Accuracy of Non-Invasive Blood Pressure Measurements at the Arm, Calf , and Finger

Prospective, Multi-center Evaluation of the Accuracy of Non-invasive Measurement of Blood Pressure Using an Arm, Calf and Finger Cuff.

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT04269382
Acronym
PANIC
Enrollment
133
Registered
2020-02-13
Start date
2020-02-22
Completion date
2021-06-02
Last updated
2021-07-01

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Hypertension, Hypotension and Shock, Intensive Care Unit

Keywords

Blood Pressure Determination

Brief summary

The present study was designed to assess, in a population of patients admitted to the intensive care unit and already carrying an indwelling arterial catheter, the ability of combined continuous (ClearSight™) and intermittent (automatic cuff) non-invasive monitoring to detect low mean BP (\<65mmHg). The intra-arterial measurement will be the reference measurement. As secondary objectives, the investigators will assess the ability of combined continuous (ClearSight™) and intermittent (automatic cuff) non-invasive monitoring to detect stage 2 hypertension, and to detect changes in BP during a cardiovascular intervention (as clinically indicated but not imposed by the study protocol). Additionally, the accuracy of both devices against the current international standard (ISO standard) will be assessed.

Detailed description

The measurement of arterial blood pressure (BP) is of paramount importance when caring for critically ill patients, as BP, especially mean BP, is the driving force of blood flow through all the organs. The reference method is the invasive method that measures BP continuously and directly in an artery (most often the radial artery), through an indwelling arterial catheter. A popular, less invasive method, most often used in emergency situations or during the first hours of care in the intensive care unit (ICU), is the non-invasive oscillometric method (NIBP) using an arm cuff (some have shown that the cuff could also be placed around the calf, just above the ankle), that provides only intermittent measurements. In between the above 2 techniques, the place of continuous non-invasive BP monitoring devices using a different technology (plethysmography and volume clamp technique) and a cuff placed around a finger is uncertain. It would be interesting to evaluate whether the 2 non-invasive techniques (using arm or finger cuff) can be advantageously combined either to replace invasive monitoring or to reinforce non-invasive monitoring (in patients in whom an arterial catheter is not envisaged). In the critically ill patient, reliable detection of low BP is often the most important task assigned to BP monitoring devices, at least in the initial phase of management. It is conceivable that if the finger cuff (ClearSight™ device) was not infallible in measuring BP, low BP could still be detected sufficiently reliably by the finger cuff (even with weak accuracy), giving an early warning. The arm measurement could then confirm hypotension in a robust, on-demand manner, since the ability of the arm cuff to detect hypotension has already been demonstrated. The ability to detect hypotension using a combined non-invasive approach (finger and arm cuff) has never been explored with the ClearSight™ device. This combined approach could similarly detect high blood pressure or a rapid change in BP as a result of cardiovascular intervention. In addition, the automatic cuff cannot always be placed around the arm (presence of wounds, burns, fractures or surgery of the upper limb, venous catheter, etc.) for intermittent NIBP measurements, and is then placed around the calf. However, this commonly observed practice exposes the NIBP measurement to less accuracy. The finger cuff could be an alternative to the calf cuff. However, the accuracy of finger measurement has never been compared to the accuracy of oscillometric measurement at the calf.

Interventions

DEVICECombined non-invasive intermittent (oscillometric arm cuff), non-invasive continuous (finger cuff and ClearSight device), and invasive continuous (intra-arterial catheter) BP measurements

Continuous (finger cuff) noninvasive and invasive BP measurements, over a 30-min period. Pairs of noninvasive/invasive BP recorded each minute). In the same time, 3 pairs of oscillometric NIBP (cuff placed around a calf), and invasive BP measurements, at 30-sec intervals. The cuff is then placed around an arm and triplets of BP (arm cuff, finger cuff, intra-arterial catheter) will be taken over a 20-min period, at 3-min intervals and every time the invasive and/or the continuous finger cuff measurements cross the threshold of 65 mmHg for the mean BP. If clinically indicated, a cardiovascular intervention that can be either vascular volume expansion or initiation of vasoactive drug therapy, can be administered during this period. Finally, the cuff placed around the arm will be once again placed around the calf to take 3 pairs of oscillometric NIBP measurement with cuff placed around a calf, and simultaneous invasive BP measurement, at 30-second intervals.

Sponsors

Centre Hospitalier Régional d'Orléans
Lead SponsorOTHER

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
DIAGNOSTIC
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Patient hospitalised in one of the three recruiting intensive care units * Patient already carrying an indwelling arterial catheter * Stable BP over a 5-min period: no variation of mean BP above 10% and no change in vasopressor therapy

Exclusion criteria

* Brachial circumference \> 42 cm. * Any local fracture, wound, ischemic injury, infection, amputation, thrombophlebitis, that could prevent inflating a cuff around the arm or a finger * Need of emergency therapy not compatible with the BP measurements schedule * Asymmetry of mean BP between right and left arm (\> 5 mmHg) * Pregnancy * Age \< 18 years * Patient under curatorship or guardianship

Design outcomes

Primary

MeasureTime frameDescription
Discriminative power of the combined BP measurements with finger cuff and arm cuff to detect intra-arterial mean BP below 65 mmHg30 minutesArea under the receiver operating characteristics curve of the combined BP measurements with finger cuff and arm cuff to detect intra-arterial mean BP below 65 mmHg

Secondary

MeasureTime frameDescription
Discriminative power of the combined BP measurements with finger cuff and arm cuff to detect intra-arterial systolic hypertension30 minutesArea under the receiver operating characteristics curve of the combined BP measurements with finger cuff and arm cuff to detect systolic intra-arterial above 140 mmHg
Discriminative power of the combined BP measurements with finger cuff and arm cuff to detect intra-arterial diastolic hypertension30 minutesArea under the receiver operating characteristics curve of the combined BP measurements with finger cuff and arm cuff to detect diastolic intra-arterial above 90 mmHg
Accuracy of continuous finger cuff measurements tested against the intra-arterial reference30 minutesAccuracy of continuous finger cuff BP measurements tested against the intra-arterial reference, according to the ISO standard
Accuracy of intermittent, oscillometric arm cuff BP measurements tested against the intra-arterial reference30 minutesAccuracy of intermittent, oscillometric arm cuff BP measurements tested against the intra-arterial reference, according to the ISO standard
Accuracy of intermittent, oscillometric calf cuff measurements tested against the intra-arterial reference30 minutesAccuracy of intermittent, oscillometric calf cuff BP measurements tested against the intra-arterial reference, according to the ISO standard