TactiFlex Paroxysmal Atrial Fibrillation IDE Trial

Safety and Effectiveness of the TactiFlex(TM) Ablation Catheter, Sensor Enabled(TM) (TactiFlex SE) for Treating Drug-refractory, Symptomatic Atrial Fibrillation.

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT04356040
Enrollment
355
Registered
2020-04-21
Start date
2020-06-29
Completion date
2022-07-01
Last updated
2023-10-31

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Paroxysmal Atrial Fibrillation

Brief summary

Prospective non-randomized parallel-assignment multi-center clinical investigation. The study design includes two subject cohorts: 1) Paroxysmal Atrial Fibrillation Main Study, and 2) Paroxysmal Atrial Fibrillation High Standard Power Substudy. Subjects in the main study cohorts are to be treated using the full range of ablation power settings in the Instructions For Use. Subjects in the High Standard Power Substudy are to be treated in the upper end of the recommended ablation power settings (40-50 Watts).

Detailed description

This clinical investigation is intended to demonstrate the safety and effectiveness of the TactiFlex(TM) Ablation Catheter, Sensor Enabled(TM) (TactiFlex SE) for treating drug-refractory, symptomatic paroxysmal atrial fibrillation. This clinical investigation will be conducted under an investigational device exemption and is intended to support market approval of the TactiFlex SE catheter worldwide. Three hundred and Fifty Five (355) subjects will be enrolled at up to 50 investigational sites worldwide. This clinical investigation is sponsored by Abbott.

Interventions

DEVICETactiFlex SE

Radiofrequency ablation with the TactiFlex SE ablation catheter. The procedure should be performed according to the TactiFlex SE investigational Instructions for Use document using the recommended ablation parameters as noted in the document.

DEVICETactiFlex SE - High Standard Power

Radiofrequency ablation with the TactiFlex SE ablation catheter. Subjects in the High Standard Power Substudy are to undergo the same study procedures as subjects in the main study, except that ablation power settings of 40-50 Watts are to be used in the left atrium, unless there is a medical reason to use a lower power.

Sponsors

Abbott Medical Devices
Lead SponsorINDUSTRY

Study design

Allocation
NON_RANDOMIZED
Intervention model
PARALLEL
Primary purpose
TREATMENT
Masking
NONE

Intervention model description

This is a prospective non-randomized parallel-assignment multi-center clinical investigation. The study design includes two subject cohorts: 1) Paroxysmal Atrial Fibrillation Main Study, and 2) Paroxysmal Atrial Fibrillation High Standard Power Substudy. Subjects in the main study cohorts are to be treated using the full range of ablation power settings in the Instructions For Use. Subjects in the Substudy are to be treated in the upper end of the recommended ablation power settings (40-50 Watts).

