Dosimetric Validation of Scans Generated by GAN From Pre-therapeutic MRI in Stereotactic Cerebral Radiotherapy

Dosimetric Validation of Scans Generated by GAN From Pre-therapeutic MRI in Stereotactic Cerebral Radiotherapy (SBRT-IRM)

Status
Completed
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT04525053
Acronym
SBRT-IRM
Enrollment
184
Registered
2020-08-25
Start date
2020-07-16
Completion date
2020-08-17
Last updated
2020-08-25

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Brain Metastases

Brief summary

Stereotactic radiotherapy (SRT) is being widely accepted as a treatment of choice for patients with a small number of brain metastases and an acceptable size, allowing a better target dose conformity resulting in high local control rates and better sparing of organs at risk. Currently, imaging for such a delivery technique requires both a recent magnetic resonance imaging (MRI) brain study for volume definition and a computed tomography (CT) scan for SRT planning. An MRI-only workflow could reduce the risk of misalignment between the two imaging modalities and shorten the delay of planning. Given the absence of a calibrated electronic density on MRI, the investigators aim to assess the equivalence of synthetic CTs generated by a generative adversarial network (GAN) for planning in the brain SRT setting.

Interventions

None listed

Sponsors

University Hospital, Brest
Lead SponsorOTHER

Study design

Observational model
COHORT
Time perspective
RETROSPECTIVE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* age ≥ 18 years old * SRT treatment for one or more brain metastasis(ses) * brain MRI and planning CT scans realized less than 14 days prior to the treatment delivery

Exclusion criteria

* MRI's field of view (FOV) judged insufficient for tumour(s) and OAR visualization

Design outcomes

Primary

MeasureTime frameDescription
Local Gamma Analysis Passing RateOnce, at the time of treatment deliveryTo be considered as comparable, 2 dose maps must have a Local Gamma Analysis Passing Rate higher than 98%. The primary outcome measure will be to compare the Local Gamma Analysis Passing Rate between the 2 dose maps (one initial dosimetric planification and one dosimetric planification extracted from the synthetic ct scan).

Secondary

MeasureTime frameDescription
Global Gamma Analysis Passing RateOnce, at the time of treatment deliveryTo be considered as comparable, 2 dose maps must have a Global Gamma Analysis Passing Rate higher than 95%. The primary outcome measure will be to compare the Global Gamma Analysis Passing Rate between the 2 dose maps (one initial dosimetric planification and one dosimetric planification extracted from the synthetic ct scan).
Planning Target Volume's coverage valuesOnce, at the time of treatment deliveryPlanning Target Volume's (PTV) coverage values will be compared between the initial dosimetric planification and the dosimetric planification extracted from the synthetic ct scan.
Organs at risk Dose ConstraintsOnce, at the time of treatment deliveryDose constraints to the Organs at Risks (OAR) will be compared between the initial dosimetric planification and the dosimetric planification extracted from the synthetic ct scan.