Left Bundle Area Versus Selective His Bundle Pacing

Left Bundle Area Versus Selective His Bundle Pacing (LEFTBASH): Single Center, Open Label, Randomized Pilot Study to Evaluate Capture Thresholds and Acute Echocardiographic Hemodynamic Characteristics

Status
Terminated
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT04093414
Acronym
LEFTBASH
Enrollment
9
Registered
2019-09-18
Start date
2019-10-07
Completion date
2022-02-22
Last updated
2022-08-17

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Atrioventricular Block, Bradycardia, Sinus Node Dysfunction

Keywords

Pacemaker, Bundle of His, Left Bundle Branch

Brief summary

This is a voluntary research study to find out which location in the heart a pacemaker wire is the most efficient for a patient's heart and for battery life. Patients who volunteer and are eligible for the study will be randomized to receive one of two positions for the wire to be screwed into, in addition to studying multiple positions in the heart during the pacemaker insertion. Enrolled patients will be in the study for 1 year. They will also have an Ultrasound of their heart performed to assess how the pacemaker wire is affecting their heart. Pacemakers are connected to the heart by wires that are screwed into the heart. The wires can be connected to the heart in different places, which can affect how well the heart beats over time. The typical position is at the tip of the heart. This position may cause the heart to beat inefficiently. Over time, this could lead to weakened heart muscle, irregular heart rhythm, and more hospitalizations. The heart has special muscle cells and fibers that carry electrical signals through and around the heart. An alternative spot to place the pacemaker wire is in an area where these special cells are grouped together (called the HIS bundle). The pacemaker wire can be connected to the heart at a location which may allow the heart to beat more efficiently when compared to putting the wire at traditional spots in the heart (called HIS bundle pacing). However, sometimes connecting the wire into the HIS bundle may cause the pacemaker battery to wear out faster. Physicians can also connect the pacemaker wired near the HIS bundle (called Left left Bundle bundle area pacing). The study physicians hope this will allow the heart to beat more efficiently without causing the battery to wear out faster. The study physicians would like to study how different wire positions change heart beat efficiency and how long the pacemaker battery lasts when the wires are placed in different locations. This study will connect the pacemaker wire at either the HIS Bundle or the left bundle area pacing, to see how effectively the heart pumps and how much battery is being used.

Detailed description

This is a single center, open label, prospective randomized pilot study to evaluate the capture threshold of His Bundle versus left bundle area pacing. Secondary analysis will focus on changes in left ventricular performance and mechanical synchrony. Symptomatic bradyarrhythmias are effectively treated with cardiac pacemakers.The amount of pacing by the lead positioned in the bottom chamber of the heart at traditional sites such as the right ventricular apex have been associated with increased rates of atrial fibrillation, heart failure, and mortality. Traditional pacing sites result in cardiac electromechanical dyssynchrony, for which alternate pacing sites to minimize these untoward effects have been sought. HIS bundle pacing, which utilizes a patient's native conduction, has demonstrated improved electrical synchrony and left ventricular function when compared to traditional pacing at the tip of the bottom chamber. Barriers to wide spread application to this technique include the His bundle anatomic location and its attendant difficulties associated with implant, as well as higher capture thresholds leading to decreased battery duration of the pacemaker. An alternative to HIS bundle pacing is placing the lead just past the HIS bundle area, which is further in the heart, and to actively fixate the lead into the interventricular septum. This is referred to as Left Bundle Pacing, as it may electrically capture the left bundle, which would simulate a patient's native conduction. The researchers will evaluate the two different pacing sites (HIS bundle and Left Bundle area sites) to determine how effectively the heart pumps with each pacing site. Patients requiring pacemaker implant will be screened for study eligibility and approached for informed consent. Baseline assessments including echocardiogram and ECG will be obtained. A standard of care echo done within 3 months of the procedure will be used as the baseline echo. Once enrolled, the researchers will randomize patients to one of two arms. One arm will fixate the ventricular pacemaker lead to the HIS bundle area while the other arm will have the leads fixed into the left bundle area. The pacemaker wires will be connected in accordance with randomization group. The day after the procedure, a chest x-ray, EKG and pacemaker evaluation will be performed. An echocardiogram will be obtained at 3 months post procedure. At 6 and 12 months post procedure, EKG rhythm strips will be obtained from patients' standard of care remote pacemaker monitoring systems.

