Agreement and Precision Study of the Nidek Mirante

Agreement and Precision Study of the Nidek Mirante OCT Compared to the Optovue RTVue XR Avanti OCT and SLO Image Comparison of the Nidek Mirante and the OPTOS P200DTx

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT04318132
Enrollment
170
Registered
2020-03-23
Start date
2020-01-25
Completion date
2021-12-30
Last updated
2023-08-24

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Corneal Disease, Glaucoma, Retinal Disease

Brief summary

This study is a prospective comparative, randomized, single center study to gather agreement and precision of the Nidek Mirante OCT with SLO and Anterior Segment Imaging capabilities in comparison to the Optovue RTVue XR Avanti OCT and Optos P200DTx in normal subjects, subjects with glaucoma, subjects with retinal disease and subjects with corneal disease.

Detailed description

This is a prospective comparative clinical study to be conducted at one clinical site located in the United States. Three Nidek Mirante devices, three Optovue RTVue XR Avanti devices and one OPTOS P200DTx will be used. For agreement and precision analysis of OCT scans and agreement analysis of image quality of OCT ACA scans, each Nidek Mirante device will be paired with one of the Optovue RTVue XR Avanti devices. One device operator will be assigned to each device pair to create three distinct operator/device configurations. For agreement analysis of SLO image quality, one OPTOS P200DTx will be included in one of three operator/device configurations.

Interventions

DEVICENidek Mirante

The Nidek Mirante Optical Coherence Tomography (OCT) including scanning laser ophthalmoscope (SLO) function is an non-contact system for imaging the fundus and for axial cross sectional imaging of anterior and posterior ocular structures.

The Avanti is an optical coherence tomography system indicated for the in vivo imaging,axial cross-sectional, and three-dimensional imaging and measurement of anterior and posterior ocular structures.

DEVICEOptos P200DTx

The P200DTx scanning laser ophthalmoscope (SLO) is indicated for use as a widefield and retinal fluorescence and autofluorescence imaging ophthalmoscope to aid in the diagnosis and monitoring of diseases and disorders that manifest in the retina.

Sponsors

Nidek Co. LTD.
Lead SponsorINDUSTRY

Study design

Allocation
RANDOMIZED
Intervention model
SINGLE_GROUP
Primary purpose
DIAGNOSTIC
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
22 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

- Normal Group: 1. Male or female subjects from 22 years of age or older who have full legal capacity to volunteer on the date the informed consent is signed; 2. Subjects who can follow the instructions by the clinical staff at the clinical site, and can attend examinations on the scheduled examination date; 3. Subjects who agree to participate in the study; 4. Subjects with normal eye examinations (without pathology other than cataract) in both eyes on the date of the study visit as observed with a +90 Diopter Fundus Lens; 5. Subjects with current best-spectacle-corrected visual acuity (BSCVA) of 20/40 or better in both eyes on the date the study visit.

Exclusion criteria

- Normal Group: 1. Subjects unable to tolerate ophthalmic imaging; 2. Subjects with ocular media not sufficiently clear to obtain acceptable OCT images in either eye; 3. Subjects with any current ocular pathology other than cataract in either eye, as determined by self-report and confirmed by the investigator assessment and/or confirmed by the investigator assessment at the study visit; 4. Subjects with (or history of) leukemia, dementia or multiple sclerosis; 5. Subject has a condition or will be in a situation which the investigator feels may put the subject at significant risk, may confound the study results, or may interfere significantly with the subject's participation in the study. Inclusion Criteria - Glaucoma Group: 1. Male or female subjects from 22 years of age or older who have full legal capacity to volunteer on the date the informed consent was signed; 2. Subjects who can follow the instructions by the clinical staff at the clinical site, and can attend examinations on the scheduled examination date; 3. Subjects who agree to participate in the study; 4. Subjects who have been diagnosed with glaucoma in the glaucoma study eye(s), with optic nerve damage as evidenced by either of the following optic disc or RNFL structural abnormalities observed via fundus exam during the study visit: 1. Diffuse thinning, focal narrowing, or notching of the optic disc rim, especially at the inferior or superior poles with or without disc hemorrhage; or 2. Optic disc neural rim asymmetry of the two eyes consistent with loss of neural tissue; 5. Subjects with a current BSCVA of 20/40 or better in the glaucoma study eye(s) on the date of the visit; 6. History of a reliable Humphrey Field Analyzer (HFA) visual field (24-2 SITA Standard, white on white) measured on the day of study visit 1, or within the previous six (6) months from study visit 1 with reliable results, defined as fixation losses \< 20%, or false positives \< 33%, or false negatives \< 33% in the glaucoma study eye(s) that is consistent with glaucomatous optic nerve damage in glaucoma study eye(s) based on at least one of the following findings: 1. On pattern deviation (PD), there exists a cluster of 3 or more points in an expected location of the visual field depressed below the 5% level, at least 1 of which is depressed below the 1% level; or 2. Glaucoma hemi-field test outside normal limits.

