Assessing Listening Effort at Different Signal-to-noise Ratios in Bone-anchored Users

Assessing Listening Effort at Different Signal-to-noise Ratios in Bone-anchored Users

Status
Terminated
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT04242940
Enrollment
14
Registered
2020-01-27
Start date
2020-02-21
Completion date
2022-05-17
Last updated
2025-05-15

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Bilateral Hearing Loss

Keywords

Bone anchored hearing systems, BAHS, Listening effort, Pupillometry, Ponto 4

Brief summary

The purpose of this study is to assess listening effort during a speech-in-noise task in bone-anchored hearing systems (BAHS) users via pupillometry.

Detailed description

Pupil dilation can be used as an objective indicator of listening effort during the execution of a task. The aim is to compare listening effort with different settings (OpenSound Navigator ON and OFF) of Ponto 4, the sound processor released by Oticon Medical AB in June 2019, in listening scenarios that differ in complexity. The study consists of three visits of about two hours each in duration, during which the patients will perform a speech-in-noise task where the listening task varies in difficulty (i.e., in terms of signal-to-noise ratio, SNR). Pupil dilation will be recorded during the listening task as an indicator of listening effort. The patients will also use Ponto 4, instead of their own Ponto device, during a field-trial period of three months in between visit 2 and 3. Subjective evaluations of Ponto 4 will be performed via three questionnaires, one regarding self-reported performance in everyday life, one regarding work-related fatigue, and one regarding overall sound processor preference (Ponto 4 vs. patient's own device). All these outcome measures are non-invasive measurements. This is a post market study and all products used are CE marked and used in clinical practice worldwide.

Interventions

DEVICEFitting of Ponto 4

Bone-anchored hearing systems (BAHS) use the body's natural ability to transfer sound through bone conduction. The sound processor picks up sound and converts it into vibrations that are transferred through the skull bone to the inner ear (cochlea). Thus, for patients with conductive or mixed hearing losses, patients with lasting hearing loss following a middle ear disease or malformations (such as microtia), the vibrations are bypassing the conductive problem in the ear canal or middle ear. The intervention in this study is audiologically fitting one/two bone-anchored sound processors (Ponto 4) to patients that are already implanted with abutments.

Sponsors

Oticon Medical
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
BASIC_SCIENCE
Masking
NONE

Masking description

The participants are however blinded to the Ponto 4 setting (OpenSound Navigator ON/OFF in Ponto 4) activated for each list of sentences.

Eligibility

Sex/Gender
ALL
Age
18 Years to 70 Years
Healthy volunteers
No

Inclusion criteria

* Minimum 18 years and maximum 70 years old. * UK English mother tongue. * Bilateral conductive or mixed hearing loss. * Users fitted unilaterally or bilaterally on abutment. * Users are currently fitted with either Ponto 3, Ponto Plus or Ponto Pro (devices with a fitting range up to 45 dB HL). * Users with at least 6 months of daily experience with one Ponto sound processor. * BC thresholds of the implanted ear(s) better (lower) than or equal to 40 dB HL (average of 0.5, 1, 2 and 3 kHz) based on BC in-situ measurement at the first visit. * People currently employed/self-employed (since the Need for Recovery questionnaire refers to work-related fatigue). * Speech recognition scores better (higher) than or equal to 50% at +4 dB SNR with OSN OFF (as assessed via the first training list at Visit 1). This is to avoid including patients that are already giving up at +4 dB SNR. * Users entitled for a sound-processor upgrade (according to the investigator) will first be selected. If there are not enough users entitled for an upgrade, then patients fitted with Ponto between 7 and 12 months before visit 1 will be considered. If there are not enough users within 7-12 months, then patients fitted with Ponto between 13 and 18 months before visit 1 will be considered.

Exclusion criteria

* Single-sided deafness (SSD) * Hearing aid or cochlear implant on the opposite ear. * Test persons with current eye diseases and/or history of eye surgery on both eyes. * If the patient takes medications that can impact the autonomic nervous system - and hence pupil dilation (e.g., drugs for Parkinson's disease, peptic ulcers, incontinence, motion sickness). * If the investigator or physician assesses that the patient is not fit for trial participation.

