Skip to content

Desaturation Validation of INVSENSOR00038

Desaturation Validation of INVSENSOR00038

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT04017611
Enrollment
13
Registered
2019-07-12
Start date
2019-07-08
Completion date
2019-07-19
Last updated
2020-10-20

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Healthy

Brief summary

This study is designed to evaluate the trending accuracy of a noninvasive regional oximetry measurement of somatic oxygen saturation. One investigational sensor (INVSENSOR00038) will be placed on the somatic site of the volunteer. The values obtained by the test sensor will be compared to the reference value obtained from an FDA-cleared sensor placed on the somatic site on the contralateral side. Data will be collected from healthy adult subjects while undergoing a desaturation procedure wherein the concentration of oxygen inhaled is slowly reduced until the subject's oxygen saturation is approximately 70%. The subject will then be returned to inhaling room air.

Interventions

DEVICEINVSENSOR00038

Noninvasive regional oximeter

Sponsors

Masimo Corporation
Lead SponsorINDUSTRY

Study design

Allocation
NA
Intervention model
SINGLE_GROUP
Primary purpose
OTHER
Masking
NONE

Eligibility

Sex/Gender
ALL
Age
18 Years to 50 Years
Healthy volunteers
Yes

Inclusion criteria

* Subject is between 18 and 50 years of age. * Subject weighs a minimum of 110 lbs and no more than 250 lbs unless subject is over 6 feet tall. * Must have a hemoglobin value greater than or equal to 11 g/dL. * Baseline heart rate ≥ 45 bpm and ≤ 85 bpm. * Blood Pressure (Systolic BP \< 140 mmHg and Diastolic BP \< 90 mmHg). * Carbon Monoxide (CO) value less than or equal to 2.0% fractional carboxyhemoglobin (FCOHb) * Subject has a physical status of ASA I or II (American Society of Anesthesiology Class I; Healthy subjects without any systemic disease at all. American Society of Anesthesiology Class II; subjects with mild systemic disease) as it applies to the systemic disease portion of the classification. * Subject is able to read and communicate in English and understands the study and risks involved.

Exclusion criteria

* Subject is pregnant. * Subject smokes (including e-cigarette use). * Subject has a BMI \> 35 and has been classified as morbidly obese or at an increased risk for participation by a medical professional. * Subject has a history of fainting (vasovagal), blacking out or losing consciousness during or after a blood draw, or has a fear of blood draws. * Subject has open wounds, inflamed tattoos or piercings and/or visible healing wounds that a medical professional renders them at an increased risk for participation.\* * Subject experiences frequent or severe headaches and/or migraine headaches, migraine auras, altitude sickness, and/or headaches accompanied by visual changes or sensitivity to light or sound. * Subject has known drug or alcohol abuse and/or uses recreational drugs. * Subject has experienced a concussion or head injury with loss of consciousness within the last year. * Subject has any chronic bleeding disorders (i.e. hemophilia). * Subject has any history of a stroke, myocardial infarction, seizures or heart attack. * Subject has any cancer or history of cancer (not including skin cancer)\*. * Subject has a known neurological and/or psychiatric disorder (i.e. schizophrenia, bipolar disorder, Multiple Sclerosis, Huntington's disease) that interferes with the subjects' level of consciousness. * Subject has any cardiac dysrhythmia(s) (i.e. atrial fibrillation) and has not received clearance by their physician to participate. * Subject has Wolff-Parkinson-White Syndrome or Stokes-Adams Syndrome. * Subject has taken anticoagulant medication within the last 30 days (excluding nonsteroidal anti-inflammatory drugs (NSAIDS)). * Subject has donated blood within the past 4 weeks. * Subject has taken opioid pain medication within 24 hours before the study. * Subject has any type of infectious disease (i.e. Hepatitis, HIV, Tuberculosis, Flu, Malaria, Measles, etc.). * Subject is taking medications known to treat any type of infectious disease. * Subject has either signs or history of peripheral ischemia or carpal tunnel syndrome. * Subject has had invasive surgery within the past year- including but not limited to major dental surgery\*, appendix\*, plastic surgery\*. * Subject has had invasive surgery within the past year- including but not limited gallbladder, major fracture repairs (involving plates/ screws), jaw surgery, urinary tract surgery, major ENT surgery, joint replacement or gynecological surgeries, heart surgery or thoracic surgery. * Subject has symptoms of congestion, head colds, flu or other illnesses. * Subject is claustrophobic and/or has generalized anxiety disorder. * Subject has been in severe car accident(s) or a similar type of accident(s) requiring hospitalization within the last 12 months. * Subject has chronic unresolved asthma, lung disease or respiratory disease. * Subject is allergic to lidocaine, latex, adhesives, or plastic. * Subject has a heart condition, insulin-dependent diabetes or uncontrolled hypertension. * Subject has delivered vaginally, has had a pregnancy terminated, a miscarriage with hospitalization or had a C-section within the last 6 months. * Subject has any medical condition which in the judgment of the investigator and/or medical staff, renders them ineligible for participation in this study (Discretion of investigator/study staff).

Design outcomes

Primary

MeasureTime frameDescription
Trending Accuracy of INVSENSOR000381-5 hoursThis study will evaluate the trending accuracy of the INVSENSOR00038 relative to the reference value obtained from the comparative FDA-cleared device.

Countries

United States

Participant flow

Participants by arm

ArmCount
INVSENSOR00038
All subjects who are enrolled into the test group and participate in data collection receive the noninvasive INVSENSOR00038 sensor. INVSENSOR00038: Noninvasive regional oximeter
12
Total12

Withdrawals & dropouts

PeriodReasonFG000
Overall StudyWithdrawn prior to completion1

Baseline characteristics

CharacteristicINVSENSOR00038
Age, Categorical
<=18 years
0 Participants
Age, Categorical
>=65 years
0 Participants
Age, Categorical
Between 18 and 65 years
12 Participants
Race/Ethnicity, Customized
African American
3 Participants
Race/Ethnicity, Customized
Asian or Pacific Islander
2 Participants
Race/Ethnicity, Customized
Caucasian
4 Participants
Race/Ethnicity, Customized
Hispanic
3 Participants
Sex: Female, Male
Female
6 Participants
Sex: Female, Male
Male
6 Participants

Adverse events

Event typeEG000
affected / at risk
deaths
Total, all-cause mortality
0 / 13
other
Total, other adverse events
0 / 13
serious
Total, serious adverse events
0 / 13

Outcome results

Primary

Trending Accuracy of INVSENSOR00038

This study will evaluate the trending accuracy of the INVSENSOR00038 relative to the reference value obtained from the comparative FDA-cleared device.

Time frame: 1-5 hours

ArmMeasureValue (NUMBER)
INVSENSOR00038Trending Accuracy of INVSENSOR000382.5 percentage of regional oxygen saturation

Source: ClinicalTrials.gov · Data processed: Feb 4, 2026