Design outcomes

Primary

MeasureTime frameDescription
Precision of the Corneal Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCT1 dayPrecision (repeatability) of the measured Thickness of the Cornea for the OCT function of the NIDEK TONOREF III.
Agreement of Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCT1 dayAgreement of the measured Macular thickness (μm) between the test device and the predicate device for the OCT function of NIDEK Mirante.
Agreement of Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCT1 dayAgreement of the measured Retinal nerve fiber layer thickness around Optic nerve head between the test device and the predicate device for the OCT function of NIDEK Mirante.
Agreement of Corneal Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCT1 dayAgreement of the measured Corneal thickness between the test device and the predicate device for the OCT function of NIDEK Mirante.
Agreement of SLO Image Quality (Likert Scale) for Nidek Mirante and OPTOS P200DTx1 dayAgreement of the SLO Image quality between the test device and the predicate device for the SLO function of NIDEK Mirante. The image quality of is graded Clinical Utility and Overall Utility for each image patterns (SLO Color Fundus, SLO B-FAF and SLO G-FAF) on a Likert scale. The grading used the 0-3 point scale, and higher scores mean a better image quality.
Agreement of the ACA Image Quality (Likert Scale) for Nidek Mirante and Optovue RTVue XR Avanti OCT1 dayAgreement of the ACA Image quality between the test device and the predicate device for the OCT function of NIDEK Mirante. The image quality of is graded Clinical Utility and Overall Utility on a Likert scale. The grading used the 0-3 point scale, and higher scores mean a better image quality.
Precision of the Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCT1 dayPrecision (repeatability) of the measured Full retinal thickness at Macula for the OCT function of the NIDEK TONOREF III.
Precision of the Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCT1 dayPrecision (repeatability) of the measured Retinal nerve fiber layer thickness around Optic nerve head at Macula for the OCT function of the NIDEK TONOREF III.

Secondary

MeasureTime frameDescription
Adverse Events1 dayAn adverse event (AE) is defined as any unfavorable and unintended sign (including an abnormal laboratory finding), symptom, or disease temporally associated with the use of an investigational device but not necessarily related to the investigational device.

Countries

United States

Participant flow

Participants by arm

ArmCount
NIDEK Mirante Comparison - Normal
Subjects without any current ocular pathology other than cataract in either eye
45
NIDEK Mirante Comparison - Retina
Subjects diagnosed with retinal pathology
47
NIDEK Mirante Comparison - Glaucoma
Subjects who have been diagnosed with glaucoma
46
NIDEK Mirante Comparison - Corneal Disease
Subjects who have been diagnosed with corneal pathologies
32
Total170

