Assessing the Accuracy of the A&D TM-2657W Oscillometric Device

Assessing the Accuracy of the A&D TM-2657W Oscillometric Device

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT04238442
Enrollment
121
Registered
2020-01-23
Start date
2019-08-14
Completion date
2020-02-05
Last updated
2020-04-10

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Blood Pressure

Brief summary

The A&D automatic blood pressure device is a new kiosk that can measure blood pressure in the general population. The investigators plan to use the existing International Standards Organization validation protocol to determine whether this device measures blood pressure accurately.

Detailed description

Methods and Analysis Study methodology will follow the ISO protocol: * 85 adult (age 18 years or greater) subjects will be included (it is estimated that approximately 140 subjects will need to be screened as some will be ineligible). 85 is the recommended sample size for validation studies using the ISO protocol. * Subjects will be recruited via an existing registry of over 250 individuals that have participated in past measurement studies, through advertisements, and from the University of Alberta Hypertension Clinic. * Specific requirements for a certain percentage of subjects with specific age, sex, upper arm circumference, and baseline blood pressure levels will be observed to ensure that a representative sample is studied (according to the ISO standards). * Subjects that are pregnant or with atrial fibrillation will be excluded. Blood pressure measurements will be taken using recommended, optimal technique. * Nine measurements will be taken in each subject, alternating between blinded two-observer auscultation with a mercury-based sphygmomanometer (the reference standard) and the TM-2657. Careful attention will be paid to ensuring proper cuffing. The first two measurements will be discarded and the latter seven measurements (4 auscultatory and TM-2657 measurements) will be used in the analysis, which will be conducted according to ISO standards. * Accuracy will be assessed according to Criterion 1 and 2 of the ISO standard. Bland-Altman plots will be generated. A plot of measurement error versus arm circumference will be generated as well.

Interventions

DEVICEA&D TM-2657W oscillometric device

Assessing the accuracy of the A&D TM-2657W oscillometric device using International Standards Organization validation protocol. This involves comparing the accuracy of the A&D device to 2 observer mercury auscultation (gold standard).

Sponsors

A&D Medical
CollaboratorUNKNOWN
University of Alberta
Lead SponsorOTHER

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
DIAGNOSTIC
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

* 18 years of age or older * normal sinus rhythm * have capacity to give consent * following the International Standards Organization criteria for: gender distribution, limb size, blood pressure distribution

Exclusion criteria

* atrial fibrillation * pregnancy

Design outcomes

Primary

MeasureTime frameDescription
Accuracy per International Standards Organization Protocol3 monthsAgreement of the device to mercury auscultation.