Interventions

DEVICESelect Secure pacing lead

A Select Secure pacing lead is placed in the bundle of His or Left bundle branch area based upon randomization

Sponsors

David Haines, MD
Lead SponsorOTHER

Study design

Allocation
RANDOMIZED
Intervention model
PARALLEL
Primary purpose
TREATMENT
Masking
NONE

Intervention model description

Single center, open label, randomized, prospective

Eligibility

Sex/Gender
ALL
Age
19 Years to No maximum
Healthy volunteers
No

Inclusion criteria

1. Over 18 years old. 2. With signed consent. 3. Pacemaker indication according to 2018 American College of Cardiology/American Heart Association/Heart Rhythm Society Guideline on the Evaluation and Management of Patients with Bradycardia and Cardiac Conduction Delay. With one or both of the following: 1. Symptomatic sinus node dysfunction. 2. Symptomatic Atrioventricular (AV) block or high degree AV block. 3. Tachy-Brady syndrome

Exclusion criteria

1. Previously implanted cardiac pacing devices except transvenous temporary pacemaker. 2. Patients who are eligible for appropriate cardiac resynchronization therapy(CRT) or implantable cardiovert defibrillator (ICD) implantation 3. Patients with prior septal myectomy 4. Patients with prior surgical or transcatheter aortic valve replacement 5. Anatomy precluding implant evaluated during the screening or identified during the procedure. 6. Those without ability to achieve selective His bundle pacing evaluated during the screening or identified during the procedure 7. Pregnant women

Design outcomes

Primary

MeasureTime frameDescription
Ventricular Capture Threshold, 3 Months Unipolar or Bipolar3 monthsVentricular capture threshold is the minimum amplitude of electrical signal from the pacemaker that consistently results in capture of the ventricular myocardium (normal contraction of the ventricle after electrical depolarization) with a 1.0 millisecond pulse width setting in unipolar or Bipolar output modes, measured in volts. During pacemaker placement, the ideal ventricular capture threshold is determined by delivering a series of pulses starting at 0.75 volts and increasing by 0.125 V with each trial until consistent ventricular contraction is achieved. Pacemaker settings are programmed using the ventricular capture threshold, adjusted to include a safety margin, and may be re-adjusted over time if clinically necessary. The measurement is automatically recorded within the pacemaker and will be interrogated from the pacemaker device at 3 months post implantation.