Withdrawals & dropouts

PeriodReasonFG000FG001FG002FG003
Overall StudyProtocol Violation0030

Baseline characteristics

CharacteristicNIDEK Mirante Comparison - RetinaNIDEK Mirante Comparison - GlaucomaNIDEK Mirante Comparison - Corneal DiseaseNIDEK Mirante Comparison - NormalTotal
Age, Categorical
<=18 years
0 Participants0 Participants0 Participants0 Participants0 Participants
Age, Categorical
>=65 years
31 Participants28 Participants11 Participants4 Participants74 Participants
Age, Categorical
Between 18 and 65 years
16 Participants18 Participants21 Participants41 Participants96 Participants
Age, Continuous69.7 years
STANDARD_DEVIATION 11.23
65.1 years
STANDARD_DEVIATION 9.2
55.4 years
STANDARD_DEVIATION 15.09
44.9 years
STANDARD_DEVIATION 15.36
59.2 years
STANDARD_DEVIATION 16.05
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants0 Participants0 Participants1 Participants1 Participants
Race (NIH/OMB)
Asian
0 Participants2 Participants1 Participants0 Participants3 Participants
Race (NIH/OMB)
Black or African American
1 Participants1 Participants1 Participants2 Participants5 Participants
Race (NIH/OMB)
More than one race
0 Participants0 Participants0 Participants1 Participants1 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants0 Participants0 Participants0 Participants0 Participants
Race (NIH/OMB)
Unknown or Not Reported
0 Participants1 Participants1 Participants0 Participants2 Participants
Race (NIH/OMB)
White
46 Participants42 Participants29 Participants41 Participants158 Participants
Region of Enrollment
United States
47 Participants46 Participants32 Participants45 Participants170 Participants
Sex: Female, Male
Female
26 Participants25 Participants20 Participants28 Participants99 Participants
Sex: Female, Male
Male
21 Participants21 Participants12 Participants17 Participants71 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
EG002
affected / at risk
EG003
affected / at risk
deaths
Total, all-cause mortality
0 / 450 / 470 / 460 / 32
other
Total, other adverse events
0 / 450 / 470 / 460 / 32
serious
Total, serious adverse events
0 / 450 / 470 / 460 / 32

Outcome results

Primary

Agreement of Corneal Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCT

Agreement of the measured Corneal thickness between the test device and the predicate device for the OCT function of NIDEK Mirante.

Time frame: 1 day

Population: The retina and glaucoma arm/group were omitted because this group did not measure the corneal thickness.

ArmMeasureValue (MEAN)Dispersion
NIDEK Mirante Comparison - NormalAgreement of Corneal Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCT16.511 µmStandard Deviation 3.9117
NIDEK Mirante Comparison - RetinaAgreement of Corneal Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCT14.842 µmStandard Deviation 8.3084
Primary

Agreement of Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCT

Agreement of the measured Macular thickness (μm) between the test device and the predicate device for the OCT function of NIDEK Mirante.

Time frame: 1 day

Population: The cornea arm/group was omitted because this group did not measure the macular thickness.

