NICU Oxygen Control Study

NICU Oxygen Control Study

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT04269161
Enrollment
48
Registered
2020-02-13
Start date
2022-05-24
Completion date
2024-06-06
Last updated
2025-07-04

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Hyperoxia, Hypoxia

Brief summary

Prematurely born infants in the hospital neonatal intensive care unit (NICU) will be included in the study. This clinical trial is a randomized crossover study to show that our automated oxygen control device performance is no worse than a NICU nurse in keeping a premature neonate's SPO2 within the prescribed range. Since subjects receive the device (automatic oxygen control) and the standard of care (manual control by a nurse), every subject serves as their own perfectly matched control. Performance measures include the average time it takes for the SpO2 to return to the desired range (primary endpoint) and the total amount of time that the SpO2 is within the desired range (secondary endpoint). The device will be applied to premature infants on respiratory support humidified high flow nasal cannula (HFNC) with oxygen controlled using a blend valve. Two groups include one that begins the study period with the device and one that begins the study period without the device. The two groups are switched between manual and automatic every 6 hours into the trial period and complete a total of 6 days. The target number of subjects is 60. We will analyze the study as a superiority trial if there is strong evidence of superiority.

Detailed description

We intend to enroll between 20-30 subjects at each of two separate institutions (University of Missouri Women's and Children's Hospital and Studer Family Children's Hospital in Pensacola, FL). The subjects will be premature infants \<34 weeks post conceptual age (PCA) requiring respiratory support. This randomized clinical trial will utilize a 24 period 2 treatment crossover design to show that the device performs no worse than (non-inferiority trial) an NICU nurse in keeping a premature neonate's SpO2 within the prescribed range while the infant is on respiratory support. Due to the nature of the study, any masking of the intervention is not possible. Upon enrollment, the subjects will be randomized into enrollment in two groups (A and B). The primary endpoint will be the mean time required to re-establish SpO2 within the prescribed range, as measured from the time an out-of-range alarm is triggered. A secondary endpoint will be proportion of time SpO2 is within the prescribed range, using an area-under-the-curve approach (with a discrete state) to account for varying time-on-test. These outcome measures are complementary because the former doesn't account for the number of alarms, while the latter does. This is important because the oxygen control device operates continuously in a proactive manner, rather than only reacting due alarms, so it is doing more than mimicking the nurse -- the second measure allows us to capture that. Group A will initially have the automatic device interface with HFNC for 6 hours. The device will have the target SpO2 parameters ordered by the treating physician input into the device. A study laptop will interface with the device, cardiopulmonary monitor, and pulse oximeter to record the data for the study. Sensors will be used to record all adjustments to the device/respiratory support equipment (i.e. blend valve and flow valve used in HFNC). These sensors will continuously record the data for later analysis. The device will constantly evaluate data sent to it from the pulse oximeter and bedside monitor recording all of the data and alarms. In response to alarms, displayed data, doctor's orders, etc., nurses will continue to apply manual inputs to make adjustments to flow and provide tactile stimulus to the subject but not adjust FiO2 unless manual mode is selected. Recorded sensor measurements and manual inputs by the nurse will be used to refine the existing models as well as new models of response in HR, RR, and SpO2 to flow adjustments and tactile stimulus. After the first 6 hours, the device will be switched to manual mode for the subject (nurse makes all adjustments for FiO2), but the laptop and sensor data logging system will continue to record data from the patient and the respiratory support equipment. This will record the information for the nurse intervention/baseline care part of the study, which continue for 6 hours. During the entire process, the bedside nurse will keep a diary of any events/interventions using the time prominently displayed on the monitoring laptop. This time-stamped diary system will allow for easier retrieval of and comparison to the data from the device and monitoring laptop. Also, the monitoring laptop will have a record of all the data, including alarms from the pulse oximeter as well as the bedside monitor to allow for easier retrieval of data related to alarm events and interventions. The laptop will also record any interventions made by the device to allow for easier retrieval of data related to device interventions. The treatment will then alternate periods of each treatment for a total of 6 days (24 6-hour periods). Group B will have the exact opposite order as group A. Group B infants will initially have the laptop interface with all of the monitors and sensor measurements. However, the nurse intervention/baseline care stage of the study will take place for the first 6 hours. Next, group B will have the device interface with their respiratory equipment, and the data will be recorded as described above for the next 6 hours of the study. The the treatment will alternate every 6 hours for a total of 6 days. This design was chosen because the premature infants should have fewer events as they grow older each day, and it will help take into account this potential bias. Also, the subjects will be randomized to group A or B in sets of 8 (i.e. in each group of 8 envelopes, 4 will be group A and 4 will be group B). During the entire study process the infants will receive normal NICU care and the parameters for the SpO2 range will be set by the physician caring for the infant. There are also built in manual overrides for the device which allow the NICU to make changes while the subject is on the device phase of the study. The device will be able to record these changes and the staff will record their manual interventions in the study diary. We have planned our sample size using a non-inferiority test for a cross-over design, based on our primary endpoint, t\_delta. For a given patient, define t\_delta = (mean elapsed time needed for device to re-establish SpO2 after alarm) - (mean elapsed time needed for nurse to re-establish SpO2 after alarm). The margin of non- inferiority will be chosen as t\_delta \> -10 sec, so that a device which is no worse than 10 sec, on average, than a nurse will be considered non-inferior. Assuming the standard deviation of t\_delta =12 and the true mean difference is zero under the alternate hypothesis, a sample size of 48 achieves 88% with alpha=0.05. If there is 16% patient drop-out before crossover, so that the final n=40, the power drops to 82%. In all analysis, a (paired) t-test will be used. Our secondary endpoint is secondary endpoint is the proportion of time SpO2 is within the prescribed range, using an area-under-the-curve approach (with a discrete state) to account for varying time-on-test. Our secondary endpoint will be analyzed in a similar manner. We will plan for one interim analysis to determine if the trial should be stopped early due to futility (strong evidence of inferiority, where a confidence interval for t\_delta lies entirely to the left of -10 and doesn't intersect -10) or for efficacy (strong evidence of superiority with margin \> +20 sec). This will be carried out when n=32 (16 subjects per site) is attained and stopping decisions will be based on O'Brien-Fleming stopping principles. The interim analysis will be carried out by an independent statistician on the University of Missouri's Data Safety and Monitoring Committee, which is also available to monitor the study for adverse events if requested by the IRB. In the event that the patient drop-out is greater than 16% before crossover, then a more complicated estimation procedure will be employed using mixed effects models; otherwise, complete cases will be used.

