Analgesia, Chronic Pain, Heat, Low Back Pain
Conditions
Keywords
chronic pain, low back pain, heat, thermal, analgesia
Brief summary
This study is a double blinded randomized active placebo controlled trial in subjects with chronic low back pain. The trial compares pain relief by a device that delivered high level pulsed heat (45 °C) to pain relief delivered by a steady heat lower temperature device (37 °C). The hypothesis is that the high temperature pulsed heat device will produce significantly better pain relief as compared to the lower level steady heat device. The secondary hypothesis is that pain relief will occur faster in the high pulsed heat device as compared to the control device.
Detailed description
1. This study will be conducted under Good Clinical Practice (GCP) guidelines and suitable for FDA submission in order to obtain claims such as, provides temporary pain relief for people with chronic lower back pain. 2. The study is designed as a randomized double blinded two arm placebo controlled trial. 3. The active arm (N=50) receives two active Series 5 heating units (Soovu Labs Inc.) that pulse heat to 45ºC (113ºF). The active Series 5 placebo arm (N=50) gets two identical appearing units (Soovu Labs Inc.) that produce low level of heat at 37ºC (96.8 ºF). For reference skin temperature is about 33 ºC and room temperature 23 ºC. 4. The study hypothesis is that in subjects with chronic low back pain, the active heating device group will show a statistical improvement in pain scores as compared to the active placebo group as measured by the primary outcome measure, the Numeric Pain Scale (NPS). 5. Subjects will be told the study is examining the pain relieving effectiveness of two different levels of heat. 6. Subjects must have chronic low back pain with a pretreatment level of pain 4 or greater on a 0-10 point scale. By definition chronic low back pain is a condition that has been present for at least 6 months and must be present more days than not over the course of a week. 7. No use of pain medications during the four hour trial. However, pain medications including schedule IIIs can be taken up to 30 minutes before the start of the trial. Medications permitted during the trial include medications such as tramadol, codeine, NSIADs, gabapentin and acetaminophen. Non-pain medications for medical indications are permitted at any time. Subjects taking schedule II opioids are excluded from the study. Schedule II opioids include oxycodone, hydromorphone, hydrocodone, fentanyl and methadone. 8. Ages 22 through 70 inclusive. 9. Exclusion criteria- sciatica if present must be rated significantly less than the non -radiating component of the low back pain, pregnancy, skin lesions such as open skin or sores, scar tissue, skins grafts, old burns over the treatment area or neuropathic sensory loss over the treatment site. 10. Pre-treatment assessments include, Numeric Pain Scale, and a pre-study questionnaire (history, treatment preferences, short form Brief Pain Inventory, and short form Magill). 11. Each treatment session lasts approximately 30 minutes with pain assessments throughout using NPS (primary outcome measure). Secondary outcome measures include measures of comfort and willingness to endorse their treatment. 12. All subjects are followed for a 4 hour post treatment period with intermittent pain assessments (NPS). 13. At the end of the post treatment period, subjects complete a post study usability questionnaire and a blinding questionnaire.
Interventions
The Soovu Labss Inc. battery powered device is a one inch diameter heating pod that attaches to the user via a ring system. The device may be programmed to deliver a wide variety of treatment algorithms. Control of the devices are through a phone-based bluetooth connection. For this study the treatment algorithm used for subjects in this arm (experimental) used a temperature that was ramped from skin temperature to a peak temperature of 45° C and cycled for 30 minutes.
The Soovu Labs Inc. battery powered device is a one inch diameter heating pod that attaches to the user via a ring system. The device may be programmed to deliver a wide variety of treatment algorithms. Control of the devices are via a phone-based bluetooth connection. For this control arm of the study the treatment algorithm held a steady temperature of 37 degrees C for a 30 minute period.
Sponsors
Study design
Masking description
Double blinded with a control condition using a low level heating (37 degree C) device. The active experimental device used heast that intermittently pulsed to 45 degrees C.