ArmMeasureGroupValue (MEAN)Dispersion
NIDEK Mirante Comparison - NormalAgreement of Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTInner Nasal21.911 μmStandard Deviation 6.0521
NIDEK Mirante Comparison - NormalAgreement of Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTOuter Inferior13.844 μmStandard Deviation 5.5757
NIDEK Mirante Comparison - NormalAgreement of Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTOuter Temporal10.267 μmStandard Deviation 6.5796
NIDEK Mirante Comparison - NormalAgreement of Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTInner Inferior22.267 μmStandard Deviation 4.7645
NIDEK Mirante Comparison - NormalAgreement of Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTCenter10.658 μmStandard Deviation 4.3789
NIDEK Mirante Comparison - NormalAgreement of Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTOuter Nasal12.733 μmStandard Deviation 6.0317
NIDEK Mirante Comparison - NormalAgreement of Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTInner Superior22.889 μmStandard Deviation 6.0386
NIDEK Mirante Comparison - NormalAgreement of Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTInner Temporal19.200 μmStandard Deviation 6.4194
NIDEK Mirante Comparison - NormalAgreement of Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTOuter Superior11.311 μmStandard Deviation 5.4306
NIDEK Mirante Comparison - RetinaAgreement of Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTInner Inferior19.739 μmStandard Deviation 5.4014
NIDEK Mirante Comparison - RetinaAgreement of Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTCenter11.032 μmStandard Deviation 6.9749
NIDEK Mirante Comparison - RetinaAgreement of Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTInner Temporal17.065 μmStandard Deviation 4.2079
NIDEK Mirante Comparison - RetinaAgreement of Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTInner Superior18.870 μmStandard Deviation 3.8965
NIDEK Mirante Comparison - RetinaAgreement of Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTInner Nasal19.043 μmStandard Deviation 4.5896
NIDEK Mirante Comparison - RetinaAgreement of Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTOuter Temporal9.587 μmStandard Deviation 5.256
NIDEK Mirante Comparison - RetinaAgreement of Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTOuter Superior10.370 μmStandard Deviation 6.0309
NIDEK Mirante Comparison - RetinaAgreement of Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTOuter Nasal10.522 μmStandard Deviation 6.4524
NIDEK Mirante Comparison - RetinaAgreement of Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTOuter Inferior11.957 μmStandard Deviation 7.4951
NIDEK Mirante Comparison - GlaucomaAgreement of Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTInner Superior20.628 μmStandard Deviation 5.8879
NIDEK Mirante Comparison - GlaucomaAgreement of Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTCenter11.415 μmStandard Deviation 4.8615
NIDEK Mirante Comparison - GlaucomaAgreement of Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTOuter Superior10.767 μmStandard Deviation 6.2786
NIDEK Mirante Comparison - GlaucomaAgreement of Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTInner Temporal17.047 μmStandard Deviation 6.2104
NIDEK Mirante Comparison - GlaucomaAgreement of Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTOuter Inferior12.581 μmStandard Deviation 6.3555
NIDEK Mirante Comparison - GlaucomaAgreement of Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTInner Inferior19.163 μmStandard Deviation 7.387
NIDEK Mirante Comparison - GlaucomaAgreement of Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTInner Nasal19.721 μmStandard Deviation 6.1965
NIDEK Mirante Comparison - GlaucomaAgreement of Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTOuter Nasal11.023 μmStandard Deviation 5.8614
NIDEK Mirante Comparison - GlaucomaAgreement of Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTOuter Temporal9.628 μmStandard Deviation 5.3275
Primary

Agreement of Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCT

Agreement of the measured Retinal nerve fiber layer thickness around Optic nerve head between the test device and the predicate device for the OCT function of NIDEK Mirante.

Time frame: 1 day

Population: The retina and cornea arm/group were omitted because this group did not measure the retinal nerve fiber layer thickness.

ArmMeasureGroupValue (MEAN)Dispersion
NIDEK Mirante Comparison - NormalAgreement of Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTTSNIT Temporal-4.515 µmStandard Deviation 8.1383
NIDEK Mirante Comparison - NormalAgreement of Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTTSNIT Inferior17.211 µmStandard Deviation 8.6067
NIDEK Mirante Comparison - NormalAgreement of Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTS/I Superior8.554 µmStandard Deviation 5.5885
NIDEK Mirante Comparison - NormalAgreement of Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTHorizontal C/D Ratio0.018 µmStandard Deviation 0.1977
NIDEK Mirante Comparison - NormalAgreement of Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTTSNIT Superior11.762 µmStandard Deviation 8.8701
NIDEK Mirante Comparison - NormalAgreement of Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTVertical C/D Ratio0.030 µmStandard Deviation 0.1944
NIDEK Mirante Comparison - NormalAgreement of Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTS/I Inferior9.725 µmStandard Deviation 5.3925
NIDEK Mirante Comparison - NormalAgreement of Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTDisc Area0.271 µmStandard Deviation 0.1995
NIDEK Mirante Comparison - NormalAgreement of Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTTSNIT Nasal12.250 µmStandard Deviation 8.7671
NIDEK Mirante Comparison - NormalAgreement of Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTCup Area0.055 µmStandard Deviation 0.2236
NIDEK Mirante Comparison - NormalAgreement of Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTWhole Chart9.233 µmStandard Deviation 4.3936
NIDEK Mirante Comparison - RetinaAgreement of Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTCup Area-0.010 µmStandard Deviation 0.2491
NIDEK Mirante Comparison - RetinaAgreement of Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTWhole Chart7.536 µmStandard Deviation 4.7895
NIDEK Mirante Comparison - RetinaAgreement of Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTS/I Inferior7.631 µmStandard Deviation 7.0398
NIDEK Mirante Comparison - RetinaAgreement of Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTS/I Superior7.426 µmStandard Deviation 7.1255
NIDEK Mirante Comparison - RetinaAgreement of Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTTSNIT Temporal1.965 µmStandard Deviation 10.0886
NIDEK Mirante Comparison - RetinaAgreement of Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTTSNIT Superior5.928 µmStandard Deviation 10.1354
NIDEK Mirante Comparison - RetinaAgreement of Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTTSNIT Nasal12.395 µmStandard Deviation 7.7614
NIDEK Mirante Comparison - RetinaAgreement of Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTTSNIT Inferior9.854 µmStandard Deviation 8.5394
NIDEK Mirante Comparison - RetinaAgreement of Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTHorizontal C/D Ratio-0.061 µmStandard Deviation 0.1358
NIDEK Mirante Comparison - RetinaAgreement of Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTVertical C/D Ratio-0.056 µmStandard Deviation 0.1401
NIDEK Mirante Comparison - RetinaAgreement of Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTDisc Area0.210 µmStandard Deviation 0.2919
Primary

