Accuracy of the NICCI™ Monitor in Children and Adolescents

Accuracy of the NICCI™ Monitor in Providing Beat-to-beat Non-invasive Blood Pressure Readings in Children and Adolescents

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT04373746
Enrollment
82
Registered
2020-05-04
Start date
2020-10-26
Completion date
2021-12-30
Last updated
2023-04-18

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Anesthesia, Surgery

Brief summary

This is a prospective study that will compare the blood pressure readings from an arterial cannula with those obtained non-invasively by the NICCI device. The study will enroll 50 patients into each of three weight ranges: 10-20 kg, 20-40 kg, and 40-80 kg. The NICCI, Pulsion/GETINGE uses novel technology to provide a continuous estimation of BP displayed as a waveform similar to that seen with an invasive arterial cannula, but is totally non-invasive by using the fingers. The patient rests their hand on the device, which is about the size of a computer mouse, and then a blood pressure cuff goes around their upper arm.

Interventions

DEVICENICCI

Non-invasive beat-to-beat blood pressure monitor capable of producing a waveform similar to an invasive arterial line.

Sponsors

Joseph D. Tobias
Lead SponsorOTHER

Study design

Allocation
NON_RANDOMIZED
Intervention model
SINGLE_GROUP
Primary purpose
DIAGNOSTIC
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
No minimum to 18 Years
Healthy volunteers
No

Inclusion criteria

* Patients requiring anesthetic care and in whom an indwelling arterial cannula will be placed for the surgical procedure.

Exclusion criteria

* Patients with history of a peripheral neurologic or neuropathic disorder * Patients in whom the upper extremity cannot be used for blood pressure monitoring * Patients in whom an invasive arterial cannula cannot be placed * Patients with vascular implants at the sites of non-invasive blood pressure measurement (fingers and upper arm of the examined arm) * Edematous patients

Design outcomes

Primary

MeasureTime frameDescription
Difference in Blood Pressure2-8 hours (average length of major surgery)The absolute difference of the SBP, DBP, and MAP values from the arterial cannula (AC) and NICCI device were calculated. Instead of the directional difference (positive or negative), the absolute difference was used to avoid affecting the data when calculating the mean and standard deviation (SD) of the differences. For instance, if the reading from the NICCI device was 5 mm Hg above or below the reading from the AC, a value of 5 mm Hg was used and not -5 mm Hg or +5 mm Hg.

Secondary

MeasureTime frameDescription
Percentage of BP Values From the NICCI Device That Were ≤ 5 mm Hg the Values Obtained From the AC2-8 hours (average length of major surgery)The absolute difference of the SBP, DBP, and MAP values from the arterial cannula (AC) and NICCI device were calculated. Instead of the directional difference (positive or negative), the absolute difference was used to avoid affecting the data when calculating the mean and standard deviation (SD) of the differences. For instance, if the reading from the NICCI device was 5 mm Hg above or below the reading from the AC, a value of 5 mm Hg was used and not -5 mm Hg or +5 mm Hg. We also calculated the percentage of BP values from the NICCI device that were ≤ 5 mm Hg, ≤ 10 mm Hg, and \> 10 mm Hg from the values obtained from the AC.
Percentage of BP Values From the NICCI Device That Were ≤ 10 mm Hg the Values Obtained From the AC2-8 hours (average length of major surgery)The absolute difference of the SBP, DBP, and MAP values from the arterial cannula (AC) and NICCI device were calculated. Instead of the directional difference (positive or negative), the absolute difference was used to avoid affecting the data when calculating the mean and standard deviation (SD) of the differences. For instance, if the reading from the NICCI device was 5 mm Hg above or below the reading from the AC, a value of 5 mm Hg was used and not -5 mm Hg or +5 mm Hg. We also calculated the percentage of BP values from the NICCI device that were ≤ 5 mm Hg, ≤ 10 mm Hg, and \> 10 mm Hg from the values obtained from the AC.
Percentage of BP Values From the NICCI Device That Were > 10 mm Hg From the Values Obtained From the AC2-8 hours (average length of major surgery)The absolute difference of the SBP, DBP, and MAP values from the arterial cannula (AC) and NICCI device were calculated. Instead of the directional difference (positive or negative), the absolute difference was used to avoid affecting the data when calculating the mean and standard deviation (SD) of the differences. For instance, if the reading from the NICCI device was 5 mm Hg above or below the reading from the AC, a value of 5 mm Hg was used and not -5 mm Hg or +5 mm Hg. We also calculated the percentage of BP values from the NICCI device that were ≤ 5 mm Hg, ≤ 10 mm Hg, and \> 10 mm Hg from the values obtained from the AC.