Design outcomes

Primary

MeasureTime frameDescription
Overall Pupil Dilation10 daysOverall pupil dilation for OSN ON and OSN OFF at +4 dB SNR a third-order polynomial function was fit to the pupil dilation data, with the OSN setting as a fixed factor, as well as the interaction of OSN with each of the polynomial time terms. The formula used is the following: Pupil Dilation = (Linear + Quadratic + Cubic) X OSN setting + (1 + Linear + Quadratic + Cubic\| Subject); where the Linear, Quadratic, and Cubic are the orthogonal terms; X indicates the interaction with OSN setting, and the terms reported in the second parenthesis are the random factors. The random factors include the effect of listeners, and of each of the time terms, to account for the variability in the time course of dilation across listeners. Outcome measure is the percentage reduction(% change) in pupil dilation with OSN ON vs OSN OFF

Secondary

MeasureTime frameDescription
Peak Pupil Dilation (PPD) for OSN ON and OSN OFF10 daysThe difference in pupil dilation between OSN ON and OSN OFF was also analyzed via Peak Pupil Dilation (PPD); thus, by investigating differences in dilation at the peak of the curve. The PPD was obtained for each subject as the maximum dilation between 7.5 and 9 s.
Speech Recognition Scores With OSN ON vs. OSN OFF Across SNRs10 daysA mixed-linear model with OSN setting (2 levels: ON and OFF) and SNR (5 levels: -8, -4, 0, +4, +8 dB) as fixed factors, and subject as random factor, was fit to the data.
Speech Spatial and Qualities of Hearing Scale (SSQ) Questionnaire Scores4 monthsSelf-reported perception before, during (at 1 month), and after field trial (at 3months). The SSQ questionnaire consists of 49 items with a scale from 0-10 where 0 represents Not at all and 10 represents Perfect. The SSQ questionnaire is comprised of three subscales domains (Speech, Spatial, and Qualities of Hearing) - that is, each of the 49 items fits into one of these subscale categories. The subscale scores AND the total scores (the total score is comprised of all of the 49 items) use the same scale of 0 to 10 where 0 represents Not at all and 10 represents Perfect. For the total scale and all subscales (again all questions and by extension, the scales range from 0 to 10) the higher the value (closer to 10) represents a better outcome than lower values (closer to 0).
Aided Sound Field Thresholds4 monthsSound field audiometry performed with the Auricle system while the patient is wearing the Ponto sound processor. This measurement is conducted at 10 frequencies from 250 Hz to 8 kHz. The Pure tone threshold is calculated as the average across .5, 1, 2, & 3 kHz)
BC in Situ Thresholds.1 hourHearing level thresholds (dB HL) as assessed in situ with Ponto 4 placed on the patient's abutment. Bone-conduction thresholds (BC) are measured as the softest sound level that is audible by the patient at 10 frequencies from 250 Hz to 8 kHz. The Pure tone threshold is calculated as the average across .5, 1, 2, & 3 kHz
Preference Questionnaire Scores.4 monthsSelf-reported preference after field trial. The preference questionnaire consists of 11 questions. The scores are from 1 to 5, where 1 means Own sound processor much better and 5 means Test sound processor \[Ponto 4\] much better.

Other

MeasureTime frameDescription
Need for Recovery (NFR) Questionnaire Scores.4 monthsSelf-reported work-related fatigue. The NFR scale is an 11-item scale assessing the effects of fatigue caused by work (e.g., I find it hard to relax at the end of a working day). Possible responses are yes or no. The total NFR score is the number of yes responses divided by the total number of items, presented as a percentage (i.e., range 0-100). The higher the score, the greater the NfR felt by the respondent. Scores with own device compared with Ponto 4 after 3months field trial are reported.
Pupil Dilation After Field Trial.4 monthsChange in pupil dilation after field trial to check for acclimatization effects. Comparison of speech recognition scores for OSN Auto and OSN OFF at start and end of field trial.

Countries

United Kingdom

Participant flow

Participants by arm

ArmCount
Ponto 4 Sound Processor
All patients will be fitted with a bone-anchored sound processor (Ponto 4), unilaterally or bilaterally. Fitting of Ponto 4: Bone-anchored hearing systems (BAHS) use the body's natural ability to transfer sound through bone conduction. The sound processor picks up sound and converts it into vibrations that are transferred through the skull bone to the inner ear (cochlea). Thus, for patients with conductive or mixed hearing losses, patients with lasting hearing loss following a middle ear disease or malformations (such as microtia), the vibrations are bypassing the conductive problem in the ear canal or middle ear. The intervention in this study is audiologically fitting one/two bone-anchored sound processors (Ponto 4) to patients that are already implanted with abutments.
14
Total14