Agreement of SLO Image Quality (Likert Scale) for Nidek Mirante and OPTOS P200DTx

Agreement of the SLO Image quality between the test device and the predicate device for the SLO function of NIDEK Mirante. The image quality of is graded Clinical Utility and Overall Utility for each image patterns (SLO Color Fundus, SLO B-FAF and SLO G-FAF) on a Likert scale. The grading used the 0-3 point scale, and higher scores mean a better image quality.

Time frame: 1 day

Population: Some patients could not be captured acceptable SLO images. The cornea arm/group was omitted because this group was not included in the SLO images assessment.

ArmMeasureGroupValue (MEAN)Dispersion
NIDEK Mirante Comparison - NormalAgreement of SLO Image Quality (Likert Scale) for Nidek Mirante and OPTOS P200DTxSLO Color Fundus: Clinical Utility2.9 score on a scaleStandard Deviation 0.21
NIDEK Mirante Comparison - NormalAgreement of SLO Image Quality (Likert Scale) for Nidek Mirante and OPTOS P200DTxSLO Color Fundus: Overall Quality3.0 score on a scaleStandard Deviation 0.17
NIDEK Mirante Comparison - NormalAgreement of SLO Image Quality (Likert Scale) for Nidek Mirante and OPTOS P200DTxSLO B-FAF: Clinical Utility2.6 score on a scaleStandard Deviation 0.48
NIDEK Mirante Comparison - NormalAgreement of SLO Image Quality (Likert Scale) for Nidek Mirante and OPTOS P200DTxSLO B-FAF: Overall Quality2.6 score on a scaleStandard Deviation 0.52
NIDEK Mirante Comparison - NormalAgreement of SLO Image Quality (Likert Scale) for Nidek Mirante and OPTOS P200DTxSLO G-FAF: Clinical Utility2.4 score on a scaleStandard Deviation 0.61
NIDEK Mirante Comparison - NormalAgreement of SLO Image Quality (Likert Scale) for Nidek Mirante and OPTOS P200DTxSLO G-FAF: Overall Quality2.4 score on a scaleStandard Deviation 0.65
NIDEK Mirante Comparison - RetinaAgreement of SLO Image Quality (Likert Scale) for Nidek Mirante and OPTOS P200DTxSLO G-FAF: Overall Quality2.4 score on a scaleStandard Deviation 0.54
NIDEK Mirante Comparison - RetinaAgreement of SLO Image Quality (Likert Scale) for Nidek Mirante and OPTOS P200DTxSLO Color Fundus: Clinical Utility2.9 score on a scaleStandard Deviation 0.27
NIDEK Mirante Comparison - RetinaAgreement of SLO Image Quality (Likert Scale) for Nidek Mirante and OPTOS P200DTxSLO B-FAF: Overall Quality2.2 score on a scaleStandard Deviation 0.67
NIDEK Mirante Comparison - RetinaAgreement of SLO Image Quality (Likert Scale) for Nidek Mirante and OPTOS P200DTxSLO G-FAF: Clinical Utility2.4 score on a scaleStandard Deviation 0.53
NIDEK Mirante Comparison - RetinaAgreement of SLO Image Quality (Likert Scale) for Nidek Mirante and OPTOS P200DTxSLO Color Fundus: Overall Quality2.8 score on a scaleStandard Deviation 0.37
NIDEK Mirante Comparison - RetinaAgreement of SLO Image Quality (Likert Scale) for Nidek Mirante and OPTOS P200DTxSLO B-FAF: Clinical Utility2.3 score on a scaleStandard Deviation 0.62
NIDEK Mirante Comparison - GlaucomaAgreement of SLO Image Quality (Likert Scale) for Nidek Mirante and OPTOS P200DTxSLO Color Fundus: Overall Quality3.0 score on a scaleStandard Deviation 0.11
NIDEK Mirante Comparison - GlaucomaAgreement of SLO Image Quality (Likert Scale) for Nidek Mirante and OPTOS P200DTxSLO B-FAF: Clinical Utility2.6 score on a scaleStandard Deviation 0.41
NIDEK Mirante Comparison - GlaucomaAgreement of SLO Image Quality (Likert Scale) for Nidek Mirante and OPTOS P200DTxSLO G-FAF: Overall Quality2.6 score on a scaleStandard Deviation 0.47
NIDEK Mirante Comparison - GlaucomaAgreement of SLO Image Quality (Likert Scale) for Nidek Mirante and OPTOS P200DTxSLO B-FAF: Overall Quality2.5 score on a scaleStandard Deviation 0.49
NIDEK Mirante Comparison - GlaucomaAgreement of SLO Image Quality (Likert Scale) for Nidek Mirante and OPTOS P200DTxSLO Color Fundus: Clinical Utility3.0 score on a scaleStandard Deviation 0.11
NIDEK Mirante Comparison - GlaucomaAgreement of SLO Image Quality (Likert Scale) for Nidek Mirante and OPTOS P200DTxSLO G-FAF: Clinical Utility2.6 score on a scaleStandard Deviation 0.41
Primary