Interventions

DEVICEAutomatic control of inspired oxygen

A device will be used to automatically adjust the blend of oxygen and air with the ability to return to manual control as needed.

Sponsors

Sacred Heart Health System
CollaboratorOTHER
University of Missouri-Columbia
Lead SponsorOTHER

Study design

Allocation
RANDOMIZED
Intervention model
CROSSOVER
Primary purpose
BASIC_SCIENCE
Masking
NONE

Eligibility

Sex/Gender
ALL
Healthy volunteers
No

Inclusion criteria

* Infants admitted to the NICU * Less than 31 weeks estimated gestational age or less than 1500 grams at birth * Currently on high flow nasal cannula or bubble CPAP * Require at least 2 adjustments to the FiO2 per shift and/or have at least 2 desaturation events per shift

Exclusion criteria

* Infants admitted to the NICU with congenital heart disease. * Infants who are set on a minimum FiO2 set point by their healthcare provider

Design outcomes

Primary

MeasureTime frameDescription
Elapsed Time to Respond to SpO2 AlarmTwenty-four study periods consisting of 6-hour manual study periods and 6-hour automatic study periods over six days for each of the subjects in the analysis.Mean in elapsed time needed to re- establish SpO2 within the desired range after an alarm. The alarms and alarm times are recorded directly from the bedside monitor and the responses are measured directly in terms of the blend valve position and SpO2 response measured by the pulse oximeter.

Secondary

MeasureTime frameDescription
Proportion of Time SpO2 is Within the Prescribed Range in a Six-hour Time BlockTwelve 6-hour manual study periods and 6-hour automatic study periods over six days for each of the subjects in the analysis.For each 6-hour time block, we calculate the proportion of time the patient stays within the prescribed SpO2 range. The proportion of time that the SpO2 is within the prescribed range is computed using an area-under-the-curve approach for data taken during automatic control by the experimental device and manual control by the nurse.