Secondary

MeasureTime frameDescription
Stroke Volume - Intrinsic Conductionat index procedure prior to lead fixationVolume of blood in milliliters pumped from the left ventricle of the heart at each beat with no pacing applied
Mechanical Dyssynchrony of Anterior Left Ventricle Myocardial Wall- Intrinsic Conductionat index procedure prior to lead fixationTime to peak systolic velocity of the anterior left ventricle myocardial wall in milliseconds with no pacing applied, elicited by tissue Doppler
Mechanical Dyssynchrony of Inferior Left Ventricle Myocardial Wall- Intrinsic Conductionat index procedure prior to lead fixationTime to peak systolic velocity of the inferior left ventricle myocardial wall in milliseconds with no pacing applied, elicited by tissue Doppler
Mechanical Dyssynchrony of Inferior-septal Left Ventricle Myocardial Wall- Intrinsic Conductionat index procedure prior to lead fixationTime to peak systolic velocity of the inferior-septal left ventricle myocardial wall in milliseconds with no pacing applied, elicited by tissue Doppler
Mechanical Dyssynchrony of Left Ventricle Anterior-septal Myocardial Wall- Intrinsic Conductionat index procedure prior to lead fixationTime to peak systolic velocity of the left ventricle anterior-septal myocardial wall in milliseconds with no pacing applied, elicited by tissue Doppler
Mechanical Dyssynchrony of Lateral Left Ventricle Myocardial Wall- Intrinsic Conductionat index procedure prior to lead fixationTime to peak systolic velocity of the lateral left ventricle myocardial wall in milliseconds with no pacing applied, elicited by tissue Doppler
Mechanical Dyssynchrony of Inferior-lateral Left Ventricle Myocardial Wall- Intrinsic Conductionat index procedure prior to lead fixationTime to peak systolic velocity of the inferior-lateral left ventricle myocardial wall in milliseconds with no pacing applied, elicited by tissue Doppler
QRS Duration - Intrinsic Conductionat index procedure prior to lead fixationDuration (time in milliseconds) of the QRS wave complex interval measured from the end of the PR interval to the end of the S wave measured on a 12-lead electrocardiogram, with intrinsic conduction (prior to implantation of the pacemaker), indicating the length of time required for the electrical depolarization of the right and left ventricles of the heart and contraction of the large ventricular muscles.
QRS Duration After Lead Fixationat index procedure following final lead fixationDuration (time in milliseconds) of the QRS wave complex interval measured from the end of the PR interval to the end of the S wave measured on a 12-lead electrocardiogram, with intrinsic conduction (prior to implantation of the pacemaker), indicating the length of time required for the electrical depolarization of the right and left ventricles of the heart and contraction of the large ventricular muscles. A normal duration is between 80-100 milliseconds. A QRS duration of greater than 120 milliseconds is considered abnormal.
Left Ventricular Ejection Fraction (LVEF) - 3 Months3 monthsLVEF is the percentage of blood pumped from the left ventricle of the heart with each beat. It is calculated as the fraction of chamber volume ejected in systole (stroke volume) in relation to the volume of the blood in the ventricle at the end of diastole (end-diastolic volume). Volumes are measured via ultrasound in an echocardiogram. A healthy LVEF ranges from 50-70%. LVEF less than 40% are considered low and indicate some degree of heart failure. LVEF less than 35% are considered dangerous and indicate a subject at risk for arrhythmia.
Left Ventricular Ejection Fraction (LVEF) - Intrinsic Conductionat index procedure prior to lead fixationPercentage of blood pumped from the left ventricle of the heart at each beat with no pacing applied
Mechanical Dyssynchrony of Anterior Left Ventricle Myocardial Wall - 3 Months3 monthsTime to peak systolic velocity of the anterior left ventricle myocardial wall in milliseconds 3 months after final pacing lead fixation, elicited by tissue Doppler
Mechanical Dyssynchrony of Inferior Left Ventricle Myocardial Wall - 3 Months3 monthsTime to peak systolic velocity of the inferior left ventricle myocardial wall in milliseconds 3 months after final pacing lead fixation, elicited by tissue Doppler
Mechanical Dyssynchrony of Left Ventricle Anterior-septal Myocardial Wall - 3 Months3 monthsTime to peak systolic velocity of the left ventricle anterior-septal myocardial wall in milliseconds 3 months after final pacing lead fixation, elicited by tissue Doppler
Mechanical Dyssynchrony of Left Ventricle Inferior-septal Myocardial Wall- 3 Months3 monthsTime to peak systolic velocity of the left ventricle inferior-septal myocardial wall in milliseconds 3 months after final pacing lead fixation, elicited by tissue Doppler
Mechanical Dyssynchrony of Lateral Left Ventricle Myocardial Wall- 3 Months3 monthsTime to peak systolic velocity of the lateral left ventricle myocardial wall in milliseconds 3 months after final lead fixation, elicited by tissue Doppler
Mechanical Dyssynchrony of Inferior-lateral Left Ventricle Myocardial Wall- 3 Months3 monthsTime to peak systolic velocity of the inferior-lateral left ventricle myocardial wall in milliseconds 3 months after final lead fixation, elicited by tissue Doppler
QRS Duration- 3 Months3 monthsDuration (time in milliseconds) of the QRS wave complex interval measured from the end of the PR interval to the end of the S wave measured on a 12-lead electrocardiogram, 3 months after final pacing lead fixation, indicating the length of time required for the electrical depolarization of the right and left ventricles of the heart and contraction of the large ventricular muscles. A normal duration is between 80-100 milliseconds. A QRS duration of greater than 120 milliseconds is considered abnormal.
Capture Threshold - 6 Months, Unipolar or Bipolar6 monthsVentricular capture threshold is the minimum amplitude of electrical signal from the pacemaker that consistently results in capture of the ventricular myocardium (normal contraction of the ventricle after electrical depolarization) with a 1.0 millisecond pulse width setting in unipolar or Bipolar output modes, whichever value is lower, measured in volts. During pacemaker placement, the ideal ventricular capture threshold is determined by delivering a series of pulses starting at 0.75 volts and increasing by 0.125 V with each trial until consistent ventricular contraction is achieved. Pacemaker settings are programmed using the ventricular capture threshold, adjusted to include a safety margin, and may be re-adjusted over time if clinically necessary. The measurement is automatically recorded within the pacemaker and will be interrogated from the pacemaker device at 3 months post implantation.
Capture Threshold - 12 Months, Unipolar or Bipolar12 monthsVentricular capture threshold is the minimum amplitude of electrical signal from the pacemaker that consistently results in capture of the ventricular myocardium (normal contraction of the ventricle after electrical depolarization) with a 1.0 millisecond pulse width setting in unipolar or Bipolar output modes, whichever value is lower, measured in volts. During pacemaker placement, the ideal ventricular capture threshold is determined by delivering a series of pulses starting at 0.75 volts and increasing by 0.125 V with each trial until consistent ventricular contraction is achieved. Pacemaker settings are programmed using the ventricular capture threshold, adjusted to include a safety margin, and may be re-adjusted over time if clinically necessary. The measurement is automatically recorded within the pacemaker and will be interrogated from the pacemaker device at 3 months post implantation.
Stroke Volume - 3 Months3 monthsVolume of blood in milliliters pumped from the left ventricle of the heart at each beat 3 months after final pacing lead fixation