Baseline characteristics

CharacteristicPonto 4 Sound Processor
Age, Continuous49.9 years
STANDARD_DEVIATION 12.4
BC in situ
Left Ear
22.61 decibels
STANDARD_DEVIATION 10.61
BC in situ
Right Ear
20.83 decibels
STANDARD_DEVIATION 9.82
Ponto fitting side
Bilateral
8 participants
Ponto fitting side
Left Side
2 participants
Ponto fitting side
Right Side
3 participants
PTA AC
Left Ear
61.96 decibels
STANDARD_DEVIATION 12.25
PTA AC
Right Ear
63.75 decibels
STANDARD_DEVIATION 12.53
PTA BC
Left Ear
26.52 decibels
STANDARD_DEVIATION 9.53
PTA BC
Right Ear
27.23 decibels
STANDARD_DEVIATION 9.62
Race and Ethnicity Not Collected— Participants
Sex: Female, Male
Female
7 Participants
Sex: Female, Male
Male
7 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 14
other
Total, other adverse events
0 / 14
serious
Total, serious adverse events
0 / 14

Outcome results

Primary

Overall Pupil Dilation

Overall pupil dilation for OSN ON and OSN OFF at +4 dB SNR a third-order polynomial function was fit to the pupil dilation data, with the OSN setting as a fixed factor, as well as the interaction of OSN with each of the polynomial time terms. The formula used is the following: Pupil Dilation = (Linear + Quadratic + Cubic) X OSN setting + (1 + Linear + Quadratic + Cubic\| Subject); where the Linear, Quadratic, and Cubic are the orthogonal terms; X indicates the interaction with OSN setting, and the terms reported in the second parenthesis are the random factors. The random factors include the effect of listeners, and of each of the time terms, to account for the variability in the time course of dilation across listeners. Outcome measure is the percentage reduction(% change) in pupil dilation with OSN ON vs OSN OFF

Time frame: 10 days

Population: Four subjects had too many eye blinks in their pupillometry recordings (\>15% of trials) and therefore their pupil data were unable to be analyzed.~Note, OSN ON, OSN OFF are not two separate arms/groups. OSN ON, OSN OFF are test conditions.

ArmMeasureValue (MEAN)Dispersion
Ponto 4 Sound ProcessorOverall Pupil Dilation58.7 percentage changeStandard Error 0.3
Secondary

Aided Sound Field Thresholds

Sound field audiometry performed with the Auricle system while the patient is wearing the Ponto sound processor. This measurement is conducted at 10 frequencies from 250 Hz to 8 kHz. The Pure tone threshold is calculated as the average across .5, 1, 2, & 3 kHz)

Time frame: 4 months

Population: A threshold was unable to be obtained for two subjects' right ears and three left ears.

ArmMeasureGroupValue (MEAN)Dispersion
Ponto 4 Sound ProcessorAided Sound Field ThresholdsAided Right Ear32.29 decibelsStandard Deviation 9.35
Ponto 4 Sound ProcessorAided Sound Field ThresholdsAided Left Ear37.95 decibelsStandard Deviation 9.12
Secondary

BC in Situ Thresholds.

Hearing level thresholds (dB HL) as assessed in situ with Ponto 4 placed on the patient's abutment. Bone-conduction thresholds (BC) are measured as the softest sound level that is audible by the patient at 10 frequencies from 250 Hz to 8 kHz. The Pure tone threshold is calculated as the average across .5, 1, 2, & 3 kHz

Time frame: 1 hour

Population: A threshold was unable to be obtained for two subjects' right ears and three left.

ArmMeasureGroupValue (MEAN)Dispersion
Ponto 4 Sound ProcessorBC in Situ Thresholds.BC in situ Right Ear20.83 decibelsStandard Deviation 9.82
Ponto 4 Sound ProcessorBC in Situ Thresholds.BC in situ Left Ear22.61 decibelsStandard Deviation 10.61
Secondary

Peak Pupil Dilation (PPD) for OSN ON and OSN OFF

The difference in pupil dilation between OSN ON and OSN OFF was also analyzed via Peak Pupil Dilation (PPD); thus, by investigating differences in dilation at the peak of the curve. The PPD was obtained for each subject as the maximum dilation between 7.5 and 9 s.

Time frame: 10 days

Population: Four subjects had too many eye blinks in their pupillometry recordings (\>15% of trials) and therefore their pupil data were unable to be analyzed.