Countries

United States

Participant flow

Participants by arm

ArmCount
Crossover Sequence 1 and 2, Automatic, Manual and Manual, Automatic
This group is comprised of Crossover Sequence 1 and Crossover Sequence 2 with all participants that completed six days in the study. The automatic oxygen control device or the nurse made adjustments to the blend of oxygen and air supplied to the participant for alternating blocks of six hours. The study continues in an automatic and manual repeating pattern for six days.
39
Total39

Withdrawals & dropouts

PeriodReasonFG000FG001
Overall StudyChange of modality and no longer eligible33
Overall StudyEquipment issue11
Overall StudyRequired procedure outside of NICU10

Baseline characteristics

CharacteristicCrossover Sequence 1 and 2, Automatic, Manual and Manual, Automatic
Age, Continuous42.897 days
STANDARD_DEVIATION 27.649
Bed type
Isolette
32 Participants
Bed type
Open Crib
7 Participants
Crossover Sequence 1 or 2
Sequence 1 - Automatic, Manual
19 Participants
Crossover Sequence 1 or 2
Sequence 2 - Manual, Automatic
20 Participants
Modality
Bubble CPAP (BCPAP)
17 Participants
Modality
High-Flow Nasal Cannula (HFNC)
22 Participants
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants
Race (NIH/OMB)
Asian
0 Participants
Race (NIH/OMB)
Black or African American
7 Participants
Race (NIH/OMB)
More than one race
2 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants
Race (NIH/OMB)
Unknown or Not Reported
0 Participants
Race (NIH/OMB)
White
30 Participants
Sex: Female, Male
Female
16 Participants
Sex: Female, Male
Male
23 Participants
Weight at start of study1702.205 grams
STANDARD_DEVIATION 531.817

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
deaths
Total, all-cause mortality
0 / 480 / 48
other
Total, other adverse events
0 / 480 / 48
serious
Total, serious adverse events
0 / 480 / 48

Outcome results

Primary

Elapsed Time to Respond to SpO2 Alarm

Mean in elapsed time needed to re- establish SpO2 within the desired range after an alarm. The alarms and alarm times are recorded directly from the bedside monitor and the responses are measured directly in terms of the blend valve position and SpO2 response measured by the pulse oximeter.

Time frame: Twenty-four study periods consisting of 6-hour manual study periods and 6-hour automatic study periods over six days for each of the subjects in the analysis.

Population: All participants that completed six days of the crossover study period were analyzed.

ArmMeasureValue (MEAN)Dispersion
Manual Mode (Nurse)Elapsed Time to Respond to SpO2 Alarm51.408 secondsStandard Deviation 66.168
Automatic Mode (Device)Elapsed Time to Respond to SpO2 Alarm41.523 secondsStandard Deviation 30.771
Comparison: A sample requirement of 48 patients was determined based on a pilot study to provide 88% power, and 0.05 significance (two sided) to show the mean difference is less than 10 seconds based on a 2x2 crossover design. Considering patient drop-out, a sample size of n=40 drops power to 82%.p-value: 0.00395% CI: [-16.51, -3.27]t-test, 2 sided
Secondary

Proportion of Time SpO2 is Within the Prescribed Range in a Six-hour Time Block

For each 6-hour time block, we calculate the proportion of time the patient stays within the prescribed SpO2 range. The proportion of time that the SpO2 is within the prescribed range is computed using an area-under-the-curve approach for data taken during automatic control by the experimental device and manual control by the nurse.

Time frame: Twelve 6-hour manual study periods and 6-hour automatic study periods over six days for each of the subjects in the analysis.

Population: All participants who completed six days in the study were included in the analysis.

ArmMeasureValue (MEAN)Dispersion
Manual Mode (Nurse)Proportion of Time SpO2 is Within the Prescribed Range in a Six-hour Time Block0.676 proportion of timeStandard Deviation 0.214
Automatic Mode (Device)Proportion of Time SpO2 is Within the Prescribed Range in a Six-hour Time Block0.695 proportion of timeStandard Deviation 0.224
Comparison: The difference in the proportion of time in the target saturation is calculated between the automatic and manual models. For each 6-hour time block, we calculate the proportion of time the patient stays within the prescribed SpO2 range.p-value: 0.0295% CI: [0.003, 0.034]t-test, 2 sided