Countries

United States

Participant flow

Participants by arm

ArmCount
Selective or Non Selective His Bundle Pacing
Pacemaker wires placed in Bundle of His Select Secure pacing lead: A Select Secure pacing lead is placed in the bundle of His or Left bundle branch area based upon randomization
3
Left Bundle Area Pacing
Pacemaker wires placed in Left Bundle Branch area Select Secure pacing lead: A Select Secure pacing lead is placed in the bundle of His or Left bundle branch area based upon randomization
6
Total9

Withdrawals & dropouts

PeriodReasonFG000FG001
3 Month Follow upPhysician Decision01
6month Follow upPhysician Decision13

Baseline characteristics

CharacteristicLeft Bundle Area PacingTotalSelective or Non Selective His Bundle Pacing
Age, Continuous78.83 years
STANDARD_DEVIATION 9.7
79.44 years
STANDARD_DEVIATION 8.95
80.67 years
STANDARD_DEVIATION 9.02
Race and Ethnicity Not Collected0 Participants
Region of Enrollment
United States
6 participants9 participants3 participants
Sex: Female, Male
Female
2 Participants3 Participants1 Participants
Sex: Female, Male
Male
4 Participants6 Participants2 Participants

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
deaths
Total, all-cause mortality
0 / 30 / 6
other
Total, other adverse events
0 / 30 / 6
serious
Total, serious adverse events
0 / 30 / 6

Outcome results

Primary

Ventricular Capture Threshold, 3 Months Unipolar or Bipolar

Ventricular capture threshold is the minimum amplitude of electrical signal from the pacemaker that consistently results in capture of the ventricular myocardium (normal contraction of the ventricle after electrical depolarization) with a 1.0 millisecond pulse width setting in unipolar or Bipolar output modes, measured in volts. During pacemaker placement, the ideal ventricular capture threshold is determined by delivering a series of pulses starting at 0.75 volts and increasing by 0.125 V with each trial until consistent ventricular contraction is achieved. Pacemaker settings are programmed using the ventricular capture threshold, adjusted to include a safety margin, and may be re-adjusted over time if clinically necessary. The measurement is automatically recorded within the pacemaker and will be interrogated from the pacemaker device at 3 months post implantation.

