Skip to content

ivWatch SmartTouch Sensor: Device Validation for Non-Infiltrated Tissues

ivWatch SmartTouch Sensor: Device Validation for Non-Infiltrated Tissues

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT04065373
Enrollment
45
Registered
2019-08-22
Start date
2019-04-25
Completion date
2019-06-21
Last updated
2019-08-22

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Infiltration of Peripheral IV Therapy

Brief summary

A single arm study of 40 adult volunteers to assess the safety and efficacy of the ivWatch SmartTouch and fiber optic sensors when observing non-infiltrated (normal) tissues at common peripheral IV therapy sites.

Interventions

DEVICEivWatch Model 400 with SmartTouch and fiber optic sensor

The ivWatch Model 400 with SmartTouch or fiber optic sensor monitored tissue at common IV sites over a 24 hour period.

Sponsors

ivWatch, LLC
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
DIAGNOSTIC
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to No maximum
Healthy volunteers
Yes

Inclusion criteria

* Informed consent * Pass health screen * 18 years or older

Exclusion criteria

* Fail health screen

Design outcomes

Primary

MeasureTime frameDescription
Normal Tissue Red Notification Rate24 HoursThe ivWatch Model 400 issues red notifications to communicate the need for a clinician to check an IV site. A red notification suggests an increased likelihood that an IV infiltration may be occurring, with a greater likelihood relative to a yellow notification. This measure describes the average number of red notifications issued by the ivWatch device in a given day when monitoring normal, non-infiltrated tissues. The 95% confidence interval for the red notification rate was calculated using a negative binomial regression model.

Secondary

MeasureTime frameDescription
Normal Tissue Yellow Notification Rate24 HoursThe ivWatch Model 400 issues yellow notifications to communicate the need for a clinician to check an IV site. A yellow notification suggests an increased likelihood that an IV infiltration may be occurring, although at a lower likelihood relative to a red notification. This measure describes the average number of yellow notifications issued by the ivWatch device in a given day when monitoring normal, non-infiltrated tissues. The 95% confidence interval for the yellow notification rate was calculated using a negative binomial regression model.
Significant Skin Irritation or Disruption to Skin Integrity24 HoursThe number of IV sites with significant skin irritation or disruption to skin integrity assessed at the end of the study. The Clopper-Pearson method was used for estimating the binomial proportion confidence interval.

Countries

United States

Outcome results

None listed

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026