ArmMeasureGroupValue (MEAN)Dispersion
Ponto 4 Sound ProcessorPeak Pupil Dilation (PPD) for OSN ON and OSN OFFPPD - OSN OFF1.80 Arbitrary unitsStandard Deviation 1.34
Ponto 4 Sound ProcessorPeak Pupil Dilation (PPD) for OSN ON and OSN OFFPPD - OSN ON1.05 Arbitrary unitsStandard Deviation 1.56
Secondary

Preference Questionnaire Scores.

Self-reported preference after field trial. The preference questionnaire consists of 11 questions. The scores are from 1 to 5, where 1 means Own sound processor much better and 5 means Test sound processor \[Ponto 4\] much better.

Time frame: 4 months

Population: One subject did not complete the preference questionnaire.

ArmMeasureGroupValue (NUMBER)
Ponto 4 Sound ProcessorPreference Questionnaire Scores.Prefer Ponto 411 participants
Ponto 4 Sound ProcessorPreference Questionnaire Scores.No preference2 participants
Ponto 4 Sound ProcessorPreference Questionnaire Scores.Prefer own device0 participants
Secondary

Speech Recognition Scores With OSN ON vs. OSN OFF Across SNRs

A mixed-linear model with OSN setting (2 levels: ON and OFF) and SNR (5 levels: -8, -4, 0, +4, +8 dB) as fixed factors, and subject as random factor, was fit to the data.

Time frame: 10 days

Population: Note that one subject did not perform the measurement at 0 dB SNR for OSN ON. Hence, the mean and std for OSN ON at 0 dB SNR is calculated for N=13, while all the other conditions are averaged across all subjects (N=14).

ArmMeasureGroupValue (MEAN)Dispersion
Ponto 4 Sound ProcessorSpeech Recognition Scores With OSN ON vs. OSN OFF Across SNRsOSN ON -8dB SNR21.66 Percentage correctStandard Deviation 16.24
Ponto 4 Sound ProcessorSpeech Recognition Scores With OSN ON vs. OSN OFF Across SNRsOSN ON -4dB SNR51.77 Percentage correctStandard Deviation 19.63
Ponto 4 Sound ProcessorSpeech Recognition Scores With OSN ON vs. OSN OFF Across SNRsOSN ON 0dB SNR79.26 Percentage correctStandard Deviation 13.96
Ponto 4 Sound ProcessorSpeech Recognition Scores With OSN ON vs. OSN OFF Across SNRsOSN ON +4dB SNR95.41 Percentage correctStandard Deviation 3.33
Ponto 4 Sound ProcessorSpeech Recognition Scores With OSN ON vs. OSN OFF Across SNRsOSN ON +8dB SNR97.41 Percentage correctStandard Deviation 2.49
Ponto 4 Sound ProcessorSpeech Recognition Scores With OSN ON vs. OSN OFF Across SNRsOSN OFF -8dB SNR5.82 Percentage correctStandard Deviation 8.27
Ponto 4 Sound ProcessorSpeech Recognition Scores With OSN ON vs. OSN OFF Across SNRsOSN OFF -4dB SNR30.85 Percentage correctStandard Deviation 25.61
Ponto 4 Sound ProcessorSpeech Recognition Scores With OSN ON vs. OSN OFF Across SNRsOSN OFF 0dB SNR66.51 Percentage correctStandard Deviation 16.31
Ponto 4 Sound ProcessorSpeech Recognition Scores With OSN ON vs. OSN OFF Across SNRsOSN OFF +4dB SNR89.45 Percentage correctStandard Deviation 6.38
Ponto 4 Sound ProcessorSpeech Recognition Scores With OSN ON vs. OSN OFF Across SNRsOSN OFF +8dB SNR95.78 Percentage correctStandard Deviation 3.95
Secondary

Speech Spatial and Qualities of Hearing Scale (SSQ) Questionnaire Scores

Self-reported perception before, during (at 1 month), and after field trial (at 3months). The SSQ questionnaire consists of 49 items with a scale from 0-10 where 0 represents Not at all and 10 represents Perfect. The SSQ questionnaire is comprised of three subscales domains (Speech, Spatial, and Qualities of Hearing) - that is, each of the 49 items fits into one of these subscale categories. The subscale scores AND the total scores (the total score is comprised of all of the 49 items) use the same scale of 0 to 10 where 0 represents Not at all and 10 represents Perfect. For the total scale and all subscales (again all questions and by extension, the scales range from 0 to 10) the higher the value (closer to 10) represents a better outcome than lower values (closer to 0).