Intervention model description
Double blinded, randomized, active placebo controlled trial
Eligibility
Inclusion criteria
1. Chronic low back pain. By definition chronic low back pain is a condition that has been present for at least 6 months on more days than not. If there is a radiating component of the low back pain the radiating component must have a pain rating less than the non-radiating component of the low back pain. 2. Subjects must have pretreatment level of pain 4 or greater. 3. Ages 22 through 70 inclusive 4. Non pain medications are permitted as needed for conditions such as hypertension, diabetes, heart disease etc. 5. Medications such as tramadol, codeine, NSIADs, gabapentin and acetaminophen are not permitted during the trial but may be included under the discretion of the study physician. 6. Subjects must have a reliable method for clinic contact and follow-up. \-
Exclusion criteria
1. Sciatica or radicular pain without non-radiating low back pain, cancer, radicular pain greater than the non-radiating component of low back pain, pregnancy, skin lesions such as open skin or sores, scar tissue, skins grafts, old burns over the treatment area. 2. Medications including oxycodone, hydromorphone, hydrocodone, fentanyl and methadone are excluded from the trial but may be included under the discretion of the study physician. Such decisions will be documented should they occur. \-
Design outcomes
Primary
| Measure | Time frame | Description |
|---|---|---|
| Pain Rating | Changes in the Numerical Pain Scale will be compared between baseline and 30 minutes after minutes of treatment. | Numerical Pain Score (NPS) Range Zero (minimum) through ten (maximum score) thirty minutes after completion of the treatment session. |
Other
| Measure | Time frame | Description |
|---|---|---|
| Pain Relief 30 Minutes After the Intervention | Changes in the Numerical Pain Scale will be compared between baseline and thirty minutes post intervention. | Numerical Pain Scale Range Zero (minimum) through ten (maximum score) In this study a lower number on the Numeric Pain Scale means the subject(s) has less pain. Less pain particularly in the experimental group is a better outcome. |
Countries
United States
Participant flow
Participants by arm
| Arm | Count |
|---|---|
| High Level Pulsed Heat Subjects randomized to this arm received a generation 5 device (Soovu Labs Inc.) that produced 30 minutes of heat. The heat was delivered as waves peaking at 45° C.
Generation 5 device Soovu Labs Inc.: The Soovu Labs Inc. battery powered device is a one inch diameter heating pod that attaches to the user via a ring system. The device may be programmed to deliver a wide variety of treatment algorithms. Control of the devices are through a phone-based bluetooth connection. | 49 |
| Low Level Steady Heat Subjects randomized to this arm received an identical device (Soovu Labs Inc.) that produced 30 minutes ot heat. The heat was delivered in a steady manner at 37° C.
Generation 5 device Soovu Labs Inc.: The Soovu Labs Inc. battery powered device is a one inch diameter heating pod that attaches to the user via a ring system. The device may be programmed to deliver a wide variety of treatment algorithms. Control of the devices are through a phone-based bluetooth connection. | 51 |
| Total | 100 |
Baseline characteristics
| Characteristic | High Level Pulsed Heat | Low Level Steady Heat | Total |
|---|---|---|---|
| Age, Categorical <=18 years | 0 Participants | 0 Participants | 0 Participants |
| Age, Categorical >=65 years | 5 Participants | 3 Participants | 8 Participants |
| Age, Categorical Between 18 and 65 years | 44 Participants | 48 Participants | 92 Participants |
| Numerical Pain Scale (NPS) | 5.56 units on a scale STANDARD_DEVIATION 0.06 | 5.55 units on a scale STANDARD_DEVIATION 0.07 | 5.55 units on a scale STANDARD_DEVIATION 0.06 |
| Race and Ethnicity Not Collected | — | — | 0 Participants |
| Region of Enrollment United States | 49 participants | 51 participants | 100 participants |
| Sex: Female, Male Female | 33 Participants | 34 Participants | 67 Participants |
| Sex: Female, Male Male | 16 Participants | 17 Participants | 33 Participants |
Adverse events
| Event type | EG000 affected / at risk | EG001 affected / at risk |
|---|---|---|
| deaths Total, all-cause mortality | 0 / 49 | 0 / 51 |
| other Total, other adverse events | 0 / 49 | 0 / 51 |
| serious Total, serious adverse events | 0 / 49 | 0 / 51 |
Outcome results
Pain Rating
Numerical Pain Score (NPS) Range Zero (minimum) through ten (maximum score) thirty minutes after completion of the treatment session.
Time frame: Changes in the Numerical Pain Scale will be compared between baseline and 30 minutes after minutes of treatment.
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| High Level Pulsed Heat | Pain Rating | -1.40 reduction in pain score on a scale | Standard Error 0.016 |
| Low Level Steady Heat | Pain Rating | -0.68 reduction in pain score on a scale | Standard Error 0.019 |
Pain Relief 30 Minutes After the Intervention
Numerical Pain Scale Range Zero (minimum) through ten (maximum score) In this study a lower number on the Numeric Pain Scale means the subject(s) has less pain. Less pain particularly in the experimental group is a better outcome.
Time frame: Changes in the Numerical Pain Scale will be compared between baseline and thirty minutes post intervention.
Population: All subjects with longstanding chronic low back pain median duration 10 years. Subjects who received the high level pulsed heat device with heat pulses up to 45 degrees C. Steady heat subjects received a device that heated to 37 degrees C at a steady level.
| Arm | Measure | Value (MEAN) | Dispersion |
|---|---|---|---|
| High Level Pulsed Heat | Pain Relief 30 Minutes After the Intervention | -1.40 units on a scale | Standard Error 0.016 |
| Low Level Steady Heat | Pain Relief 30 Minutes After the Intervention | -0.68 units on a scale | Standard Error 0.019 |