COVID-19
Conditions
Keywords
POC, IVD, SARS-CoV-2, IgG, IgM, Antigen, Antibody
Brief summary
Collection of nasal/nasopharyngeal/throat swabs and blood samples from patients presenting at their designated care facility displaying symptoms of COVID-19 and undergoing a SOC SARS-CoV-2 test or those who have tested positive in the past to aid development, calibration and performance evaluation for the LumiraDx POC test.
Detailed description
CAPTURE is a two-stage sample collection study to complete the development, calibration and performance evaluation phases of the LumiraDx Point of Care (POC) device for the detection of the SARS-CoV-2 virus and the SARS-CoV-2 IgG/IgM status of infected patients. The study is split into two arms targeting separate populations throughout each stage. Stage 1: Consists of sample collection only, to facilitate the in-house development and calibration of the LumiraDx SARS-CoV-2 assays. Stage 2: Consists of sample collection to facilitate the performance evaluation of the LumiraDx SARS-CoV-2 studies. In this stage the testing will be completed onsite using the LumiraDx POC device.
Interventions
This is a prospective sample collection and performance evaluation study which involves no therapeutic intervention. Sample types include: * Nasal/Throat/Nasal Throat Combo/ Nasopharyngeal (as per sites SOC) Swab * Whole Venous Blood
This is a prospective sample collection and performance evaluation study which involves no therapeutic intervention. Sample types include: * Whole Venous Blood * Fingerstick Capillary Sample
Sponsors
Study design
Eligibility
Inclusion criteria
1. Patient is at least 18 years of age. 2. Patient meets either of the below categories: 1. Patient is presenting with symptoms indicative of COVID-19 and will be completing SOC sampling for SARS-CoV-2 PCR testing. (Eligible for Arm A) 2. Patient has had a confirmed positive SARS-CoV-2 PCR test result in the past. (Eligible for Arm B)
Exclusion criteria
1. The patient does not have the capacity to consent as determined by the Research Team (and the impartial witness if applicable). 2. The patient is deemed to be unsuitable for research at the research teams' discretion.
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| Performance of the LumiraDx Assay versus reference methods with regards to clinical sensitivity and specificity for the detection of the SARS-COV-2 IgG/IgM and viral antigen in patient samples. | 1 Year approx | Primary Outcome - To collect sufficient SARS-CoV-2 samples (nasal/nasopharyngeal/throat swabs and blood samples) to aid development, calibration and performance evaluation of the LumiraDx POC test. Results will be used to asses performance of the LumiraDx assay versus reference methods with regards to the clinical sensitivity and specificity for the detection of the SARS-COV-2 IgG/IgM and viral antigen in patient samples. |
Countries
United Kingdom
Contacts
University College London Hospitals