Countries

United States

Participant flow

Participants by arm

ArmCount
10-40 kg
Patients undergoing major surgery that weigh between 10-40 kg. NICCI: Non-invasive beat-to-beat blood pressure monitor capable of producing a waveform similar to an invasive arterial line.
24
40-80 kg
Patients undergoing major surgery that weigh between 40-80 kg. NICCI: Non-invasive beat-to-beat blood pressure monitor capable of producing a waveform similar to an invasive arterial line.
44
Total68

Baseline characteristics

Characteristic10-40 kg40-80 kgTotal
Age, Categorical
<=18 years
24 Participants44 Participants68 Participants
Age, Categorical
>=65 years
0 Participants0 Participants0 Participants
Age, Categorical
Between 18 and 65 years
0 Participants0 Participants0 Participants
Age, Continuous11 years
STANDARD_DEVIATION 4
15 years
STANDARD_DEVIATION 2
14 years
STANDARD_DEVIATION 3
ASA physical class
1
1 Participants8 Participants9 Participants
ASA physical class
2
10 Participants22 Participants32 Participants
ASA physical class
3
10 Participants13 Participants23 Participants
ASA physical class
4
3 Participants1 Participants4 Participants
BMI17 kg/m^2
STANDARD_DEVIATION 2
22 kg/m^2
STANDARD_DEVIATION 4
20 kg/m^2
STANDARD_DEVIATION 4
Height134 centimeters
STANDARD_DEVIATION 19
162 centimeters
STANDARD_DEVIATION 9
153 centimeters
STANDARD_DEVIATION 19
Patient position
Lateral
0 Participants1 Participants1 Participants
Patient position
Prone
18 Participants40 Participants58 Participants
Patient position
Supine
6 Participants3 Participants9 Participants
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants0 Participants0 Participants
Race (NIH/OMB)
Asian
0 Participants1 Participants1 Participants
Race (NIH/OMB)
Black or African American
0 Participants6 Participants6 Participants
Race (NIH/OMB)
More than one race
1 Participants2 Participants3 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants0 Participants0 Participants
Race (NIH/OMB)
Unknown or Not Reported
0 Participants0 Participants0 Participants
Race (NIH/OMB)
White
23 Participants35 Participants58 Participants
Region of Enrollment
United States
24 participants44 participants68 participants
Sex: Female, Male
Female
12 Participants23 Participants35 Participants
Sex: Female, Male
Male
12 Participants21 Participants33 Participants
Weight30 kilograms
STANDARD_DEVIATION 8
56 kilograms
STANDARD_DEVIATION 10
48 kilograms
STANDARD_DEVIATION 15

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
deaths
Total, all-cause mortality
0 / 240 / 44
other
Total, other adverse events
0 / 240 / 44
serious
Total, serious adverse events
0 / 240 / 44

Outcome results

Primary

Difference in Blood Pressure

The absolute difference of the SBP, DBP, and MAP values from the arterial cannula (AC) and NICCI device were calculated. Instead of the directional difference (positive or negative), the absolute difference was used to avoid affecting the data when calculating the mean and standard deviation (SD) of the differences. For instance, if the reading from the NICCI device was 5 mm Hg above or below the reading from the AC, a value of 5 mm Hg was used and not -5 mm Hg or +5 mm Hg.

Time frame: 2-8 hours (average length of major surgery)

ArmMeasureGroupValue (MEAN)Dispersion
10-40 kgDifference in Blood PressureSBP11 mm HgStandard Deviation 9
10-40 kgDifference in Blood PressureDBP10 mm HgStandard Deviation 7
10-40 kgDifference in Blood PressureMAP10 mm HgStandard Deviation 7
40-80 kgDifference in Blood PressureSBP10 mm HgStandard Deviation 8
40-80 kgDifference in Blood PressureDBP9 mm HgStandard Deviation 7
40-80 kgDifference in Blood PressureMAP9 mm HgStandard Deviation 7
Secondary

Percentage of BP Values From the NICCI Device That Were > 10 mm Hg From the Values Obtained From the AC

The absolute difference of the SBP, DBP, and MAP values from the arterial cannula (AC) and NICCI device were calculated. Instead of the directional difference (positive or negative), the absolute difference was used to avoid affecting the data when calculating the mean and standard deviation (SD) of the differences. For instance, if the reading from the NICCI device was 5 mm Hg above or below the reading from the AC, a value of 5 mm Hg was used and not -5 mm Hg or +5 mm Hg. We also calculated the percentage of BP values from the NICCI device that were ≤ 5 mm Hg, ≤ 10 mm Hg, and \> 10 mm Hg from the values obtained from the AC.

Time frame: 2-8 hours (average length of major surgery)

Population: The study team analyzed pairs of SBP, DBP, and MAP values from the NICCI device and the arterial canula.

