A Study With Technolas® TENEO 317 Model 2 Excimer Laser to Treat Participants With Myopia or Myopic Astigmatism

A Study to Investigate the Safety and Effectiveness of the Technolas® TENEO 317 Model 2 Excimer Laser for Laser In Situ Keratomileusis (LASIK) Surgery to Treat Myopia or Myopic Astigmatism

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT04111757
Enrollment
168
Registered
2019-10-01
Start date
2019-07-25
Completion date
2021-08-06
Last updated
2024-06-14

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Myopia, Myopic Astigmatism

Keywords

LASIK, Excimer Laser

Brief summary

The primary objective of this study is to collect safety and effectiveness data for the Technolas Teneo 317 Model 2 excimer laser for LASIK correction in participants with myopia and myopic astigmatism.

Detailed description

Technolas TENEO 317 Model 2 excimer laser is a scanning excimer laser that operates at 193 nm ultraviolet wavelength to photoablate corneal tissue in order to achieve a refractive change. It is planned to enroll and treat up to 334 participant eyes, with an expectation that 300 study eyes will complete post-surgical follow-up for 6 months or until the point of achieved refractive stability. When a cohort of at least 300 eyes has achieved refractive stability at 6 months post-surgery, all other treated eyes that have not reached the 6-month examination may be discontinued at the Sponsor's request. Or, if 300 eyes achieve refractive stability at 9 months post-surgery, all other treated eyes that have not reached the 9-month examination may be discontinued

Interventions

DEVICETechnolas® TENEO 317 Model 2

LASIK eye surgery

Sponsors

Bausch & Lomb Incorporated
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
TREATMENT
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
22 Years to No maximum
Healthy volunteers
No

Inclusion criteria

* Are 22 years of age or older. * Have read, understood, and signed an informed consent form (ICF). * Have demonstrated stable refraction (for example, a change of ≤0.5 D in sphere and cylinder) for a minimum of 12 months prior to surgery, verified by consecutive refractions and/or medical records or prescription history. * Have myopic refractive error with or without astigmatism; sphere between -1.0 D and -10.00 D, cylinder between 0.0 D and -3.0 D; with a manifest refraction spherical equivalent (MRSE) between -1.0 D and -11.50 D. * Have uncorrected distance visual acuity (UDVA) of 20/40 or worse. * Have manifest distance best spectacle corrected visual acuity (BSCVA) of 20/25 (logarithm of the minimum angle of resolution \[logMAR\] 0.1) or better in an operative eye. * Have equal to or less than 0.50 D spherical equivalent (SE) difference between cycloplegic and manifest refractions at Visit 1 (pre-operative). * Have normal corneal topography as determined by the Investigator. * Have discontinued use of contact lenses for at least 2 weeks (for hard or toric lenses) or 3 days (for soft contact lenses) prior to the pre-operative examination, and through the day of surgery. * All contact lens wearers must demonstrate a stable refraction (within ±0.5 D), as determined by MRSE, on 2 consecutive examinations at least 1 week apart, in an eye to be treated and the axis of cylinder should not differ by more than 15 degrees from the baseline cycloplegic refraction. * Have the ability to lie flat without difficulty. * Are willing and able to comply with the schedule for all post-surgery follow-up visits.

Exclusion criteria

* Participants for whom the combination of their baseline corneal thickness and the planned operative parameters for the LASIK procedure would result in treatment depth less than 250 microns from corneal endothelium. * Eyes for which the baseline manifest subjective refraction exhibits a difference greater than 0.50 D in sphere power, or a difference greater than 0.50 D in cylinder power, or a difference in cylinder axis of more than 15 degrees compared to the baseline cycloplegic subjective refraction. For manifest cylinder of less than 0.50 D, the difference in cylinder axis will not be taken into consideration. * Participants for whom the pre-operative assessment of the cornea indicates that one or both eyes are not suitable candidates for treatment based upon the Investigator's medical judgment. * Have evidence of retinal vascular disease. * Have a history of or have active corneal disease or infection (for example, recurrent corneal erosion syndrome, herpes simplex, or herpes zoster keratitis) in either eye. * Have a known sensitivity to any study medication. * Have central corneal scars affecting visual acuity or unstable keratometry with irregular mires in an eye considered for eligibility. * Have keratoconus, subclinical or forme fruste keratoconus, corneal dystrophy, or other corneal irregularity (for example, irregular astigmatism). * Have visually significant or progressive cataract in an eye considered for eligibility. * Had previous intraocular or corneal surgery in an eye considered for eligibility that might confound the outcome of the study or increase the risk to the participant. * Use chronic medications by any administration route that may increase risk to the participant or may confound the outcome of the study, including those known to affect wound healing (for example, corticosteroids, antimetabolites). * Are known to have acute or chronic disease or illness (for example, dry eye, cataract, glaucoma, immuno-compromised, rheumatoid arthritis, clinically significant atopic disease, acne rosacea) that would increase operative risk or may confound the results of the study. * Are taking medications contraindicated for LASIK such as isotretinoin (Accutane) or amiodarone hydrochloride (Cordarone). * Are known to be pregnant, lactating, or who plan to become pregnant during the course of the study. * Have known sensitivity to medications used for standard LASIK. * Have the presence of systemic disease likely to affect wound healing, for example, autoimmune disease, systemic connective tissue disease, diabetes, or severe atopic disease. * Are participating in any other ophthalmic clinical trial within 30 days of screening or during this clinical trial. * Have an ocular muscle disorder including a strabismus or nystagmus, or other disorder affecting fixation. * Have a history of or evidence of glaucoma or are a glaucoma suspect. * Have eyes with mesopic pupil size \>7.0 millimeters (mm). * Have a Schirmer's pre-operative test without anesthesia \<4 mm/5 minutes.