Time frame: 3 months

ArmMeasureValue (MEAN)Dispersion
Selective or Non Selective His Bundle PacingVentricular Capture Threshold, 3 Months Unipolar or Bipolar0.66 voltsStandard Deviation 0.14
Left Bundle Area PacingVentricular Capture Threshold, 3 Months Unipolar or Bipolar0.65 voltsStandard Deviation 0.13
Secondary

Capture Threshold - 12 Months, Unipolar or Bipolar

Ventricular capture threshold is the minimum amplitude of electrical signal from the pacemaker that consistently results in capture of the ventricular myocardium (normal contraction of the ventricle after electrical depolarization) with a 1.0 millisecond pulse width setting in unipolar or Bipolar output modes, whichever value is lower, measured in volts. During pacemaker placement, the ideal ventricular capture threshold is determined by delivering a series of pulses starting at 0.75 volts and increasing by 0.125 V with each trial until consistent ventricular contraction is achieved. Pacemaker settings are programmed using the ventricular capture threshold, adjusted to include a safety margin, and may be re-adjusted over time if clinically necessary. The measurement is automatically recorded within the pacemaker and will be interrogated from the pacemaker device at 3 months post implantation.

Time frame: 12 months

ArmMeasureValue (MEAN)Dispersion
Selective or Non Selective His Bundle PacingCapture Threshold - 12 Months, Unipolar or Bipolar1.31 VoltsStandard Deviation 0.61
Left Bundle Area PacingCapture Threshold - 12 Months, Unipolar or Bipolar1.12 VoltsStandard Deviation 0
Secondary

Capture Threshold - 6 Months, Unipolar or Bipolar

Ventricular capture threshold is the minimum amplitude of electrical signal from the pacemaker that consistently results in capture of the ventricular myocardium (normal contraction of the ventricle after electrical depolarization) with a 1.0 millisecond pulse width setting in unipolar or Bipolar output modes, whichever value is lower, measured in volts. During pacemaker placement, the ideal ventricular capture threshold is determined by delivering a series of pulses starting at 0.75 volts and increasing by 0.125 V with each trial until consistent ventricular contraction is achieved. Pacemaker settings are programmed using the ventricular capture threshold, adjusted to include a safety margin, and may be re-adjusted over time if clinically necessary. The measurement is automatically recorded within the pacemaker and will be interrogated from the pacemaker device at 3 months post implantation.

Time frame: 6 months

Population: 4 patients completed data collection for 6 month follow-up

ArmMeasureValue (MEAN)Dispersion
Selective or Non Selective His Bundle PacingCapture Threshold - 6 Months, Unipolar or Bipolar1.18 VoltsStandard Deviation 0.79
Left Bundle Area PacingCapture Threshold - 6 Months, Unipolar or Bipolar0.93 VoltsStandard Deviation 0.08
Secondary

Left Ventricular Ejection Fraction (LVEF) - 3 Months

LVEF is the percentage of blood pumped from the left ventricle of the heart with each beat. It is calculated as the fraction of chamber volume ejected in systole (stroke volume) in relation to the volume of the blood in the ventricle at the end of diastole (end-diastolic volume). Volumes are measured via ultrasound in an echocardiogram. A healthy LVEF ranges from 50-70%. LVEF less than 40% are considered low and indicate some degree of heart failure. LVEF less than 35% are considered dangerous and indicate a subject at risk for arrhythmia.