Eligibility

Sex/Gender
ALL
Age
18 Years to 99 Years
Healthy volunteers
No

Inclusion criteria

A patient will be eligible for clinical trial participation if he/she meets the following criteria: 1. Plans to undergo a catheter ablation procedure due to symptomatic paroxysmal atrial fibrillation 2. Physician's note indicating recurrent self-terminating atrial fibrillation 3. One electrocardiographically documented atrial fibrillation episode within 12-months prior to informed consent/enrollment. Documented evidence of the atrial fibrillation episode must either be continuous atrial fibrillation on a 12-lead ECG or include at least 30 seconds of atrial fibrillation from another ECG device. 4. At least 18 years of age 5. Able and willing to comply with all trial requirements 6. Informed of the nature of the trial, agreed to its provisions and has provided written informed consent as approved by the Institutional Review Board/Ethics Committee of the respective clinical trial site. A patient will be excluded from enrollment in the clinical trial if he/she meets any of the following criteria: 1. Persistent or long-standing persistent atrial fibrillation 2. Active systemic infection 3. Known presence of cardiac thrombus 4. Hypertrophic cardiomyopathy 5. Arrhythmia due to reversible causes including thyroid disorders, acute alcohol intoxication, and other major surgical procedures in the 90-day period preceding procedure 6. Myocardial infarction, acute coronary syndrome, percutaneous coronary intervention, or valve or coronary bypass grafting surgery within 90 days of procedure 7. Left atrial diameter \> 5.0 cm measured within 180 days of procedure (echocardiography or computerized tomography) 8. Left ventricular ejection fraction \< 35% measured within 180 days of procedure (echocardiography or computerized tomography) 9. New York Heart Association class III or IV 10. Previous left atrial surgical or catheter ablation procedure 11. Left atrial surgical procedure or incision with resulting scar (including left atrial appendage closure device) 12. Previous tricuspid or mitral valve replacement or repair 13. Heart disease in which corrective surgery is anticipated within 180 days after the procedure 14. Bleeding diathesis or suspected pro-coagulant state 15. Contraindication to long term anti-thromboembolic therapy 16. Presence of any condition that precludes appropriate vascular access 17. Renal failure requiring dialysis 18. Known sensitivity to contrast media (if needed during the procedure) that cannot be controlled with pre-medication 19. Severe pulmonary disease (e.g., restrictive pulmonary disease, constrictive or chronic obstructive pulmonary disease) or any other disease or malfunction of the lungs or respiratory system that produces severe chronic symptoms 20. Women who are pregnant or breastfeeding 21. Presence of other anatomic or comorbid condition that, in the investigator's opinion, could limit the patient's ability to participate in the clinical trial or to comply with follow up requirements, or impact the scientific soundness of the clinical trial results 22. Patient is currently participating in another clinical trial or has participated in a clinical trial within 30 days prior to screening that may interfere with this clinical trial 23. Patient is unlikely to survive the protocol follow up period of 12-months after the procedure 24. Body mass index \> 40 kg/m2 25. Presence of other medical, social, or psychological conditions that, in the investigator's opinion, could limit the subject's ability to participate in the clinical investigation or to comply with follow-up requirements, or impact the scientific soundness of the clinical investigation results. 26. Individuals without legal authority 27. Individuals unable to read or write 28. Patients who have had a ventriculotomy or atriotomy within the preceding 4 weeks of procedure, 29. Patients with prosthetic valves, 30. Patients with a myxoma, 31. Patients with an interatrial baffle or patch as the transseptal puncture could persist and produce an iatrogenic atrial shunt 32. Patient unable to receive heparin or an acceptable alternative to achieve adequate anticoagulation 33. Stroke or transient ischemic attack within the last 90 days 34. Stent, constriction, or stenosis in a pulmonary vein. 35. Rheumatic heart disease 36. Severe mitral regurgitation (regurgitant volume ≥ 60 mL/beat, regurgitant fraction ≥ 50%, and/or effective regurgitant orifice area ≥ 0.40cm2).

Design outcomes

Primary

MeasureTime frameDescription
Number of Participants With Device or Procedure-related Serious Adverse EventsThrough 12 monthsTthe rate of device and/or procedure-related serious adverse events with onset within 7-days of any ablation procedure that uses the TactiFlex SE catheter (initial or repeat procedure performed 31-80 days of initial procedure) that are defined below: * Atrio-esophageal fistula1 * Cardiac tamponade/perforation1 * Death * Heart block * Myocardial infarction * Pericarditis * Phrenic nerve injury resulting in diaphragmatic paralysis * Pulmonary edema * Pulmonary vein stenosis1 * Stroke/cerebrovascular accident * Thromboembolism * Transient ischemic attack * Vagal nerve injury/gastroparesis * Vascular access complications (including major bleeding events) 1. Atrio-esophageal fistula, cardiac tamponade/perforation and pulmonary vein stenosis will be evaluated through 12-months.
KM Rate of Freedom From Atrial Fibrillation, Atrial Flutter or Atrial Tachycardia RecurrenceThrough 12 monthsFreedom from documented (symptomatic or asymptomatic) Atrial Fibrillation/Atrial Flutter/Atrial Tachycardia episodes of \>30 seconds duration that are documented by 12-lead ECG, transtelephonic monitoring or Holter monitor after the initial catheter ablation procedure through 12-months of follow-up. Recurrence during the 90-day blanking period will not be considered a treatment failure. One repeat procedure will be allowed during the blanking period. Failure to achieve acute procedural success during the last ablation procedure with the TactiFlex SE catheter will constitute failure. After the 90-day blanking period, use of Class I or III anti-arrhythmic drugs will not count as a therapy failure provided that only previously failed drugs are taken at doses that do not exceed the previously failed dose.

