Skip to content

Comparison of Fingerstick Versus Venous Sample for Troponin I.

Sample and Method Comparison With Minicare® Point-of-care Device for Cardiac Troponin I Assay at the Emergency Department.

Status
Completed
Phases
Unknown
Study type
Observational
Source
ClinicalTrials.gov
Registry ID
NCT04153006
Enrollment
131
Registered
2019-11-06
Start date
2019-09-03
Completion date
2020-04-13
Last updated
2022-06-10

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Chest Pain, Heart Attack, Myocardial Infarction, Myocardial Ischemia

Keywords

Troponin, Point-of-care, Minicare

Brief summary

This study is a prospective, observational, cohort study aiming to compare point-of-care high-sensitive troponin I testing from different sample types with central laboratory (CL) HS cTnI plasma samples. A registry of all included patients and their troponin results (POC, CL and HS cTnT) will be made to compare these testing methods.

Detailed description

Point-of-care (POC) troponin testing, defined as laboratory testing near a patient location with rapid availability of results, has attracted much interest in the emergency department setting (ED) and seems feasible. These devices might enable earlier decisions, reduce stay at the ED and improve patient flow. While an elevated troponin in patients with suspected acute coronary syndrome (ACS) confirms diagnosis and initiates adequate treatment, ruling out ACS aids in proper patient dismissal. A next step could be ruling out myocardial infarction by the general practitioner (GP) or fast responder using an on-site POC troponin test. However, drawing venous blood might not be easily available to every GP, especially not during peak hours. The Minicare cTnI is a bedside system which requires capillary blood, venous whole blood or plasma. The results of the troponin will be given within 10 minutes. It is a very sensitive troponin test, the most clinically sensitive available POC for c-Troponin. Currently Minicare prepares for a high sensitive troponin analysis targeting a reliable result within an hour after onset of chest pain. The objective of this study is to determine if high sensitive troponin testing by Minicare (POC HS cTnI) has the same analytical performance as standard high sensitivity troponin I testing in our central hospital laboratory ( CL HS cTnI ARCHITECT immunoassay analyzer, Abbott).

Interventions

DEVICEPOC troponin analysis.

Point-of-care (POC) high sensitive troponin I (HS cTnI) analysis in whole blood, plasma and capillary whole blood with the Minicare® device.

Sponsors

Siemens Healthcare Diagnostics Inc
CollaboratorINDUSTRY
Minicare® B.V.
CollaboratorUNKNOWN
VieCuri Medical Centre
Lead SponsorOTHER

Study design

Observational model
COHORT
Time perspective
RETROSPECTIVE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
No

Inclusion criteria

• Patients 18 years or older with chest pain suspected of ACS.

Exclusion criteria

* ST elevation myocardial infarction and out of hospital cardiac arrest. * Patients with sudden onset tachycardia and a frequency of 110 bpm or higher (supraventricular or ventricular) * Patients who are hemodynamically unstable or in which an acute non-coronary diagnosis is suspected, e.g. pulmonary embolism, thoracic aortic dissection etc. * Patients recently already admitted for the same set of symptoms at a previous healthcare institution before being transferred to the participating clinical site. * Patients not willing or not able to provide informed consent due to their medical condition as judged by the physician.

Design outcomes

Primary

MeasureTime frameDescription
Method and sample comparison30 daysThe primary objective is to compare the analytical performance (method and sample comparison) of Minicare® high sensitive troponin I testing (POC, different sample types) and conventional venipuncture troponin I test in our central hospital laboratory (CL) with the Abbott Architect. This comparison will comprise: * Minicare® POC (different sample types) vs conventional HS cTnI CL test (Method comparison) * Minicare® capillary vs. Minicare® venipuncture and vs. Minicare® plasma (Sample comparison) The analyses linked to the primary objectives are: * The agreement between POC (three sample types) and CL testing by using the Bland-Altman method. * The agreement between different POC sample types by using the Bland-Altman method. The primary objective will be achieved by taking capillary and (an extra) venous blood samples (one per timepoint) from every patient presenting at the cardiac ED with chest pain suspected for ACS.

Secondary

MeasureTime frameDescription
Final patients diagnoses.30 daysTo compare final patients diagnosis (ACS vs. no ACS) and treatment based on POC venous troponin testing versus CL plasma troponin I testing versus HS cTnT plasma testing (regular patient care).
Overview baseline characteristics.30 daysTo create an overview of baseline characteristics of the population.
Major adverse cardiac event (MACE)30 daysMACE is is defined as a composite of cardiac death and myocardial infarction.
Fals-positive and fals-negatives.30 daysTo compare the proportion of false positive and false negative results between the different POC samples and if there is a significant difference.
Safety analysis of POC troponin.30 daysTo create an overview of the sensitivity, specificity, negative predictive value (NPV) and positive predictive value (PPV) of POC and standard laboratory troponin I testing.
Linear regression and Pearson's correlation.30 days* The relationship between POC (three sample types) and CL by linear regression and Pearson's correlation. * The relationship between POC sample types by linear regression and Pearson's correlation.

Countries

Netherlands

Outcome results

None listed