Skip to content

Comparison of Imaging Quality Between Spectral Photon Counting Computed Tomography (SPCCT) and Dual Energy Computed Tomography (DECT)

Comparison of Imaging Quality Between Spectral Photon Counting Computed Tomography (SPCCT) and Dual Energy Computed Tomography (DECT)

Status
Recruiting
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT04328181
Acronym
SPEQUA
Enrollment
339
Registered
2020-03-31
Start date
2021-01-29
Completion date
2026-01-29
Last updated
2024-04-04

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Diabetic Foot Ulcer, Coronary Artery Disease, Parenchymatous; Pneumonia, Kidney Stone, Inner Ear Disease, Brain Stroke, Joint Diseases, Diabetes, Adrenal Incidentaloma, Hyperaldosteronism, Macroadenoma, Interstitial Lung Disease, Intracranial Arteriovenous Malformations

Keywords

Diagnostic, SPCCT

Brief summary

This pilot study wants to determine to which extent SPCCT allows obtaining images with improved quality and diagnostic confidence when compared to standard Dual Energy CT (DECT), both with and without contrast agent injection. Depending on the anatomical structures/organs to be visualized during CT examinations, different scanning protocols are performed with quite variable ionizing radiation doses. Therefore, in order to obtain the most extensive and representative results of the improvement in image quality between SPCCT and DECT that will be performed CT imaging on several body regions and structures, including diabetic foot, diabetic calcium coronary scoring, adrenal glands, coronary arteries, lung parenchyma, kidney stones, inner ear, brain and joints, earl/temporal bone, colorectal carcinosis.

Interventions

For the dual-layer spectral CT scanner (IQon®, Philips, Amsterdam, The Netherlands), the following parameters will be used for the acquisition: * Tube potential 120 kVp; * Tube current time product of 150 mAs; * Gantry revolution time 0.33 s; * Automatic exposure control (angular and longitudinal) combined xyz-axis; * Beam collimation geometry 64 x 0.625 mm - scan field-of-view - 22 cm;

DEVICEDECT (Dual Energy CT)

The SPCCT scanner is a prototype spectral photon-counting computed tomography system derived from a modified clinical CT system with a field-of-view (FOV) of 168 mm in-plane, and a z-coverage of 20 mm. It is equipped with energy-sensitive photon-counting detectors relying on the direct conversion high band gap semiconductor of cadmium zinc telluride (CZT). The electronics comprises 5 rate counters with 5 different configurable energy thresholds. The SPCCT scanner has the following parameters: * Base platform : iCT scanner * Tube Voltage (kVp) : 120, capability for imaging at 80 and 100 kVp * Tube current (mA) : 10 - 300 * Spatial Resolution : \> 30 lp/cm, ≤ 250 µm * Z-coverage isocenter (mm) : 20 mm * FOV (mm) : 500 * Minimum rotation time : 0.33 s/rotation * Acquisition modes : Axial, Helical, Step & Shoot

Sponsors

Hospices Civils de Lyon
Lead SponsorOTHER

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
DIAGNOSTIC
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Patients presenting following one of following medical conditions: 1. Asymptomatic type 1 or 2 diabetes as regard to cardiovascular risks 2. Diabetic foot ulcer 3. Adrenal glands lesions: Adrenal incidentaloma or hyperaldosteronism or macroadenoma 4. Urinary stone(s) 5. Known Coronary artery disease: Stent imaging or control of calcified plaques 6. Diffuse Interstitial Lung Diseases: Idiopathic Pulmonary Fibrosis, Hypersensitivity Pneumonitis, Ground Glass Opacification, Chronic Thromboembolic Pulmonary Hypertension 7. Conductive hearing loss 8. Brain stroke (late or post thrombectomy) 9. Intracranial arteriovenous malformation treated with coils or Onyx 10. Joints diseases in haemophilia 11. Ear/temporal bone 12. Colorectal carcinosis 13. Dissection aortique de type A opérée ou de type B non opérée, hématome intramural * Patient has accepted to participate to the study and has signed the written consent; * Pre-menopausal women only: Negative urinary pregnancy test on the day of imaging before the administration of study drug; * Patient is affiliated to the French social security

Exclusion criteria

* Contraindication to the use of iodine containing contrast media (including subjects with suspicion for/or known to have NSF) (if injection); * History of severe allergic or anaphylactic reaction to any allergen including drugs and contrast agents (as judged by the investigator, taking into account the intensity of the event); * History of delayed major or delayed cutaneous reaction to Iomeron injection * Estimated Glomerular Filtration Rate (eGFR) value \< 30 mL/min/1.73 m2 derived from a serum creatinine result within 1 month before the imaging for examinations with contrast agent. * Any subject on hemodialysis or peritoneal dialysis; * Suspected clinical instability or unpredictability of the clinical course during the study period (e.g. due to previous surgery); * Pregnant or nursing (including pumping for storage and feeding); * Patient under guardianship, curatorship or safeguard of justice.

Design outcomes

Primary

MeasureTime frameDescription
quality of the imagesDay 8A single four-point scale will be used (1: unacceptable, 2: usable under limited conditions, 3: probably acceptable, 4: fully acceptable) based on the European guidelines on quality criteria for computed tomography

Secondary

MeasureTime frameDescription
Subjective image quality gradedDay 8It will be calculated on a five-point scale (1: poor, 2: fair, 3: average, 4: good, 5: excellent) for each following criterion: noise, artifacts and sharpness.
CT Dose Index volumic (CTDIvol)Day 8To determine the radiation dose delivered to the patients during the DECT and SPCCT imaging procedures. The CTDI is an estimation of the dose delivered to the organs for each acquired section that is based on acquisition parameters of a water phantom with a 32 cm diameter. The value is expressed in milligray (mGy).
Dose Length Product (DLP)Day 8To determine the radiation dose delivered to the patients during the DECT and SPCCT imaging procedures. The DLP is obtained as follows: CTDI \* length of body explored = value in mGy.cm.
Equivalent dose (mSv)Day 8To determine the radiation dose delivered to the patients during the DECT and SPCCT imaging procedures. The equivalent dose is obtained by multiplying the DLP to the specific organ conversion factor.
Quantitatively image quality : NoiseDay 8The noise by selecting regions of interest (ROI) will calculated.
Diagnostic confidence gradedDay 8The diagnostic confidence grade will be calculated on a four-point scale (1: insufficient, 2: poor, 3: average, 4: good).
Quantitatively image quality : contrast-to-noise ratioDay 8The contrast-to-noise ratio (CNR) by selecting regions of interest (ROI) will calculated.
Depiction of anatomical structures of interestDay 8Depiction of anatomical structures of interest will be graded on a four-point scale (1: visualization just possible, 2: unclear borders but different structures already visible, 3: very good visualization, well-defined anatomy, 4: perfect delineation of anatomy).
Radiation doseDay 8An average radiation dose delivered to the patients for each clinical application will be calculated.
Statistical comparison between SPCCT and DECTDay 8Statistical comparison between SPCCT and DECT will be performed over all images and anatomical structures globally and also for each clinical application of interest.
Quantitatively image quality : DensityDay 8The density (HU) by selecting regions of interest (ROI) will calculated.

Countries

France

Contacts

Primary ContactPhilippe DOUEK, Pr
douek@creatis.insa-lyon.fr04 72 07 18 83
Backup ContactAdeline MANSUY
adeline.mansuy@chu-lyon.fr04 72 11 51 70

Outcome results

None listed

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026