Design outcomes

Primary

MeasureTime frameDescription
Percentage of Eyes That Achieve MRSE Predictability Within ±0.50 Diopter (D) and ±1.00 DAssessed at Month 3 to Month 9Predictability is defined as the difference between the attempted and achieved manifest refraction spherical equivalent (MRSE) at the time of refractive stability.
Percentage of Eyes Targeted for Emmetropia That Achieve Uncorrected Distance Visual Acuity (UDVA) of 20/40 or BetterAssessed at Month 3 to Month 9Percentage of eyes targeted for emmetropia that achieve UDVA of 20/40 or better at the point at which refractive stability is reached.

Countries

United States

Participant flow

Participants by arm

ArmCount
Technolas TENEO 317 Model 2
One or both eyes of participants will undergo LASIK surgery with the Technolas® TENEO 317 Model 2 (version 1.28 US software) Excimer Laser on Day 0. Technolas® TENEO 317 Model 2: LASIK eye surgery
168
Technolas TENEO 317 Model 2
One or both eyes of participants will undergo LASIK surgery with the Technolas® TENEO 317 Model 2 (version 1.28 US software) Excimer Laser on Day 0. Technolas® TENEO 317 Model 2: LASIK eye surgery
333
Total501

Baseline characteristics

CharacteristicTechnolas TENEO 317 Model 2
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
0 Participants
Age, Categorical
Between 18 and 65 years
168 Participants
Age, Continuous32.5 Years
STANDARD_DEVIATION 6.63
Ethnicity (NIH/OMB)
Hispanic or Latino
19 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
149 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
0 Participants
Race (NIH/OMB)
American Indian or Alaska Native
1 Participants
Race (NIH/OMB)
Asian
18 Participants
Race (NIH/OMB)
Black or African American
3 Participants
Race (NIH/OMB)
More than one race
3 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants
Race (NIH/OMB)
Unknown or Not Reported
3 Participants
Race (NIH/OMB)
White
140 Participants
Region of Enrollment
United States
168 Participants
Sex: Female, Male
Female
101 Participants
Sex: Female, Male
Male
67 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 168
other
Total, other adverse events
0 / 168
serious
Total, serious adverse events
0 / 168

Outcome results

Primary

Percentage of Eyes Targeted for Emmetropia That Achieve Uncorrected Distance Visual Acuity (UDVA) of 20/40 or Better

Percentage of eyes targeted for emmetropia that achieve UDVA of 20/40 or better at the point at which refractive stability is reached.

Time frame: Assessed at Month 3 to Month 9

ArmMeasureValue (NUMBER)
Technolas TENEO 317 Model 2Percentage of Eyes Targeted for Emmetropia That Achieve Uncorrected Distance Visual Acuity (UDVA) of 20/40 or Better304 Eyes
Primary

Percentage of Eyes That Achieve MRSE Predictability Within ±0.50 Diopter (D) and ±1.00 D

Predictability is defined as the difference between the attempted and achieved manifest refraction spherical equivalent (MRSE) at the time of refractive stability.

Time frame: Assessed at Month 3 to Month 9

ArmMeasureGroupValue (NUMBER)
Technolas TENEO 317 Model 2Percentage of Eyes That Achieve MRSE Predictability Within ±0.50 Diopter (D) and ±1.00 D±0.50 Diopter287 Eyes
Technolas TENEO 317 Model 2Percentage of Eyes That Achieve MRSE Predictability Within ±0.50 Diopter (D) and ±1.00 D±1.00 Diopter302 Eyes

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026