ArmMeasureGroupValue (NUMBER)
10-40 kgPercentage of BP Values From the NICCI Device That Were > 10 mm Hg From the Values Obtained From the ACSBP43 percentage of NICCI readings
10-40 kgPercentage of BP Values From the NICCI Device That Were > 10 mm Hg From the Values Obtained From the ACDBP40 percentage of NICCI readings
10-40 kgPercentage of BP Values From the NICCI Device That Were > 10 mm Hg From the Values Obtained From the ACMAP44 percentage of NICCI readings
40-80 kgPercentage of BP Values From the NICCI Device That Were > 10 mm Hg From the Values Obtained From the ACSBP40 percentage of NICCI readings
40-80 kgPercentage of BP Values From the NICCI Device That Were > 10 mm Hg From the Values Obtained From the ACDBP33 percentage of NICCI readings
40-80 kgPercentage of BP Values From the NICCI Device That Were > 10 mm Hg From the Values Obtained From the ACMAP44 percentage of NICCI readings
Secondary

Percentage of BP Values From the NICCI Device That Were ≤ 10 mm Hg the Values Obtained From the AC

The absolute difference of the SBP, DBP, and MAP values from the arterial cannula (AC) and NICCI device were calculated. Instead of the directional difference (positive or negative), the absolute difference was used to avoid affecting the data when calculating the mean and standard deviation (SD) of the differences. For instance, if the reading from the NICCI device was 5 mm Hg above or below the reading from the AC, a value of 5 mm Hg was used and not -5 mm Hg or +5 mm Hg. We also calculated the percentage of BP values from the NICCI device that were ≤ 5 mm Hg, ≤ 10 mm Hg, and \> 10 mm Hg from the values obtained from the AC.

Time frame: 2-8 hours (average length of major surgery)

Population: The study team analyzed pairs of SBP, DBP, and MAP values from the NICCI device and the arterial canula.

ArmMeasureGroupValue (NUMBER)
10-40 kgPercentage of BP Values From the NICCI Device That Were ≤ 10 mm Hg the Values Obtained From the ACSBP57 percentage of NICCI readings
10-40 kgPercentage of BP Values From the NICCI Device That Were ≤ 10 mm Hg the Values Obtained From the ACDBP60 percentage of NICCI readings
10-40 kgPercentage of BP Values From the NICCI Device That Were ≤ 10 mm Hg the Values Obtained From the ACMAP56 percentage of NICCI readings
40-80 kgPercentage of BP Values From the NICCI Device That Were ≤ 10 mm Hg the Values Obtained From the ACSBP60 percentage of NICCI readings
40-80 kgPercentage of BP Values From the NICCI Device That Were ≤ 10 mm Hg the Values Obtained From the ACDBP67 percentage of NICCI readings
40-80 kgPercentage of BP Values From the NICCI Device That Were ≤ 10 mm Hg the Values Obtained From the ACMAP56 percentage of NICCI readings
Secondary

Percentage of BP Values From the NICCI Device That Were ≤ 5 mm Hg the Values Obtained From the AC

The absolute difference of the SBP, DBP, and MAP values from the arterial cannula (AC) and NICCI device were calculated. Instead of the directional difference (positive or negative), the absolute difference was used to avoid affecting the data when calculating the mean and standard deviation (SD) of the differences. For instance, if the reading from the NICCI device was 5 mm Hg above or below the reading from the AC, a value of 5 mm Hg was used and not -5 mm Hg or +5 mm Hg. We also calculated the percentage of BP values from the NICCI device that were ≤ 5 mm Hg, ≤ 10 mm Hg, and \> 10 mm Hg from the values obtained from the AC.

Time frame: 2-8 hours (average length of major surgery)

Population: The study team analyzed pairs of SBP, DBP, and MAP values from the NICCI device and the arterial canula.

ArmMeasureGroupValue (NUMBER)
10-40 kgPercentage of BP Values From the NICCI Device That Were ≤ 5 mm Hg the Values Obtained From the ACSBP32 percentage of NICCI readings
10-40 kgPercentage of BP Values From the NICCI Device That Were ≤ 5 mm Hg the Values Obtained From the ACDBP35 percentage of NICCI readings
10-40 kgPercentage of BP Values From the NICCI Device That Were ≤ 5 mm Hg the Values Obtained From the ACMAP31 percentage of NICCI readings
40-80 kgPercentage of BP Values From the NICCI Device That Were ≤ 5 mm Hg the Values Obtained From the ACSBP35 percentage of NICCI readings
40-80 kgPercentage of BP Values From the NICCI Device That Were ≤ 5 mm Hg the Values Obtained From the ACDBP39 percentage of NICCI readings
40-80 kgPercentage of BP Values From the NICCI Device That Were ≤ 5 mm Hg the Values Obtained From the ACMAP31 percentage of NICCI readings

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026