Comparing Pressure Based and Dilution Based Total Lung Capacity Measurements in COPD Patients for Lung Function Diagnostics

Mono-centred, Controlled Cross-over Study Comparing Pressure Based, and Dilution Based Total Lung Capacity Measurements in COPD Patients and Healthy Subjects Using Two Commercially Available Devices for Lung Function Diagnostics

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT04531293
Enrollment
128
Registered
2020-08-28
Start date
2020-12-31
Completion date
2022-04-11
Last updated
2022-06-06

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

COPD

Keywords

COPD, DLCO, spirometry, alveolar volume, body plethysmography

Brief summary

Total Lung Capacity (TLC) of the lungs can be determined using a pressure based method (i.e. by body plethysmography) or it can be determined using a dilution based method (i.e. by single breath dilution during a CO diffusing capacity test). The results of the two methods differ, especially in patients with higher grades of COPD. The study investigates an improved analysis for the determination of TLC using the dilution method; this method has been suggested in the new ATS/ERS DLCO (diffusion capacity of lung for carbon monoxide) standard published in 2017. This study is based on approx. 120 participants. So far, there has not been a study using commercially available equipment. Spirometry, lung volume measurement by body plethysmography and DLCO measurements will be performed on reference device (Masterscreen Body/Diff, Vyaire, USA), Spirometry and DLCO will also be performed on the device under test (EasyOne Pro LAB, ndd Medizintechnik AG, Switzerland) in a crossover design. Both devices are CE marked and FDA approved. The tests performed on both devices are standardized tests performed routinely with patients.

Interventions

DEVICERespiratory Analysis System

Measurement of total lung capacity using total-breath method

Sponsors

ndd Medizintechnik AG
Lead SponsorINDUSTRY

Study design

Allocation
NON_RANDOMIZED
Intervention model
CROSSOVER
Primary purpose
OTHER
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

3 groups * Healthy subjects, FEV1/FEV1(pred) \> 80% * COPD GOLD classification 1 or 2, FEV1/FEV1(pred) ≥ 50% * COPD GOLD classification 3 or 4, FEV1/FEV1(pred) \< 50%

Exclusion criteria

* Smoked during 2 hours prior to measurements Additional

Design outcomes

Primary

MeasureTime frameDescription
Comparison of Total Lung Capacity (TLC) measurement using total-breath method and standard methodsingle measurement on day 1Difference in Total Lung Capacity (TLC) using total-breath method (according to ATS/ERS standard 2017, EasyOne Pro) and TLC using normal dilution method (according to ATS/ERS standard 2005, EasyOne Pro) in relation to TLC using body plethysmography (Vyaire).

Secondary

MeasureTime frameDescription
Dependency between primary outcome and COPD disease severitysingle measurement on day 1Correlation between ( TLC(dilution, total-breath, EasyOne Pro) - TLC(dilution, standard 2005, EasyOne Pro) ) / TLC(body, Vyaire) and FEV1(EasyOne Pro) / FEV1(predicted)
Device comparison using the same method for both devices.single measurement on day 1Total Lung Capacity (TLC) (dilution, standard 2005, EasyOne Pro) / TLC(dilution, standard, Vyaire). Determines accuracy and comparability of both devices in measurement of TLC.

Countries

Switzerland

Outcome results

None listed