Agreement of the ACA Image Quality (Likert Scale) for Nidek Mirante and Optovue RTVue XR Avanti OCT

Agreement of the ACA Image quality between the test device and the predicate device for the OCT function of NIDEK Mirante. The image quality of is graded Clinical Utility and Overall Utility on a Likert scale. The grading used the 0-3 point scale, and higher scores mean a better image quality.

Time frame: 1 day

ArmMeasureGroupValue (MEAN)Dispersion
NIDEK Mirante Comparison - NormalAgreement of the ACA Image Quality (Likert Scale) for Nidek Mirante and Optovue RTVue XR Avanti OCTClinical Utility2.4 score on a scaleStandard Deviation 0.39
NIDEK Mirante Comparison - NormalAgreement of the ACA Image Quality (Likert Scale) for Nidek Mirante and Optovue RTVue XR Avanti OCTOverall Utility2.2 score on a scaleStandard Deviation 0.47
NIDEK Mirante Comparison - RetinaAgreement of the ACA Image Quality (Likert Scale) for Nidek Mirante and Optovue RTVue XR Avanti OCTOverall Utility2.1 score on a scaleStandard Deviation 0.45
NIDEK Mirante Comparison - RetinaAgreement of the ACA Image Quality (Likert Scale) for Nidek Mirante and Optovue RTVue XR Avanti OCTClinical Utility2.2 score on a scaleStandard Deviation 0.42
NIDEK Mirante Comparison - GlaucomaAgreement of the ACA Image Quality (Likert Scale) for Nidek Mirante and Optovue RTVue XR Avanti OCTClinical Utility2.3 score on a scaleStandard Deviation 0.43
NIDEK Mirante Comparison - GlaucomaAgreement of the ACA Image Quality (Likert Scale) for Nidek Mirante and Optovue RTVue XR Avanti OCTOverall Utility2.2 score on a scaleStandard Deviation 0.45
NIDEK Mirante Comparison - Corneal DiseaseAgreement of the ACA Image Quality (Likert Scale) for Nidek Mirante and Optovue RTVue XR Avanti OCTClinical Utility2.2 score on a scaleStandard Deviation 0.56
NIDEK Mirante Comparison - Corneal DiseaseAgreement of the ACA Image Quality (Likert Scale) for Nidek Mirante and Optovue RTVue XR Avanti OCTOverall Utility2.1 score on a scaleStandard Deviation 0.54
Primary

