracheostomyAerosol Particle Concentrations With Size of 3-5 Micrometer at 1 Foot Away From the PatientLarge-volume Nebulizer (Cool Aerosol) With Trach Collar6000 particles/m3
Spontaneous Breathing Patients With TracheostomyAerosol Particle Concentrations With Size of 3-5 Micrometer at 1 Foot Away From the PatientVenturi-adapter With Trach Collar5700 particles/m3
Spontaneous Breathing Patients With TracheostomyAerosol Particle Concentrations With Size of 3-5 Micrometer at 1 Foot Away From the PatientLarge-volume Nebulizer (Cool Aerosol) With T-piece and a Filter6000 particles/m3
Spontaneous Breathing Patients With TracheostomyAerosol Particle Concentrations With Size of 3-5 Micrometer at 1 Foot Away From the PatientHigh-flow High Humidity Device With a Scavenger or a Surgical Mask6700 particles/m3
Spontaneous Breathing Patients With TracheostomyAerosol Particle Concentrations With Size of 3-5 Micrometer at 1 Foot Away From the PatientHigh-flow High Humidity Oxygen Device With Tracheostomy Adapter6700 particles/m3
Primary

Aerosol Particle Concentrations With Size of 5-10 Micrometer at 1 Foot Away From the Patient

aerosol particle concentrations (the concentrations of particles inside the room air, the unit is particles per cubic meters) at 1 foot away from the patient

Time frame: 5 minutes after using the device

ArmMeasureGroupValue (MEDIAN)
Spontaneous Breathing Patients With TracheostomyAerosol Particle Concentrations With Size of 5-10 Micrometer at 1 Foot Away From the PatientHigh-flow High Humidity Oxygen Device With Tracheostomy Adapter3500 particles/m3
Spontaneous Breathing Patients With TracheostomyAerosol Particle Concentrations With Size of 5-10 Micrometer at 1 Foot Away From the PatientLarge-volume Nebulizer (Cool Aerosol) With Trach Collar3200 particles/m3
Spontaneous Breathing Patients With TracheostomyAerosol Particle Concentrations With Size of 5-10 Micrometer at 1 Foot Away From the PatientVenturi-adapter With Trach Collar3900 particles/m3
Spontaneous Breathing Patients With TracheostomyAerosol Particle Concentrations With Size of 5-10 Micrometer at 1 Foot Away From the PatientLarge-volume Nebulizer (Cool Aerosol) With T-piece and a Filter3500 particles/m3
Spontaneous Breathing Patients With TracheostomyAerosol Particle Concentrations With Size of 5-10 Micrometer at 1 Foot Away From the PatientHigh-flow High Humidity Device With a Scavenger or a Surgical Mask3200 particles/m3
Secondary

Patient Comfort With Different Oxygen Devices

Patients would scale their comfort on a 5-point Likert scale, 1 was the most uncomfortable, and 5 was the most comfortable.

Time frame: 5 minutes after using the device

ArmMeasureGroupValue (MEAN)Dispersion
Spontaneous Breathing Patients With TracheostomyPatient Comfort With Different Oxygen DevicesHigh-flow High Humidity Oxygen Device With Tracheostomy Adapter3.5 score on a scale (comfort scores 1-5)Standard Deviation 1.5
Spontaneous Breathing Patients With TracheostomyPatient Comfort With Different Oxygen DevicesLarge-volume Nebulizer (Cool Aerosol) With Trach Collar3.8 score on a scale (comfort scores 1-5)Standard Deviation 1.3
Spontaneous Breathing Patients With TracheostomyPatient Comfort With Different Oxygen DevicesVenturi-adapter With Trach Collar3.5 score on a scale (comfort scores 1-5)Standard Deviation 1.2
Spontaneous Breathing Patients With TracheostomyPatient Comfort With Different Oxygen DevicesLarge-volume Nebulizer (Cool Aerosol) With T-piece and a Filter3.3 score on a scale (comfort scores 1-5)Standard Deviation 1.3
Spontaneous Breathing Patients With TracheostomyPatient Comfort With Different Oxygen DevicesHigh-flow High Humidity Device With a Scavenger or a Surgical Mask3.5 score on a scale (comfort scores 1-5)Standard Deviation 1.2