Patent Hemostasis With Vasoband Versus TR Band

Randomized COmparison of Isolated Radial Artery ComPrEssioN Versus Radial and Ipsilateral Ulnar Artery Compression in Achieving Patent Hemostasis: OPEN-Radial Trial

Status
Completed
Phases
NA
Study type
Interventional
Source
ClinicalTrials.gov
Registry ID
NCT04002791
Acronym
OPEN-Radial
Enrollment
253
Registered
2019-07-01
Start date
2019-03-16
Completion date
2019-10-16
Last updated
2020-11-13

For informational purposes only — not medical advice. Sourced from public registries and may not reflect the latest updates. Terms

Conditions

Radial Artery Injury at Wrist and Hand Level

Brief summary

In patients referred for cardiac catheterization to be performed using transradial access, a randomized comparison will be performed with the primary endpoint of patency of radial artery at the time of application of hemostatic compression band. The single radial only TR-Band (Terumo, Japan) will be compared to the dual balloon Vasoband (Vasoinnovations, USA) capable of ipsilateral ulnar artery compression.

Detailed description

Introduction: TRA is increasing in utilization across the world due to its established benefits of reduction in access site complications, cost, and increase in patient comfort (1). The most common structural consequence of TRA is radial artery occlusion (RAO). It occurs in 2-10% of patients undergoing TRA (2). Several pharmacologic as well as non-pharmacologic practices have been associated with RAO. Maintaining a hardware profile smaller than the inner diameter of radial artery has been associated with better radial artery patency rates after the procedure (3). Use of unfractionated heparin and other parenteral anticoagulants administered during the procedure has also been associated with reduction in RAO, with a dose-dependent effect with better reduction with higher doses of heparin (4,5). One of the most effective intervention for reduction of RAO is avoidance of flow cessation in radial artery during hemostatic compression (6). This technique, called patent hemostasis, has been shown to unequivocally and sizably reduce the incidence of RAO. In patients with small diameter radial arteries as well as low blood pressures, the ability to achieve and maintain patent hemostasis has been observed to be lower than those otherwise. In view of the co-dominant nature of forearm arterial circulation, the radial and ulnar arteries are interdependent and compression of ipsilateral ulnar artery has been shown to increase flow as well as peak flow velocity in the radial artery (7). Ipsilateral ulnar artery compression during hemostatic compression of a punctured radial artery, has been shown to sizably and significantly reduce the incidence of RAO compared to patent hemostasis (8). Randomized and observational datasets utilizing this technique published so far have been using two devices to achieve radial and ulnar compression, with difficult ergonomics leading to frequent dislodgment of ulnar compression during the radial artery hemostatic process. The investigators intend to study the comparative efficacy of the first dedicated dual compression device in achieving patent hemostasis compared to the current popular radial artery compression device. Patients and Methods: Patients referred for diagnostic cardiac catheterization using TRA will be screened. Those providing written informed consent, will be enrolled in the study. A transradial cardiac catheterization will be performed using the standard technique as described below. Catheterization Technique: After sterile preparation as well as local anesthetic infiltration, 5 French hydrophilic introducer sheath will be introduced in the radial artery using standard puncture technique (9). Once the introducer is in the radial artery lumen, 200 µg of nitroglycerin, 2.5 mg of verapamil and 5000 IU of unfractionated heparin will be administered intra-arterially. 5 French or lower profile hardware will be used as per operator discretion to complete the procedure. Once the procedure is completed, the patient will be randomized to: Group 1, which will include patients receiving hemostatic compression using the current standard of care TR band (Terumo Corporation, Japan). The band will be applied according to the instructions for use, with optimal pressure applied to achieve full hemostasis. Group 2, will consist of patients who will receive Vaso-band (VasoInnovations, Inc, USA), applied according to instructions for use, with ulnar balloon inflated for the first 60 minutes of radial artery hemostatic compression, with the radial balloon inflated to obtain full hemostasis, after the sheath is removed. Primary Study End-point: Radial artery patency at 0-15 minutes after application of radial artery hemostatic compression. Monitoring of Patency: The radial artery patency will be monitored immediately after application of the band using reverse Barbeau test in both groups, and patency will be re-evaluated at 15 minutes after band application, 30 minutes after band application, 60 minutes after band application, as well as upon removal of the band at 120 minutes. If longer duration of application of pressure with the band is necessary in either group to achieve hemostasis, the bands will be applied for longer duration and the adequacy of hemostasis will be evaluated every 15 minutes after 120-minute duration and the band removed once adequately adequate hemostasis has been achieved. Radial patency will be assessed at 60 minutes after the band is removed. Baseline demographic as well as procedural data, pharmacologic data, as well as elaborately data will be collected on every patient. Statistical Analysis Plan: Categorical variables will be expressed as proportions, and differences between the two groups will be assessed using chi-square test, or Pearson test as deemed appropriate. Numeric variables will be assessed for normality of the distribution, and will be expressed as mean and standard deviation for normally distributed variables, and median as well as interquartile range for those without normal distribution. Parametric tests will be used to compare differences between the two groups for normally distributed numeric variables, and nonparametric tests will be used in those variables with a non-normal distribution. Multivariable analysis if necessary, will be performed using forward selection binary logistic regression for categorical dependent variable and linear regression for numeric variables. Received operative characteristic derived c-statistic will be used to assess the stability of the model. Sample Size Calculation: Based on the available literature, using the current standard of care TR band, patent hemostasis defined as patency of radial artery during hemostatic compression at 15 minutes into compression is achievable in 75% of patients undergoing TRA. The only published randomized trial using prophylactic ipsilateral ulnar artery compression during radial artery hemostatic compression has shown 96% achievement of patent hemostasis at 15 minutes after onset of compression. Using this using these data, using a chi-square model, a sample size of 250 to 300 patients equally divided between two groups, is necessary to achieve a 90% power with an alpha editor of 0.05. Expecting a 10% crossover rate will be accommodated a sample size of 200 patients will be needed. After randomizing 200 patients, an interim analysis will be performed. If event rates are low, the sample-size will be increased to 350 patients.