Time frame: 4 months

Population: Two subjects did not complete the SSQ at the 1 month timepoint.

ArmMeasureGroupValue (MEAN)Dispersion
Ponto 4 Sound ProcessorSpeech Spatial and Qualities of Hearing Scale (SSQ) Questionnaire ScoresBefore Trial - Speech4.98 score on a scaleStandard Deviation 1.6
Ponto 4 Sound ProcessorSpeech Spatial and Qualities of Hearing Scale (SSQ) Questionnaire ScoresBefore Trial - Spatial3.87 score on a scaleStandard Deviation 1.35
Ponto 4 Sound ProcessorSpeech Spatial and Qualities of Hearing Scale (SSQ) Questionnaire ScoresBefore Trial - Qualities5.98 score on a scaleStandard Deviation 1.63
Ponto 4 Sound ProcessorSpeech Spatial and Qualities of Hearing Scale (SSQ) Questionnaire ScoresBefore Trial - Total4.94 score on a scaleStandard Deviation 1.33
Ponto 4 Sound ProcessorSpeech Spatial and Qualities of Hearing Scale (SSQ) Questionnaire Scores1 month - Speech5.67 score on a scaleStandard Deviation 1.46
Ponto 4 Sound ProcessorSpeech Spatial and Qualities of Hearing Scale (SSQ) Questionnaire Scores1 month - Spatial4.87 score on a scaleStandard Deviation 1.66
Ponto 4 Sound ProcessorSpeech Spatial and Qualities of Hearing Scale (SSQ) Questionnaire Scores1 month - Qualities6.67 score on a scaleStandard Deviation 1.13
Ponto 4 Sound ProcessorSpeech Spatial and Qualities of Hearing Scale (SSQ) Questionnaire Scores1 month - Total5.74 score on a scaleStandard Deviation 1.34
Ponto 4 Sound ProcessorSpeech Spatial and Qualities of Hearing Scale (SSQ) Questionnaire Scores3 months - Speech5.88 score on a scaleStandard Deviation 1.35
Ponto 4 Sound ProcessorSpeech Spatial and Qualities of Hearing Scale (SSQ) Questionnaire Scores3 months - Spatial4.83 score on a scaleStandard Deviation 1.45
Ponto 4 Sound ProcessorSpeech Spatial and Qualities of Hearing Scale (SSQ) Questionnaire Scores3 months - Qualities6.46 score on a scaleStandard Deviation 1.12
Ponto 4 Sound ProcessorSpeech Spatial and Qualities of Hearing Scale (SSQ) Questionnaire Scores3 months - Total5.72 score on a scaleStandard Deviation 1.23
Other Pre-specified

Need for Recovery (NFR) Questionnaire Scores.

Self-reported work-related fatigue. The NFR scale is an 11-item scale assessing the effects of fatigue caused by work (e.g., I find it hard to relax at the end of a working day). Possible responses are yes or no. The total NFR score is the number of yes responses divided by the total number of items, presented as a percentage (i.e., range 0-100). The higher the score, the greater the NfR felt by the respondent. Scores with own device compared with Ponto 4 after 3months field trial are reported.

Time frame: 4 months

ArmMeasureGroupValue (MEAN)Dispersion
Ponto 4 Sound ProcessorNeed for Recovery (NFR) Questionnaire Scores.Own device47.4 score on a scaleStandard Deviation 37.3
Ponto 4 Sound ProcessorNeed for Recovery (NFR) Questionnaire Scores.Ponto 4 (3 months)47.4 score on a scaleStandard Deviation 34
Other Pre-specified

Pupil Dilation After Field Trial.

Change in pupil dilation after field trial to check for acclimatization effects. Comparison of speech recognition scores for OSN Auto and OSN OFF at start and end of field trial.

Time frame: 4 months

ArmMeasureGroupValue (MEAN)Dispersion
Ponto 4 Sound ProcessorPupil Dilation After Field Trial.OSN Auto at start94.7 percent correct speech scoresStandard Deviation 5.1
Ponto 4 Sound ProcessorPupil Dilation After Field Trial.OSN Auto at end94.4 percent correct speech scoresStandard Deviation 2.7
Ponto 4 Sound ProcessorPupil Dilation After Field Trial.OSN OFF at start89.5 percent correct speech scoresStandard Deviation 6.4
Ponto 4 Sound ProcessorPupil Dilation After Field Trial.OSN OFF at end86.9 percent correct speech scoresStandard Deviation 7.6