Precision of the Corneal Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCT

Precision (repeatability) of the measured Thickness of the Cornea for the OCT function of the NIDEK TONOREF III.

Time frame: 1 day

Population: The retina and glaucoma arm/group were omitted because these groups did not measure the thickness of the cornea.

ArmMeasureValue (NUMBER)
NIDEK Mirante Comparison - NormalPrecision of the Corneal Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCT0.535 %CV Repeatability
NIDEK Mirante Comparison - RetinaPrecision of the Corneal Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCT0.878 %CV Repeatability
Primary

Precision of the Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCT

Precision (repeatability) of the measured Full retinal thickness at Macula for the OCT function of the NIDEK TONOREF III.

Time frame: 1 day

Population: The cornea arm/group was omitted because this group did not measure the Full retinal thickness at Macula.

ArmMeasureGroupValue (NUMBER)
NIDEK Mirante Comparison - NormalPrecision of the Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTInner Nasal0.465 %CV Repeatability
NIDEK Mirante Comparison - NormalPrecision of the Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTOuter Inferior0.865 %CV Repeatability
NIDEK Mirante Comparison - NormalPrecision of the Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTOuter Temporal0.764 %CV Repeatability
NIDEK Mirante Comparison - NormalPrecision of the Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTInner Inferior0.502 %CV Repeatability
NIDEK Mirante Comparison - NormalPrecision of the Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTCenter1.034 %CV Repeatability
NIDEK Mirante Comparison - NormalPrecision of the Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTOuter Nasal0.415 %CV Repeatability
NIDEK Mirante Comparison - NormalPrecision of the Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTInner Superior0.606 %CV Repeatability
NIDEK Mirante Comparison - NormalPrecision of the Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTInner Temporal0.425 %CV Repeatability
NIDEK Mirante Comparison - NormalPrecision of the Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTOuter Superior0.682 %CV Repeatability
NIDEK Mirante Comparison - RetinaPrecision of the Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTInner Inferior0.806 %CV Repeatability
NIDEK Mirante Comparison - RetinaPrecision of the Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTCenter1.163 %CV Repeatability
NIDEK Mirante Comparison - RetinaPrecision of the Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTInner Temporal0.863 %CV Repeatability
NIDEK Mirante Comparison - RetinaPrecision of the Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTInner Superior0.701 %CV Repeatability
NIDEK Mirante Comparison - RetinaPrecision of the Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTInner Nasal0.608 %CV Repeatability
NIDEK Mirante Comparison - RetinaPrecision of the Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTOuter Temporal0.971 %CV Repeatability
NIDEK Mirante Comparison - RetinaPrecision of the Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTOuter Superior0.776 %CV Repeatability
NIDEK Mirante Comparison - RetinaPrecision of the Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTOuter Nasal0.593 %CV Repeatability
NIDEK Mirante Comparison - RetinaPrecision of the Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTOuter Inferior1.075 %CV Repeatability
NIDEK Mirante Comparison - GlaucomaPrecision of the Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTInner Superior0.531 %CV Repeatability
NIDEK Mirante Comparison - GlaucomaPrecision of the Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTCenter1.031 %CV Repeatability
NIDEK Mirante Comparison - GlaucomaPrecision of the Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTOuter Superior0.784 %CV Repeatability
NIDEK Mirante Comparison - GlaucomaPrecision of the Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTInner Temporal0.433 %CV Repeatability
NIDEK Mirante Comparison - GlaucomaPrecision of the Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTOuter Inferior1.012 %CV Repeatability
NIDEK Mirante Comparison - GlaucomaPrecision of the Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTInner Inferior0.678 %CV Repeatability
NIDEK Mirante Comparison - GlaucomaPrecision of the Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTInner Nasal0.441 %CV Repeatability
NIDEK Mirante Comparison - GlaucomaPrecision of the Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTOuter Nasal0.528 %CV Repeatability
NIDEK Mirante Comparison - GlaucomaPrecision of the Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTOuter Temporal0.786 %CV Repeatability
Primary