Interventions

DEVICEVasoBand

Dual-bladder radial artery hemostatic compression device capable of ipsilateral ulnar artery compression

Sponsors

Total Cardiovascular Solutions
Lead SponsorOTHER

Study design

Allocation
RANDOMIZED
Intervention model
PARALLEL
Primary purpose
PREVENTION
Masking
NONE

Intervention model description

Randomized comparison of patent hemostasis achievement with single bladder radial artery compression device with a dual-bladder ulnar compression capable radial artery compression device.

Eligibility

Sex/Gender
ALL
Age
18 Years to 99 Years
Healthy volunteers
No

Inclusion criteria

* Patients referred for cardiac catheterization undergoing transradial access

Exclusion criteria

* Previous ipsilateral radial artery cannulation, oral systemic anticoagulation therapy, thrombolytic therapy in the previous 7 days.

Design outcomes

Primary

MeasureTime frameDescription
Number of Patients Who Achieved Patent Hemostasis0-15 minutesPatent hemostasis achievement at 0-15 minutes after onset of radial artery hemostatic compression

Secondary

MeasureTime frameDescription
Number of Participants With Radial Artery Occlusion One Hour After Removal of Hemostatic Compression0-60 minutesAbsence of Patency of radial artery at 1-hour after removal of hemostatic compression

Countries

India

Participant flow

Participants by arm

ArmCount
Group 1: TR Band Arm
Patients will receive hemostatic compression using the current standard of care TR band (Terumo Corporation, Japan). The band will be applied according to the instructions for use, with optimal pressure applied using Patent hemostasis protocol to achieve full hemostasis. VasoBand: Dual-bladder radial artery hemostatic compression device capable of ipsilateral ulnar artery compression
127
Group 2: Vaso-band Arm
Patients will receive Vaso-band (VasoInnovations, Inc, USA), applied with ulnar balloon inflated with 15 ml of air, and the radial balloon inflated after the sheath is removed, to apply optimal pressure for obtaining full hemostasis. Ulnar balloon will be deflated after 60 minutes of radial artery hemostatic compression. VasoBand: Dual-bladder radial artery hemostatic compression device capable of ipsilateral ulnar artery compression
126
Total253

Baseline characteristics

CharacteristicGroup 1: TR Band ArmGroup 2: Vaso-band ArmTotal
Age, Continuous57 years
STANDARD_DEVIATION 10
59 years
STANDARD_DEVIATION 11
58 years
STANDARD_DEVIATION 10.7
Barbeau Test Pattern A70 Participants75 Participants145 Participants
Barbeau Test Pattern B53 Participants46 Participants99 Participants
Barbeau Test Pattern C4 Participants5 Participants9 Participants
Diabetes49 Participants46 Participants95 Participants
Height165 Centimeter165 Centimeter165 Centimeter
Hemoglobin13.5 Gram/Dl13.2 Gram/Dl13.3 Gram/Dl
Hypertension80 Participants79 Participants159 Participants
Platelets270 Cells x1000/ml250.5 Cells x1000/ml261 Cells x1000/ml
Previous CABG1 Participants4 Participants5 Participants
Previous PCI12 Participants12 Participants24 Participants
Race/Ethnicity, Customized
INDIAN
127 Participants126 Participants253 Participants
Serum Creatinine0.91 mg/dl0.91 mg/dl0.91 mg/dl
Sex: Female, Male
Female
39 Participants33 Participants72 Participants
Sex: Female, Male
Male
88 Participants93 Participants181 Participants
Weight72 Kilograms71 Kilograms71 Kilograms
Wrist Circumference21 cm20 cm20 cm

Adverse events

Event typeEG000
affected / at risk
EG001
affected / at risk
deaths
Total, all-cause mortality
0 / 1270 / 126
other
Total, other adverse events
0 / 1270 / 126
serious
Total, serious adverse events
16 / 1272 / 126

Outcome results

Primary

Number of Patients Who Achieved Patent Hemostasis

Patent hemostasis achievement at 0-15 minutes after onset of radial artery hemostatic compression

Time frame: 0-15 minutes

ArmMeasureValue (COUNT_OF_PARTICIPANTS)
Group 1: TR Band ArmNumber of Patients Who Achieved Patent Hemostasis94 Participants
Group 2: Vaso-band ArmNumber of Patients Who Achieved Patent Hemostasis122 Participants
Secondary

Number of Participants With Radial Artery Occlusion One Hour After Removal of Hemostatic Compression

Absence of Patency of radial artery at 1-hour after removal of hemostatic compression

Time frame: 0-60 minutes

ArmMeasureValue (COUNT_OF_PARTICIPANTS)
Group 1: TR Band ArmNumber of Participants With Radial Artery Occlusion One Hour After Removal of Hemostatic Compression13 Participants
Group 2: Vaso-band ArmNumber of Participants With Radial Artery Occlusion One Hour After Removal of Hemostatic Compression2 Participants