Secondary

MeasureTime frameDescription
KM Rate of Freedom From Symptomatic Atrial Fibrillation, Atrial Flutter or Atrial Tachycardia RecurrenceThrough 12 monthsFreedom from documented symptomatic atrial fibrillation/atrial flutter/atrial tachycardia episodes of \>30 seconds duration that are documented by 12-lead ECG, transtelephonic monitoring or Holter monitor after the initial catheter ablation procedure through 12-months of follow-up. Recurrence during the 90-day blanking period (≤90 days post-initial procedure) will not be considered a treatment failure. One repeat procedure will be allowed for ablation 31-80 days after the initial procedure and will not be considered a treatment failure. Failure to achieve acute procedural success during the last ablation procedure with the TactiFlex SE catheter will constitute failure. After the 90-day blanking period, use of Class I or III anti-arrhythmias drugs will not count as a therapy failure provided that only previously failed Class I or III anti-arrhythmias drugs are taken at doses that do not exceed the previously failed dose.
KM Rate of Freedom From Atrial Fibrillation, Atrial Flutter or Atrial Tachycardia Recurrence With Only 1 Ablation Procedure12 monthsFreedom from documented (symptomatic or asymptomatic) Atrial Fibrillation/Atrial Flutter/Atrial Tachycardia episodes of \>30 seconds duration that are documented by 12-lead ECG, transtelephonic monitoring or Holter monitor after the initial catheter ablation procedure through 12-months of follow-up. Recurrence during the 90-day blanking period (≤90 days post-initial procedure) will not be considered a treatment failure. Any repeat ablation procedure in the left atrium will be considered a treatment failure. Failure to achieve acute procedural success during the ablation procedure with the TactiFlex SE catheter will constitute failure. After the 90-day blanking period, use of Class I or III anti-arrhythmic drugs will not count as a therapy failure provided that only previously failed Class I or III anti-arrhythmic drugs are taken at doses that do not exceed the previously failed dose.
KM Rate of Freedom From Atrial Fibrillation, Atrial Flutter or Atrial Tachycardia Recurrence, Without Anti-arrhythmic DrugsThrough 12 monthsFreedom from documented (symptomatic or asymptomatic) Atrial Fibrillation/Atrial Flutter/Atrial Tachycardia episodes of \>30 seconds duration that are documented by 12-lead ECG, transtelephonic monitoring or Holter monitor after the initial catheter ablation procedure through 12-months of follow-up (9 months after a 90-day blanking period). Recurrence during the 90-day blanking period (≤90 days post-initial procedure) will not be considered a treatment failure. One repeat procedure will be allowed for ablation 31-80 days after the initial procedure and will not be considered a treatment failure. Failure to achieve acute procedural success during the last ablation procedure with the TactiFlex SE catheter will constitute failure. After the 90-day blanking period, any use of a Class I or III anti-arrhythmic drugs will count as an effectiveness failure.

Countries

Australia, Austria, Canada, Czechia, Germany, Hong Kong, Italy, Netherlands, Taiwan, United States

Participant flow

Participants by arm

ArmCount
Paroxysmal AF - Main Study
TactiFlex SE: Radiofrequency ablation with the TactiFlex SE ablation catheter. The procedure should be performed according to the TactiFlex SE investigational Instructions for Use document using the recommended ablation parameters as noted in the document.
284
Paroxysmal AF HSP Sub-Study
TactiFlex SE - HSP: Radiofrequency ablation with the TactiFlex SE ablation catheter. Subjects in the HSP Substudy are to undergo the same study procedures as subjects in the main study, except that ablation power settings of 40-50 Watts are to be used in the left atrium, unless there is a medical reason to use a lower power.
50
Total334

Withdrawals & dropouts

PeriodReasonFG000FG001
Overall StudyDeath20
Overall StudyDid not meet Inclusion/Exclusion criteria70
Overall StudyEnrollment Pause100
Overall StudyLost to Follow-up10
Overall StudyPhysician Decision01
Overall StudyPI left institution30
Overall StudyProcedure aborted prior to device insertion into the vasculature10
Overall StudyWithdrawal by Subject131