Time frame: 3 months

Population: Echocardiogram data not collected for any subject for any outcome in this study

Secondary

Left Ventricular Ejection Fraction (LVEF) - Intrinsic Conduction

Percentage of blood pumped from the left ventricle of the heart at each beat with no pacing applied

Time frame: at index procedure prior to lead fixation

Population: Echocardiogram data not collected for any subject for any outcome in this study.

Secondary

Mechanical Dyssynchrony of Anterior Left Ventricle Myocardial Wall - 3 Months

Time to peak systolic velocity of the anterior left ventricle myocardial wall in milliseconds 3 months after final pacing lead fixation, elicited by tissue Doppler

Time frame: 3 months

Population: Echocardiogram data not collected for any subject for any outcome in this study

Secondary

Mechanical Dyssynchrony of Anterior Left Ventricle Myocardial Wall- Intrinsic Conduction

Time to peak systolic velocity of the anterior left ventricle myocardial wall in milliseconds with no pacing applied, elicited by tissue Doppler

Time frame: at index procedure prior to lead fixation

Population: Echocardiogram data not collected for any subject for any outcome in this study

Secondary

Mechanical Dyssynchrony of Inferior-lateral Left Ventricle Myocardial Wall- 3 Months

Time to peak systolic velocity of the inferior-lateral left ventricle myocardial wall in milliseconds 3 months after final lead fixation, elicited by tissue Doppler

Time frame: 3 months

Population: Echocardiogram data not collected for any subject for any outcome in this study

Secondary

Mechanical Dyssynchrony of Inferior-lateral Left Ventricle Myocardial Wall- Intrinsic Conduction

Time to peak systolic velocity of the inferior-lateral left ventricle myocardial wall in milliseconds with no pacing applied, elicited by tissue Doppler

Time frame: at index procedure prior to lead fixation

Population: Echocardiogram data not collected for any subject for any outcome in this study

Secondary

Mechanical Dyssynchrony of Inferior Left Ventricle Myocardial Wall - 3 Months

Time to peak systolic velocity of the inferior left ventricle myocardial wall in milliseconds 3 months after final pacing lead fixation, elicited by tissue Doppler

Time frame: 3 months

Population: Echocardiogram data not collected for any subject for any outcome in this study

Secondary

Mechanical Dyssynchrony of Inferior Left Ventricle Myocardial Wall- Intrinsic Conduction

Time to peak systolic velocity of the inferior left ventricle myocardial wall in milliseconds with no pacing applied, elicited by tissue Doppler

Time frame: at index procedure prior to lead fixation

Population: Echocardiogram data not collected for any subject for any outcome in this study

Secondary

Mechanical Dyssynchrony of Inferior-septal Left Ventricle Myocardial Wall- Intrinsic Conduction

Time to peak systolic velocity of the inferior-septal left ventricle myocardial wall in milliseconds with no pacing applied, elicited by tissue Doppler

Time frame: at index procedure prior to lead fixation

Population: Echocardiogram data not collected for any subject for any outcome in this study

Secondary

Mechanical Dyssynchrony of Lateral Left Ventricle Myocardial Wall- 3 Months

Time to peak systolic velocity of the lateral left ventricle myocardial wall in milliseconds 3 months after final lead fixation, elicited by tissue Doppler

Time frame: 3 months

Population: Echocardiogram data not collected for any subject for any outcome in this study

Secondary

Mechanical Dyssynchrony of Lateral Left Ventricle Myocardial Wall- Intrinsic Conduction

Time to peak systolic velocity of the lateral left ventricle myocardial wall in milliseconds with no pacing applied, elicited by tissue Doppler

Time frame: at index procedure prior to lead fixation

Population: Echocardiogram data not collected for any subject for any outcome in this study

Secondary

Mechanical Dyssynchrony of Left Ventricle Anterior-septal Myocardial Wall - 3 Months

Time to peak systolic velocity of the left ventricle anterior-septal myocardial wall in milliseconds 3 months after final pacing lead fixation, elicited by tissue Doppler