Precision of the Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCT

Precision (repeatability) of the measured Retinal nerve fiber layer thickness around Optic nerve head at Macula for the OCT function of the NIDEK TONOREF III.

Time frame: 1 day

Population: The retina and cornea arm/group were omitted because these groups did not measure the retinal nerve fiber layer thickness around the Optic nerve head.

ArmMeasureGroupValue (NUMBER)
NIDEK Mirante Comparison - NormalPrecision of the Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTTSNIT Temporal4.514 %CV Repeatability
NIDEK Mirante Comparison - NormalPrecision of the Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTS/I Superior3.550 %CV Repeatability
NIDEK Mirante Comparison - NormalPrecision of the Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTTSNIT Superior3.708 %CV Repeatability
NIDEK Mirante Comparison - NormalPrecision of the Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTHorizontal C/D Ratio9.380 %CV Repeatability
NIDEK Mirante Comparison - NormalPrecision of the Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTWhole Chart2.268 %CV Repeatability
NIDEK Mirante Comparison - NormalPrecision of the Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTVertical C/D Ratio7.394 %CV Repeatability
NIDEK Mirante Comparison - NormalPrecision of the Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTTSNIT Nasal6.730 %CV Repeatability
NIDEK Mirante Comparison - NormalPrecision of the Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTDisc Area5.892 %CV Repeatability
NIDEK Mirante Comparison - NormalPrecision of the Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTS/I Inferior3.950 %CV Repeatability
NIDEK Mirante Comparison - NormalPrecision of the Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTCup Area14.148 %CV Repeatability
NIDEK Mirante Comparison - NormalPrecision of the Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTTSNIT Inferior3.813 %CV Repeatability
NIDEK Mirante Comparison - RetinaPrecision of the Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTCup Area10.047 %CV Repeatability
NIDEK Mirante Comparison - RetinaPrecision of the Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTTSNIT Inferior6.015 %CV Repeatability
NIDEK Mirante Comparison - RetinaPrecision of the Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTTSNIT Nasal7.814 %CV Repeatability
NIDEK Mirante Comparison - RetinaPrecision of the Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTTSNIT Superior6.495 %CV Repeatability
NIDEK Mirante Comparison - RetinaPrecision of the Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTTSNIT Temporal7.703 %CV Repeatability
NIDEK Mirante Comparison - RetinaPrecision of the Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTWhole Chart3.531 %CV Repeatability
NIDEK Mirante Comparison - RetinaPrecision of the Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTS/I Inferior5.563 %CV Repeatability
NIDEK Mirante Comparison - RetinaPrecision of the Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTS/I Superior5.900 %CV Repeatability
NIDEK Mirante Comparison - RetinaPrecision of the Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTHorizontal C/D Ratio6.725 %CV Repeatability
NIDEK Mirante Comparison - RetinaPrecision of the Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTVertical C/D Ratio5.912 %CV Repeatability
NIDEK Mirante Comparison - RetinaPrecision of the Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCTDisc Area7.553 %CV Repeatability
Secondary

Adverse Events

An adverse event (AE) is defined as any unfavorable and unintended sign (including an abnormal laboratory finding), symptom, or disease temporally associated with the use of an investigational device but not necessarily related to the investigational device.

Time frame: 1 day

ArmMeasureValue (COUNT_OF_UNITS)
NIDEK Mirante Comparison - NormalAdverse Events0 eyes
NIDEK Mirante Comparison - RetinaAdverse Events0 eyes
NIDEK Mirante Comparison - GlaucomaAdverse Events0 eyes
NIDEK Mirante Comparison - Corneal DiseaseAdverse Events0 eyes