Baseline characteristics

CharacteristicParoxysmal AF - Main StudyParoxysmal AF HSP Sub-StudyTotal
Age, Continuous65.0 years67.0 years65.0 years
Body Mass Index28.4 kg/m^228.35 kg/m^228.4 kg/m^2
CHA2DS2Vasc Score2 units on a scale2 units on a scale2 units on a scale
Ethnicity (NIH/OMB)
Hispanic or Latino
5 Participants2 Participants7 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
264 Participants45 Participants309 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
15 Participants3 Participants18 Participants
Left Atrial Diameter40.0 Millimeters40.0 Millimeters40.0 Millimeters
Left Ventricular Ejection Fraction60.0 Percentage55.0 Percentage60.0 Percentage
New York Heart Association Functional Classification
Class I
54 Participants1 Participants55 Participants
New York Heart Association Functional Classification
Class II
27 Participants5 Participants32 Participants
New York Heart Association Functional Classification
Class III or IV
0 Participants0 Participants0 Participants
New York Heart Association Functional Classification
No Heart Failure
203 Participants44 Participants247 Participants
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants0 Participants0 Participants
Race (NIH/OMB)
Asian
10 Participants2 Participants12 Participants
Race (NIH/OMB)
Black or African American
3 Participants1 Participants4 Participants
Race (NIH/OMB)
More than one race
1 Participants1 Participants2 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants0 Participants0 Participants
Race (NIH/OMB)
Unknown or Not Reported
8 Participants2 Participants10 Participants
Race (NIH/OMB)
White
262 Participants44 Participants306 Participants
Region of Enrollment
Australia
31 participants0 participants31 participants
Region of Enrollment
Austria
29 participants0 participants29 participants
Region of Enrollment
Canada
3 participants0 participants3 participants
Region of Enrollment
Czechia
7 participants0 participants7 participants
Region of Enrollment
Germany
22 participants0 participants22 participants
Region of Enrollment
Hong Kong
3 participants0 participants3 participants
Region of Enrollment
Italy
2 participants0 participants2 participants
Region of Enrollment
Taiwan
2 participants0 participants2 participants
Region of Enrollment
United States
185 participants50 participants235 participants
Sex: Female, Male
Female
104 Participants16 Participants120 Participants
Sex: Female, Male
Male
180 Participants34 Participants214 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
deaths
Total, all-cause mortality
2 / 2840 / 50
other
Total, other adverse events
5 / 2843 / 50
serious
Total, serious adverse events
65 / 2848 / 50

Outcome results

Primary

KM Rate of Freedom From Atrial Fibrillation, Atrial Flutter or Atrial Tachycardia Recurrence

Freedom from documented (symptomatic or asymptomatic) Atrial Fibrillation/Atrial Flutter/Atrial Tachycardia episodes of \>30 seconds duration that are documented by 12-lead ECG, transtelephonic monitoring or Holter monitor after the initial catheter ablation procedure through 12-months of follow-up. Recurrence during the 90-day blanking period will not be considered a treatment failure. One repeat procedure will be allowed during the blanking period. Failure to achieve acute procedural success during the last ablation procedure with the TactiFlex SE catheter will constitute failure. After the 90-day blanking period, use of Class I or III anti-arrhythmic drugs will not count as a therapy failure provided that only previously failed drugs are taken at doses that do not exceed the previously failed dose.

Time frame: Through 12 months

Population: All treated subjects

ArmMeasureValue (NUMBER)
Main StudyKM Rate of Freedom From Atrial Fibrillation, Atrial Flutter or Atrial Tachycardia Recurrence72.9 percentage
High Standard Power Sub-StudyKM Rate of Freedom From Atrial Fibrillation, Atrial Flutter or Atrial Tachycardia Recurrence71.8 percentage
Primary

Number of Participants With Device or Procedure-related Serious Adverse Events

Tthe rate of device and/or procedure-related serious adverse events with onset within 7-days of any ablation procedure that uses the TactiFlex SE catheter (initial or repeat procedure performed 31-80 days of initial procedure) that are defined below: * Atrio-esophageal fistula1 * Cardiac tamponade/perforation1 * Death * Heart block * Myocardial infarction * Pericarditis * Phrenic nerve injury resulting in diaphragmatic paralysis * Pulmonary edema * Pulmonary vein stenosis1 * Stroke/cerebrovascular accident * Thromboembolism * Transient ischemic attack * Vagal nerve injury/gastroparesis * Vascular access complications (including major bleeding events) 1. Atrio-esophageal fistula, cardiac tamponade/perforation and pulmonary vein stenosis will be evaluated through 12-months.

Time frame: Through 12 months

Population: The primary safety analysis population for the primary safety endpoint included all treated subjects who have completed their 7-day follow-up visit; or crossed the end of the 7-day visit window without the visit but with a primary safety endpoint event.