Time frame: 3 months

Population: Echocardiogram data not collected for any subject for any outcome in this study

Secondary

Mechanical Dyssynchrony of Left Ventricle Anterior-septal Myocardial Wall- Intrinsic Conduction

Time to peak systolic velocity of the left ventricle anterior-septal myocardial wall in milliseconds with no pacing applied, elicited by tissue Doppler

Time frame: at index procedure prior to lead fixation

Population: Echocardiogram data not collected for any subject for any outcome in this study

Secondary

Mechanical Dyssynchrony of Left Ventricle Inferior-septal Myocardial Wall- 3 Months

Time to peak systolic velocity of the left ventricle inferior-septal myocardial wall in milliseconds 3 months after final pacing lead fixation, elicited by tissue Doppler

Time frame: 3 months

Population: Echocardiogram data not collected for any subject for any outcome in this study

Secondary

QRS Duration- 3 Months

Duration (time in milliseconds) of the QRS wave complex interval measured from the end of the PR interval to the end of the S wave measured on a 12-lead electrocardiogram, 3 months after final pacing lead fixation, indicating the length of time required for the electrical depolarization of the right and left ventricles of the heart and contraction of the large ventricular muscles. A normal duration is between 80-100 milliseconds. A QRS duration of greater than 120 milliseconds is considered abnormal.

Time frame: 3 months

Population: 1 patient in the left bundle area pacing group had data missing for this outcome

ArmMeasureValue (MEAN)Dispersion
Selective or Non Selective His Bundle PacingQRS Duration- 3 Months108.67 millisecondsStandard Deviation 19.22
Left Bundle Area PacingQRS Duration- 3 Months123 millisecondsStandard Deviation 46.26
Secondary

QRS Duration After Lead Fixation

Duration (time in milliseconds) of the QRS wave complex interval measured from the end of the PR interval to the end of the S wave measured on a 12-lead electrocardiogram, with intrinsic conduction (prior to implantation of the pacemaker), indicating the length of time required for the electrical depolarization of the right and left ventricles of the heart and contraction of the large ventricular muscles. A normal duration is between 80-100 milliseconds. A QRS duration of greater than 120 milliseconds is considered abnormal.

Time frame: at index procedure following final lead fixation

Population: One patient in left bundle area pacing group had missing data for this outcome

ArmMeasureValue (MEAN)Dispersion
Selective or Non Selective His Bundle PacingQRS Duration After Lead Fixation129.33 milllisecondsStandard Deviation 35.8
Left Bundle Area PacingQRS Duration After Lead Fixation129.8 milllisecondsStandard Deviation 10.83
Secondary

QRS Duration - Intrinsic Conduction

Duration (time in milliseconds) of the QRS wave complex interval measured from the end of the PR interval to the end of the S wave measured on a 12-lead electrocardiogram, with intrinsic conduction (prior to implantation of the pacemaker), indicating the length of time required for the electrical depolarization of the right and left ventricles of the heart and contraction of the large ventricular muscles.

Time frame: at index procedure prior to lead fixation

Population: One patient in HIS bundle pacing group had no data for this outcome.

ArmMeasureValue (MEAN)Dispersion
Selective or Non Selective His Bundle PacingQRS Duration - Intrinsic Conduction145 millisecondsStandard Deviation 21.21
Left Bundle Area PacingQRS Duration - Intrinsic Conduction124.83 millisecondsStandard Deviation 26.63
Secondary

Stroke Volume - 3 Months

Volume of blood in milliliters pumped from the left ventricle of the heart at each beat 3 months after final pacing lead fixation

Time frame: 3 months

Population: Echocardiogram data not collected for any subject for any outcome in this study

Secondary

Stroke Volume - Intrinsic Conduction

Volume of blood in milliliters pumped from the left ventricle of the heart at each beat with no pacing applied

Time frame: at index procedure prior to lead fixation

Population: Echocardiogram data not collected for any subject for any outcome in this study