Cuff Pressure, Endotracheal Tube
Conditions
Brief summary
This study was designed to evaluate the effect of change of head position on the cuff pressure of TaperGaurd endotracheal tube , compared to cylindrical endotracheal tube during oral surgery.
Detailed description
The patient in cylindrical group (n=26) were intubated with cylindrical shaped endotracheal tube (Unomedical, Kedah, Malaysia) and the patient in tapergaurd group (n=26) was intubated with TaperGuard endotracheal tube (Covidien, Athlone, Ireland). The group allocations were informed to the attending anesthesiologist prior to anesthesia. Tracheal intubation was performed with size of inner diameter 7.0 mm endotracheal tube for male and 6.5 mm endotracheal tube for female by the same anesthesiologist. The anesthesiologist measured cuff pressure and distance from endotracheal tube tip to carina during study period.This study was designed to evaluate the effect of change of head position on the cuff pressure of TaperGaurd endotracheal tube, compared to cylindrical endotracheal tube during oral surgery.
Interventions
TaperGuard endotracheal tube is intubated
endotracheal tube is intubated
Sponsors
Study design
Eligibility
Inclusion criteria
* Patient who were underwent nasal intubation
Exclusion criteria
* history of difficult intubation, limited neck movement, respiratory diseases and body mass index more than 35 kg/m2
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| Cuff pressure | 5 minutes after change of head position | Cuff pressure is measured with a manometer (Mallinckrodt Medical, Hennef, Germany). |
Countries
South Korea