ArmMeasureValue (COUNT_OF_PARTICIPANTS)
Main StudyNumber of Participants With Device or Procedure-related Serious Adverse Events12 Participants
High Standard Power Sub-StudyNumber of Participants With Device or Procedure-related Serious Adverse Events2 Participants
Secondary

KM Rate of Freedom From Atrial Fibrillation, Atrial Flutter or Atrial Tachycardia Recurrence With Only 1 Ablation Procedure

Freedom from documented (symptomatic or asymptomatic) Atrial Fibrillation/Atrial Flutter/Atrial Tachycardia episodes of \>30 seconds duration that are documented by 12-lead ECG, transtelephonic monitoring or Holter monitor after the initial catheter ablation procedure through 12-months of follow-up. Recurrence during the 90-day blanking period (≤90 days post-initial procedure) will not be considered a treatment failure. Any repeat ablation procedure in the left atrium will be considered a treatment failure. Failure to achieve acute procedural success during the ablation procedure with the TactiFlex SE catheter will constitute failure. After the 90-day blanking period, use of Class I or III anti-arrhythmic drugs will not count as a therapy failure provided that only previously failed Class I or III anti-arrhythmic drugs are taken at doses that do not exceed the previously failed dose.

Time frame: 12 months

Population: All treated subjects

ArmMeasureValue (NUMBER)
Main StudyKM Rate of Freedom From Atrial Fibrillation, Atrial Flutter or Atrial Tachycardia Recurrence With Only 1 Ablation Procedure71.5 percentage
High Standard Power Sub-StudyKM Rate of Freedom From Atrial Fibrillation, Atrial Flutter or Atrial Tachycardia Recurrence With Only 1 Ablation Procedure71.8 percentage
Secondary

KM Rate of Freedom From Atrial Fibrillation, Atrial Flutter or Atrial Tachycardia Recurrence, Without Anti-arrhythmic Drugs

Freedom from documented (symptomatic or asymptomatic) Atrial Fibrillation/Atrial Flutter/Atrial Tachycardia episodes of \>30 seconds duration that are documented by 12-lead ECG, transtelephonic monitoring or Holter monitor after the initial catheter ablation procedure through 12-months of follow-up (9 months after a 90-day blanking period). Recurrence during the 90-day blanking period (≤90 days post-initial procedure) will not be considered a treatment failure. One repeat procedure will be allowed for ablation 31-80 days after the initial procedure and will not be considered a treatment failure. Failure to achieve acute procedural success during the last ablation procedure with the TactiFlex SE catheter will constitute failure. After the 90-day blanking period, any use of a Class I or III anti-arrhythmic drugs will count as an effectiveness failure.

Time frame: Through 12 months

Population: All treated subjects

ArmMeasureValue (NUMBER)
Main StudyKM Rate of Freedom From Atrial Fibrillation, Atrial Flutter or Atrial Tachycardia Recurrence, Without Anti-arrhythmic Drugs64.0 percentage
High Standard Power Sub-StudyKM Rate of Freedom From Atrial Fibrillation, Atrial Flutter or Atrial Tachycardia Recurrence, Without Anti-arrhythmic Drugs65.8 percentage
Secondary

KM Rate of Freedom From Symptomatic Atrial Fibrillation, Atrial Flutter or Atrial Tachycardia Recurrence

Freedom from documented symptomatic atrial fibrillation/atrial flutter/atrial tachycardia episodes of \>30 seconds duration that are documented by 12-lead ECG, transtelephonic monitoring or Holter monitor after the initial catheter ablation procedure through 12-months of follow-up. Recurrence during the 90-day blanking period (≤90 days post-initial procedure) will not be considered a treatment failure. One repeat procedure will be allowed for ablation 31-80 days after the initial procedure and will not be considered a treatment failure. Failure to achieve acute procedural success during the last ablation procedure with the TactiFlex SE catheter will constitute failure. After the 90-day blanking period, use of Class I or III anti-arrhythmias drugs will not count as a therapy failure provided that only previously failed Class I or III anti-arrhythmias drugs are taken at doses that do not exceed the previously failed dose.

Time frame: Through 12 months

Population: All Treated Subjects

ArmMeasureValue (NUMBER)
Main StudyKM Rate of Freedom From Symptomatic Atrial Fibrillation, Atrial Flutter or Atrial Tachycardia Recurrence80.4 percentage
High Standard Power Sub-StudyKM Rate of Freedom From Symptomatic Atrial Fibrillation, Atrial Flutter or Atrial Tachycardia Recurrence75.9 percentage